ESG-400 Maintenance and On-Site Manual Issue 6 March 2018.pdf
Page 1
j
MAINTENANCE / ON-SITE - MANUAL
Electro Surgical Generator
ESG-400
7.022.211 / ISSUE 6
Page 2
ESG-400
INTRODUCTION
Introduction
The intended use depends on the approval of the country. Refer to the instructions for use of the electrosurgical
unit.
Maintenance instructions
This maintenance manual contains essential information on using and maintaining this electrosurgical
generator safely and effectively. Instructions for the operation of this electrosurgical generator and related
danger, warnings and cautions concerning electrosurgery are beyond the scope of this maintenance manual.
Before using and maintaining, thoroughly review this manual and the instructions for use or maintenance
manuals of all equipment which will be used during maintenance. Use the equipment as instructed. Keep this
manual in a safe, accessible location. If you have any questions or comments about any information in this
maintenance manual, contact Olympus.
Signal words
The following signal words are used throughout this maintenance manual:
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in
death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury. It may also be used to alert against unsafe practices
or potential equipment damage.
NOTE
Indicates additional helpful information.
d
User qualifications
The user must have received appropriate training in using this electrosurgical generator. The following
instructions are for use by qualified personnel only. Use of this maintenance manual by other individuals is
prohibited. The training will be provided by authorized representatives of Olympus during installation and
commissioning.
NOTE
The user must have received appropriate training in using, servicing,
adjustment, updating and upgrading this electrosurgical generator.
Federal Law of the USA restricts this device to use by, or on the order of, a physician.
7.022.211 / ISSUE 6
2 / 109
Introduction
Page 3
ESG-400
Precautions
High frequency leakage current or spark discharge may cause user burns.
Follow the dangers, warnings and cautions given below when handling and servicing this electrosurgical unit.
This information is to be supplemented by the dangers, warnings and cautions given in each chapter.
User-related error prevention
WARNING
Improper use
The safety and effectiveness of electrosurgical interventions depend not only on the design of
the equipment used, but also to a major extent on factors which are under the control of the
user. It is therefore extremely important to read, understand and follow the instructions supplied
with the electrosurgical generator and the accessories in order to ensure safety and
effectiveness.
Always use the electrosurgical generator as outlined in this maintenance manual. Improper use
will not only impede functions and prevent optimum performance, but may cause equipment
damage and / or complications. Before each use, always inspect the equipment as outlined in
this maintenance manual.
WARNING
Annual safety checks / Inspection
The electrosurgical generator and the footswitch must undergo a safety check in yearly intervals
in accordance with the national statutory regulations (refer to chapter 5 “Inspection”).
Environmental conditions
CAUTION
Interference of the unit with other equipment
Be sure that this electrosurgical unit is not used adjacent to or stacked with other equipment
(other than the components of this electrosurgical unit or system) to avoid electromagnetic
interference.
Before use, thoroughly confirm the compatibility of all equipment.
To ensure electrical safety, the electrosurgical unit should not be used in conjunction with:
•
Electrical equipment whose safety against leakage current is not guaranteed.
•
Electrosurgical equipment whose safety in combined use is not guaranteed.
The electrosurgical generator complies with the electromagnetic compatibility (EMC) standard.
Nevertheless, when the electrosurgical generator is active it may disturb neighboring electronic
equipment. If an auxiliary computer system is in use together with the electrosurgical generator
and endoscopic imaging techniques, the image on the monitor might freeze or blackout. Follow
the instructions in “Electromagnetic Compatibility (EMC) information” in the Appendix of the
instruction for use regarding electromagnetic ambient conditions.
Never loop the cords (active cord, bipolar cord, neutral electrode cord) or bundle cords together
with cords belonging to other medical equipment. The high frequency signals or spark discharge
noise generated by the unit may interfere with the operation of other medical equipment.
Do not use the electrosurgical unit in a location exposed to strong electromagnetic radiation
(microwave or short-wave medical treatment equipment, Magnetic Resonance Imaging, radio or
mobile phone equipment). Electrosurgical unit malfunction can occur.
CAUTION
Unsuitable temperature and humidity
The electrosurgical generator should only be used under the conditions as described in chapter
1-3 (Limitations). Use under other conditions may impede normal performance and / or result in
equipment damage.
