OLYMPUS CLV-U20
EVIS UNIVERSAL LIGHT SOURCE
t;1J (}Aj
r--------- WARNING
The user of this equipment should be thoroughly trained in the applicable procedure.
Furthermore, failure to read and thoroughly understand the contents of this instruction
manual may result in serious injury to the patient and/or user. It is essential to follow
the instructions contained in this and other manuals which pertain to any equipment
and accessories used in conjunction with this equipment. Possible injuries related
to endoscopic procedures may include electric shock, explosion, burns, perforation,
hemorrhage, etc.
Fai/ure to follow these instructions may also result in damage to and/or malfunction
of this equipment.
CAUTION
Federal (USA) law restricts this device to sale by or on the order of a physician,
OLYMPUS·
III m; ~
fIJJtT
'IJ, f
J I ()() ,) 1-
Page 3
The Olympus CLV~U20 has been designed for endoscopic diagnosis, treatment and photo documentation in
conjunction with Olympus endoscopes, accessories and other ancillary equipments. Do not use the instrument
for any purpose other than its intended application.
Please read this entire manual carefully before using the instrument. It contains pertinent information on the
proper care and handling of your videoscope. Although videoscopes by nature are delicate instruments, proper
handling and cleaning, as described in this manual, will greatly reduce the need for costly repair and maximize
the life of your new instrument.
This manual describes the recommended procedures for preparing and inspecting the equipment prior to use.
It does not describe how an actual procedure is to be performed in detail. Nor does it attempt to acquaint a
beginner with endoscopic technique and the medical aspects of endoscopy. This instrument should be used
only by physicians who have received thorough previous training in the art of endoscopy.
The safety and performance of an endoscopic system depends not only on the endoscope but also on any
ancillary equipment used with it. Safety precautions must be exercised when handling electrical equipment
to prevent operator/patient shock.
If you have any questions concerning the material contained in this manual or concerning the operation or
safety of the equipment, please contact your Olympus representative or the nearest Olympus office.
Page 4
SAFETY SYMBOLS
Replace the Lamp and Filter-2 following the instructions (Section 9-1 Replacement of Lamps;
9-2 Replacement of Filter-2l.
The lamp and heat sinks are extremely hot immediately after use. Allow sufficient time for
them to be cooled before lamp replacement.
Page 5
CONTENTS
1
FEATURES AND MAIN SPECIFICATIONS
1
1-1
1-2
1-3
1
2
4
Features
Main Specifications
Safety Precaution
1
2
DESCRIPTION OF CONTROLS
6
2
3
CONSTRUCTION
9
3
4
STANDARD SET
10
4
PREPARATION FOR USE
11
5
5-1
5-2
5-3
5-4
5-5
5-6
5-7
11
12
13
13
14
15
15
5
6
7
8
9
Installation of the Light Source
Connection to AC Mains Supply..................................................................................
Preparation of the Videoscope
Preparation for Air Feeding/Water Feeding and Suction
Connecting the Endoscope to the Light Source
Preparation for Still Photography................................................................................
Preparation for Cinematography, CCTV and Polaroid Photography........................
16
6
OPERATING THE LIGHT SOURCE
24
7
7-1
7-2
7-3
7-4
7-5
24
25
26
27
27
INSPECTION OF ENDOSCOPIC SYSTEM
Use in Combination with the Videoscope
With Fiberscope
With Rigid Scopes
Cinematography and CCTV
After Use
CARE AND STORAGE
28
8-1
8-2
28
28
Care after Use
Storage
MAINTENANCE
9-1
9-2
9-3
9-4
9-5
Replacement of Lamps
Replacement of Filter·2
Resetting the Circuit Breaker
Cleaning the Ventilation Grills
Periodic Inspection
29
8
9
29
31
31
32
32
10
TROUBLESHOOTING GUIDE
33
10
11
ENDOSCOPIC SYSTEM CHART (EVIS/OES SYSTEMS)
35
11
11-1
11-2
35
37
OES System Chart
EVIS Endoscopic Video Information System
Page 6
1
(j)
z
o
~
U
1. By use of the mode selector, the instrument serves as a possible and universal light source
for OES-fiberscopes, Olympus rigid scopes for EVIS 100 System (instantaneous single-plate
color chip system), and for EVIS 200 System (sequencing system).
L.1.
U
UJ
0-
(j)
Z
;:;:
2'
o
z
<t
(j)
UJ
a::
::::>
f-
<t
UJ
L.1.
2. By use of OES system, EVIS 100 system and EVIS 200 system, the insertion position of the
scope point can be checked by the transillumination function.
3 When using EVIS 100 or EVIS 200 system, by use of the Foot Switch, the anc'illary equipments
can be released.
4. Combined with OES fiberscope, SC16-10 and rigid scope and SC16-10R, the following
advantages can be offered.
