OrthoScan
OrthoScan-HD Flat Detector Service Manual
Service Manual
64 Pages
Preview
Page 1
OrthoScan-HD Flat Detector Mini C-Arm Imaging System
Service Manual [email protected] Phone: 480-503-8010 fax: 480-503-8011 www.orthoscan.com 8212 East Evans Road Scottsdale, AZ 85260 PROPRIETARY AND CONFIDENTIAL – The information contained in this document is the sole property of OrthoScan, Inc. Any reproduction in part or as a whole without the express written permission of OrthoScan, Inc. is prohibited. PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 1 of 64
TABLE OF CONTENTS Introduction ... 1 Product Specifications and Components ... 2 ORTHOSCAN HD FLAT DETECTOR PRODUCT SPECIFICATIONS ... 2.1 ORTHOSCAN MAJOR COMPONENTS ... 2.2 FLEX ARM COMPONENTS AND FOCAL SPOT LOCATION DIAGRAM ... 2.3 ELECTRONIC CHASSIS COMPONENTS ... 2.4
Removal/Replacement Instructions... 3 EMI PLATE REMOVAL ... 3.1 ELECTRICAL CHASSIS REMOVAL ... 3.2 VIDEO CCA REMOVAL ... 3.3 MONITOR KEYBOARD ASSEMBLY, REMOVAL AND ADJUSTMENT ... 3.4 MONITOR ARM REMOVAL AND ADJUSTMENT ... 3.5 PRINTER REMOVAL ... 3.6 INFRARED FOOTSWITCH CONNECTIONS AND INFORMATION ... 3.7 X-RAY TUBE (HVPS) ... 3.8 COLLIMATOR ALIGNMENT ... 3.9 GENERATOR TESTING ... 3.10 FLAT DETECTOR REPLACEMENT AND COVER REMOVAL ... 3.11 FLAT DETECTOR POWER SUPPLY ... 3.12 SYSTEM SETTINGS AND CALIBRATIONS ... 3.13 FLEX ARM TENSION ADJUSTMENTS ... 3.14 Laser Alignment ... 3.15
System Set-Up ... 4 DICOM SETUP ... 4.1 STORAGE PROVIDER SETUP ... 4.2 WORKLIST PROVIDER SETUP ... 4.3 DICOM PRINT PROVIDER SETUP... 4.4
System QC Documentation ... 5 RADIATION QC DOCUMENT... 5.1 RADIATION QC PROCEDURE ... 5.2
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 2 of 64
1 Introduction
Purpose This service manual provides service information for various major components and sub-assemblies of the OrthoScan-HD Flat Detector mini c-arm imaging system.
Scope This document, diagrams, and schematics contained herein are written for technical personnel, such as biomedical engineers.
Precautions Repair, maintenance, calibration or other types of service should be carried out by suitably qualified personnel. OrthoScan provides service directly. Service by unauthorized service personnel will VOID any remaining warranty.
ELECTROMAGNETIC COMPATIBILITY This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the manufacturer’s instructions, may cause or receive interference. If this equipment seems to cause or receive interference when turned on (which may be determined by turning equipment off and on), try to correct the interference by one or more of the following steps: relocate the equipment with respect to other equipment in the room, or plug the system into a different outlet so that the circuit is not shared with the other equipment, or use other measures, as required. Call the phone number listed on the cover of this manual if you have questions concerning electrical power for the system. Changes or modifications to this equipment not expressly approved by the agency responsible for compliance could void the authority of the owner/user to operate the equipment. If any modification is made, the owner/user is responsible for compliance of the equipment.
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 3 of 64
2 Product Specifications and Components
2.1 OrthoScan-HD Flat Detector Product Specifications NOTE: Product design and specifications are subject to change without notice. DEVICE CLASSIFICATIONS Equipment Classification (IEC 60601)
Class 1 Device
Water ingress protection classification
Ordinary Equipment
Mode of operation
Continuous Operation
Electric shock protection classification
Type B applied part Not suitable for use with flammable anesthetics
HD Mini C-arm X-ray System
Class II (FDA) / Class IIb (EU) / Class III (Canada)
POWER REQUIREMENTS Nominal rated line voltage
100 / 120 / 230 V~ (factory preset)
Frequency
50 Hz or 60 Hz
Line voltage regulation
+ 10% daily line reg.