7.022.211 / ISSUE 6
3 / 109
Introduction
Page 4
ESG-400
Accessories
WARNING
Mechanical stress
Do not apply excessive bending, straining, or squeezing force to any cords. It may cause
malfunction.
CAUTION
Non-compatible accessories and accessory damage
The electrosurgical generator shall only be used with compatible accessories. When connecting
accessories (cords, electrodes, HF instruments) avoid output settings where the maximum
output voltage of the electrosurgical generator may exceed the rated accessory voltage (refer to
“Mode characteristics”, “Output characteristics” in the Appendix of the instruction for use, and the
instruction manual of the accessory). For a list of compatible neutral electrodes, refer to
“Specifications” in the Appendix of the instruction for use.
Before use, the electrosurgical unit and accessories must be examined for damage. All
communication cables and its plugs must be free of scratches and cracks. Cables and
accessories with damaged insulation or connections must not be used.
Electric shock
WARNING
Grounding failure
To prevent the risk of electric shock, the housing of the electrosurgical unit must be grounded.
Always connect the power cord plug to a properly grounded wall outlet. Do not use a
3-pin / 2-pin adapter, as it can impair safe operation of the unit.
WARNING
User shock
To prevent user shock, malfunction and damage of the electrosurgical unit, keep liquids away
from all electrical equipment.
When taking measurements or troubleshooting of the electrosurgical unit, take appropriate
precautions, such as using isolated tools and equipment, using the “one hand rule,” etc.
CAUTION
Injury during servicing
When the housing is opened, there is a danger of electric shock. The unit must only be serviced
by authorized technicians.
Burns
WARNING
User
The maximum output voltage characteristics of the electrosurgical generator are shown in the
diagrams in “Output characteristics” in the Appendix of the instruction for use. When setting the
power level, first set it to a low level and increase it gradually. If the output is initially set to a high
level, the electrode’s insulation may be damaged and cause user and / or patient burns.
However, certain modes may present an unacceptable risk at low output power settings. For
example, with the PulseCut fast mode or PulseCut slow mode, the risk of an excessive thermal
effect rises if the output power setting is too low. Therefore, it is recommended that you perform
basic testing before using the electrosurgical generator. If the instruction manual of the HF
instrument to be used stipulate a rated voltage, the output should be set so that it does not
exceed that voltage.
High frequency, high voltage signals that can cause severe burns are present in the
monopolar / bipolar sockets described in this maintenance manual. Take appropriate precautions
when testing and troubleshooting this area of the electrosurgical unit.
7.022.211 / ISSUE 6
4 / 109
Introduction
Page 5
ESG-400
Fire / Explosion
DANGER
Ignitable anaesthetics / fire supporting gases
The risk of flammable gases or other materials being ignited exists with any contact of electrical
energy. Precautionary measures must be taken to keep flammable materials and substances
away from an active electrosurgical unit (do not use flammable anesthetics, nitrous oxide or
oxygen). Otherwise, explosion or fire may result and cause serious injuries. This electrosurgical
unit is not explosion-proof. Do not use the unit within an explosion zone.
WARNING
Ignitable cleaning- and disinfection agents
Flammable agents used for cleaning and disinfection must be allowed to evaporate before the
electrosurgical unit is used and serviced.
Non-flammable agents should be used for cleaning and disinfection wherever possible.
WARNING
Risk of fire
Disconnect the power plug before changing the fuses! Replace fuses as marked. The fuses
must only be replaced by authorized technicians.
Hazards and complications
WARNING
Contamination
The electrosurgical unit may be contaminated with infections materials; therefore, all surfaces of
the unit’s housing should be cleaned before servicing according to chapter 1-8 (Cleaning).
WARNING
Output performance
Should any abnormal output be suspected during operation, immediately terminate the use of
the equipment by releasing the footswitch. If the footswitch does not react, switch off the
electrosurgical unit. Otherwise, malfunction of the equipment may cause an unintended increase
in output.
WARNING
Service persons
Take additional precautions for service technicians, when using the unit’s service operation
mode (see chapter 15, Service operation mode).
CAUTION
Unit defect
To prevent electrosurgical unit damage, never short-circuit electrodes (accessories, neutral
electrodes).
In the event of a defect or malfunction in the unit, an undesirably high output power may be
emitted.
DANGER
Unit defect
Never use the electrosurgical unit if an abnormality is suspected.
