BCS
  • Home
  • Contact Us
  • Sign In

OLYMPUS

OLYMPUS Leakage Tester MB-155 Instructions

Revision Ver GR3093 21

OLYMPUS Leakage Tester

4 Pages

USA: CAUTION:  INSTRUCTIONS  Federal law restricts this device to sale by or on the order of a physician... LEAKAGE TESTER  MB-155  1 Intended Use The Leakage Tester is designed to check watertightness of Olympus endoscopes in conjunction with Olympus Light Source or Maintenance Unit (MU-1)... 2 Symbols The meaning(s) of the symbol(s) shown on the back cover of this instruction manual are as follows: Manufacturer  Authorized representative in the European Community  3 Signal Words The following signal words are used throughout this manual: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury... t may also be used to alert against unsafe practices or potential equipment damage... 4 Specifications Medical Devices Directive  Insufflation function  This device complies with the requirements of Directive 93/42/EEC concerning medical devices... lassification: Class I 2...  × 104 Pa or less  Performance  5 Inspection 1... nspect that the Connector is not markedly deformed... nspect that the Coiled Tube is not cracked... nspect that the Connector Cap is not deformed, nor is it wet inside... onnector  Coiled Tube  Connector Cap  Water drops inside the Connector Cap may enter the endoscope through the Venting Connector of the endoscope’s Light Guide Connector Section or Water-resistant Cap and damage the internal construction... emove them by shaking the Connector Cap several times... –1–   4... nspect that the Guide Pin inside the Connector Cap is not missing, broken, or deformed... onnector Cap  Guide Pin  6 Operation Refer to the instructions manual of the Light Sources, Maintenance Unit and Endoscopes... 6...  Connecting to the Light Source (or Maintenance Unit)  1... urn the power switch ON... hen using the Light Source, set the air feeding pressure to “HIGH” or “3” using the air regulator switch... 3... • When using a Light Source other than OES/EVIS, turn ON the switch and ignition switch (if applicable)...  Adjust the Light Source brightness to the minimum light output... nsert the Leakage Tester’s Connector fully into the output socket, then make sure that the air pump’s operating noise is heard...   6...  When using a Light Source other than OES/EVIS, make sure that air is emitted through the air pipe inside the output socket... Connecting to the Endoscope  1... ightly depress the pin inside the Connector Cap and ensure that air is being emitted through the Coiled Tube...  Repeat Section 6... above, if air is not emitted... ake sure that no water drop is present inside the Connector Cap... 2... heck that the Venting Connector of the endoscope’s Light Guide Connector Section or Water-resistant Cap is free from water drops...   4... Wipe dry the Venting Connector with a piece of gauze to prevent the endoscope internal construction from getting damaged... lace the Connector Cap over the Venting Connector, aligning the Pin on the Connector with the keyway on the Connector Cap... epress and rotate the Connector Cap clockwise (approximately 90) until no further rotation is possible... 5... ull the Connector Cap outward and make sure that it does not come off...  • •  6...  At this point, note an expansion of the Bending Section’s covering due to increased internal pressure... hen attaching the Leakage Tester, always connect it to the Light Source first... eversing Sections 6... and 6... (i...  connecting to the endoscope first) will damage the endoscope... onnect it to the Venting Connector of the endoscope’s Light Guide Connector Section or Water-resistant Cap before putting the endoscope in water; otherwise water will leak into the endoscope... Leakage Testing  1... mmerse entire endoscope in water... bserve the endoscope carefully for about 1 minute... f no bubbling is observed from the endoscope, the endoscope is watertight... roceed to Section 6...  • •  Some initial bubbling may be observed due to air trapped in crevices on the endoscope’s outer surface... his is normal...  thorough observation may be achieved by turning over the endoscope in water... ontinuous bubbling from a given area indicates a leak... O NOT USE ENDOSCOPE... 3... f continuous bubbling in observed from a given area, this indicates a leak...  • • •  Remove the endoscope from the water immediately... urn OFF Light Source and disconnect the Leakage Tester from the Light Source... ait approximately 30 seconds (or until the covering of the Bending Section returns to its normal shape); then disconnect the opposite end of Leakage Tester from the endoscope... O NOT USE THE ENDOSCOPE... nd send it to Olympus... –2–
File Type: PDF
File Size: 639KB
File Name: olympus_leakage_tester_mb-155_instructions__ver_gr3.pdf
Available for download

Available for download in the BioClinical Services library

If you already belong to a hospital with access, please login via your intranet.

