cll-v1_instructions_for_use__april_2012.pdf
Page 1
INSTRUCTIONS FOR USE
OLYMPUS CLL-V1
WA97020A
Page 3
3
Content
1 General Information............................................................... 5
1.1 Signal Words................................................................. 5
1.2 Conventions................................................................... 5
1.3 User Instructions........................................................... 5
1.4 Manufacturer................................................................. 5
2 Safety Information.................................................................. 6
2.1 Intended Use................................................................. 6
2.2 User Qualification.......................................................... 6
2.3 Environment of Use....................................................... 6
2.4 General Warnings and Cautions................................... 7
3 Upon Delivery........................................................................ 8
4 Product Description............................................................... 9
4.1 Symbols......................................................................... 9
4.2 Warranty...................................................................... 11
5 Preparation.......................................................................... 12
5.1 Safety Information for Preparation............................... 12
5.2 Inspection.................................................................... 12
5.3 Installing the Foot Holders........................................... 13
5.4 Connecting the Light Source to the Mains Supply....... 13
5.5 Connecting the Light-Guide Cable.............................. 14
5.6 Testing the Light Source.............................................. 14
6 Use...................................................................................... 15
6.1 Safety Information for Use........................................... 15
7 Troubleshooting................................................................... 17
8 After Use.............................................................................. 18
9 Cleaning and Disinfection.................................................... 19
9.1 Safety Information for Cleaning and Disinfection........ 19
9.2 Manual Cleaning and Disinfection............................... 19
10 Maintenance........................................................................ 21
10.1 Annual Inspection........................................................ 21
10.2 Replacing the Fuses.................................................... 21
11 Transport and Storage......................................................... 22
12 Repair and Disposal............................................................ 23
12.1 Repair.......................................................................... 23
12.2 Disposal....................................................................... 23
W7089150
OLYMPUS CLL-V1
Page 4
4
13 Compatible Equipment........................................................ 24
14 Technical Data..................................................................... 25
14.1 Specifications.............................................................. 25
14.2 Ambient Conditions..................................................... 25
14.3 Safety.......................................................................... 25
14.4 Power Cable................................................................ 26
15 Electromagnetic Compatibility............................................. 27
OLYMPUS CLL-V1
W7089150
Page 5
5
1 General Information
1.1 Signal Words
The following signal words are used throughout this document.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury.
NOTICE
Indicates a property damage message.
1.2 Conventions
This is the safety alert symbol. It is used to alert the user to potential physical
injury hazards. Observe all safety messages that follow this symbol to avoid
possible injury.
1. A numeration indicates a sequence of actions.
2.
• Bullet points indicate individual actions or different options for action.
-- Dashes indicate the listing of data, options or objects.
1) Numbers with right parenthesis name elements in illustrations.
1.3 User Instructions
• Before use, thoroughly read these instructions for use and the instructions for
use of all other products that will be used during the procedure.
• If the required instructions for use are missing, immediately contact an
Olympus representative.
• Keep the instructions for use in a safe, accessible location.
1.4 Manufacturer
Olympus Winter & Ibe GmbH
Kuehnstr. 61
22045 Hamburg
Germany
W7089150
OLYMPUS CLL-V1
Page 6
6
2 Safety Information
2.1 Intended Use
Light source for use with a compatible telescope, camera head, video system
center, and other ancillary equipment. The light source is to be used for
endoscopic diagnosis, treatment and video observation in the following fields:
-- Bronchoscopy
-- Otorhinolaryngology
-- Urology
-- Gynaecology
-- General surgery
-- Gastroenterology
• Do not use for any other purposes.
2.2 User Qualification
Medical use
This product is only intended to be used by a trained physician or trained
qualified medical personnel under the supervision of a physician.
These instructions for use do not explain or discuss clinical procedures.
Cleaning and disinfection
Cleaning and disinfection of the product may only be performed by qualified
hygiene personnel.
Repair
Repair of the product may only be performed by trained qualified servicing
personnel that has been authorized by Olympus.
Otherwise, Olympus cannot be held responsible for the safety and performance
of the product.
2.3 Environment of Use
Medical use
This product is only intended to be used in hospitals and medical rooms with
appropriate endoscopic equipment.
Location
This product must not be used in a flammable atmosphere.
Cleaning and disinfection
This product must be cleaned and reprocessed according to the applicable
national and local standards and regulations.
OLYMPUS CLL-V1
W7089150
Page 7
7
2.4 General Warnings and Cautions
The following warnings and cautions apply to the general handling of the
product. This information is to be supplemented by the warnings and cautions
given in each chapter in this document or in the instructions for use of any
product being used with this product.