Repair and Maintenance
CAUTION
Repair
Repairs must only be carried out by Olympus or a firm authorized by Olympus.
CAUTION
Maintenance
Preventive maintenance (inspection / periodic safety check) must only be carried out by a
qualified person / technician.
7.022.211 / ISSUE 6
5 / 109
Introduction
Page 6
ESG-400
Copyright
©2011 Olympus Winter & Ibe GmbH. All rights reserved.
Unauthorized reproduction or distribution in part or in whole is prohibited.
Trademarks
OLYMPUS is a registered trademark of the Olympus Corporation.
The company names, product names, and proprietary technical terms in this document are the trademarks or
registered trademarks of their respective owners.
7.022.211 / ISSUE 6
6 / 109
Introduction
Page 7
ESG-400
CONTENT
INTRODUCTION ................................................................................................................ 2
CONTENT .......................................................................................................................... 7
CHAPTER 1: PRODUCT SPECIFICATIONS .................................................................... 9
1 OUTLINE ...................................................................................................................... 10
1-1
1-2
1-3
Intended Use .................................................................................................................................... 10
Compatibility..................................................................................................................................... 10
Expected service life ........................................................................................................................ 10
2 FEATURES ................................................................................................................... 11
2-1
2-2
Application Modes ............................................................................................................................ 11
Accessories ...................................................................................................................................... 12
3 LIMITATIONS ................................................................................................................ 13
4 SPECIFICATIONS ........................................................................................................ 14
4-1
4-2
4-3
4-4
4-5
4-6
4-7
4-8
ELECTROSURGICAL GENERATOR ESG-400 (REF: WB91051W) .............................................. 14
Power cords (4.5 m angled plug) ..................................................................................................... 15
Footswitch (REF: WB50402W, double pedal) ................................................................................. 16
Footswitch (REF: WB50403W, single pedal, optional) .................................................................... 16
Neutral electrode cable “P-cord” (REF: MAJ-814, optional) ............................................................ 16
Communication cable 0.25 m (REF: MAJ-1871, optional) .............................................................. 17
Communication cable 10 m (REF: MAJ-1872, optional) ................................................................. 17
Adapter for UHI-2/3 (REF: MAJ-1873, optional) .............................................................................. 17
5 NAME AND FUNCTION OF EACH PART .................................................................... 18
5-1 Symbols and descriptions ................................................................................................................ 18
5-1-1
Safety related symbols ............................................................................................................ 18
5-1-2
Front panel .............................................................................................................................. 19
5-1-3
Touch screen ........................................................................................................................... 19
5-1-4
Rear panel ............................................................................................................................... 21
5-2 Front panel ....................................................................................................................................... 22
5-3 Rear panel ........................................................................................................................................ 24
5-4 Bottom panel .................................................................................................................................... 25
5-5 All screen ......................................................................................................................................... 25
5-6 Set screen ........................................................................................................................................ 26
5-7 Mode screen .................................................................................................................................... 27
5-8 Footswitch with two pedals .............................................................................................................. 28
5-9 Footswitch with one pedal (optional)................................................................................................ 28
5-10 Neutral electrode cable “P-cord” (optional) ...................................................................................... 29
6 CONNECTOR ............................................................................................................... 30
6-1
6-2
6-3
6-4
6-5
6-6
6-7
Docking Connector .......................................................................................................................... 30
Monopolar Standard 1 ...................................................................................................................... 30
Monoploar Standard 2 (Erbe) ........................................................................................................... 30
Bipolar Standard 3............................................................................................................................ 30
Monopolar Neutral Electrode ........................................................................................................... 31
Foot switch 1 (SIP/SOP) .................................................................................................................. 31
Foot switch 2 (SIP/SOP) .................................................................................................................. 32
7 SYSTEM DIAGRAM ..................................................................................................... 33
8 CLEANING, STORAGE AND DISPOSAL .................................................................... 34
8-1
8-2
8-3
Cleaning ........................................................................................................................................... 34
Storage ............................................................................................................................................. 35
Disposal of the unit .......................................................................................................................... 35
CHAPTER 2: BLOCK DESCRIPTION............................................................................. 37