Contact Us to Register Your InterestAlready a member? Log in hereAbout Us

olympus_leakage_tester_mb-155_instructions__ver_gr3.pdf

Page 1

USA: CAUTION:

INSTRUCTIONS

Federal law restricts this device to

sale by or on the order of a physician.

LEAKAGE TESTER

MB-155

1 Intended Use

The Leakage Tester is designed to check watertightness of Olympus endoscopes in conjunction with Olympus Light

Source or Maintenance Unit (MU-1).

2 Symbols

The meaning(s) of the symbol(s) shown on the back cover of this instruction manual are as follows:

Manufacturer

Authorized representative in the European Community

3 Signal Words

The following signal words are used throughout this manual:

Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It

may also be used to alert against unsafe practices or potential equipment damage.

4 Specifications

Medical Devices Directive

Insufflation function

This device complies with the

requirements of Directive 93/42/EEC

concerning medical devices.

Classification: Class I

2.84 × 104 Pa or less

Performance

5 Inspection

1. Inspect that the Connector is not markedly deformed.

2. Inspect that the Coiled Tube is not cracked.

3. Inspect that the Connector Cap is not deformed, nor is it wet inside.

Connector

Coiled Tube

Connector

Cap

Water drops inside the Connector Cap may enter the endoscope through the Venting Connector of the

endoscope’s Light Guide Connector Section or Water-resistant Cap and damage the internal construction.

Remove them by shaking the Connector Cap several times.

–1–

Page 2

4. Inspect that the Guide Pin inside the Connector Cap is not missing, broken, or deformed.

Connector Cap

Guide Pin

6 Operation

Refer to the instructions manual of the Light Sources, Maintenance Unit and Endoscopes.

6.1

Connecting to the Light Source (or Maintenance Unit)

1. Turn the power switch ON.

2. When using the Light Source, set the air feeding pressure to “HIGH” or “3” using the air regulator switch.

3.

• When using a Light Source other than OES/EVIS, turn ON the switch and ignition switch (if applicable).

• Adjust the Light Source brightness to the minimum light output.

Insert the Leakage Tester’s Connector fully into the output socket, then make sure that the air pump’s operating

noise is heard.

•

6.2

When using a Light Source other than OES/EVIS, make sure that air is emitted through the air pipe inside the

output socket.

Connecting to the Endoscope

1. Lightly depress the pin inside the Connector Cap and ensure that air is being emitted through the Coiled Tube.

• Repeat Section 6.1 above, if air is not emitted.

Make sure that no water drop is present inside the Connector Cap.

2.

3. Check that the Venting Connector of the endoscope’s Light Guide Connector Section or Water-resistant Cap is

free from water drops.

•

4.

Wipe dry the Venting Connector with a piece of gauze to prevent the endoscope internal construction from

getting damaged.

Place the Connector Cap over the Venting Connector, aligning the Pin on the Connector with the keyway on the

Connector Cap. Depress and rotate the Connector Cap clockwise (approximately 90) until no further rotation is

possible.

5. Pull the Connector Cap outward and make sure that it does not come off.

•

•

•

6.3

At this point, note an expansion of the Bending Section’s covering due to increased internal pressure.

When attaching the Leakage Tester, always connect it to the Light Source first. Reversing Sections 6.1 and

6.2 (i.e. connecting to the endoscope first) will damage the endoscope.

Connect it to the Venting Connector of the endoscope’s Light Guide Connector Section or Water-resistant

Cap before putting the endoscope in water; otherwise water will leak into the endoscope.

Leakage Testing

1. Immerse entire endoscope in water.

2. Observe the endoscope carefully for about 1 minute. If no bubbling is observed from the endoscope, the

endoscope is watertight. Proceed to Section 6.4.