• Photographs under stable exposure can be taken in short distances.
• Date is visible in the viewfinder and imprinted on film.
• Remote shutter release using a foot switch frees the endoscopist from finger-tip shutter
operation.
• Indicators in the camera's viewfinder predict exposure level before taking photographs and
the exposure level is given when taking photographs.
5. 1.5 times the flash output of the maximum illumination intensity of the light source is
available.
6. Flat switches and display panel can be easily wiped clean.
7. An emergency lamp turns on instantly if the Xenon lamp burns out, to provide illumination
bright enough for endoscope withdrawal. A light on the front panel indicates at a glance if
the emergency lamp is not installed, is installed improperly or is burned out, or is lighting
instead of the Xenon lamp.
8. Air feeding pressure can be adjusted in three levels.
9. With no scope and even when Xenon lamp illuminates, illumination light automatically dims
to prevent dazzling and the air feeding pump simultaneously stops.
10. Combined with function expansion devices (under development), light source can be more
functionated.
Page 7
]-2 Main Specifications
"m
~
!
Lamp life
! Approx. 300 hours on continuous use' (over 20,000 flashes)
L
Switching
I
' Light output adjustment
1-----"--
Color conversion
Filter-2 is replaceable.
S
~
z
(j)
"
Halogen lamp (without mirror) 24V 150W
-
-~~
-
~
o
z
(j)
-----
Adjustment method (Automatic)
Auto exposure sensitivity Setting (Auto)
(i
----- - - - - - - - - - - - - - -
Approx. 500 hours.
--
Servodiaphragm method
I
,I
17 steps: 1 step 1/4 EV (F; full admission)
Exposure decreases from INDEX NUMBER "Full" to "5".
Auto/Manual; time control by light source shutter
Exposure control
--
---_._--
-------
Exposure light amount
Flash output (approx. 1.5 times visual light output)
Shutter speed
1/4 to 1/100 sec.
i
------- ---- --------
Auto exposure sensitivity setting
1 sec.
_.
- ---------------9 steps: 1 step ~ 1/2 EV. Exposure decreases from INDEX "1"
to "5" (or from "+2" to "-2") with 1/2 EV per step.
Manual exposure sensitivity setting
10 steps: 1 step
Shortest recycling time
----------- ----
_._---~~
F~edlng
Pump pressure (at output socketl
~
-0
m
Q
-
I
Still
Photography
z
o
---- ---- - - - - - - - - -
Emergency lamp life
• Cine- TV
::>
Light-path diaphragm control
Cooling
-
(j)
regu~at_or
Forced-air cooling
I Emergency lamp
• Videoscope
m
I
Ignition
Visual.-observatjon
Xenon short-arc lamp (ozone-free) 300W
Lamp
l-------
Illumination
C
:0
112 EV, except for 1 EV between "F" (Fulll and "1 ".
--- -----------------------
Linear type pump
High:
less than 0.5 kgflcm' (at flow rate 0)
Medium: more than 1.6 £ Imin. (at 0.09 kgf/cm 2 )
0.075 - 0.15 kgf/cm' (at flow rate 0)
Low:
At "HIGH" the pressure is equivalent to that avai!able by the other light
i source than OES/EVIS.
*On intermittent use, the lamp life may differ slightly.
Page 8
(f)
z
o
~
'd
li-
Method
Water Feeding
(for lens wash)
_____ L.. .•
Water container
e;
W
0-
z
<i'
Front Panel
i
~
(f)
w
0:
o
~
W
li-
Indication of Absent lamp, Burn-out or In operation
Setting of filter 1 and 2
,
o
z
-0:
Capacity 250 cc. use of 2/3 full or less.
Emergency lamp
j-----! Filters
(f)
Air pressurization of water container (detachable)
Memorization of selected setting
Selected setting are retained even when the light source is switched
off, by internal battery.
Available types
Videoscope (EVIS 100/200) OES fiberscope, Rigid scopes OES Ultra
thin fibers copes with rigid scope type eyepiece or fibers copes for
ultrasonic survey (GF-UM3, GF-UM20, JF-UM3, CF-UM3), OSF, OGF
--
Compatible
Endoscopes
e-----
-
Compatible
i
I
Cameras
-------
' Olympus medical camera 5C16 series
~C1610,4,3R,3R2, lOR!, SCP-10, OM-1 N/SC35
Available types
Type of protection against electric shock
Classification
(Electro-medical
equipment)
Degree of protection against electric shock
L-!
I
!0!:!<I~'
6peralng
f····
Environments
..
"'
!
I Type
BF (Body Floating). Incorporates insulated patient attachment
(endoscope). Application to the heart directly should not be attempted.