1 BLOCK DESCRIPTIONS .............................................................................................. 38
1-1
1-2
1-3
1-4
1-5
Motherboard ..................................................................................................................................... 39
HVPS Board ..................................................................................................................................... 41
Generator board ............................................................................................................................... 41
Relay Board ..................................................................................................................................... 41
Front Panel....................................................................................................................................... 42
CHAPTER 3: REPAIR SYSTEM ...................................................................................... 43
7.022.211 / ISSUE 6
7 / 109
Content
Page 8
ESG-400
1 ESG-400 MAIN UNIT .................................................................................................... 44
2 BOARD COMPATIBILITY ............................................................................................. 44
3 OPTIONAL ACCESSORIES ......................................................................................... 44
3-1
3-2
3-3
WB50402W (Footswitch with two pedals) ....................................................................................... 44
WB50403W (Footswitch with one pedal) ......................................................................................... 44
MAJ-814 (Neutral electrode cable “P-cord”) .................................................................................... 44
4 PRECAUTIONS ON FUNCTION AND OPERATION SETTINGS ................................. 44
4-1
General Precautions ........................................................................................................................ 44
CHAPTER 4: TROUBLESHOOTING .............................................................................. 45
1 GENERAL ..................................................................................................................... 46
2 NEUTRAL ELECTRODE OPERATION ........................................................................ 47
3 ERROR SCREEN, CODES AND MEASURES ............................................................. 48
3-1
3-2
What to do when no error code is displayed .................................................................................... 50
What to do when an error code is displayed .................................................................................... 54
CHAPTER 5: INSPECTION ............................................................................................. 73
1 JIGS, TOOLS, AND MEASURING EQUIPMENT FOR INSPECTION .......................... 74
2 INSPECTION PROCEDURES ...................................................................................... 75
2-1 Visual inspection of the electrosurgical generator and accessories ................................................ 76
2-2 Verifying the contact quality monitor function .................................................................................. 80
2-3 Checking the DC resistance (according to IEC 60601-2-2) ............................................................. 82
2-4 Checking the earth resistance (according to IEC 60601-1 and IEC 62353) .................................... 82
2-5 Checking the earth leakage current (according to IEC 60601-1) .................................................... 82
2-6 Checking the patient leakage current (according to IEC 60601-1) .................................................. 83
2-7 Checking the current and power consumption and output waveform.............................................. 84
2-8 Checking the high frequency leakage current (according to IEC 60601-2-2) .................................. 86
2-8-1
Measurement of the monopolar high frequency leakage current under loaded condition ...... 86
2-8-2
Measurement of the monopolar high frequency leakage current under unloaded condition .. 88
2-8-3
Measurement of the bipolar high frequency leakage current under loaded condition ............ 90
2-8-4
Measurement of the bipolar high frequency leakage current under unloaded condition ........ 92
2-9 Checking the output power .............................................................................................................. 94
2-10 Checking for certain features and error messages .......................................................................... 94
2-11 Final test ........................................................................................................................................... 95
2-11-1
Self test.................................................................................................................................... 95
2-11-2
Display and sound check ........................................................................................................ 95
2-11-3
Functionality of push buttons .................................................................................................. 95
2-11-4
Communication test................................................................................................................. 96
2-11-5
Restore of output power settings ............................................................................................ 96
2-12 Inspection label (For applicable markets) ........................................................................................ 96
3 INSPECTION CARD ..................................................................................................... 97
CHAPTER 6: DEVICE MENU ........................................................................................ 102
1 SAFETY TEST ............................................................................................................ 103
2 SOFTWARE VERSION ............................................................................................... 104
CHAPTER 7: REVISION HISTORY ............................................................................... 106
3 REVISION HISTORY .................................................................................................. 107
7.022.211 / ISSUE 6
8 / 109
Content
Page 9
ESG-400
CHAPTER 1: PRODUCT SPECIFICATIONS
1 OUTLINE ...................................................................................................................... 10
1-1
1-2
1-3
Intended Use .................................................................................................................................... 10
Compatibility..................................................................................................................................... 10
Expected service life ........................................................................................................................ 10
2 FEATURES ................................................................................................................... 11
2-1
2-2
Application Modes ............................................................................................................................ 11
Accessories ...................................................................................................................................... 12