WARNING
Risk of injury to the patient
Other electrical devices can interfere with the function of the OLYMPUS CLL-V1.
• Do not place the OLYMPUS CLL-V1 next to, on or under other electrical
devices. If the OLYMPUS CLL-V1 must be placed next to, on or under other
electrical devices, monitor the OLYMPUS CLL-V1 during the procedure
to make sure that the OLYMPUS CLL-V1 does not switch off and that the
brightness does not change.
WARNING
Risk of injury to the patient and/or the user
Light emitted from the light-guide cable or endoscope can damage the retina.
• Make sure that the light-guide cable or endoscope does not point at the
patient’s or user’s eye.
• Do not look directly into the light emitted from the light-guide cable or
endoscope.
WARNING
Risk of injury to the patient and/or the user
There is a risk of injury to the patient due to malfunction of the equipment.
• Always have spare equipment available.
CAUTION
Risk of injury to the patient and/or the user
Using incompatible equipment may lead to injury of the patient and/or the user
as well as damage to the product.
• Only use compatible equipment as listed in the chapter “Compatible
Equipment” in this document.
W7089150
OLYMPUS CLL-V1
Page 8
8
3 Upon Delivery
• Make sure that the shipping case contains the following items:
-- Light source
-- 4 foot holders
-- Stencil for foot holders
-- 2 spare fuses
-- Instructions for use
• Contact an Olympus representative or an authorized service center if any
items are missing or damaged.
OLYMPUS CLL-V1
W7089150
Page 9
9
4 Product Description
Light source, front
1)
2)
3)
4)
Power switch with power indicator
Socket for light-guide cable
Brightness control buttons (- and +)
Brightness display
Light source, rear panel
1)
2)
3)
4)
Ventilation slots
Potential equalization conductor
Power supply with fuse holder
Identification plate
4.1 Symbols
This section gives an explanation for each symbol used on the product and on
the packing of the product.
Catalog number
W7089150
OLYMPUS CLL-V1
Page 10
10
Quantity
Serial number
Manufacturer
Follow instructions for use
Caution, consult accompanying documents
Indicates the range of humidity to which the medical device can be safely
exposed
Indicates the temperature limits to which the medical device can be safely
exposed
CE certification mark – symbol for the compliance with the Medical Device
Directive 93/42/EEC
Potential equalization
Fuse
Type BF applied part
cTUVus marking
Symbol to indicate compliance with EMC requirements by the Japanese Ministry
of Health, Labour, and Welfare
OLYMPUS CLL-V1
W7089150
Page 11
11
In accordance with European Directive 2002/96/EC on Waste Electrical and
Electronic Equipment, this symbol indicates that the product must not be
disposed off as unsorted municipal waste, but should be collected separately
4.2 Warranty
Any warranty claims towards Olympus are forfeited if the user or unauthorized
persons attempt repair or modification of the product. No warranty is provided
for any damage due to misuse of the product.
W7089150
OLYMPUS CLL-V1
Page 12
12
5 Preparation
5.1 Safety Information for Preparation
WARNING
Risk of injury to the patient
There is a risk of an electric shock when using endoscopic equipment in
combination or when using endoscopic equipment in combination with
energized endotherapy devices. The patient leakage currents may be additive.
• Before each use, check the applied part classification of the endoscopic
equipment and especially the energized endotherapy devices used during the
procedure.
• Only use endoscopic equipment in combination or endoscopic equipment in
combination with energized endotherapy devices that meet at least the same
requirements of applied part classification.
This is particularly important if a type CF applied part endoscope is used. In
this case a type CF applied part energized endotherapy device should be
used to minimize total leakage current.
• For the applied part classification, refer to the respective instructions for use.
CAUTION
Risk of injury to the patient
Blocking the ventilation slots can cause overheating. As a result, the light will
first be dimmed and eventually be switched off.
• Check that the ventilation slots are not blocked.
5.2 Inspection
Inspection regarding cleaning and disinfection
• Make sure that the product has been properly cleaned and disinfected.
• Visually inspect the product thoroughly. The product should have no signs of
visual contamination and should be visually clean.
Inspecting the light source
• Visually inspect the light source. Check that:
-- the technical condition permits a safe use.
-- all labels are clearly visible.
-- the air inlets on the bottom side of the light source and the ventilation slots are
clean.
Inspecting the electrical contacts
• Check the electrical contacts of all equipment which will be used during the
procedure for corrosion.
If any electrical contacts are corroded, replace the equipment.
OLYMPUS CLL-V1
W7089150
Page 13
13
5.3 Installing the Foot Holders
NOTICE
Risk of damage to the product
If the OLYMPUS CLL-V1 falls, it will be damaged.
