ucr_endoscopic_co2_regulation_unit_maintenance_manu.pdf
Page 1
MAINTENANCE MANUAL
ENDOSCOPIC CO2 REGULATION UNIT
UCR
ISSUE3
2012/03
144P-1843
Page 3
UCR
CONTENTS
Chapter 1: Introduction ............................................................. 1-1
1 Introduction ............................................................................................................ 1-1
2 Applicable Unit ....................................................................................................... 1-1
3 Copyright ................................................................................................................ 1-1
4 Trademarks ............................................................................................................. 1-1
Chapter 2: Product Outline ....................................................... 2-1
1 Product outline ....................................................................................................... 2-1
2 Characteristics ....................................................................................................... 2-1
3 Restrictions ............................................................................................................ 2-2
Chapter 3: Specifications .......................................................... 3-1
Chapter 4: Installation and Connection ................................... 4-1
Chapter 5: Care, Storage, and Disposal................................... 5-1
Chapter 6: Inspection ................................................................ 6-1
1 Inspection Check Sheet ......................................................................................... 6-1
Chapter 7: Safety Check............................................................ 7-1
1 Exterior leakage current ........................................................................................ 7-1
2 Ground leakage current ......................................................................................... 7-1
3 Ground resistance ................................................................................................. 7-2
4 Safety Check Sheet ............................................................................................... 7-3
Chapter 8: Block Diagram ......................................................... 8-1
Chapter 9: Troubleshooting ...................................................... 9-1
ISSUE3
1
CONTENTS
Page 5
UCR
Chapter 1: Introduction
1. Introduction
Repair and maintenance of this product requires highly specialized knowledge and techniques.
We recommend that you contact an Olympus service center in your area if a problem develops with the
product.
If repairs or modifications are made by personnel not authorized by Olympus, the warranty is void, and
Olympus shall not be liable for damage that occurs to or as a result of use of the modified product.
The information contained in this manual is subject to change at any time without notice. Later editions of
this manual will contain revised and updated material.
2. Applicable Unit
• ENDOSCOPIC CO2 REGULATION UNIT UCR AC100-120V with English
• ENDOSCOPIC CO2 REGULATION UNIT UCR AC100-240V with English
• ENDOSCOPIC CO2 REGULATION UNIT UCR AC100-240V with Symbol
3. Copyright
© 2007 Olympus Medical Systems Corp. All rights reserved.
Unauthorized reproduction or distribution in part or in whole is prohibited.
4. Trademarks
OLYMPUS is a registered trademark of Olympus Corporation.
The company names, product names, and proprietary technical terms in this document are the trademarks or registered trademarks of their respective owners.
ISSUE3
1-1
Introduction
Page 7
UCR
Chapter 2: Product Outline
1. Product outline
(1)
Intended use
The endoscopic CO2 regulation unit has been designed to be used with Olympus
endoscopes and ancillary equipment for CO2 gas and water feeding.
(2)
Coverage of combination
The product can be combined with wide use video scope for upper digestive tract,
duodenal videoscope, small intestine videoscope, large intestine videoscope, and sigmoid
videoscope from, 150 series, EVIS EXERA 145/160 series, EVIS EXERA Ⅱ 165/180
series, EVIS 240 series, PCF-S, and (scopes in this document refer to EVIS 140 series EVIS LUCERA 260 series ).
2. Characteristics
(1)
The product can be combined with the scopes listed in the coverage of combinations and
perform CO2 gas supply instead of air supply.
(2)
The CO2 gas supply volume is equivalent to the air supply volume of conventional light
source.
(3)
Preparation is easy as the main body and CO2 cylinder can be connected with a specific
high pressure hose.
ISSUE3
(4)
Gas supply start/stop can be operated through a single button.
(5)
Residual pressure of the gas cylinder is displayed in the supply pressure display bar
graph. Flow display is also available while CO2 gas is output from the main body.
(6)
Timer function is also available which buzzes and automatically switches off after a
specified period of time.
(7)
Flow can be adjusted by combined the product with the optional low flow gas tube (MAJ1742, MAJ-1816).
(8)
The product is small sized, compact, and can be mounted on our mobile workstation NP1
and WP1.
2-1
Product Outline
Page 8
UCR
3. Restrictions
(1)
The product shall be used in medical institutions under supervision of a medical doctor.
(2)
The outer case needs grounding for safety purpose.
