1. INTRODUCTION 1.1 Safety Precautions
Please follow these safety precautions: • Verify that the AC Adapter input voltage matches the local AC power supply. • Use the balance only in dry locations. • Do not operate the balance in hostile environments. • Do not drop loads on the platform. • Do not place the balance upside down on the platform or platform mounting cone. • Service should be performed only by authorized personnel.
1.2 Controls
No
Button: O/T - On Off Yes Unit Menu No Print Cal
Functions: Short Press: Long Press: Short press (Menu): Short Press: Long Press: Short press (Menu): Long press (Menu): Short Press: Long Press:
Yes
Turns balance on, sets display to zero Turns balance off Selects or accepts setting Steps through active units and modes Enters Menu Steps through available settings Exit menu or abort out of menu item Sends data Initiates Span Calibration
2. INSTALLATION 2.1 Package Contents
0.1g and 0.01g Models Instruction Manual Power Adapter Balance Pan Pan Support Wind-Ring (InCal models only) Warranty Card
0.001g and 0.0001g Models Instruction Manual Power Adapter Balance Pan Glass Doors and Panels Warranty Card
Pioneer™
EN-2
2.2 Install Components 0.1g and 0.01g Models
1) Install Wind-Ring (InCal only)
2) Install Pan Support
3) Install Pan
0.001g and 0.0001g Models
1) Install Side Doors - Insert fully into Top Frame then down over retainer.
2) Install Panels - Insert bottom edge in groove then press until locked.
3) Install Top Door
4) Install Pan
EN-3
Pioneer™
2.3 Level Balance Level the balance on a firm, steady surface. Avoid locations with excessive air current, vibrations, heat sources or rapid temperature changes.
2.4 Connect Power
For use with CSA Certified (or equivalent approved) power source, which must have a limited circuit output.
2.5 Initial Calibration Without InCal – Power on the balance by pressing 0/T. Press and hold Print/Cal until [CAL] is displayed. The display flashes the calibration mass needed. To select the alternate calibration weight press No. Put the calibration mass on the pan. The display flashes [BUSY], then [CLEAr Pan]. Remove the mass. When calibration is complete, [DONE] is displayed. InCal – Press and hold Print/Cal until [CAL] is displayed. The display flashes [BUSY], when calibration is complete, [DONE] is displayed. Note: Calibrations should be performed after a warm up time of 60 minutes.
3. OPERATION Count, APW Optimization, Percentage or specific units of measure must be activated in the MODE or UNIT menus if they are not initially available.
3.1 Weighing Mode Repeatedly press Unit until the desired unit icon is displayed. Press 0/T to zero the balance and then place objects to be weighed on the pan.
3.2 Count Mode – Use the Count mode to count parts of uniform weight. To access Count Mode, press Unit until the display shows [Count]. Establish an Average Piece Weight (APW) – Each time a new type of part is to be counted, the nominal weight of one piece (APW) must be established using a small quantity of pieces. With [CLr.APwW] displayed, press No to use the previously saved APW, or press Yes to establish a new APW. The display indicates the number of pieces to be used to establish the new APW. If a different sample size is preferred, press No until the desired sample size is displayed (5, 10, 20, 50 or 100). Put the specified number of pieces on the pan. Press Yes to accept new APW or No to abort. Count – Place the quantity to be counted on the pan. APW Optimization – Since the weight of each piece varies slightly, APW Optimization may be used to increase the
Pioneer™
EN-4
accuracy of the count. The balance automatically recalculates the Average Piece Weight when the number of parts on the pan is less than three times the original sample size. The display shows [APwW.OPT] each time the APW is optimized.
3.3 Percent Mode – Use this mode to measure the weight of a sample as a percentage of a reference weight. To access the Percent Mode, press Unit until the display shows [PErCEnt]. Establish a new Reference Weight - With Clear reference [CLr.rEF] displayed, Press No to use the previously saved Reference Weight. Press Yes to establish a new Reference Weight. Put the reference sample on the pan and press Yes to accept or No to abort. Percent – Place the object(s) to be compared to the reference weight on the pan.