• Protect the OLYMPUS CLL-V1 against falls. Install the foot holders that are
delivered with the OLYMPUS CLL-V1 as described below.
1. Put the stencil on the surface where the OLYMPUS CLL-V1 will be set up.
2. Remove the protective foil from the foot holders.
3. Place the foot holders in the cut-outs of the stencil. The adhesive side must
face down.
4. Remove the stencil.
5. Place the OLYMPUS CLL-V1 on the foot holders. Make sure that the feet of
the OLYMPUS CLL-V1 fit into the foot holders.
5.4 Connecting the Light Source to the Mains Supply
According to IEC 60601-1 requirements, the OLYMPUS CLL-V1 is equipped with
a potential equalization conductor.
• If local safety regulations require, connect the OLYMPUS CLL-V1 to a potential
equalization line.
1. Connect the power cable to the light source.
2. Connect the other end of the power cable to the mains socket.
Connect the light source only to a grounded alternating current mains supply
that complies with the requirements of IEC 60320-1 / C13 and the information
given in the chapter “Technical Data”.
The power cable must not be longer than 4.5 m.
W7089150
OLYMPUS CLL-V1
Page 14
14
5.5 Connecting the Light-Guide Cable
1. Connect the light source connector to the light source.
The light source connector must click into position.
2. Check the light-guide cable for visible damage. Refer to the instructions for
use of the light-guide cable.
3. Test the light transmission of the light-guide cable. Refer to the instructions
for use of the light-guide cable.
5.6 Testing the Light Source
The light source keeps the settings from the previous use.
Functional test
1. Switch on the light source.
All indicators and displays light up for a few seconds. The light output is
activated.
2. Check the quality of the endoscopic image. Refer to the instructions for use
of the endoscope, light-guide cable, camera control unit and the monitor.
3. Check the brightness display of the light source.
OLYMPUS CLL-V1
W7089150
Page 15
15
6 Use
6.1 Safety Information for Use
WARNING
Risk of injury to the patient
A malfunction of the OLYMPUS CLL-V1 can lead to injury of the patient.
• If the OLYMPUS CLL-V1 shows any signs of malfunction or unexpected
behavior during use, contact an Olympus representative or an authorized
service center. Do not use the OLYMPUS CLL-V1.
WARNING
Risk of injury to the patient
The endoscope, the video equipment, and the OLYMPUS CLL-V1 are classified
as type BF instruments according to IEC 60601-1. Using the endoscope can
have a serious effect on the cardiac function (e.g. ventricular fibrillation).
• Never let the endoscope or any of the instruments that are used during the
procedure come in contact with the heart or any tissue near the heart.
CAUTION
Risk of injury to the patient
Light sources emit large amounts of energy. As a result the connectors of the
endoscopic equipment and the distal end of the endoscope become hot. There
is a risk of:
-- thermal injury to the patient’s tissue (e.g., from prolonged exposure to the
intense illumination in cavities with small lumens, or if the distal end of the
endoscope is placed into close proximity to the tissue).
-- burns to the patient’s or user’s skin.
-- burns or thermal damage to surgical equipment (e.g., surgical drapes, plastic
materials, etc.).
• Do not place the endoscopic equipment on the patient’s skin, on flammable
materials, or on heat-sensitive materials.
• Set the output power of the light source to the minimum level that is necessary
for a sufficient illumination of the target area. Avoid prolonged exposure to
intense illumination.
• Switch off the light source if not used.
CAUTION
Risk of injury to the patient
Light sources emit large amounts of energy. As a result, the temperatures in
front of the light emission surface increase. There is a risk of thermal injury to
the patient’s tissue.
• Do not use high intensity illumination in cavities with small lumens.
• Do not use high intensity illumination when the distal end of the endoscope is
placed into close proximity to the tissue.
• If the distal end is soiled, remove the endoscope from the trocar tube and
clean the distal end.
WARNING
Risk of injury to the patient
LED light offers a different color reproduction than conventional light sources.
This can cause incorrect diagnoses.
W7089150
OLYMPUS CLL-V1
Page 16
16
• Perform a white balance before using the OLYMPUS CLL-V1.
Additionally, perform a white balance after changing the endoscope.
OLYMPUS CLL-V1
W7089150
Page 17
17
7 Troubleshooting
Problem
Cause
Countermeasure
The power indicator and the
brightness display do not illuminate.
No power supply.
Check whether the light source has been switched on.
Check whether the power cable of the light source has been properly connected.
Replace the fuse. Refer to section “Replacing the Fuses”.
If these measures do not solve the problem, contact an Olympus representative.