3 LIMITATIONS ................................................................................................................ 13
4 SPECIFICATIONS ........................................................................................................ 14
4-1
4-2
4-3
4-4
4-5
4-6
4-7
4-8
ELECTROSURGICAL GENERATOR ESG-400 (REF: WB91051W) .............................................. 14
Power cords (4.5 m angled plug) ..................................................................................................... 15
Footswitch (REF: WB50402W, double pedal) ................................................................................. 16
Footswitch (REF: WB50403W, single pedal, optional) .................................................................... 16
Neutral electrode cable “P-cord” (REF: MAJ-814, optional) ............................................................ 16
Communication cable 0.25 m (REF: MAJ-1871, optional) .............................................................. 17
Communication cable 10 m (REF: MAJ-1872, optional) ................................................................. 17
Adapter for UHI-2/3 (REF: MAJ-1873, optional) .............................................................................. 17
5 NAME AND FUNCTION OF EACH PART .................................................................... 18
5-1 Symbols and descriptions ................................................................................................................ 18
5-1-1
Safety related symbols ............................................................................................................ 18
5-1-2
Front panel .............................................................................................................................. 19
5-1-3
Touch screen ........................................................................................................................... 19
5-1-4
Rear panel ............................................................................................................................... 21
5-2 Front panel ....................................................................................................................................... 22
5-3 Rear panel ........................................................................................................................................ 24
5-4 Bottom panel .................................................................................................................................... 25
5-5 All screen ......................................................................................................................................... 25
5-6 Set screen ........................................................................................................................................ 26
5-7 Mode screen .................................................................................................................................... 27
5-8 Footswitch with two pedals .............................................................................................................. 28
5-9 Footswitch with one pedal (optional)................................................................................................ 28
5-10 Neutral electrode cable “P-cord” (optional) ...................................................................................... 29
6 CONNECTOR ............................................................................................................... 30
6-1
6-2
6-3
6-4
6-5
6-6
6-7
Docking Connector .......................................................................................................................... 30
Monopolar Standard 1 ...................................................................................................................... 30
Monoploar Standard 2 (Erbe) ........................................................................................................... 30
Bipolar Standard 3............................................................................................................................ 30
Monopolar Neutral Electrode ........................................................................................................... 31
Foot switch 1 (SIP/SOP) .................................................................................................................. 31
Foot switch 2 (SIP/SOP) .................................................................................................................. 32
7 SYSTEM DIAGRAM ..................................................................................................... 33
8 CLEANING, STORAGE AND DISPOSAL .................................................................... 34
8-1
8-2
8-3
Cleaning ........................................................................................................................................... 34
Storage ............................................................................................................................................. 35
Disposal of the unit .......................................................................................................................... 35
7.022.211 / ISSUE 6
9 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 10
ESG-400
1 Outline
1-1
Intended Use
The intended use depends on the approval of the country. Refer to the instructions for use of the electrosurgical
generator.
1-2
Compatibility
This product can be used in combination with the products listed in compatibility table.
1-3
Expected service life
The expected service life is 10 years.
7.022.211 / ISSUE 6
10 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 11
ESG-400
2 Features
The ESG-400 is a reusable, non-sterile electrosurgical generator with different mono- and bipolar cutting and
coagulation modes. The maximum output power is 320 W.
On the front side it features a touch screen display that displays the connection status of accessories and
peripherals connected to the electrosurgical generator. It is also used to show and modify the output settings
(e.g. mode, output power, effect) as well as to control other functions (e.g. save settings).
In addition the ESG-400 has a bipolar socket, two monopolar sockets, a neutral electrode socket, and a
universal socket to connect applicators with instrument recognition. The power switch turns the generator on
and off.
Two contact quality indicators (one for split and one for non-split electrodes) are green illuminated if neutral
electrodes are correctly connected. Three additional push buttons allow recalling a previously saved setting
(Select Procedure), to assign the footswitches to specific output sockets (Footswitch), and to control several
other functions (Menu), e.g. select language, touch tone control, output volume, or brightness.
On the rear panel the volume control, a ventilation hole, the equipotential bonding point, the AC power socket,
and two footswitch sockets can be found. Furthermore, for the connection of peripheral equipment 26-pin plugs
respectively 14-pin plugs can be connected to the LINK-IN or to the LINK-OUT socket.
On the bottom panel, a docking socket is featured. It can be used to connect the ESG-400 directly to the
USG-400 and upcoming devices. The ESG-400 is compatible with the new USG-400 ultrasonic generator to
enable the use of combined (US + HF) instruments.
