SERVICE MANUAL LFT5508 Issue 3
VIACLIN Alternating Pressure Mattress Overlay Model 2502
LFT5508 - AMENDMENT RECORD
REVISION
CHANGE NOTE
TECH. NOTE
1 2 3
025/02
011/02
023/02
-
-
-
-
-
BRIEF DETAILS
SIGNATURE
DATE
To have a common flow restrictor assembly for 2503 & 5202 Replace part number MEC483 with MEC14753
K.Wilson
3/9/02
K.Wilson
16/9/02
K.Wilson
3/10/02
K.Wilson
22/10/02
Redrawn instruction diagrams to show overlay straps around mattress. Reformated Viaclin service manual 2502 so that it is similar in structure as the newer manuals. Changed text ‘5.5 ltr/min’ to ‘6.5 ltr/min’. Added note”EXISTING 2501 OR 2502 THAT REQUIRE A REPLACEMENT POWERUNIT WILL BE REPLACED BY A 2503 POWER UNIT (SA242) AND AN APPROPRIATE ADAPTER (REF: MOD KIT MOD14831)”. See IU 088/02
POLICY STATEMENT Pegasus Ltd has a policy of continuous product improvement and reserves the right to change details presented in this Service Manual without notice.
Tel: +44 (0) 23 9278 4200 Fax:: +44 (0) 23 9278 4250 E-mail: [email protected] Web Site: www.pegasus-uk.com
LIST OF CONTENTS Section
Title
1.
SAFETY STATEMENT
2.
INTRODUCTION General System Description
3.
SETTING UP/OPERATION Setting Up Profiling and Knee-Breaking Moving the Bed and Power Cuts CPR (Cardio Pulmonary Resuscitation) Deflation, Removal, Carrying and Storage Alarms and Fault Finding
4.
CONSTRUCTION OF PRODUCT Major Elements of the System Mattress Power Unit Technical Details
5.
INFECTION CONTROL AND CLEANING Infection Control Cleaning Guidelines
6.
EQUIPMENT MAINTENANCE General Power Unit Mattress
7.
TEST EQUIPMENT
8.
PARTS LIST Power Unit Mattress
LTF5508 VIACLIN MODEL 2502
i
Issue 3 Oct 02
SECTION 1 - SAFETY STATEMENT 1.1.
The VIACLIN system must be used in accordance with the manufacturer's instructions.
1.2.
The mains electricity supply within the building where the VIACLIN system is to be used must comply with IEE Regulations.
1.3.
Only personnel trained or formally approved by Pegasus Ltd in operation and maintenance of Pegasus systems may perform maintenance, modification or repair work on Pegasus Ltd Power Control Units.
1.4.
Unqualified personnel attempting to work on Pegasus Ltd Power Control Units risk serious injury to themselves and others, and possibly death by electrocution. Do not work, or attempt to work on Pegasus Power Units unless you are properly qualified to do so.
1.5.
Pegasus Ltd systems are designed to comply with all relevant electrical safety, manufacturing and performance standards published by IS0, IEE & BSI.
1.6
Materials used for cleaning procedures may be subject to COSHH regulations - manufacturer's Instructions for Use must be followed at all times.
1.7
Materials used for maintenance and repair may be subject to COSHH regulations - manufacturer's Instructions for Use must be followed at all times.
LFT5508 VIACLIN MODEL 2502
1.1
Issue 3 Oct 02
SECTION 2 - INTRODUCTION 2.1
GENERAL
The VIACLIN Alternating Pressure Mattress Overlay is designed for use whenever patients have some mobility problems and/or who are vulnerable to pressure ulcer development and/or who have existing tissue damage. It gives gentle, dynamic support to bedridden and immobile people in recumbent, supine and knee-break postures, providing alternatively support then pressure relief to each part of the underside of the patient. The bed can be repositioned whilst the Mattress Overlay is still inflated. The system is suitable for use in hospitals, nursing homes and in the community.