The power indicator is flashing.
Error in light source.
Switch off the light source. Switch on the light source.
If these measures do not solve the problem, contact an Olympus representative.
The light source does not emit light.
No power supply or brightness
setting too low.
Check whether the light source has been switched on.
Check whether the power cable of the light source has been properly connected.
Check whether the light-guide cable has been properly connected to the light
source.
Increase the brightness setting.
Switch off the light source. Switch on the light source.
If these measures do not solve the problem, contact an Olympus representative.
Brightness decreases during the
procedure.
Light source gets hot.
Switch off the light source.
Allow the light source to cool down.
If these measures do not solve the problem, contact an Olympus representative.
W7089150
OLYMPUS CLL-V1
Page 18
18
8 After Use
CAUTION
Risk of injury to the user
The connectors on the light-guide cable become hot. There is a risk of burns
when touching the connectors.
• Let the light-guide cable cool down after use.
Preparing the light source for cleaning and disinfection
1. Switch off the power switch.
2. Disconnect the power cable.
3. Make sure that the light source has completely cooled down to room
temperature to avoid any risk of ignition or explosion.
4. Disconnect the light-guide cable from the light source.
OLYMPUS CLL-V1
W7089150
Page 19
19
9 Cleaning and Disinfection
9.1 Safety Information for Cleaning and Disinfection
WARNING
Risk of injury to the patient and/or user
Liquids entering the OLYMPUS CLL-V1 or its accessories during reprocessing
can cause a malfunction.
• Make sure that no liquids enter the OLYMPUS CLL-V1 or its accessories
during reprocessing.
WARNING
Risk of infection for the user
Patient debris and reprocessing chemicals are hazardous.
• Wear personal protective equipment to guard against dangerous chemicals
and potentially infectious material.
• During cleaning and disinfection, wear appropriate personal protective
equipment, such as eye wear, face mask, moisture-resistant clothing and
chemical-resistant gloves that fit properly and are long enough so that no part
of the skin is exposed.
• Always remove contaminated protective equipment before leaving the
reprocessing area.
NOTICE
Risk of damage to the product
Cleaning can scratch the surfaces and deform or corrode the connectors and
electrical contacts.
• Do not clean the connectors or electrical contacts of the OLYMPUS CLL-V1.
• Do not wipe the surfaces with an abrasive material. The surfaces will be
scratched.
9.2 Manual Cleaning and Disinfection
• When using 70 % alcohol, make sure that the system has cooled down to
room temperature.
• Refer to national and local guidelines regarding when using alcohol as
disinfectant.
Cleaning the surfaces
Perform the following cleaning procedure immediately after use. If cleaning is
delayed, organic debris will solidify and it may be difficult to clean the surfaces.
1. Remove all dust and soiling with a soft cloth moistened with neutral
detergent solution (approved by the manufacturer for cleaning plastic and
metal surfaces).
Do not exceed maximum recommended levels of concentration given by the
manufacturer of the solution.
2. Dry the surfaces with a clean, lint-free cloth.
Disinfecting the surfaces
• Make sure to disinfect the equipment regularly.
W7089150
OLYMPUS CLL-V1
Page 20
20
• Wipe surfaces with a cloth that has been moistened with a disinfectant.
Use an alcohol based disinfectant according to the field of applications.
The disinfectant must be approved by the manufacturer for the surface
disinfection of medical devices and for the material to be disinfected.
• Refer to local/national guidelines regarding the approval of using alcohol as
disinfectant.
• Refer to the instructions given by the manufacturer of the disinfectant. Make
sure not to exceed the manufacturer’s specification on temperature and
concentration.
• Never immerse the equipment in liquid.
OLYMPUS CLL-V1
W7089150
Page 21
21
10 Maintenance
10.1 Annual Inspection
WARNING
Risk of injury to the patient and/or user
When the OLYMPUS CLL-V1 is opened, it can be damaged.
• Do not open the OLYMPUS CLL-V1. There are no user-serviceable parts
inside. For service, contact an authorized service center.
To determine and evaluate the safety status of the medical system, the system
operator must perform a technical safety check at least every 12 months.
Inspecting the light source
1. Clean the ventilation slots on the back of the light source.
2. Clean the air inlets in the bottom of the light source.
3. Check that the ventilation slots are not covered.
4. Check that the light source is properly connected to the mains supply.
5. If necessary, replace the fuse.
6. Check the function of the system.
Checking the LED operating hours
To check the LED operating hours, the user must access the service mode of
the light source:
1. Simultaneously press both brightness control buttons (- and +) while
switching on the light source.
2. Keep both brightness control buttons pressed until the power indicator starts
flashing. The service mode is active.