------
The light source should be kept away from the zone of risk of flammable
gases.
oW","" '~"~""""'""
Voltage (AC)
: Power requirements
I
'
i
Ambient temperature
-
Relative humidity
Atmospheric pressure
Class I (3-pin cord)
I
I
110V
Frequency
I
I nput: at observation
at flash
!
120V
40°C (50 -
6A
6A
!
W~ight
240V
-----
5A
7A
I
3A
4A
---
Within ±10%
104°F)
--
i 3070 -
85%
106 kPa (700 -
1060 mbar)
,.~-----,-"
Dimensions &
220V
50/60Hz
Voltage fluctuation
10 -
I
587 (W) X 165 {HI X 537 {DI mm {max.!, 20 kg (44 Ibl
---
Page 9
l_l_-3_S_a--'-rfi_e_ty_P--'-_re_c_a_u_tio_n_s
-----)::===========
."
m
Be sure to connect the power plug
to a hospital grade receptacle (a wall
mains outlet).
~
~\,
,
1\
"
Powe'
PIU:~~
Hospital grade
receptacle
(wall mains outlet)
~
CD The CLV-U20 light source should be used only in a medical facility under the supervision of a
trained physician.
C
:u
m
CIJ
@ The light source is designed (floating) to function integrally with insulated patient attachement
(endoscopes) to prevent operator/patient shock. Endoscopes and cameras to be used with it
should not be grounded in order to prevent shocks which may be caused by leakage current
from other electrical apparatus applied to the patient The endoscopist should wear rubber
gloves as an added precaution.
»
z
o
;;:
:J:':
z
CIJ
® The housing of the light source must be grounded securely and effectively. Do not defect
-0
m
n
the line cord ground connection.
Jl
@) The light source/endoscope combination should never be applied to the heart directly.
n
@ Never install and operate the light source where there is the risk of flammable gases.
o
@ Keep liquids away from all electrical equipment to prevent operator shock and instrument
damage. Do not use the light source if spilled fluids have entered the unit.
CIJ
~
z
Flam",a',le gases
Spillable liquids
Page 10
(f)
z
o
i=
«
u
u:
U
LlJ
0-
(f)
Z
;;:
2'
Cl
(j) The light source/endoscope combination should not be used in conjunction with:
1. Electrical apparatus whose safety against leakage current is not guaranteed.
2. Electrosurgical equipment whose safety in linked usage is not guaranteed.
®
,._----------.---.---------- .... -.. _..... _----------------------.---,
The light source should be used in accordance with the operating conditions described in
this manual; otherwise, improper performance, compromised safety, or instrument damage
may result.
® Repairs should be carried out only by Olympus service personnel.
z
«
(f)
LlJ
Cpmbinedusewith equipment whose
:
safety
not guaranteed.
.,,electrical
._._ •• c. __
•
, _, is
•••.•.••••
~
.__._,
.:
a:
:::>
f-
«
LlJ
Rubber gloves
CL
~
:. ----,----_._------
--- -. -_._--- ,... ..
.~------------------_.
---- ...:
Page 11
2
FRONTPA~
Exposure Sensitivity Selector
Press to set exposure
sensitivity.
Mode Selector
Press to set AUTO or MANUAL
for stili photographs Including
polaroid.
*
Air Regulator
Output Connector
*
Press to adjust airflow pressure.
Transmits light, electrical
signals and air through the
endoscope.
"non".
With a special-purpose filter {optional) for
visual observation and photography, press
"filter-2" to give the illumination of "filter2".
"fllter-2" Is attached with a light
cutting filter {when Installed) and can
be replaced with a special-purpose
filter on request.
With EVIS 200 System, "filter-1"
cannot be used.
AUTO/MANUAL Exposure Selector
Press one of the push-buttons
indicating system names to
select one system.
Filter Switch
Each press to select "filter-1 ", "filter-2" and
";\-'TI~R_
11 1
@@CLV-U20
Z
:rJ
o
Transillumination Switch
AUX
FOOT SW.
Lamp Life Meter
Power Switch
Press the upper portion of the
power switch to turn on power
source and the power switch
indicator turns on simultaneously.
Lamp Ignition Switch
Press to turn on Xenon lamp.
Press to turn 0 N the transillumination indicator.
The light from the endoscope
becomes brighter for 9
seconds and automatically
return.
With EVIS 200 system,
the images on the monitor get over-exposed in
black or white.
The illumination returns
to the original level shortly after pressing the
switch again while using
transillumination.
*
IGNITION
Indicates cumulative working
hours {approx.) of Xenon lamp.
Connector for Rigid
Scope's Camera Cord
Connects with the camera
cord for SC35 and SC16-10R.
Footswith Connector
Connects with a foot release
MB-332.
*
AUTO/MANUAL Brightness Selector
Press to set "AUTO/MANUAL" of
brightness.