3.4 Weigh Below Feature
Remove Weigh Below Cover
Attach wire or string to Hook
Suspend sample
4. SETTINGS 4.1 Menu Navigation .END. .CAL.
.SETUP.
.MODE.
.UNIT.
.PRINT.
.RS232.
.LOC.
InCal*
AZSM
COUNT
GRAM
OUT STA
BAUD
CAL
SPAN
FILTER
MILLIGRAM
A. PRINT
PARITY
SETUP
LINEAR**
A.TARE
PERCENT
KILOGRAM
CONTENT
HAND.SH
MODE
CAL ADJ
GRAD
END.MOD
OUNCE
LINE FO
END.R232
UNIT
END.CAL
GBL RST
POUND
END.PRT
APWOPT
END.SET
* InCal models only ** Not available in InCal models
NO YES
NOT ALL UNITS SHOWN
RS232
END.UNIT
END.LOCK
Enter Menu – When the balance is on, press and hold Unit/Menu until [mMENU] appears. Release the button and the Calibrate [.CAL.] menu will display. Menu Navigation – Select menus, menu items and settings through use of the Yes and No buttons. Solid arrows point to the content displayed when Yes is pressed, Dashed lines when No is pressed. Changing Settings – To select the displayed Setting, press Yes. To move to the next Setting, press No. Exit Menu – When [.END.] is displayed, press Yes to exit the menu function, or press No to return to the Cal menu. Note: Press and hold No at any time to exit quickly.
EN-5
Pioneer™
4.2 Calibration Menu [.CAL.] InCal or Span calibration should be performed daily and when the room temperature changes. InCal [InCal] calibrates the balance using an internal mass. Span Calibration [SpaN] uses two weight values: zero and a weight between 50% and 100% of the capacity of the balance. Linearity calibration [LIN] uses three weight values zero, 50% of capacity and full capacity. Generally this calibration is not required unless testing shows that the linearity error exceeds the Linearity tolerance in the Specification table. (Not available in InCal models) Calibration Adjust [CAL Adj] allows adjustment to the result of the internal calibration by +/- 99 divisions. (InCal models only)
4.3 Setup Menu [.setup.] Automatic Zero-Setting [azsmM] – Environmental changes can cause the display to drift. The Automatic ZeroSetting Mechanism (AZSM) is designed to keep the balance set at zero, despite these slight changes. (OFF, SET.5d, SET 1d, SET 2d, SET 5d) Filter [filter] – Use the low setting (SET LO) when environmental disturbances are not present. Use the medium setting (SET MED) for normal environments. Use the high setting (SET HI) when vibrations or air currents are present. Auto Tare [A-tare] – The initial item placed on the balance is assumed to be a container so it is zeroed out. The next item is then weighed. When the pan is cleared the balance resets, waiting for a container. (SET OFF, SET ON) Grad [Grad] – Select the readability displayed. Reducing the readability may be needed for approval. (SET 1d, SET [1]d, SET 10d) Global Reset [Gbl rst] – Resets all settings to factory default values. (RESET)
4.4 Mode Menu [.mMode.] Count Mode [Count] – (SET ON or OFF) Average Piece Weight (APW) Optimization [aPwW.OPT] – (SET ON, SET OFF) Percent Mode [PerCent] – (SET ON, SET OFF)
4.5 Unit Menu [.Unit.] The Unit menu is used to enable or disable a specific unit. (SET ON, SET OFF) The unit is indicated by a small character next to Unit in the display (g = grams). The default setting is Grams SET ON and all other units SET OFF. T-Units – When unit [t] displays; press Yes to show the T-Unit settings; SET OFF, SET TT (Taiwan Tael), SET TH (Hong Kong Tael), SET TS (Singapore Tael), SET TO (Tola) or SET TI (Tical). M-Units – When unit [m] is displayed, press Yes to show the M-Unit settings; SET OFF, SET MO (Momme) or SET ME (Mesghal). Custom Unit – Custom Unit (C) is used to create a