olympus_exera_iii_bronchovideoscope_bf_series_repro.pdf
Page 1
REPROCESSING MANUAL
INSTRUCTIONS
Chapter 1
General Policy
1
Chapter 2
Function and Inspection of the
Accessories for Reprocessing
9
Compatible Reprocessing
Methods and Chemical Agents
23
Reprocessing Workflow for
Endoscopes and Accessories
35
Reprocessing the Endoscope
(and Related Reprocessing
Accessories)
43
Chapter 6
Reprocessing the Accessories
79
Chapter 7
Reprocessing Endoscopes and
Accessories Using an
Automated Endoscope
Reprocessor/Washer-Disinfector
91
Storage and Disposal
93
Chapter 3
Chapter 4
EVIS EXERA III BRONCHOVIDEOSCOPE
OLYMPUS BF-XP190
OLYMPUS BF-P190
OLYMPUS BF-Q190
OLYMPUS BF-H190
OLYMPUS BF-1TH190
Chapter 5
Chapter 8
Accessories:
• Sterilization cap (MAJ-1538)
• Suction cleaning adapter (MAJ-222)
• Single use single-ended cleaning brush (BW-400B)
• Channel cleaning brush (BW-15B)
• Single use combination cleaning brush (BW-411B)
• Suction connector cleaning brush (BW-15SH)
• Channel-opening cleaning brush (MH-507)
• Biopsy valve (MD-495*1)
• Suction valve (MAJ-207*1)
• Mouthpiece (MA-651)
*1 These products may not be available in some areas.
MAJ-1538
MAJ-222
BW-400B
BW-15B
MH-507
BW-411B
BW-15SH
MD-495
MAJ-207
MA-651
Refer to the endoscope’s companion manual, the “OPERATION MANUAL” with your endoscope model listed on the
cover, for operation information.
USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
Page 3
Contents
Contents
Chapter 1 General Policy ....................................................................... 1
1.1
Instructions ............................................................................................................... 1
1.2
Importance of reprocessing .................................................................................... 2
1.3
Signal words ............................................................................................................. 2
1.4
Warnings and cautions ............................................................................................ 3
1.5
Reprocessing before the first use .......................................................................... 7
1.6
Reprocessing and storage after use ...................................................................... 7
1.7
Reprocessing before patient procedure ................................................................ 8
Chapter 2 Function and Inspection of the Accessories for
Reprocessing ..........................................................................9
2.1
Sterilization cap (MAJ-1538) .................................................................................... 9
2.2
Single use single-ended cleaning brush (BW-400B) ........................................... 10
2.3
Suction connector cleaning brush (BW-15SH) .................................................... 12
2.4
Channel cleaning brush (BW-15B) ........................................................................ 14
2.5
Channel-opening cleaning brush (MH-507) ......................................................... 16
2.6
Single use combination cleaning brush (BW-411B) ............................................ 18
2.7
Suction cleaning adapter (MAJ-222) .................................................................... 21
Chapter 3 Compatible Reprocessing Methods and
Chemical Agents ...................................................................23
3.1
Compatibility summary .......................................................................................... 23
Microbiological efficacy ........................................................................................................
Material durability ................................................................................................................
List of compatible methods validated in terms of microbiological efficacy and
material durability ................................................................................................................
List of compatible methods validated in terms of material durability ....................................
24
24
25
27
3.2
Water (for reprocessing) ........................................................................................ 29
3.3
Detergent solution .................................................................................................. 29
3.4
Disinfectant solution .............................................................................................. 30
3.5
Rinse water ............................................................................................................. 30
3.6
Alcohol .................................................................................................................... 30
3.7
Ethylene oxide gas sterilization ............................................................................ 31
3.8
Steam sterilization (autoclaving) .......................................................................... 33
BF-190 Series REPROCESSING MANUAL
i
Page 4
Contents
Chapter 4 Reprocessing Workflow for Endoscopes and
Accessories...........................................................................35
4.1
Summary of reprocessing workflow ..................................................................... 35
4.2
Workflow for manually cleaning and disinfecting endoscopes and
accessories ...............................................................................................................36
4.3
Workflow for cleaning and disinfecting endoscopes and accessories
using an AER/WD .....................................................................................................38
4.4
Workflow for manually cleaning and sterilizing endoscopes and
accessories ...............................................................................................................40
Chapter 5 Reprocessing the Endoscope
(and Related Reprocessing Accessories) ..........................43
5.1
Summary of reprocessing the endoscope ........................................................... 43
5.2
Preparing the equipment for reprocessing .......................................................... 45
Equipment needed .............................................................................................................. 45
5.3
Precleaning the endoscope ................................................................................... 47
Equipment needed ..............................................................................................................