For other combinations than CINE/
TV, "MANUAL" should be selected to
avoid thermal mucosal damage as
weI! as to protect your eyes. Select
also the minimum necessary light
level.
With EVIS 100/EVIS 200 Systems, the
selector is automatically set to
"AUTO".
*
*
z
-i
F Ll<::l
10+2
ON
~
o
o
PHOTO
IN EX ADJUS
POWER
:rJ
()
ILL
~
CJ1
()
o
-n
r-----+---W'\--------I%:\--\--+f------4--....
OLYMPUS
o
m
Brightness Control
Press to set CINEnv exposure level (at
"AUTO") or to adjust light level (at
"MANUAL)
r
CJ1
Page 12
I
(/)
--.J
a:
I-
Z
u
Turn on to set air
regulator switch.
z
~
o
t
(/)
UJ
o
OLYMPUS
ONEJ'
OFF
0
\
I
~~~1\2:T1 U
PH
Turn on to indicate
which system is used.
BRIGHT ESS
VISHf-L
MODE
300
':-1 \J
00
~IE~g) cB
c
5""-2
MAN.D
I
/
IGNITION
/
~.
,DAlTO
[AUTO)
MAN
6@AUX
@FOOT SY.
of!~l--1
c;INE'TV
MAX~~
c
oB
Lamp is turned on when "fi1ter-1"
is selected.
1=+2
~
I I
\
0
I~Bt~=A -NST
OUTPUT
==--"
Turn on to indicate the
brightness control is set at
"AUTO".
\\
\
TILL
~ CLV-U20
pOwER
"MANUAL".
A single indicator turns on
in accordance with index
number.
o"a:
u
on to indicate the
brightness control is set at
Turn
Turn on to illuminate for the combination excluding SC35 and OES
fiberscope, and also at "MANUAL"
exposure.
o
o
Multiple LEDs turn on with light level
while at MANUAL. At AUTO, a single
LED indicator turns on with exposure
sensitivity setting.
Turn on to indicate at AUTO
exposure. (SC16-10/0ES fiberscope
SC16-10R/rigid scope combination
only).
DESIGNATION AND FUNCTIOt<j
o~
~2 U U
L- ~ 3
r
~.
MP
9~Ar 0~ t'---
AUT
[ MAN.
I
~
~
I
Filter selection switch
~
Lamp illuminates
when transmitted
illumination is used.
1
Lamp turns on to warn of absent or faulty
Xenon lamp and that the spare lamp
(halogen) is lighting instead of Xenon
lamp.
exposure
indicate at "MANUAL"
"WIM"""
Press AUTO/MANUAL
exposure selector to
indicate at "AUTO"
exposure.
~
!LAf,!P!L1FE
\
selector to
exposure.
Lamp is turned on when "filter-2"
~ is selected.
ILLUM
Cl
o
~
S~RE
c
0\J
MIN~
J.------
I
I
Press brightness selector
to indicate at "AUTO"
brightness control.
Press brightness selector
to indicate at "MANUAL"
brightness control.
Page 13
MAIN BODY
I
Water Container
Receiver
Ratings Plate
o
m
(f)
n
Ventilation Grill
:IJ
::'1
6
z
o-n
Connector for optional
function expansion
devices
Connects to a power
unit for use with
function
expansion
devices (under develop~
ment).
n
•
Ventilation Grill
o
..,z
:IJ
o
r
(f)
Connector for CV-100/200
Connects to light source
cable (MD-434) for CV100 and CV-200.
Handle
Lowered when lamp
access cover is used.
Circuit Breaker
Protects the light source and the
power source in hospital facilities
in an emergency.
Please use circuit breakers
(15A~20A) for
hospital
facilities.
Lamp access cover
Opens to allow replacement of Xenon lamp,
emergency lamp and
filter-2.
*
Power Cord
Cord Hanger
Used for winding the power
cord.
Page 14
3
lEVIS 100/200 SYSTEM
Syslem Diagram for Light Source
I
R.G.B. Filter
Motor
Filter
Output Connector
•
•
r---"
z
t==='~
Power Plug
Xenon lamp
Xenon I
'::==1-rYnt;jr---l~~-
power
Power
switch
supply
I
o
EVIS 100/200
f=
u
::>
ex:
Videoscope
Videoscope
Cable 100/200
I-
I
I
I
I
1 - - - - - DC power
(J)
Z
~-"-;'=_
I
o
u
supply
I
...--- Connection (optional)
I
I
....---Connector for CV-lOO/200
I
CCD
Photodocumentation
Equipment
I
I
I
Light source cable
I
.--- IMD-434)
I
I
I
I
Panel displaY
switch
-~
I OES System I
EVIS Video System Center
Power Plug
CV-1 OO/CV- 200
F=:::::\'\;:::::==~
Keyboard
Isolation Transformer
Page 15
4
CD
Main Body (CLV~U201 .