2-1
Application Modes
Monopolar Cut:
•
PureCut (Cutting of varying tissue structures; 3 Effects)
•
BlendCut (Cutting of varying tissue structures; 3 Effects)
•
PulseCut slow (Intermittent cutting; 3 Effects)
•
PulseCut fast (Intermittent cutting; 3 Effects)
Monopolar Coagulation:
•
SoftCoag (Coagulation of tissue with little sticking and carbonization; 3 Effects)
•
ForcedCoag (Fast and effective coagulation; 3 Effects)
•
SprayCoag (Contact-free surface coagulation with little penetration depth; 3 Effects)
•
PowerCoag (Fast and effective coagulation with increased dissection capabilities; 3 Effects)
7.022.211 / ISSUE 6
11 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 12
ESG-400
Bipolar Cut:
•
BipolarCut (All bipolar cutting procedures of tissue structures; 3 Effects)
•
SalineCut (Cutting in conductive fluid; 3 Effects; only available via UNIVERSAL socket)
•
PK PureCut (Cutting of varying tissue structures; 3 Effects; only available via UNIVERSAL socket)
•
PK SoftCut (Cutting of varying tissue structures; 3 Effects; only available via UNIVERSAL socket)
•
PK LoopCut (Cutting of varying tissue structures, especially fibroid tissue; 3 Effects; only available via
UNIVERSAL socket)
•
PK MorceCut (Cutting of varying tissue structures, especially fibroid tissue; 3 Effects; only available
via UNIVERSAL socket)
Bipolar Coagulation:
•
BiSoftCoag
•
AutoCoag (Coagulation with little sticking and carbonization; 3 Effects)
•
SalineCoag (Coagulation in conductive fluid; 3 Effects; only available via UNIVERSAL socket )
•
HardCoag (Controlled tissue coagulation; 3 Effects)
•
RFCoag (Controlled deep tissue coagulation; with and without RCAP)
•
FineCoag (Coagulation of tissue with little sticking and carbonization; 1 Effect)
•
PK Coag (Coagulation of tissue with little sticking and carbonization; 3 Effects)
•
PK SoftCoag (Coagulation of tissue with little sticking and carbonization; 3 Effects)
•
PK AutoCoag (Controlled tissue coagulation; 1 Effect)
(Coagulation with little sticking and carbonization; 3 Effects)
The modes have preset power levels that may be customized by the user in a defined range.
2-2
Accessories
Footswitch Double Pedal (WB50402W): It has a blue pedal that is used to activate the selected coagulation
mode and a yellow pedal that is used to activate the selected cutting mode.
Footswitch Single Pedal (optional; WB50403W): It has a blue pedal that is used to activate the selected
coagulation mode
P-Cord (optional; MAJ-814): The P-cord is used to connect a patient plate to the ESG-400.
7.022.211 / ISSUE 6
12 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 13
ESG-400
3 Limitations
(1) Use this product under the supervision of a doctor at a medical facility.
(2) Do not use this product in combination with the products other than those designated by Olympus.
(3) This product shoul be used, transported or stored in the following environment.
Operation
environment
Transportation and
storage
environment
7.022.211 / ISSUE 6
Temperature
+ 10…+ 40°C
Relative humidity
30…85%, non-condensing
Atmospheric pressure
70…106 kPa
Temperature
- 25…+ 60°C
Relative humidity
10…85%, non-condensing
Atmospheric pressure
50…106 kPa
13 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 14
ESG-400
4 Specifications
4-1
ELECTROSURGICAL GENERATOR ESG-400 (REF: WB91051W)
Power supply
Voltage range
100…120 V~ / 220…240 V~
Frequency
50 / 60 Hz
Maximum input power
1500 VA
Power fuse
10 A (only FST-series from
Schurter)
Power connection line
IEC 60320-1 / C13
Maximum length: 4.5 m
Terminal
for
equalization
Size, weight
packaging
Classification
potential Yes
and Width x Depth x Height
370 × 465 × 156 mm
Volume
25752 cm³
Weight of generator
12.5 kg
Weight of packaging
2.3 kg
Type of packaging
Cardboard
and
expanded
polypropylene material
Protection class according CF, Class I
to IEC 60601-1
Classification according to IIb
MDD 93/42/EEC
Output
High frequency functions
Monopolar / Bipolar
High frequency
430 kHz ±20%