Preparation ..........................................................................................................................
Wipe the insertion section ...................................................................................................
Aspirate water ......................................................................................................................
Detach the suction tube from the endoscope ......................................................................
Detach the endoscope from the light source .......................................................................
5.4
47
48
48
49
50
50
Leakage testing of the endoscope ........................................................................ 50
Equipment needed .............................................................................................................. 50
Detach the accessories from the endoscope ...................................................................... 51
Perform the leakage test ..................................................................................................... 52
5.5
Manually cleaning the endoscope and accessories ........................................... 54
Equipment needed ..............................................................................................................
Clean the external surface ...................................................................................................
Brush the channels ..............................................................................................................
Aspirate detergent solution through the instrument channel and the suction channel ........
Immerse the endoscope and the suction cleaning adapter in detergent solution ................
Remove detergent solution from all channels .....................................................................
Dry external surfaces ...........................................................................................................
5.6
Manually disinfecting the endoscope and accessories ...................................... 66
Equipment needed ..............................................................................................................
Preparation ..........................................................................................................................
Immerse the endoscope and the suction cleaning adapter in disinfectant solution .............
Remove the endoscope and the suction cleaning adapter from disinfectant solution .........
5.7
54
55
55
62
64
65
65
66
66
67
69
Rinsing the endoscope and accessories following disinfection ....................... 70
Equipment needed .............................................................................................................. 71
Rinse the endoscope and accessories ................................................................................ 71
Alcohol flush ........................................................................................................................ 73
ii
BF-190 Series REPROCESSING MANUAL
Page 5
Contents
5.8
Sterilizing the endoscope and accessories ......................................................... 74
Ethylene oxide gas sterilization of the endoscope and accessories .................................... 74
Steam sterilization (autoclaving) of the accessories ............................................................ 76
STERRAD® 50/100S/200/NX sterilization of the endoscope and accessories ................. 76
5.9
Presoaking the endoscope .................................................................................... 78
Equipment needed .............................................................................................................. 78
Chapter 6 Reprocessing the Accessories .......................................... 79
6.1
Summary of reprocessing the accessories ......................................................... 79
Equipment needed .............................................................................................................. 80
6.2
Manually cleaning the accessories ...................................................................... 81
Disassembling the suction valve (MAJ-207) ........................................................................ 81
Manually cleaning the accessories ...................................................................................... 82
6.3
Manually disinfecting the accessories ................................................................. 86
6.4
Rinsing the accessories following disinfection .................................................. 87
Rinse the accessories ......................................................................................................... 87
Alcohol flush ........................................................................................................................ 88
6.5
Sterilizing the accessories .................................................................................... 89
Ethylene oxide gas sterilization ........................................................................................... 89
Steam sterilization (autoclaving) .......................................................................................... 90
Chapter 7 Reprocessing Endoscopes and Accessories
Using an Automated Endoscope
Reprocessor/Washer-Disinfector ........................................91
7.1
Reprocessing endoscopes and accessories using an automated endoscope
reprocessor/washer-disinfector.............................................................................. 91
Chapter 8 Storage and Disposal ......................................................... 93
8.1
Warnings and cautions: storage and disposal .................................................... 93
8.2
Storing the disinfected endoscope and accessories .......................................... 94
8.3
Storing the sterilized endoscope and accessories ............................................. 95
8.4
Disposal .................................................................................................................. 95
BF-190 Series REPROCESSING MANUAL
iii
Page 6
Contents
iv
BF-190 Series REPROCESSING MANUAL
Page 7
1.1 Instructions
Chapter 1
1.1
General Policy
Ch.1
Instructions
• This manual contains the reprocessing methods recommended by Olympus for the
endoscopes and accessories listed on the front cover.