(including Xenon lamp and emergency lampl
ClJ Water Container
@ Vinyl Dust Cover (MD-4481 .
@ Foot Holder (MD-5121 .
1 set
(MD~431 I.
CD
==~
(fJ
-1
Do not handle rolighly.
J>
Z
o
J>
:0
o
(fJ
m
-1
Place for easy use
at any time.
/
*
Store the instruction manual in the vinyl bag and retain for easy use at any time.
Page 16
5
Refer also to the instruction manuals supplied with the endoscope and photographic attachments.
r-=--:---:--:--:--:---~:c--c::---c-~-c::---------------------------,
! 5-1
Installation of the Light Source
CD Ensure the safety precautions in 1-3, page 4 are met.
..
@ Place the light source on a stable surface in a horizontal position.
Ventilation must not be blocked.
Ensure the stability of the cart when the light source is placed on it.
Do not use the side handle of the light source as a foot support.
Please refer also to the instruction manuals for TC-Vl, CV-l00 and CV-200 when
setting to TC-Vl (Compact Video Trolleyl.
When combined with CV-l00/200 and always set the light source below the video
processor.
*
*
*
*
*
LlJ
(J)
:::>
a::
o
u.
Place the light source
on astable base.
{Prevention of overheating]
Z
o
@ Always set the light source using the attached foot holder (MD-5l21 when placed on sliding
surface or in case the light source moves while in use.
~
a::
:;:
LlJ
a::
"D
Ventilation must not be blocked.
Please refer to each instruction 0
manual for products.
place the light source in
a horizontal position.
Use the foot holders for stable
placement of the light source.
i
Page 17
1_5_-_2_C_o_n_n_ec_ti_·o_n_f_o_A_eC
__
.M_az_·n_s_eS_eu_'P_'P_ly
~
~):===========
CD Make sure the power switch is in the off position.
@ Connect the power cord properly to the "HOSPITAL GRADE" receptacle or the wall mains
outlet that meets the input requirements indicated on the ratings plate on the rear panel of
the light source.
Power switch should be in the "OFF"
position.
Always connect the power cord
properly to the "HOSPITAL
GRADE" receptacle or the wall
mains outlet.
@ The power-cord plug must be connected to a grounded hospital grade outlet or a wall rna-Ins
outlet 13-cord outletl.
Do not use converter adapter from three pin power plug to two pin power plug for
safety precautions.
Connect the light source directly to hospital grade receptable or wall mains outlet
without connecting to table tap for safety.
Firmly set power cord during inspection. Do not apply excess force to power cord by
bending, pulling, twisting and pressing.
When combined with TC-V1 (compact video trolley), do not connect the power cord
of light source to the secondary outlet of TC-V1 to prevent malfunction due to the
subsequent power voltage lowering. Be sure to connect the power cord directly to
hospital grade receptacle.
Refer also to the instruction manuals for combined instruments being used.
*
*
*
*
The converter adapter from three pin plug
to two pin plug cannot be safely grounded
and electrical shock
may
v
m
"
~
":to;
is
z
o
"
"
c
en
m
*
Ratings plate
~-J®
~
Hospital grade receptacle
Power plug
{wall mains outlet)
[Shock prevention]
Do not connect the power cord for the light
source to the outlet for TC-V1.
Page 18
5-3 Preparation of the Videoscope
Select
EVI~
100 or EVIS 200.
I'" !,i;g~I_~'" :"'~r,,:~'i.!~
',8' ,-"
I
"
I"
-":"'. J~IJ~IJ,~~
-1-"",l:J
1§]~['J:lJ1
I
w
(f)
::J
a:
o
CL
z
o
~
.:g;"o'
-8i "'"
@.QQT'W.
' """ J""">e' I
'---J
I n CV-100(200
~CV-100/200connecto~
III~ Light source connE!c~,or
J~
=
I! I
"""
a:
rnU
Light source
<>:
0-
Refer to the instruction manuals of the Vjdeoscope, Compace Video Trolley, and EVIS Video
System Center ICV-l00/CV-2001 in use.
CD Prepare the EVIS Video System following the instructions found In the manuals of the EVIS
Videos cope, Compact Video Trolley, and EVIS Video System Center CV-l00/CV-200,
@ Ensure that the power switch is 'OFF' and then connect the light source cable IMD-434)
option) between the 'light source' connector on the back of the CV-100/200, and the 'CV100/200' connector on the back of the light source.
@ Use the mode switch and select the EVIS Video System to be used:
When using the EVIS 100, select ~.
When using the EVIS 200, select ~.
* When EVIS 100 or 200 is selected, photo sensitivity will not be indicated,
c~
w
a:
0-
5-4 Preparati,Qn forAir Feeding/Water Feeding and Suction
CD When water feeding is needed, fill the container approximately 2/3 full with clean water,
Tighten the cap securely and then hand the container on the container receiver.