(3)
Use is prohibited in inflammable atmosphere.
(4)
Power supply
(5)
a)
Voltage
AC100 - 240V
b)
Power input
40VA
c)
Frequency
50/60Hz
d)
Voltage fluctuation
within ± 10%
Environment
a)
Operation environment
Ambient temperature
10 - 40 ℃
Relative humidity
30 - 85%
Atmospheric pressure 700 - 1060hPa
b)
Transport environment
Ambient temperature
-25 - 70 ℃
Relative humidity
10 - 90%
(6)
Applicable to CO2 for medical use.
(7)
Pipeline of gas for medical use
a)
Supply pressure
Upper limit 1400kPa Operation environment
Pressure provided in ISO 7396-1 (Pipeline facility of gas for medical use): 2002
Lower limit 342.3kPa
b)
Flow
40L/min or more (0 ℃ , 101.3kPa equivalent)
Product Outline
2-2
ISSUE3
Page 9
UCR
Chapter 3: Specifications
Refer to the “Specifications“ of Instruction Manual.
ISSUE3
3-1
Specifications
Page 11
UCR
Chapter 4: Installation and Connection
Refer to the “Installation and Connection“ of Instruction Manual.
ISSUE3
4-1
Installation and Connection
Page 13
UCR
Chapter 5: Care, Storage and Disposal
Refer to the “Care, Storage and Disposal“ of Instruction Manual.
ISSUE3
5-1
Care, Storage and Disposal
Page 15
UCR
Chapter 6: Inspection
Refer to the “Inspection“ of Instruction Manual.
1. Inspection Check Sheet
Please copy and use check sheet on next page.
A check item may change with the destination and specifications.
There may be a check item which is on this list and is not in the instruction manual.
Please refer to the instruction manual(Chapter of Inspection or Inspection Before Use) for an
actual check item and the check procedure.
There may be a blank column in a Check Sheet. Please utilize this blank column, if needed.
ISSUE3
6-1
Inspection
Page 16
Inspection
6-2
the power supply
the gas supply
insufflation
gas feeding at gas tube
gas and water feeding
Inspector
Inspection of
Inspection of
Inspection of
Inspection of
Inspection of
Power OFF
the power supply
the gas supply
insufflation
gas feeding at gas tube
gas and water feeding
Inspection of the gas tube (MAJ-1741), low flow gas tube
(MAJ-1742), and extra low flow gas tube (MAJ-1816)
Inspection of the cylinder hose (MAJ-1080, MAJ-1081, MAJ-1082)
Inspection of the medical gas pipeline adapter (MAJ-1084,
MAJ-1085)
Day
Inspector
Inspection of
Inspection of
Inspection of
Inspection of
Inspection of
Power OFF
Inspection of the gas tube (MAJ-1741), low flow gas tube
(MAJ-1742), and extra low flow gas tube (MAJ-1816)
Inspection of the cylinder hose (MAJ-1080, MAJ-1081, MAJ-1082)
Inspection of the medical gas pipeline adapter (MAJ-1084,
MAJ-1085)
Day
Serial number:
UCR INSPECTION CHECK SHEET
16
1
17
2
18
3
19
4
20
5
21
6
22
7
23
8
24
9
25
10
26
11
27
12
28
13
29
14
Year & Month:
30
15
/
31
.
UCR
ISSUE2
Page 17
UCR
Chapter 7: Safety Check
1. Exterior leakage current
Standard
The exterior leakage current must not exceed the values shown below.
Item
Exterior leakage current
Standard
Normal condition
Single Fault
100μA
500μA
(The UCR with CSA symbol on rear panel : 300μA )
Inspection procedure
Measure the leakage current at the potential equlization terminal.
For more information of the mesurement procedure, refer to the Manual for the leakage
current tester.
Potential equalization terminal
2. Ground leakage current
Standard
The ground leakage current must not exceed the values shown below.
item
Ground leakage current
ISSUE3
Standard
Normal condition
Single Fault
500μA
7-1
1000μA
Safety Check
Page 18
UCR
Inspection procedure
Measure the leakage current at the ground terminal of the AC inlet.
For more information of the mesurement procedure, refer to the Manual for the leakage
current tester.
3. Ground resistance
Standard
The ground groud resistance must not exceed 0.1 [ohms]
Inspection procedure
Apply 25A AC at 50 Hz or 60 Hz (under no load: do not exceed 6V) for 5–10 seconds
between the ground terminal of the AC inlet and the screw farthest from the AC inlet.