• This instruction manual contains essential information on reprocessing endoscopes and
accessories safely and effectively.
• Before reprocessing, thoroughly review this manual and the manuals of the reprocessing
equipment and chemicals that will be used for reprocessing. Reprocess all the devices as
instructed.
• Note that the complete instruction manual set for the endoscope and accessories consists of
this manual and the “OPERATION MANUAL” with your endoscope model listed on the cover.
Both manuals accompanied the endoscope at shipment.
• Keep this manual and all related manuals in a safe and accessible location (e.g., in the
reprocessing area).
• If you have any questions or comments about any information in this manual, or if a problem
that cannot be solved occurs while reprocessing, contact Olympus.
Terms used in this manual
AER/WD:
AER and WD are abbreviations for Automated Endoscope Reprocessor and
Washer-Disinfector, which are used for reprocessing the endoscopes and accessories.
Disinfection:
Some countries classify the disinfection level and define “high-level disinfection”, and other
countries do not. The term “disinfection” is used, including the meaning “high-level
disinfection”, in this manual. For disinfectant solution, follow the instructions as described
in Section 3.4, “Disinfectant solution”.
BF-190 Series REPROCESSING MANUAL
1
Page 8
1.2 Importance of reprocessing
Ch.1
1.2
Importance of reprocessing
The medical literature reports incidents of cross-contamination resulting from improper reprocessing. It
is strongly recommended that all individuals engaged in reprocessing closely observe all instructions
given in this manual and the manuals of all ancillary equipment, and have a thorough understanding of
the following items:
• Professional health and safety policies of your hospital
• Instruction manuals for the endoscope, accessories, and all the other reprocessing
equipment
• Structure and handling of the endoscope and accessories
• Handling of pertinent chemicals
When selecting appropriate methods and conditions for cleaning and disinfection and sterilization,
follow the policies at your institution, applicable national laws and standards, and professional society
guidelines and recommended practices, in addition to the instructions given in this manual.
1.3
Signal words
The following signal words are used throughout this manual:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor
or moderate injury. It may also be used to alert against unsafe practices or potential
equipment damage.
NOTE
2
Indicates additional helpful information.
BF-190 Series REPROCESSING MANUAL
Page 9
1.4 Warnings and cautions
1.4
Ch.1
Warnings and cautions
WARNING
• An insufficiently reprocessed endoscope and/or accessory may pose an infection
control risk to the patients and/or operators who touch them.
• The endoscope reprocessor, video system center, light source, front panels of
equipment, and/or mouthpiece may cause an infection control risk. Perform proper
cleaning and disinfection as described in their respective instruction manuals. A
tap, basin, and/or nozzle of pharyngeal anesthetic spray that medical personnel
come in contact with may cause an infection control risk as well. Perform proper
replacement, cleaning, and disinfection.
• All disinfection methods (whether performed manually or by an AER/WD) and all
sterilization methods (whether performed by ethylene oxide gas or steam) require
thorough prior cleaning of the instruments being reprocessed. If the instruments
are not adequately cleaned prior to disinfection/sterilization, these processes will
be ineffective. Immediately after each patient procedure and before
disinfection/sterilization, thoroughly clean the endoscope and the accessories used
with the endoscope.
• All channels of the endoscope and all accessories used with the endoscope during
the patient procedure must be reprocessed after each patient procedure, even if
the channels or accessories were not used during the patient procedure.
Insufficient reprocessing of these components may pose an infection control risk to
patients and/or operators.
• Residual disinfectant solution may cause adverse reactions in patients. Therefore,
rinse all external surfaces and channels of the endoscope and accessories
thoroughly with water to remove residual disinfectant solution following disinfection.
• The results of sterilization depend on various factors. These factors include how
the equipment was packaged and the placing and loading of the package in the
sterilization device. Verify the sterilization process using biological and/or chemical
indicators. Follow the guidelines for sterilization issued by national authorities,
professional organizations and infection control professionals, including the
frequency of the above verification, as well as the instruction manual for the
sterilization device.