Change the water daily,
Use only clean water (distilled or sterile) to prevent clogging of the water channel.
If a poor connection with the scope has caused water to spill onto the equipment,
wipe it completely dry with a soft cloth, etc.
*
*
*
2/3
T
CV Connect the feed tube of the water container to the connector on the scope.
-L
@ If suction is needed, connect the tube from the suction pump with the scope suction
connector.
Check the suction pump carefully for leakage and electrical safety.
*: Use only electrically insulated suction tube.
*
Water feed tube
Suction tube
I
Page 19
5-5 Connecting the Endoscope to the Light Source
II
Vide,osCO!lle Cc,nne[:tion
CD Turn OFF the light source.
(2) Turn OFF the EVIS Video System Center ICV-l00/2001.
@ Push the scope into the scope socket until a click is felt.
@ Attach the videoscope cable to the electical connector.
Videoscope cable
1. Align the dots on the cable and the scope, and push together firmly.
2. Rotate the ring of the videoscope cable while pushing the cable and the scope together
until a click is felt.
Be sure that the power switch of the EVIS Video System Center remains 'OFF' when
attaching or removing the videoscope cable. The CCO may be destroyed if power is
left 'ON' during attachment.
"m
:D
*
~
:D
~
* To avoid damage to the ceo, do not touch the electrical contacts inside the electrical
o
z
connector.
":Do
C
(f)
CD Securely connect the scope into the scope socket until a click is felt.
[Connection of videoscopeJ
I ~~~&~~~
or' fiberscope lfor ultrasonil:. surVE
CD Securely connect the scope into the scope socket until a click is felt.
II
Connection
CD Securely connect the light guide of the rigid scope into the scope socket.
[Connection of fiberscope]
Do not attach or detach the videoscope
cable with power ON.
m
Page 20
I 5-6 Preparation for Still Photography
Refer also to the instruction manuals for endoscopes and photodocumentation equipment to
be used with the light source.
CD Refer to the following table.
Endoscope
f--
Fiberscope
Camera
X • FP contact
SC16-10
-
SC16-4
I--
N
w
(J)
Camera cord
SC35
::J
0:
o"-
SC16-lOR
[Camera cord connection]
I
!
X
Film
Shutter speed
-
No. 1610-0
-
No. 1604-0
Ektachrom !
(ISO/ASA 200)
1/4 sec.
-
-
Filter switch
I
No. 1610-0
IS
z
o
Rigid
~
0:
«
Il.
SC16-3R
SC16-RR2
X
OM-1N
SC35
X
-
OFF
(Filter-1/2
indicator
not lighted.)
No. 1604-0
f
w
0:
1/4 sec.
:
Ektachrom
IISO/ASA 400)
Il.
CD For rigid scope, camera cord (EC-26, 27 or 28) is plugged into rigid scope's camera connector
of light source.
(2) Foot release is plugged into the connector for foot release when necessary.
This application is only possible for the combination of OES fiberscope/SC16-10 and
rigid scope/SC16-10R.
*
15-7
Preparation for Cinematography,
ccrv and l'olaroid® Photography
Refer also to the instruction manuals for endoscopes, photography devices and OES TV
system (OTV-F2/S2) to be used with the light source.
CD For cinematography, refer to the endoscope instruction manual.
(2) For cinematography, prepare cine camera with a C-mount adapter (e.g. Baulieu 16R) and
daylight type film (lSO/ASA 160).
® For CCTV, refer to the OES TV System instruction manual.
@ For polaroid® photography, refer to the instruction manuals for photography instruments.
SCP-lO film is a high speed color film (TYPE 779 ISO/ASA 640).
Page 21
6
Before each operation, the equipment should be inspected according to the following
procedure. Should the slightest irregularity or abnormality be suspected, do not use the
equipment; contact the Olympus service center.
D
{~~mJuwlamp~~ro~n.J
Turn ON the power switch: The power indicator and LED (Light-Emitting Diode) indicators on
the Iront panel light up and the cooling Ian will be heard.
If the light source does not show any sign of being activated, make sure the lamp
access cover is closed tightly. Then inspect the circuit breaker on the rear panel of
the light source accordihg to 9-3 Resetting the Circuit Breaker, page 30.
O~YM"U5
OCLV_U'O
*
Power switch
[Connection to power supply]
II
CD
Igniting the I ;,,~.~
Lamp access cover
Lamp access cover should
be. closed tightly,
{Caution when spare lamp is on.)
Make sure the pointer in the lamp life meter is positioned in the green zone.
If the pointer has reached the red zone, replace the lamp with a new one according
to 9-1 Replacement 01 Lamps, page 28.
*
Xenon lamp indicator
@ Press the Lamp Ignition switch and make sure the illumination light is transmitted through
the endoscope.