Current
< 4A at 120 V~, ≤5A at 100 V~, ≤2A at 230 V~ 100 V~: Bel Fuse 5ET 5 or Littelfuse 0213 0005(P) 120 V~: Bel Fuse 5ET 4 or Littlefuse 0213 0004(P) 230 V~: Bel Fuse 5ET 2 or Littlefuse 0213 0002(P)
Mains Fuses LASER SPECIFICATION Laser Classification
Class 2 (IEC 60825-1)
Max Power Output
3.5mWatts
Wavelength
635 nm
X-RAY SPECIFICATIONS Duty cycle at 75 kVp / 0.1mA
Continuous duty
X-ray source
Superior SXR-80
X-ray tube window
0.75 mm Beryllium
Beam filtration
2.5 mm Aluminum equivalent
Focal spot
50 microns
Field of view
Operator selectable 11cm x 8cm (Magnification Mode) or 15cm x 12cm (Normal Mode)
Rated peak tube potential
80 kVp
Tube kVp range
40kVp to 75kVp (Boost Mode: 40kVp to 78kVp)
Tube current range
40 to 100 µA (Boost Mode: 40µA to 160µA)
Deviation from indicated values
kV + 7%, µA + 10%
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 4 of 64
2 Product Specifications and Components ENVIRONMENTAL SPECIFICATIONS Storage Conditions
Temperature Range 32˚F (0˚C) to 131˚F (55˚C) Relative Humidity of >10% to < 100% (no condensation) Atmospheric Pressure of 500hPa to 1060hPa
Operating Temperature (Ambient)
15º to 32ºC (59º to 90ºF)
Operating Humidity (Ambient)
20% – 80% relative humidity, non-condensing
IMAGING SPECIFICATIONS Image Receptor
Complementary Metal Oxide Semiconductor (CMOS) Flat Detector
Image Processing
Digital
Monitor
Diagonal viewing size monochrome LCD
Video System
1.5k by 2k CMOS receptor
Video Output Signal
VGA & DVI
WEIGHTS AND DIMENSIONS Source to Detector Distance (SID)
45 cm (17.72 in)
Source to Skin Distance (SSD)
10 cm (3.94 in)
C-arm Free Space
35 cm (14 in)
C-arm Depth of Arc
49 cm (19.2 in)
Height
147 cm (58 in)
Width
79 cm (31 in)
Weight
180 kg (400 lb)
Shipping Weight
194 kg (430 lb)
Ground Clearance
18 cm (7 in)
Power Cord Length
610 cm (20 ft)
Footprint
79 cm x 101 cm (40 in x 31 in)
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 5 of 64
2 Product Specifications and Components
2.2 OrthoScan Major Components The major components of the OrthoScan-HD Flat Detector mini c-arm imaging system are shown below: G
H
I E
A C
F J N K L
B
D
D M
A B C D E F G
X-Ray Source Assembly Flat Detector Collimator Orbital Lock Flex Arm Keyboard X-Ray On Light
H I J K L M N
Flat-Screen Monitor Power Switch Printer (Optional) Retractable Power Cable Back Panel Connections Brake Cabinet
Back Panel Connections A B C D
A
B
C
Ground Probe Ethernet Port for PACS VGA Output Fuses
D
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 6 of 64
2 Product Specifications and Components
2.3 Flex Arm Components and Focal Spot Location Diagram Flex arm is comprised of the wrist joint, forearm, elbow joint and shoulder as depicted below.
Wrist
Forearm
Elbow Shoulder
Focal Spot Location Diagram
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 7 of 64
2 Product Specifications and Components
2.4 Electronic Chassis Components 915-0013-03 Video CCA
A
910-0010-04 Video CCA
A
B
C E
B
A- Video CCA B- Solid state hard drive (902-0010-20x) C- Wireless footswitch CCA (240-0054) if equipped D- EMI filter (230-0013) E- Power control board (910-0009)
D
Common to both Video CCA’s
Chassis Front- with Video CCA removed
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 8 of 64
2 Product Specifications and Components
Chassis rear view
Common to both Video CCA’s
I
H
G
F
F- Torroid isolation transformer (230-0011) G- 12v power supply (260-0035) H- 6v power supply (260-0033) I- 5v/24v power supply (260-0032)
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 9 of 64
3 Removal/Replacement Instructions
Caution! Ensure unit is unplugged prior to servicing! Service personnel must be properly grounded while working with and around sensitive electrical components.