And then measure the resistance.
For more information of the mesurement procedure, refer to the Manual for the leakage
current tester.
Screw
Safety Check
7-2
ISSUE3
Page 19
UCR
4. Safety Check Sheet
Date of inspection
Product
Serial number
Inspected by
Checked by
Electrical Safety Test
Item
Standard
Normal Condition
7.1
Exterior leakage current
7.2
Ground leakage current
7.3
Ground Resistance
ISSUE3
Single Fault
Normal Condition
Single Fault
-
7-3
Result
100 μA
500 μA
(UCR with CSA symbol : 300μA )
500 μA
1000 μA
0.1Ω
Safety Check
Page 21
ISSUE3
8-1
㪥㫆㫀㫊㪼㩷㪽㫀㫃㫋㪼㫉
㪫㫀㫄㪼㫉㩷㫊㫎㫀㫋㪺㪿
㪪㫐㫊㫋㪼㫄㩷㪺㫆㫅㫅㪼㪺㫋㫆㫉
㪞㪸㫊㩷㫊㫌㫇㫇㫃㫐㩷㫊㫆㫌㫉㪺㪼
㫊㪼㫃㪼㪺㫋㫀㫆㫅㩷㫊㫎㫀㫋㪺㪿
㪚㪦㪉㩷㪞㪘㪪㩷㪠㪥㪣㪜㪫
㪘㪚㩷㪈㪇㪇㪄㪉㪋㪇㪭
㪧㫆㫎㪼㫉㩷㫊㫎㫀㫋㪺㪿
㪫㪦㪬㪚㪟㪶㪧㪘㪥㪜㪣
㪣㪠㪞㪟㪫㪶㪪㪰㪪㪫㪜㪤
㩷㪠㫅㫇㫌㫋
㪻㪼㫋㪼㪺㫋㫀㫆㫅
㪺㫀㫉㪺㫌㫀㫋
㪞㪘㪪㪪㪶㪣㪜㪪㪪
㪭㪘㪣㪭㪜㪶㪦㪥
㪭㪘㪣㪭㪜㪶㪦㪥
㪝㪣㪦㪮㪶㪘㪣㪘㪩㪤
㪝㪣㪦㪮㪶㪣㪜㪪㪪
㪭㪘㪣㪭㪜㪶㪦㪥
㪭㪘㪣㪭㪜㪶㪦㪥
㪫㫀㫄㪼㫉㩷㫊㫀㪾㫅㪸㫃
㫊㪼㫃㪼㪺㫋㫀㫆㫅
㪺㫀㫉㪺㫌㫀㫋
㪫㫀㫄㪼㫉㩷㫊㫀㪾㫅㪸㫃
㪚㫆㫅㫍㪼㫉㫋㪼㫉