* If the Xenon
[Inspection of lamp-life meter]
lamp fails to ignite, the emergency lamp will automatically turn on,
along with the emergency lamp indicator (red) on the front panel. In such a case,
turn off the power switch and repeat Steps IIJ and B above. lithe Xenon lamp still
fails to ignite, replace the lamp with a new one.
* If
the emergency lamp indicator is lighted, replace lamp(s) referring to 10
Troubleshooting Guide, page 32.
Ol.VMf'VS
r"'CLV-U20
z
i
Xenon lamp is defective.
(f)
"m
':1
6
z
o
"
I~~~
i @.ux
Ilgnition
of Xenon lamp
[Lighting of lamp]
m
Z
o
o(f)
(")
o
n"
(f)
-<
(f)
-4
m
Endoscope
distal end
/
Illumination
(~ht
'5:
Page 22
Inspection of turning on
and off alternatively
D
InsJlecti<1O
CD EVIS 100/200 System is automatically set to "AUTO". Check the CINE. TV and the AUTO
indicators being lit.
@ For OES System, press the AUTO/MANUAL Brightness Selector. At each pressing, a peep
sounds and AUTO/MANUAL is switched alternately. Check the CINE. TV indicator at AUTO
and the Illumination Light at MANUAL being illuminated and then is set to MANUAL.
If the CINE ·TV switch is pressed without connecting to TV camera, AUTO/MANUAL
Brightness Selector can function and the illumination light output may be minimized.
*
Q) Press the Brightness Control switch ,,,,, Ii",). At each pressing, a peep sounds and the LED
light goes up (downl one step with increased (decreasedl illumination light output.
Keep the switch depressed: the peep sounds continuously, and the LED light goes up
(down) sequentially with increasing (decreasing) light output until "MAX" ("MIN") is reached.
Once "MAX" ("MIN") has been reached, pressing the switch
) dose not cause
any further change.
If the endoscope is disconnected, the LED indicator light will still change as described
above, but the actual output brightness remains at a low level until the endoscope is
*
*
Ell
:;e
connected.
* The
Ll.J
switch should not be pressed with a pointed or hard object since this will
damage the rubber covering.
f0Ul
>Ul
u
0:
o
u
Ul
o
o
Z
CD Dip the fiberscope distal end in clean water.
lAdjustment of output intensity]
(2) Press the air Regulator switch: A peep sounds, and the LED indicators turn on sequentially.
Air flow rate (identifiable by bubble amounts when the air/water valve on the fiberscope is
occluded, or by pump noise) varies with each setting.
The air pump does not operate when the endoscope is disconnected.
Ll.J
LL
o
Z
o
*
B
Q) Set the pressure level to "HIGH" by pressing the switch. Make sure the water is emitted
through the distal end nozzle.
Ll.J
CL
Ul
Z
[Adjustment of air feeding pressure)
Do not push the button with an edged
object.
Page 23
EVIS Video System
CD
Connect the videos cope and the EVIS Video System Center by following their respective
instruction manuals.
Q) Use the mode selector to select the EVIS Video System to be used with.
@ The light control will display 'AUTO' automatically.
@) Test to see if the light coming from the light guide fiber at the distal end of the scope
becomes stronger, in relation to the distance between the scope and a viewed object. Also
check to see that the light from the fiber varies when the CINEfTV value is adjusted between
1 and 5 {with the distance to the viewed object held constant}.
Increase the distance
gradually.
~
CD
OES System
~~
yv'b
v
5-60mm
Connect the videoscope and turn on the lamp.
Q) Press the Transilliumination Switch, and confirm that the light becomes brighter, and then
returns to the original leve! approximately nine seconds later.
With EVIS 200 system, monitor images will become over-exposed in black and white.
When pressing the Transillumination Switch under transillumination, the light returns
to the origianl level.
[Inspection of the brightness!
*
*
z
(fJ
g"
m
i
[Mounting the camera]
Check the indicator being
illuminated alternately.
The illumination light
gets brighter and is
automatically reset.
CD Attach an unloaded camera to the scope eyepiece by following the instruction with the
scope and camera.
o
z
o
"m
Z
o
o(fJ
@ Test the operation of the ~ exposure control switch (a 'peep' should be heard), and then
set it to 'AUTO'.
When the mode switch is set to 1~6~1 or ~ , a flashing will be seen in the SC16-101
10R camera viewfinder, and photography will be impossible. When using a camera,
the mode switch must be set to~.
CO
o
*
"n
(fJ
-<
(fJ
-4
m
;;;:
[Transmitted illumination]
Page 24
@ Selection of the photo sensitivity is made by referring to the endoscope and camera
combinations listed in the following table.