The bar graph on the display indicates the LED operating hours. A flashing LED
is equivalent to at least 500 operating hours; a continuously lit LED is equivalent
to at least 1000 operating hours.
The LED can only be replaced by an authorized service center.
The LED must be serviced after 2000 operating hours.
10.2 Replacing the Fuses
Removing the fuse holder
1. Switch off the light source.
2. Remove the power cable from the light source.
3. Remove the fuse holders.
4. Replace the fuses.
The standard set contains two spare fuses (T 3.15 A 250 V).
Reinserting the fuse holder
1. Reinsert the fuse holder.
2. Reconnect the power cable.
3. Switch on the light source.
W7089150
OLYMPUS CLL-V1
Page 22
22
11 Transport and Storage
Shipment
• Use the original cardboard packing for the transport of the product.
If this is not possible, wrap each component individually in sufficient paper or
sheets of foamed material and place them in a cardboard box.
• If applicable, use the respective transport device (e.g., instrument tray,
protective tube, etc.) of the product.
Storage
• Store the light source in a clean and dry condition.
• Store the light source in a well-ventilated place.
• Do not expose the light source to direct sunlight or other sources of ultraviolet
light.
• Do not store the light source in a location where liquids may splash.
• Do not store the light source under environmental conditions such as:
-- high or low temperatures
-- high or low humidity
-- dust
-- salty or sulfurous air
-- elevated levels of ozone
OLYMPUS CLL-V1
W7089150
Page 23
23
12 Repair and Disposal
12.1 Repair
Authorized service center
Repairs may only be performed by qualified servicing personnel that has been
authorized by Olympus. Contact your Olympus representative or an authorized
service center for repair and warranty information.
Warranty claims
Any warranty claims towards Olympus are forfeited if the user or unauthorized
persons attempt repair or modification of the product. Service centers do not
accept warranty claims for damage caused by inadequate packaging.
12.2 Disposal
When disposing of the product or any of its components, follow all applicable
national and local laws and guidelines.
W7089150
OLYMPUS CLL-V1
Page 24
24
13 Compatible Equipment
Use only compatible equipment as listed in this chapter. Olympus recommends
the use of Olympus equipment only. Only this combination ensures a proper
illumination and color reproduction of the endoscopic image.
If combinations are used that are not listed in this chapter, the user takes the full
responsibility.
Future equipment may also be compatible. For more information, contact an
Olympus representative.
Some of the products listed in this chapter may not be available in all sales
territories.
LF-V, ENF-V…, CYF-V…, URF-V…
ENF-…, LF-…, CYF-…,
URF-P…, HYF-XP, CHF-…
WA03200A
WA03210A
CLL-V1
BRIGHTNESS
MIN
MAX
WA97020A
Storz
Wolf
A3213
Wolf
A3214
Circon/ACMI
A3215
ENDOEYE
OLYMPUS CLL-V1
W7089150
Page 25
25
14 Technical Data
14.1 Specifications
Dimensions
Height.............................................................................. 111 mm
Width................................................................................ 295 mm
Depth............................................................................... 404 mm
Weight.........................................................................6.3 ± 0.2 kg
LED
Operating hours of LED.............................................. 2000 hours
Mode of operation........................................................ continuous
Mains connection
Voltage...................................................................... 100 to 240 V
Frequency....................................................................... 50/60 Hz
Power consumption.............................................................55 VA
Fuse...................................................................... T 3.15 A 250 V
14.2 Ambient Conditions
Operating conditions
Temperature.......................................10 to 40 °C (50 to 104 °F)
Relative humidity......................................................... 30 to 85 %
Atmospheric pressure......................................... 700 to 1060 hPa
Storage conditions
Temperature.......................................10 to 40 °C (50 to 104 °F)
Relative humidity......................................................... 30 to 75 %
Atmospheric pressure......................................... 500 to 1060 hPa
Transport conditions
Temperature.................................... -29 to 70 °C (-20 to 158 °F)
Relative humidity......................................................... 10 to 95 %
Atmospheric pressure......................................... 500 to 1060 hPa
14.3 Safety
Protection class according to IEC 60601-1...............................BF
Protection class
against electrical shock......................Class I Medical Equipment
Ingress protection rating........................................................IPX0
W7089150
OLYMPUS CLL-V1
Page 26
26
14.4 Power Cable
Length
WA95621A, WA95622A, WA95623A................................... 4.5 m
Plug type
WA95621A..................................................... Type E/F (CEE 7/7)
WA95622A................................................... Type B (NEMA 5-15)
WA95623A........................................................Type G (BS 1363)
OLYMPUS CLL-V1
W7089150