• Establish an internal system of identifying contaminated versus reprocessed
endoscopes and accessories to prevent both mix-ups and cross-contamination.
Some national or professional guidelines recommend separating dirty
(contaminated) area, clean area, and storage area. Touching a reprocessed
endoscope and/or accessories with contaminated gloves or placing them on a
contaminated hanger or surface, including letting them touch the floor, will
recontaminate them.
BF-190 Series REPROCESSING MANUAL
3
Page 10
1.4 Warnings and cautions
Ch.1
WARNING
• Prior to each patient procedure, confirm that the endoscope and accessories have
been properly reprocessed and stored. If there are any doubts or questions,
reprocess them again before the patient procedure, following the instructions given
in this manual.
• Perform a leakage test on the endoscope after each precleaning procedure. Do not
use the endoscope if a leak is detected. Use of an endoscope with a leak may
cause a sudden loss of the endoscopic image, damage to the bending mechanism,
or other malfunctions. Use of a leaking endoscope may also pose an infection
control risk.
Leakage tester
Figure 1.1
• Store alcohol in an airtight container. Alcohol stored in an open container may
cause a fire hazard and may result in a loss of efficacy due to evaporation.
• The accessories listed on the front cover of this manual cannot be refurbished or
repaired and are intended to be replaced once they show any signs of wear and
tear. Should any irregularity be observed, use a replacement accessory instead.
Using defective accessories may cause equipment malfunction, reduce the efficacy
of reprocessing, present a risk to patients and/or operators, or damage the
endoscope and/or accessories.
• Single-use brushes, such as the single use single-ended cleaning brush
(BW-400B) and the single use combination cleaning brush (BW-411B), are
designed for cleaning only one endoscope and its related accessories. Dispose of
the single-use brush immediately after use. Using a single-use brush to clean
multiple endoscopes and/or accessories may reduce its cleaning efficacy and may
damage the brush leading to brush breakage or the endoscope and/or accessory
damage.
4
BF-190 Series REPROCESSING MANUAL
Page 11
1.4 Warnings and cautions
Ch.1
WARNING
• Patient debris and used reprocessing chemicals pose infection control risks. To
guard against contact with dangerous chemicals and potentially infectious material,
wear appropriate personal protective equipment during reprocessing. Such
protective equipment should include appropriate eyewear, face mask, cap,
moisture-resistant clothing, shoe covers, and chemical-resistant gloves that fit
properly and are long enough to prevent skin exposure.
• The reprocessing room must be adequately ventilated to minimize the risks from
chemical vapors.
• Always remove contaminated personal protective equipment before leaving the
reprocessing area to prevent contamination from spreading.
• Only Olympus-recommended or Olympus-endorsed AERs/WDs have been
validated by Olympus. When using an AER/WD that is not recommended by
Olympus, the manufacturer of the AER/WD is responsible for validating
compatibility of the AER/WD with each Olympus endoscope and accessory.
• Before using an AER/WD, confirm that it is capable of reprocessing the endoscope
including all channels and accessories. Be sure to attach all required
connectors/adapters. Otherwise, insufficient reprocessing may pose an infection
control risk. If you are uncertain as to the ability of your AER/WD to reprocess the
endoscope including all channels and accessories, contact the manufacturer of the
AER/WD for specific instructions and information on compatibility and the required
connectors/adapters.
• Instructions provided in this manual are not valid for Olympus devices repaired by a
non-Olympus facility. The Olympus recommended reprocessing procedures have
not been validated for reprocessing devices repaired by a non-Olympus facility. In
the event that your device has been repaired by a non-Olympus facility, contact the
repair facility for instructions regarding reprocessing.
• Prions, which are the pathogenic agents of the Creutzfeldt-Jakob disease (CJD)
cannot be destroyed or inactivated by the reprocessing methods stated in this
instruction manual. When using the endoscope and accessories on patients with
CJD or variant Creutzfeldt-Jakob disease (vCJD), be sure to use them for such
patients only, or immediately dispose of them after use in an appropriate manner to
prevent the usage of exposed devices on other patients. For methods to handle
CJD, follow the respective guidelines in your country.