Maximum high frequency
320 W
power
All modes
25% duty cycle
(e.g. 10 s activated / 30
deactivated)
s
RFCoag (with or without 100% duty cycle
RCAP)
7.022.211 / ISSUE 6
14 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 15
ESG-400
Sockets
MONOPOLAR 1
3-pin (∅ 4 mm),
Valleylab standard;
coaxial (∅ 8 mm),
Bovie standard
MONOPOLAR 2
3-pin (∅ 4 mm),
Valleylab standard;
coaxial (∅ 5 / 9 mm),
Erbe standard
BIPOLAR
2-pin (∅ 4 mm,
pin spacing 28.8 mm),
Valleylab standard;
coaxial
(∅inner 8 mm, ∅outer 4 mm),
Erbe standard
UNIVERSAL
7-pin, Olympus standard
Neutral electrode
Single or split, 10 mm plug
Contact
quality Allowable resistance range 10...155 Ω ±15 Ω
monitor (CQM)
for
split
type
neutral
electrodes
Allowable resistance range < 10 Ω ±5 Ω
for non-split type neutral
electrodes
4-2
Power cords (4.5 m angled plug)
Power cords
WA95621A
Many European countries
Type E/F
WA95622A
USA, Canada
countries
and
other
Type B
WA95623A
United Kingdom
countries
and
other
Type G
7.022.211 / ISSUE 6
15 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 16
ESG-400
4-3
Footswitch (REF: WB50402W, double pedal)
Classification
Size, weight
packaging
4-4
and Width x Depth x Height
350 × 185 × 65 mm
Weight of footswitch
1.9 kg
Length of cord
4m
Weight of packaging
0.5 kg
Type of packaging
Cardboard material
Footswitch (REF: WB50403W, single pedal, optional)
Classification
Size, weight
packaging
4-5
Protection class according IPX8 (except the plug section)
to IEC 60529
Protection class according IPX8 (except the plug section)
to IEC 60529
and Width x Depth x Height
175 × 185 × 50 mm
Weight of footswitch
1.6 kg
Length of cord
4m
Weight of packaging
0.5 kg
Type of packaging
Cardboard material
Neutral electrode cable “P-cord” (REF: MAJ-814, optional)
Size
7.022.211 / ISSUE 6
Weight
0.14 kg
Length of cord
3.1 m
16 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 17
ESG-400
4-6
Communication cable 0.25 m (REF: MAJ-1871, optional)
Size
4-7
0.05 kg
Length of cord
0.25 m
Communication cable 10 m (REF: MAJ-1872, optional)
Size
4-8
Weight
Weight
0.5 kg
Length of cord
10 m
Adapter for UHI-2/3 (REF: MAJ-1873, optional)
Size
Width x Depth x Height
100 × 77 × 42 mm
Weight
0.35 kg
Compatible cables
7.022.211 / ISSUE 6
MAJ-1871, MAJ-1872
17 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 18
ESG-400
5 Name and Function of each part
5-1
5-1-1
Symbols and descriptions
Safety related symbols
Output insulated from earth
(connection for neutral electrode)
Caution, read
instructions
Refer to
instructions
Defibrillation proof type CF applied
part (cardiac application)
Type plate
Date of
manufacture
CE marking
Non-ionizing electromagnetic radiation
Waste electrical and electronic
equipment
cTUVus marking
Manufacturer
Fuse rating
Potential equalization terminal
7.022.211 / ISSUE 6
18 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 19
ESG-400
5-1-2
Front panel
Power on / off
Neutral electrode – non-split type
Neutral electrode – split type
Select procedure
Footswitch
Menu
BIPOLAR socket
MONOPOLAR 1 socket
UNIVERSAL socket
MONOPOLAR 2 socket
5-1-3
Touch screen
Double footswitch
Single footswitch
Autostart
Plus
Minus
Return
OK
Cancel
7.022.211 / ISSUE 6
19 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 20
ESG-400
Save procedure
Delete procedure
Languages
Touch tone on
Touch tone off
Software version
Safety test
Service
Volume
Brightness
Select procedure (in title line)
Menu (in title line)
Toggle
Previous
Next
Numeric
Alphabetic
Uppercase / lowercase
Backspace
Caution
Communication indicator
7.022.211 / ISSUE 6
20 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 21
ESG-400
Reset
RCAP
Resistance Controlled Automatic Power
Reference to BIPOLAR socket
Reference to MONOPOLAR 1 socket
Reference to UNIVERSAL socket
Reference to MONOPOLAR 2 socket
5-1-4
Rear panel
Volume
Footswitch
LINK-IN
LINK-OUT
7.022.211 / ISSUE 6
LINK-IN socket
LINK-OUT socket
21 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 22
ESG-400
5-2
Front panel
1.