㪫㪄㪝㪝㩷㫉㪼㫊㪼㫋㩷㫊㫀㪾㫅㪸㫃
㪠㫅㫇㫌㫋㩷㪻㪼㫋㪼㪺㫋㫀㫆㫅㩷㫊㫀㪾㫅㪸㫃
㪫㪄㪝㪝㩷㫆㫌㫋㫇㫌㫋
㫊㫀㪾㫅㪸㫃
㪞㫉㪼㪼㫅䋺㪌
㪩㪼㪻䋺㪈
㪪㪼㪺㫆㫅㪻
㪛㪼㪺㫆㫄㫇㫉㪼㫊㫊㫆㫉
㪡㪉㪇㪉
㪭㪸㫃㫍㪼
㪛㪚
㪂㪈㪉㪭
㪡㪈㪇㪉
㪂㪈㪉㪭
㪠㫅㫇㫌㫋㩷㪻㪼㫋㪼㪺㫋㫀㫆㫅㩷㪹㫌㫑㫑㪼㫉
㫊㫀㪾㫅㪸㫃䋯㪫㫀㫄㪼㫉㩷㪹㫌㫑㫑㪼㫉
㫊㫀㪾㫅㪸㫃㩷㪾㪼㫅㪼㫉㪸㫋㫀㫆㫅㩷㪺㫀㫉㪺㫌㫀㫋
㪪㫌㫇㫇㫃㫐㩷㫇㫉㪼㫊㫊㫌㫉㪼㩷㫉㪼㪻㫌㪺㫋㫀㫆㫅
㫎㪸㫉㫅㫀㫅㪾㩷㪹㫌㫑㫑㪼㫉㩷㫊㫀㪾㫅㪸㫃
㪾㪼㫅㪼㫉㪸㫋㫀㫆㫅㩷㪺㫀㫉㪺㫌㫀㫋
㪙㫃㫀㫅㫂䋯㫄㪸㫃㪽㫌㫅㪺㫋㫀㫆㫅
㫎㪸㫉㫅㫀㫅㪾㩷㪹㫌㫑㫑㪼㫉㩷㫊㫀㪾㫅㪸㫃
㪾㪼㫅㪼㫉㪸㫋㫀㫆㫅㩷㪺㫀㫉㪺㫌㫀㫋
㪂㪈㪉㪭
㪧㫆㫎㪼㫉㩷㪣㪜㪛
㪞
㪚㪦㪉㩷㪞㪘㪪㩷㪦㪬㪫㪣㪜㪫
㪬㪧㪬㪚㪩㪈㪚㪩㪇㪇
㪂㪌㪭
㪙㫌㫑㫑㪼㫉
㪞㫉㪼㪼㫅
㪙㫌㫑㫑㪼㫉㩷㫊㫆㫌㫅㪻
㪾㪼㫅㪼㫉㪸㫋㫀㫆㫅
㪺㫀㫉㪺㫌㫀㫋
䋫䋵㪭
㪞㪼㫅㪼㫉㪸㫋㫀㫆㫅㩷㪺㫀㫉㪺㫌㫀㫋
㪂㪈㪉㪭
㪠㫅㫇㫌㫋㩷㪻㪼㫋㪼㪺㫋㫀㫆㫅
㪹㫌㫑㫑㪼㫉㩷㫊㫀㪾㫅㪸㫃
㪫㫀㫄㪼㫉㩷㪹㫌㫑㫑㪼㫉㩷㫊㫀㪾㫅㪸㫃
㪪㫌㫇㫇㫃㫐㩷㫇㫉㪼㫊㫊㫌㫉㪼
㫉㪼㪻㫌㪺㫋㫀㫆㫅㩷㫎㪸㫉㫅㫀㫅㪾
㪹㫌㫑㫑㪼㫉㩷㫊㫀㪾㫅㪸㫃
㪤㪸㫃㪽㫌㫅㪺㫋㫀㫆㫅㩷㫎㪸㫉㫅㫀㫅㪾㩷㪹㫌㫑㫑㪼㫉㩷㫊㫀㪾㫅㪸㫃
㪙㫃㫀㫅㫂㩷㫊㫀㪾㫅㪸㫃
㪞㪸㫊㩷㫊㫌㫇㫇㫃㫐
㪻㫀㫊㫇㫃㪸㫐㩷㪺㫀㫉㪺㫌㫀㫋
㪞㪸㫊㩷㫊㫌㫇㫇㫃㫐
㪻㫀㫊㫇㫃㪸㫐㩷㪣㪜㪛
㪞㫉㪼㪼㫅䋯㪩㪼㪻
㪝㫃㫆㫎㩷㪻㫀㫊㫇㫃㪸㫐㩷㪣㪜㪛
㪝㪣㪦㪮㪶㪘㪣㪘㪩㪤㪶㪙㪘㪩
㪛㪧
㪪㪼㫅㫊㫆㫉
㪝㫃㫆㫎㩷㫍㫆㫃㫌㫄㪼
㪻㪼㫋㪼㪺㫋㫀㫆㫅㩷㪺㫀㫉㪺㫌㫀㫋
㪪㪼㫇㪸㫉㪸㫋㫆㫉
㪝㫃㫆㫎㩷㫍㫆㫃㫌㫄㪼㩷㪻㪼㫋㪼㪺㫋㫀㫆㫅㩷㫊㫀㪾㫅㪸㫃
䋨㪝㪣㪦㪮㪶㪣㪜㪪㪪䋩
㪪㫆㫃㪼㫅㫆㫀㪻㩷㩷㫍㪸㫃㫍㪼
㪻㫉㫀㫍㫀㫅㪾㩷㪺㫀㫉㪺㫌㫀㫋
㪝㫃㫆㫎㩷㪭㫆㫃㫌㫄㪼
㫄㪸㫃㪽㫌㫅㪺㫋㫀㫆㫅㩷㫎㪸㫉㫅㫀㫅㪾
㪻㫀㫊㫇㫃㪸㫐㩷㪺㫀㫉㪺㫌㫀㫋
㪪㫌㫇㫇㫃㫐㩷㫇㫉㪼㫊㫊㫌㫉㪼