BF-190 Series REPROCESSING MANUAL
5
Page 12
1.4 Warnings and cautions
Ch.1
WARNING
• The endoscope and accessories may be damaged by published methods for
destroying or inactivating prions. For information on the durability of Olympus
equipment against a particular reprocessing method, contact Olympus. In general,
Olympus cannot guarantee the effectiveness, safety, and durability of reprocessing
methods not described in this reprocessing manual. If you choose to use a
reprocessing method not recommended in this manual, the local institution and/or
physicians must assume responsibility for its safety and efficacy. Make sure to
carefully inspect each piece of endoscopic equipment for irregularities (damage)
prior to each patient procedure. Do not use the equipment if any irregularity is
found.
• Good quality control practices typically require appropriate documentation. Items
such as local SOPs (Standard Operating Procedures), confirmation of operator
training, routine testing of the disinfectant’s MEC (Minimal Effective Concentration),
confirmation of the disinfectant’s use-life, etc. should be documented as performed.
CAUTION
• Before immersing the endoscope in reprocessing fluids, confirm that the
sterilization cap (MAJ-1538) is not attached to the endoscope. If the sterilization
cap is attached, the reprocessing fluids will be able to penetrate the inside of the
endoscope, and it can be damaged.
• Do not turn the insertion tube rotation ring in reprocessing fluids. The reprocessing
fluids will be able to penetrate the inside of the endoscope, and it can be damaged.
• When aerating or irrigating the endoscope channels, the air or water pressure must
not exceed 0.2 MPa (2 kgf/cm2, 29 psig). Higher pressures may cause damage to
the endoscope.
• Store spare accessories in their original packaging to prevent damage.
• To prevent damage, do not apply excessive force to the endoscope and
accessories during reprocessing.
• Vapors from disinfectant solution and alcohol may damage electronic devices such
as computers. Properly manage the quality and durability of the devices used in
reprocessing rooms and the ventilation performance of the rooms.
6
BF-190 Series REPROCESSING MANUAL
Page 13
1.5 Reprocessing before the first use
1.5
Ch.1
Reprocessing before the first use
New endoscopes, repaired endoscopes, accessories, and the carrying case for endoscopes are not
reprocessed prior to shipping from Olympus, regardless of whether those instruments are for new
purchase, demo, or loaner purposes. Reprocess all such endoscopes and accessories received from
Olympus according to the instructions given in this manual before storage and before using them in a
patient procedure.
1.6
Reprocessing and storage after use
WARNING
• Do not reuse rinse water.
• Disinfectant solution is only effective when used according to the disinfectant
manufacturer’s instructions. Follow the manufacturer’s instructions regarding
activation (if required), concentration, temperature, contact time and use life
required to achieve disinfection.
• If the disinfectant solution is reused, check its efficacy by proper methods, such as
using a test strip, according to the disinfectant manufacturer’s recommendations
prior to use.
• Do not reuse alcohol.
• Alcohol is not a sterilant or high-level disinfectant.
• To maintain sterility of equipment following sterilization, use sterile packaging and
wraps according to national guidelines.
BF-190 Series REPROCESSING MANUAL
7
Page 14
1.7 Reprocessing before patient procedure
Ch.1
1.7
Reprocessing before patient procedure
WARNING
• Improper storage practices, such as not thoroughly drying external and internal
surfaces (lumens) prior to storage, will lead to an infection control risk.
• Improper handling, such as touching a reprocessed endoscope and/or accessories
with contaminated gloves, placing a reprocessed device on a contaminated hanger
or surface, allowing devices to touch the floor, etc. will recontaminate the device.
NOTE
Some national or professional guidelines recommend reprocessing endoscopes
prior to their first use of the day.
Confirm that the endoscope and accessories have undergone proper reprocessing following their last
use and that they have been stored properly. Check the storage period of reprocessed endoscopes,
and check for surface contamination (e.g., dust). Check the expiration date marked on all items and
check for tears or breaches in the sterile packaging. If there are any doubts or questions concerning
whether a device is contaminated, reprocess it again following the instructions given in this manual.