Illuminated
Endoscope
"..,
§j~~l
@,.,
:~~~i;r::£
": jO,:I: \7-;;
17
Rigid
scope
OES
Camera
I
- -
:~ -~ . "'~'"" ~
1
@",m.
0
J~
Repeat pressing exposure sensitivity selector:
ADJUST (compensation) LEDs turn on sequentially.
(Set the value desired: normally to "0".)
Shortly after pressing the selector, the exposure
sensitivity index will be illuminated. This is a
normal phenomenon.
When set to "+" indices, the exposure light
SC16-10
Auto exposure
(5C16-10 + QES fiberscope)
0
*
*
SC16-lOR
A
U
T
0
-
Illuminated
Ell
SC16-4
SCP-l0
OM-1N
SC35
0
2:
LU
--
fUl
>u
0::
o
i
I
Z
i
e-
Ul
o
0
I
u
LU
,
Auto exposure
(a camera other than 5C16-10 + DES
or non DES scope)
u..
o
z
o
0
Illuminated
E;
LU
"Z
0
SC16-3R
SC16-3R2
OM-1N
SC35
intensity becomes increased.
When set to "-" indices, the intensity becomes
decreased.
Repeat pressing exposure sensitivity selector:
INDEX LEDs turn on sequentially.
(Set the value desired.)
The smaller the index number, the greater
the amount of flash.
*
I
!
Ul
o
Inspection (Setting)
I
,
!
M
Repeat pressing exposure sensitivity selector:
A !
I INDEX LED2 turn on sequentially_
N
F:= maximum exposure
U
For each step increase of the index numA
ber, exposure decreases 50% of the previL
ous step.
*
Ul
* When the camera is not mounted on the endoscope, the "INDEX" indicators always
turn on.
the SC16-10/0ES fiberscope combination: the "INDEX" indicators turn on
momentarily, before the ADJUST indicators turn on: this is a normal phenomenon.
If the "INDEX" indicators turn on (instead of the ADJUST indicators) in the middle of
the procedure, check the individual components for correct connection.
* With
Manual exposure
Page 25
{Shutter release inhibition}
Illuminated
...}~~':J'TV ~
"6 0
j:\7
~
"'""'0
Set to minimum light output
(j) Set the light level control switch to 'MANUAL', and adjust the light level to the minimum
position (dark).
@ Place the distal end of the scope near to a suitable object, and press the camera shutter or
foot switch release. The illumination should become strong for a moment, and then return to
the original state. A buzzer should also sound.
Do not press the shutter release while the buzzer is sounding, as uneven or no
exposure may result.
If nothing happens when the shutter release is pressed, disconnect all the connections,
and re-connect them securely.
*
*
Do not activate shutter release while buzzer
is sounding
[Setting of illumination light]
CD Verify that the length of the intensification of the illumination when the camera shutter
release is pressed varies with the distance, by testing with the distal end held at varying
distances (between 5 and 60 mm) from the view object.
@ Ensure that the length of the light emitting time becomes shorter when the sensitivity
switch is altered from +2 to -2 (or 1 to 5). and the distance is held constant.
If using combinations other than the OES fiberscopes and SC16-10, and rigid scopes
and the SC16-10R, images taken at short distances may be overexposed. If this
occurs, turn the filter switch 1 'ON'.
*
z
(f)
'1J
m
(')
:::!
o
z
o-n
m
Z
[Check of the amount of flash]
o
o
(f)
(')
o'1J
n
(f)
-<
(f)
-l
m
;;
ga
5-60mm
[Inspection of auto exposure]
Page 26
Set exposure sensitivity selector at
"AUTO".
@ If auto exposure is desired, set the exposure control switchs to 'AUTO' and use the settings listed
here.
Camera
Endoscop~
.
SC16-10
SC16-4! OM-1N/SC35-l
SC-P--1-0~~-S~C-16-10R[SC16-3~/3R2
OESscop~ [~~I-lt-y-P_~-. _s~~:~~A~D~J_O-U_-S~T~~_C~~_IN_~_E_X_-,-_IND3_EX_*_-+_IN_~_EX
Rigid
Scop~
I
*OM-1N/SC35:
**SC16-3R/3R2:
-I<
If------'-----I
ADJUST
INDEX
3
S~t
S~t
o
UJ
f-
(/)
>(/)
u
The OM-1N/SC35 settings refer to use with the A10-M2 IOES scope) only.
0::
o
u
(/)
o
o
z
UJ
LL
o
Z
o
E
UJ
CL
(/)
Z
minus side (sensitivity: larger number). Underexposures must be corrected on the plus
side (sensitivity: smaller number).
® If manual exposures are desired, set the exposure control switch to 'MANAL', and set the sensitivity
constant as needed.
INDEX**
3
sync contact to X and 1/4
sync contact to X.
* The above table gives average values, but if the pictures are overexposed, correct to the
2:
I
s~c.