3.1 EMI Plate Removal Tools needed: Ratchet 3/8” socket 1/4 ” nut driver Small Phillips screwdriver Large Phillips screwdriver Nut Driver 1. Remove back panel by loosening the 4 spring loaded screws 2. Remove Monitor connections as depicted below. (See Section 3.6 for printer removal)
USB Panel Mount
Power Switch
USB to Monitor Hub Monitor Power
HDMI
3. Remove EMI shield by loosening 2 self tapping screws (401-0940) using the 3/8” ratchet
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 10 of 64
3 Removal/Replacement Instructions
4. Disconnect HDMI cable from pass through on EMI plate
915-0013-03 Video CCA
OR
HDMI Pass Through
910-0010-04 Video CCA
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 11 of 64
3 Removal/Replacement Instructions
5. Disconnect 4-position power switch receptacle connector and 2-pos monitor power connection Monitor Power Pass Through Connector
Common to both Video CCA’s Power Switch Pass Through Connector
6. Disconnect three USB cables from location on Video CCA 3 USB Connections
915-0013-03 Video CCA
910-0010-04 Video CCA
OR
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 12 of 64
3 Removal/Replacement Instructions
7. Disconnect the three printer power cables from the 920-0063 cable to the 250-0058 Black Outlet Connector. *Brown = Live Blue = Neutral Grn/Yellow stripe = Ground
Printer Power Supply Connectors
Common to both Video CCA’s
8. Disconnect green & black 2-pos X-Ray light connector from “J2” location on Video CCA
915-0013-03 Video CCA
910-0010-04 Video CCA
OR
9. The EMI plate can be removed and set aside.
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 13 of 64
3 Removal/Replacement Instructions
3.2 Electrical Chassis Removal Caution! Ensure unit is unplugged prior to servicing! Service personnel must be properly grounded while working with and around sensitive electrical components.
Tools: Ratchet 10mm socket 3/8” socket 1/4” nut driver Small Phillips head screwdriver Large Phillips head screwdriver 10mm Nut Driver 7/16” open end wrench 1. If possible login to the OrthoScan System Configuration and record all settings in the ‘ABS/Mask Setup’, ‘MagMode Setup’, ‘System Configuration’, ‘Storage Provider Setup’, ‘Worklist Provider Setup’, and ‘Print Provider Setup’ tabs, prior to removing chassis. 2. Remove the EMI plate as described in section 3.1 3. Refer to section 3.13 for masking procedures if installing new chassis. 4. Check to see if unit is equipped with a wireless receiver. If present disconnect the wireless footswitch sensor cable. When reassembling make sure pins are aligned.
Wireless Footswitch Sensor Cable
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 14 of 64
3 Removal/Replacement Instructions
5. Disconnect ground wires Ground Wires
Common to both Video CCA’s
6. Remove cord reel with three bolts using a 10mm wrench (white arrows) and swing cord reel out of the way.
Common to both Video CCA’s
7. Remove the four 7/16” nuts that hold the cord reel mount.
Common to both Video CCA’s
* 910-0010-04 board pictured. Procedure is the same for 915-0013-03 board. PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 15 of 64
3 Removal/Replacement Instructions
7. Disconnect all cable connections indicated in the pictures below.
915-0013-03 Video CCA
Flat Detector Cable
HVPS Power Connection
PROPRIETARY AND CONFIDENTIAL
HVPS Signal Cables
920-0086 Cable Connect to Shutdown
050-0013 Rev B Page 16 of 64
3 Removal/Replacement Instructions
915-0013-03 Video CCA (continued) Tube Head Fluoro Switch
Wireless Print/Save
Wired Footswitch
Wireless Fluoro
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 17 of 64
3 Removal/Replacement Instructions
915-0013-03 Video CCA (continued) DVI Cables
5 position Red Connector
CAT6 Cable
Fan Power Cables
Serial/Tube Head Controls
Flat Detector Power Cable
Proceed to step 8 PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 18 of 64
3 Removal/Replacement Instructions
910-0010-04 Video CCA
Flat Detector Cable
VGA cable
HVPS Signal Cables
Tube Head Fluoro Switch
Wireless Print/Save Wireless Fluoro
Smart Shutdown Cable Wired Footswitch
PROPRIETARY AND CONFIDENTIAL
050-0013 Rev B Page 19 of 64
3 Removal/Replacement Instructions
910-0010-04 Video CCA (continued) HVPS Power Connection
Serial/Tube Head Controls
5 position Red Connector Disconnect fan power cables.
CAT6 Connector
PROPRIETARY AND CONFIDENTIAL
Flat Detector Power Cable
050-0013 Rev B Page 20 of 64