8
BF-190 Series REPROCESSING MANUAL
Page 15
2.1 Sterilization cap (MAJ-1538)
Chapter 2
Function and Inspection of
the Accessories for
Reprocessing
Ch.2
Certain accessories are required for reprocessing the endoscope. This chapter describes the function
of these accessories. It also describes how to inspect these accessories before using them to
reprocess the endoscope.
2.1
Sterilization cap (MAJ-1538)
Connector
Figure 2.1
Function
When performing gas sterilization (e.g., ethylene
oxide gas sterilization, hydrogen peroxide low
temperature plasma), the sterilization cap must be
attached to the venting connector on the endoscope
connector.
Venting connector
Sterilization cap
Figure 2.2
Inspection
Confirm that the sterilization cap is free from scratches, flaws, and debris.
BF-190 Series REPROCESSING MANUAL
9
Page 16
2.2 Single use single-ended cleaning brush (BW-400B)
2.2
Single use single-ended cleaning brush
(BW-400B)
Ch.2
Endoscope model: BF-XP190
Metal tip
Bristles
Brush head
Shaft
Figure 2.3
Function
The single use single-ended cleaning brush is used to brush the inside of the instrument
channel of the endoscope.
Suction cylinder
Instrument channel port
Suction channel
Figure 2.4
10
BF-190 Series REPROCESSING MANUAL
Instrument channel
Page 17
2.2 Single use single-ended cleaning brush (BW-400B)
Inspection
CAUTION
Do not reprocess single use single-ended cleaning brush prior to use. The brush
may be damaged.
1
Confirm that the brush head and the metal tip of the distal end are securely attached.
Check the brush head for loose or missing bristles.
2
Check the bristles for damage. If the bristles are crushed, gently straighten them with
your gloved fingertips.
3
4
Check for bends, scratches, and other damage to the shaft.
Visually check for debris on the shaft and/or the bristles of the brush head. If there is
debris on the brush, immerse the brush in the water as described in Section 3.2,
“Water (for reprocessing)” and clean the brush until no debris is observed on the
brush.
BF-190 Series REPROCESSING MANUAL
11
Ch.2
Page 18
2.3 Suction connector cleaning brush (BW-15SH)
2.3
Ch.2
Suction connector cleaning brush (BW-15SH)
Endoscope model: BF-XP190
Bristles
Shaft
Metal tip
Brush head
Figure 2.5
Function
The suction connector cleaning brush is used to brush the inside of the suction channel of the
endoscope, and the inside of the suction valve (MAJ-207) and the biopsy valve (MD-495).
Suction cylinder
Instrument channel port
Suction channel
Instrument channel
Figure 2.6
Suction valve (MAJ-207)
Figure 2.7
12
BF-190 Series REPROCESSING MANUAL
Biopsy valve (MD-495)
Page 19
2.3 Suction connector cleaning brush (BW-15SH)
Inspection
1
Confirm that the brush head and the metal tip of the distal end are securely attached.
Check the brush head for loose or missing bristles.
2
Check the bristles for damage. If the bristles are crushed, gently straighten them with
your gloved fingertips.
3
4
Check for bends, scratches, and other damage to the shaft.
Visually check for debris on the shaft and/or the bristles of the brush head. If there is
debris on the brush, immerse the brush in the water as described in Section 3.2,
“Water (for reprocessing)” and clean the brush until no debris is observed on the
brush.
NOTE
The suction connector cleaning brush does not need to be reprocessed prior to its
first use.
BF-190 Series REPROCESSING MANUAL
13
Ch.2
Page 20
2.4 Channel cleaning brush (BW-15B)
2.4
Ch.2
Channel cleaning brush (BW-15B)
Endoscope model: except BF-XP190
Metal tip
Bristles
Brush head
Shaft
Figure 2.8
Function
The channel cleaning brush is used to brush the inside of the instrument channel and suction
channel of the endoscope and the inside of the suction valve (MAJ-207) and the biopsy valve
(MD-495).