•

•

Some initial bubbling may be observed due to air trapped in crevices on the endoscope’s outer surface. This

is normal.

A thorough observation may be achieved by turning over the endoscope in water.

Continuous bubbling from a given area indicates a leak. DO NOT USE ENDOSCOPE.

3. If continuous bubbling in observed from a given area, this indicates a leak.

•

•

•

•

Remove the endoscope from the water immediately.

Turn OFF Light Source and disconnect the Leakage Tester from the Light Source.

Wait approximately 30 seconds (or until the covering of the Bending Section returns to its normal shape);

then disconnect the opposite end of Leakage Tester from the endoscope.

DO NOT USE THE ENDOSCOPE. And send it to Olympus.

–2–

Page 3

6.4

1.

2.

3.

4.

Disconnecting the Leakage Tester

Remove the entire endoscope from the water and turn OFF the Light Source (or Maintenance Unit).

Disconnect the Leakage Tester from the Light Source (or Maintenance Unit).

Wait approximately 30 seconds (or until the covering of the Bending Section returns to its normal shape).

Disconnect the Leakage Tester from the Venting Connector on the endoscope’s Light Guide Connector Section or

Water-resistant Cap by depressing and rotating it counterclockwise.

•

•

5.

Do not detach the Leakage Tester before the endoscope has been removed from the water.

When detaching the Leakage Tester, always disconnect it from the Light Source first. Failure to follow this

exact procedure will not allow the instrument to depressurize and damage will result.

Thoroughly dry the Leakage Tester.

7 Maintenance and Storage

7.1

Leakage Tester

1. Clean the Leakage Tester with a piece of gauze moistened with 70% ethyl or isopropyl alcohol.

2. Remove any moisture from the Connector Cap.

7.2

Endoscope and Light Source (or Maintenance Unit)

Clean and store each instrument according to its instruction manual.

Leakage

Tester

Light source

Connector

Connector Cap

Connector

Coiled Tube

Connector

Cap

Attach/detach in the air

(and not under water).

Gauze

Light Guide

Connector Section

Make sure pump noise is heard.

Venting

Connector

Depress

Make sure air is emitted

through coiled tube.

Pin

Venting

Connector

The covering of the

Bending Section

Thoroughly wipe dry.

Slight expansion of the covering is OK.

–3–

Never disconnect leakage tester

while fiberscope is under water.

Page 4

Manufactured by

2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan

Fax: (042)646-2429 Telephone: (042)642-2111

Distributed by

(Premises/Goods delivery) Wendenstrasse 14-18, 20097 Hamburg, Germany

(Letters) Postfach 10 49 08, 20034 Hamburg, Germany

Fax: (040)23773-4656 Telephone: (040)23773-0

KeyMed House, Stock Road, Southend-on-Sea, Essex SS2 5QH, United Kingdom

Fax: (01702)465677 Telephone: (01702)616333

3500 Corporate Parkway, P.O. Box 610, Center Valley, PA 18034-0610, U.S.A.

Fax: (484)896-7128 Telephone: (484)896-5000

5301 Blue Lagoon Drive, Suite 290 Miami, FL 33126-2097, U.S.A.

Fax: (305)261-4421 Telephone: (305)266-2332

117071, Moscow, Malaya Kaluzhskaya 19, bld. 1, fl.2, Russia

Fax: (095)958-2277 Telephone: (095)958-2245

Olympus-Tower, 114-9 Samseong-Dong, Gangnam-Gu, Seoul 135-090 Korea

Fax: (02)6255-3494 Telephone: (02)6255-3210

A8F, Ping An International Financial Center, No. 1-3, Xinyuan South Road,

Chaoyang District, Beijing, 100027 P.R.C.

Fax: (86)10-5976-1299 Telephone: (86)10-5819-9000

491B, River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373

Fax: 6834-2438 Telephone: 6834-0010

31 Gilby Road, Mount Waverley, VIC., 3149, Australia

Fax: (03)9543-1350 Telephone: (03)9265-5400

GR3093 21

–4–

©1997 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved.

Printed in Japan 201300716 *0000

Previous Page Next Page
© 2021 BioClinical Services Ptd Ltd
Terms of Use
Privacy Statement