unit of measure not provided with the balance. The Custom Unit is defined by a factor, a multiplier (E) and a least significant digit (LSD). The balance will use this to convert grams to a custom unit of measure. (Example: 1 gram = 0.257206 Avoirdupois Dram, using a 4100g x 0.01g balance) To create a custom unit, press Yes when unit [c] is displayed. Factor – The Factor (F) is a value from 0.1000000 to 1.999999. When the Factor is displayed, the first digit is flashing. Press Yes to accept its value and activate the next digit, or No to edit. When editing, press No until the desired value appears, then press Yes to accept. Repeat until all digits have been accepted. When the Factor flashes on the display, press Yes to accept or No to re-edit. (Example: F = 0.257206) Multiplier (E) – The settings are, [E 0] (Fx1), [E 1] (Fx10), [E 2] (Fx100), [E 3] (Fx1000), [E-3] (F/1000), [E-2] (F/100), and [E-1] (F/10). Press No to display the next setting, Yes to accept. (Example: E = 0). Note: The multiplier selections are limited when the capacity in grams is exceeded.
Pioneer™
EN-6
LSD – The Least Significant Digit (LSD) is the number of displayed divisions (d) by which the weight is incremented. The values are 1d, 2d, 5d, 10d, 100d or 0.5d. Press No to go to the next setting, press Yes to accept. (Example LSD = 1d) Note: LSD options are limited if the readability in grams is exceeded. The example custom unit will display [0.26 C] when 1g is placed on the pan.
4.6 Print Menu [.Print.] Output Stable [Out.Stab] – Data will only be sent when the Stable indicator is on. This setting works with manual button pressing or continuous and interval Auto Print. (SET ON, OFF) Auto Print [A.Print] – Data will be continuously sent when [Continu] is set. Interval [Inter] sends data every 1 to 3600 seconds. When Stable [Stable] will send data when the balance detects a new stable reading. This can be a weight value only [LOad] or it can also include a stable zero [L+zero]. [OFF] disables automatic printing. Content [CONteNt] – The content in the data transmission can be modified. Each of the following settings can be set on or off. Number Only [numMber] will only send the numeric result. Balance ID [Bal. ID] will add the Balance serial number for traceability purposes. Reference [refer] will add reference information relevant to the current mode. GLP [GLP] will send additional items to allow proper documentation of laboratory results. Line Format [LinE Fo] – Single line format [single] will put all the data in one line separating each output with a comma (,). Multi line format [mMulti] will put each data output on a new line. Multi +4 [mM-4LF] will add 4 line spaces between each output.
^^^^^^^^^^^^^^^^^^^^^^^^^^
User ID: ………………... Bal ID: 123456789 Proj ID: ……………………… Time: ……:…….:……. Date: …../ ….. / ….. 120.01 g
Multi Line Format with 4 line feeds Line Feed-2 Line Feed-3 Line Feed-4 GLP (ON) Balance ID (ON) GLP (ON) GLP (ON) GLP (ON) Result
^^^^^^^^^^^^^^^^^^^^^^^^^^
4.7 RS232 Menu [.rs232.] Baud [BAUD] – The RS232 baud rate can be set to 600, 1200, 2400, 4800, 9600 and 19200. Parity [Parity] – Parity can be sent to 7 bits-even parity [7 eveN], 7bits-odd parity [7 odd], 7bits-no parity [7 No] or 8bit-no parity [8 No]. Handshake [Handsh] – Handshake can be set to off [Off], X on – X off [on-off], or hardware [HardwWr].
4.8 Lock Menu [.LOC.] When a Lock Menu item is SET ON the indicated menu cannot be changed. [Loc Cal] – Calibration, [Loc SEt] – Setup, [Loc.mMod] – Mode, [Loc.Unit] – Unit, [Loc Prt] – Print, [Loc 232] – RS232.