Suction cylinder
Instrument channel port
Suction channel
Instrument channel
Figure 2.9
Suction valve (MAJ-207)
Figure 2.10
14
BF-190 Series REPROCESSING MANUAL
Biopsy valve (MD-495)
Page 21
2.4 Channel cleaning brush (BW-15B)
Inspection
1
Confirm that the brush head and the metal tip of the distal end are securely attached.
Check the brush head for loose or missing bristles.
2
Check the bristles for damage. If the bristles are crushed, gently straighten them with
your gloved fingertips.
3
4
Check for bends, scratches, and other damage to the shaft.
Visually check for debris on the shaft and/or the bristles of the brush head. If there is
debris on the brush, immerse the brush in the water as described in Section 3.2,
“Water (for reprocessing)” and clean the brush until no debris is observed on the
brush.
NOTE
The channel cleaning brush does not need to be reprocessed prior to its first use.
BF-190 Series REPROCESSING MANUAL
15
Ch.2
Page 22
2.5 Channel-opening cleaning brush (MH-507)
2.5
Channel-opening cleaning brush (MH-507)
Ch.2
Shaft
Bristles
Brush handle
Brush head
Figure 2.11
Function
The channel-opening cleaning brush is used to brush the endoscope’s suction cylinder,
instrument channel port, and both the inside and outside of the suction valve (MAJ-207).
Suction cylinder
Instrument channel port
Figure 2.12
16
BF-190 Series REPROCESSING MANUAL
Page 23
2.5 Channel-opening cleaning brush (MH-507)
Piston
Ch.2
Spring
Main body
Valve
Suction valve (MAJ-207)
Figure 2.13
Inspection
1
2
Check the brush head for loose or missing bristles.
3
4
Check for bends, scratches, and other damage to the shaft.
Check the bristles for any damage. If the bristles are crushed, gently straighten them
with your gloved fingertips.
Visually check for debris on the shaft and/or the bristles of the brush head. If there is
debris on the brush, immerse the brush in the water as described in Section 3.2,
“Water (for reprocessing)” and clean the brush until no debris is observed on the
brush.
NOTE
The channel-opening cleaning brush does not need to be reprocessed prior to its
first use.
BF-190 Series REPROCESSING MANUAL
17
Page 24
2.6 Single use combination cleaning brush (BW-411B)
2.6
Single use combination cleaning brush
(BW-411B)
Ch.2
Bristles
Tip
Channel cleaning
brush part
Handle
Shaft
Bristles
Channel-opening
cleaning brush part
Figure 2.14
Function
The channel cleaning brush part of the single use combination cleaning brush is used to brush
the inside of the instrument channel (except BF-XP190) and suction channel of the endoscope,
and the inside of the suction valve (MAJ-207) and the biopsy valve (MD-495). The
channel-opening cleaning brush part of the single use combination cleaning brush is used to
brush the endoscope’s suction cylinder, instrument channel port, and both the inside and
outside of the suction valve.
Suction cylinder
Instrument channel port
Suction channel
Figure 2.15
18
BF-190 Series REPROCESSING MANUAL
Instrument channel
Page 25
2.6 Single use combination cleaning brush (BW-411B)
Ch.2
Suction valve (MAJ-207)
Biopsy valve (MD-495)
Figure 2.16
Piston
Spring
Main body
Valve
Suction valve (MAJ-207)
Figure 2.17
BF-190 Series REPROCESSING MANUAL
19
Page 26
2.6 Single use combination cleaning brush (BW-411B)
Inspection
CAUTION
Ch.2
Do not reprocess single use combination cleaning brush prior to use. The brush
may be damaged.
20
1
2
Remove the brush from its packaging just prior to use.
3
Check the channel cleaning brush and channel-opening cleaning brush parts for
loose or missing bristles.
4
Check the bristles of the channel cleaning brush and the channel-opening cleaning
brush parts for any damage. If the bristles are crushed, gently straighten them with
your gloved fingertips.
5
Check for bends, scratches, and other damage to the shaft.
Confirm that the channel cleaning brush part and tip at the distal end are securely
attached.
BF-190 Series REPROCESSING MANUAL