NSK
Varios Combi Pro Reprocessing Instruction
4 Pages

Preview
Page 1
Reprocessing Instruction in accordance with ISO17664
Varios Combi Pro 1 Preparation, Cleaning and Disinfection ・After cleaning remove any remaining moisture using a dry cloth or compressed air ensuring it is dry internally and externally. ・Cover the product with a cloth when drying with compressed air to prevent water aerosol from being produced. ・Use a nationally approved disinfectant with proven bactericidal, fungicidal, and virucidal properties.
Ultrasonic Handpiece Irrigation Bottle
15sec.
3.5L/min
38°C
Minuten Wipes (ALPRO)
45sec. Distilled water / Purified water
3sec.
Manual 15sec.
38°C
Water Connector
3.5L/min
2sec. 3times
500lx WL-Adapter 01 WL-clean (ALPRO) (ALPRO)
Water Connector
1min
Minuten Wipes (ALPRO)
WL-Adapter 01 (ALPRO)
3sec. 1time
WL-cid (ALPRO)
3sec.
Water Connector
1time WL-dry (ALPRO)
2min
3sec.
WL-Adapter 01 (ALPRO)
1time WL-Blow (ALPRO)
Automated
Glass Rod
ISO 15883-1 Minuten Wipes (ALPRO)
500lx
Adapter (ADS 2 : Miele / 73900 : Melag)
Minuten Wipes (ALPRO)
Handpiece Cord
500lx Minuten Wipes (ALPRO)
Cloth moistened with water
Powder Handpiece
Minuten Wipes (ALPRO)
45sec.
3sec.
ISO 15883-1 Minuten Wipes (ALPRO)
500lx
Adapter (ADS 2 : Miele / 73900 : Melag)
1 / week
3min 30-50°C
NOTICE
It is recommended to cover the ultrasonic bath during operation.
■Nozzle Wrench and Nozzle Tip Remover(Perio) 15sec.
38°C
3.5L/min 500lx Minuten Wipes (ALPRO)
Irrigation Bottle Irrigation Bottle
15sec.
38°C
3.5L/min
Minuten Wipes (ALPRO)
Chamber and Handpiece Hose (For Prophy/Perio use)
Chamber
500lx Cloth moistened with water
Minuten Wipes (ALPRO)
Control Unit and Foot Control
Cloth moistened with water
Minuten Wipes (ALPRO)
Tip See OM-Z0314E for instructions on reprocessing the Varios Tip.
2 Packaging, Sterilizing and Drying
Type
Z221A080
Class N
Steam sterilization
Class B
Temperature
132°C
134°C
FullCycleTime
≥15min
3-18min
ISO 11607-1
DryingTime
≥30min
≥30min
Use a sterilization pouch that conforms to ISO 11607-1. Seal the pouch. Maximum temperature for NSK instruments is 135°C.
Perform steam sterilization using the conditions above. Use a steam sterilizer to perform sterilization.
CAUTION Plasma
EO
3 Storage
OM-DE0945EN 000
2020.07.03 00001 N