EN-7
Pioneer™
4.9 Sealing Access to the Balance Settings The Menu Lock switch prevents changes to the Lock Menu. The switch can be secured using paper seals, wire seals or plastic ties.
Un-Locked
Locked with Plastic Tie
4.10 Approved “M” Models Balances marked with an “M” have been calibrated and sealed at the factory for use as an EEC approved balance. The Cal, Setup, Mode and Unit menus will be locked.
4.11 Local Approval The InCal models can be approved by local weights and measures authorities. The calibration must be verified and the Menu settings must be set appropriately before being locked using the Lock Menu. The Menu Lock switch on the bottom of the balance can then be sealed in the locked position. The supplied capacity label is to be placed below the display.
5.0 MAINTENANCE 5.1 Troubleshooting Symptom
Possible Cause
Cannot turn on Poor accuracy
No power to balance Verify connections and voltage. Improper calibration Perform calibration Unstable environment Move balance to suitable location Unstable environment Move the balance to suitable location Incorrect calibration masses Use correct calibration masses Mode not enabled Enter menu and enable mode Units not enabled Enter menu and enable units Average Piece Weight too small Add additional samples Time out Pan has load during power on Remove weight from pan and re-zero. Pan was removed prior to power on Install pan and re-zero. Weight on pan exceeds capacity Remove weight from the pan Pan was removed during weighing Re-install pan Factory calibration data corrupted Contact the authorized dealer Factory calibration data corrupted Perform calibration EEPROM Checksum error Contact the authorized dealer Reference Weight is too small Add additional samples Reference Weight is too low for accurate Add additional samples or continue to weigh parts counting or percent weighing. with less accurate results.
Cannot calibrate Cannot access mode Cannot access unit Err 5 Err 7.0 Err 8.1 Err 8.2 Err 8.3 Err 8.4 Err 9.5 Err 9.8 Error 53 REF Err LOWREF
Remedy
5.2 Service Information If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. Please visit our web site, www.ohaus.com to locate the Ohaus office nearest you.
Pioneer™
EN-8
5.3 Accessories Security device Auxiliary Display Density Determination Kit Printer – Thermal Printer - Impact Cable Kit – Thermal Printer Cable Kit - Impact Printer Data collection Software
76288-01 PAD7 Contact Ohaus Contact Ohaus Contact Ohaus Contact Ohaus Contact Ohaus SW12W
(0.1mg and 1mg models only)
6. TECHNICAL DATA Ambient conditions – The technical data is valid under the following ambient conditions: Ambient temperature: 10° C to 30° C Relative humidity: 15 % to 80 % at 31° C non-condensing, decreasing linearly to 50% at 40° C Height above sea level: Up to 2000 m Operability is assured at ambient temperatures between 5° C and 40° C Power – AC adapter, Balance power input 8-14.5 VAC, 50/60Hz 4VA or 8-20 VDC, 4W Protection – Protected against dust and water, Pollution degree: 2, Installation category: Class II
6.1 Drawings 0.01g and 0.1g Models
0.0001g and 0.001g Models
EN-9
Pioneer™
6.2 Specifications
External Calibration Models MODEL PA64 PA114 Capacity (g) Readability (g) Repeatability (g) Linearity (g) Tare Range Stabilization Span calibration mass (g) Lin. calibration masses (g) Pan Diameter Net Weight
65
PA214
PA213
PA413 PA512 PA2102
PA4102
PA4101
110 210 0.0001 0.0001(std.dev.) ± 0.0002
210 410 510 2100 4100 4100 0.001 0.01 0.1 0.001(std.dev.) 0.01(std.dev.) 0.1sd ± 0.002 ± 0.02 ± 0.2 To capacity by subtraction 3 Seconds 50 or 100 or 100 or 200 or 200 or 1000 or 50 or 60 2000 or 4000 100 200 200 400 500 2000 20, 50, 100, 200, 200, 1000, 100, 200 2000, 4000 50 100 200 400 500 2000 3.5in / 9cm 4.7in / 12cm 7.1in / 18cm 10.1 lb / 4.6 kg 10 lb/ 4.5 kg