2.
12.
3.
11.
4.
10.
6.
5.
7.
8.
9.
1.
Power switch
This switch turns the electrosurgical generator on and off.
2.
BIPOLAR socket
This socket connects the plug of a bipolar HF instrument (applied part).
3.
MONOPOLAR 2 socket
This socket connects the plug of a monopolar HF instrument (applied part).
4.
MONOPOLAR 1 socket
This socket connects the plug of a monopolar HF instrument (applied part).
5.
Touch-screen
Displays the connection status of the accessories and peripherals connected to the electrosurgical
generator. It is also used to show and modify the output settings (e.g. mode, output power, effect) as
well as to control other functions (e.g. save procedures, delete procedures).
6.
FOOTSWITCH push button
This button is used to open the “Footswitch screen” to assign one or two footswitch(es) or the autostart
function to a specific output socket.
7.
SELECT PROCEDURE push button
This button is used to open the “Select Procedure screen” to recall saved settings.
8.
MENU push button
This button is used to open the “Menu screen” to control several functions (save or delete a procedure,
control the touch tone, output volume and brightness as well as other functions).
7.022.211 / ISSUE 6
22 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 23
ESG-400
9.
Contact quality monitor indicator for split neutral electrode
This indicator illuminates green if a split neutral electrode is connected and the contact resistance is
within an acceptable range. The indicator illuminates red if the split neutral electrode is not connected
or not applied properly (e.g. bad contact quality or partly dislocated) or no neutral electrode is
connected (in both cases the activation of monopolar output is disabled).
10.
Contact quality monitor indicator for non-split neutral electrode
This indicator illuminates green if a non-split neutral electrode is connected.
11.
Neutral electrode socket
This socket connects the plug of a neutral electrode for monopolar application (applied part).
12.
UNIVERSAL socket
This socket connects the plug of an Olympus HF instrument with HF instrument recognition (applied
part).
7.022.211 / ISSUE 6
23 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 24
ESG-400
5-3
Rear panel
1.
2.
7.
3.
1.
5.
6.
4.
1.
Footswitch sockets
This socket connects the plug of a single or double pedal footswitch.
2.
Volume control
This knob is used for adjusting the output volume.
3.
Ventilation hole
Holes for air ventilation via a cooling fan; there are also ventilation holes on each side of the
electrosurgical generator.
4.
Equipotential bonding point
To increase electrical safety, this point is used for potential equalization. All equipment housings that
come into contact with the patient are electrically connected in order to prevent low-frequency
electrical currents from endangering the patient in the event of a defect in the conventional protective
conductor system.
5.
AC power socket
This socket serves as a connection to the mains power supply via a power cord
6.
LINK-OUT socket
This socket connects the plug (14-pin) of a cable connected to peripheral equipment.
7.
LINK-IN socket
This socket connects the plug (26-pin) of a cable connected to peripheral equipment.
NOTE
7.022.211 / ISSUE 6
The touch-screen messages may depend on the
language setting of the electrosurgical generator.
For a detailed explanation of the different types of
sockets, refer to chapter 6 “Connection of neutral
electrode” and chapter 3.7 “Connection of HF
instruments”.
24 / 109
Chapter 1: PRODUCT SPECIFICATIONs
Page 25
ESG-400
5-4
Bottom panel
1.
1.
5-5
Docking socket
This socket connects the plug (7-pin) of a docking connector to connect peripheral equipment. For
more details, see chapter 1-6-1.
All screen
1.
2.
3.
4.
5.
6.
7.
8.
9.
12.
11.
10.
1.
Reference to output sockets indicator
This indicator shows the corresponding output socket where the same symbol is printed on the front
panel.
2.
Output socket name
The name of the corresponding output socket is displayed here.
3.
Autostart indicator
This symbol indicates if the autostart function is assigned to the corresponding output socket. Blank if
7.022.211 / ISSUE 6
25 / 109
Chapter 1: PRODUCT SPECIFICATIONs