㪻㫀㫊㫇㫃㪸㫐㩷㪹㪸㫉㩷㪾㫉㪸㫇㪿
㪫㫀㫄㪼㫉㩷㫊㪼㫋㫋㫀㫅㪾㩷㫊㫀㪾㫅㪸㫃
㪪㪟㪦㪩㪫㪆㪣㪦㪥㪞
㪫㫀㫄㪼㫉㩷㫊㫀㪾㫅㪸㫃
㪪㪟㪦㪩㪫㪆㪣㪦㪥㪞
㪫㫀㫄㪼㫉㩷㪺㫀㫉㪺㫌㫀㫋
㪫㫀㫄㪼㫉㩷㫊㪼㫋㫋㫀㫅㪾
㪺㫀㫉㪺㫌㫀㫋
㪪㫌㫇㫇㫃㫐㩷㫊㫆㫌㫉㪺㪼
㪪㫌㫇㫇㫃㫐㩷㫇㫉㪼㫊㫊㫌㫉㪼
㪻㫀㫊㫇㫃㪸㫐㩷㫊㪼㫃㪼㪺㫋㫀㫆㫅
㪺㫀㫉㪺㫌㫀㫋
㪝㫀㫉㫊㫋
㪛㪼㪺㫆㫄㫇㫉㪼㫊㫊㫆㫉
㪚㫆㫄㫇㪸㫉㪸㫋㫆㫉
㫆㫌㫋㫇㫌㫋
㪞㪘㪪㪪㪶㪣㪜㪪㪪
㪧㪸㫉㪸㫄㪼㫋㪼㫉
㪺㫆㫄㫄㫌㫅㫀㪺㪸㫋㫀㫆㫅
㪺㫀㫉㪺㫌㫀㫋
㪪㪫㪘㪩㪫㪆㪪㪫㪦㪧
㫊㫎㫀㫋㪺㪿
㪡㪉㪇㪈
㪝㪣㪦㪮㪶㪘㪣㪘㪩㪤
㪝㪣㪦㪮㪶㪣㪜㪪㪪
㪭㪘㪣㪭㪜㪶㪦㪥
㪞㪘㪪㪪㪶㪣㪜㪪㪪
㪧㪦㪮㪜㪩
㪡㪉㪇㪇
㪂㪌㪭
㪞㪸㫊㩷㫊㫌㫇㫇㫃㫐㩷㫊㫆㫌㫉㪺㪼
㫊㪼㫃㪼㪺㫋㫀㫆㫅㩷㪺㫀㫉㪺㫌㫀㫋
㪜㫏㫋㪼㫉㫅㪸㫃㩷㫀㫅㫇㫌㫋
㩿㪫㪦㪬㪚㪟㪶㪧㪘㪥㪜㪣㪀
㪦㫇㪼㫉㪸㫋㫀㫆㫅㩷㫃㫆㪺㫂
㫊㫀㪾㫅㪸㫃
㩿㪣㪠㪞㪟㪫㪶㪪㪰㪪㪫㪜㪤㪀
㪡㪈㪇㪇
㪪㫌㫇㫇㫃㫐㩷㫇㫉㪼㫊㫊㫌㫉㪼
㫉㪼㪻㫌㪺㫋㫀㫆㫅㩷㫊㫀㪾㫅㪸㫃
㪪㫌㫇㫇㫃㫐
㫇㫌㫉㪼㫊㫊㫌㫉㪼
㪻㪼㫋㪼㪺㫋㫀㫆㫅
㪺㫀㫉㪺㫌㫀㫋
㪡㪈㪇㪈
㪧㫉㪼㫊㫊㫌㫉㪼
㪪㪼㫅㫊㫆㫉
UCR
Chapter 8: Block Diagram
Block Diagram
Page 23
UCR
Chapter 9: Troubleshooting
Refer to the “Troubleshooting“ of Instruction Manual.
ISSUE3
9-1
Troubleshooting
Page 26
2951 Ishikawa-cho, Hachioji-shi, Tokyo Japan
Engineering 2
Service Engineering Depertment
Service Division
PRINTED IN JAPAN
FIRST EDITION : 08.2007 W141B