InCal Models (C and CM) MODEL* PA64C PA114C PA214C
PA213C PA413C PA512C PA2102C PA4102C PA4101C
Max (g) 65 110 210 210 310 510 2100 4100 4100 Min (g) 0.01 0.02 0.5 5 d= (g) 0.0001 0.001 0.01 0.1 e= (g) 0.001 0.01 0.1 Approval Class 1 Class 2 Repeatability 0.0001(std.dev.)(g) 0.001(std.dev.) 0.01(std.dev.)(g) Linearity (g) ± 0.0002 ± 0.0003 ± 0.002 ± 0.02 Tare Range To capacity by subtraction Stabilization 3 Seconds Span cal 100 or 100 or 200 or 200 or 50 or 60 50 or 100 1000 or 2000 2000 or 4000 mass (g) 200 200 400 500 Pan dia. 3.5in / 9cm 4.7in / 12cm 7.1in / 18cm Net Weight 10.1 lb / 4.6 kg 10.0 lb / 4.5 kg * Factory approved models will have an “M” added to the model number (ex. PA512CM).
6.3 Communication 6.3.1 Commands The RS232 Interface allows a computer to control the balance as well as receiving data such as the displayed weight. The balance will return “ES” for invalid commands. Command Function IP Immediate Print of displayed weight. P Print displayed weight (uses stable on/off menu setting). CP Continuous Print. xP Interval Print x = Print Interval (1-3600 sec) T Same as pressing Zero Key. ON Turns balance ON.
Pioneer™
EN-10
Communication Commands (continued) Command Function OFF Turns balance OFF. PSN Show Serial Number. PV Version: Print product name, software revision and LFT ON (if LFT is set ON). PU Print current mode/unit x# Set PC ref wt (x) in grams P# Print PC ref wt x% Set % ref wt (x) in grams P% Print % ref wt 6.3.2 RS232 (DB9) Pin Connections Pin 2: Balance transmit line (TxD) Pin 3: Balance receive line (RxD) Pin 5: Ground signal (GND) Pin 7: Clear to send (hardware handshake) (CTS) Pin 8: Request to send (hardware handshake) (RTS)
GND
RxD TxD
5
1 9
6
CTS
6.3 Compliance
RTS
Compliance to the following standards is indicated by the corresponding mark on the product. Mark
Standard This product conforms to the EMC Directive 2004/108/EC, the Low Voltage Directive 2006/95/EC and the Non-Automatic Weighing Directive 2009/23/EC. The complete declaration of Conformity is available from Ohaus Corporation. AS/NZS4251.1; AS/NZS4252.1
CAN/CSA-C22.2 No. 1010.1-92; UL Std. No. 3101-1
Important notice for verified weighing instruments Weighing Instruments verified at the place of manufacture bear one of the preceding marks on the packing label and the green ‘M’ (metrology) sticker on the descriptive data plate. They may be put into service immediately. Weighing Instruments to be verified in two stages have no green ‘M’ (metrology) on the descriptive data plate and bear one of the preceding identification marks on the packing label. The second stage of the initial verification must be carried out by an authorized and certified service organization established within the European Community or by the National Notified Body. The first stage of the initial verification has been carried out at the manufacturer’s work. It comprises all tests according to the adopted European Standard EN 45501:1992, paragraph 8.2.2. If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the re-verification period and inform the respective Weights and Measures authority.
EN-11 Disposal
Pioneer™ In conformance with the European Directive 2002/96 EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. For disposal instructions in Europe, refer to www.ohaus.com/weee. Thank you for your contribution to environmental protection.
ISO 9001 Registration In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On May 15, 2003, Ohaus Corporation, USA, was re-registered to the ISO 9001:2000 standard.
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to Ohaus. This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages. As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for further details.