2.2
SYSTEM DESCRIPTION
The system consists of a power unit and a mattress overlay. Audio and visual alarms activate if a fault condition exists for more than three minutes (compressor failure, overlay or airpipe failure). An audio alarm sounds if the mains supply fails (disconnected or switched off). These alarms alerts medical staff to the malfunction. The specially constructed overlay comprises a series of pairs of transverse cells, which are inflated and deflated alternately. Air to inflate the overlay is provided by an electronically controlled power unit. The air pressure in the overlay is continuously monitored, and the system is completely self-regulating. The overlay is connected to the power unit by means of an airpipe which features quick release connections at the power unit end, for convenience and rapid deflation, to allow cardiac resuscitation procedures. Operator controls are purposely kept to a minimum. The overlay is positioned on top of the normal bed mattress.
LFT5508 VIACLIN MODEL 2502
2.1
Issue 3 Oct 02
SECTION 3 - SETTING UP/OPERATION 3.1
SETTING UP
CAUTION:
Use only on top of the existing bed mattress.
Remove the bedding from the bed.
Ensure there are no sharp objects to damage the Overlay. Place the Overlay on top of the bed mattress with the umbilical at the foot end of the bed. Tuck the two end flaps under the bed mattress.
Secure the Overlay straps to the mattress. Do not secure to the bed frame (only secure Overlay straps loosely to the moving parts of the bed frame, (if applicable).
LFT5508 VIACLIN MODEL 2502
3.1
Issue 3 Oct 02
Hang the power unit on the bed or place on the floor.
Connect the three airpipe connectors on to the power unit connectors. Plug the power unit into a suitable mains socket and switch on.
The system will take approximately 20 minutes to reach full inflation pressure, at which time the green NORMAL light will come on and the amber light will extinguish. During this start up phase the red FAULT light may come on and the audible alarm sound, if this occurs turn the power unit off and back on again to reset the alarm.
LFT5508 VIACLIN MODEL 2502
3.2
Issue 3 Oct 02
!!ALARM!!
11
12
1
VIACLIN
10
2
8
4
Alternating Pressure Overlay
3
9
7
6
5
Once inflated, check the securing straps and tighten if required. For maximum benefit, use only one sheet loosely placed over the mattress. If the sheet is tucked in ensure that it is left loose over the mattress to avoid hammocking. Once a patient is placed on the mattress the system will adjust automatically to the patient characteristics (weight/position) to provide optimum clinical benefit for each patient. The mattress may feel firm at first until it has adjusted to the individual. Allow approximately 15 minutes for this automatic adjustment to be completed.
If the mattress does not perform as described above, please refer to Section F of this guide ‘Alarms and Fault Finding’. 3.2
PROFILING AND KNEE-BREAKING
The system may be used while the bed is in the profiled or knee-break position: 1.
With the bed in its present position, loosen or release the Overlay securing straps.
2.
Change the bed to the required position (profile or knee-break).
3.
Re-secure the securing straps.
NOTE: Straps must be attached to moving parts of the bed frame to avoid jamming the bed mechanism or stretching the overlay.
LFT5508 VIACLIN MODEL 2502
3.3
Issue 3 Oct 02
3.3
MOVING THE BED AND POWER CUTS
CAUTION:
The patient will not receive the benefit of the VIACLIN system while it is disconnected and the power unit is switched off. Therefore the system should be reconnected as soon as possible once the bed has been moved to its new position.
If you need to move the bed with the Mattress Overlay still inflated or in the event of a major mains power failure, carry out the following procedure: 1.
OVERLAY a b
2.
Disconnect the two red airpipes from the power unit and quickly connect together. Disconnect the black airpipe from the power unit.
POWER UNIT a
b.
Switch off and reset the power-down alarm. Disconnect from the mains.
The overlay will remain inflated for over 24 hours. The bed can now be moved.
LFT5508 VIACLIN MODEL 2502
3.4
Issue 3 Oct 02
3.4
CPR (CARDIO PULMONARY RESUSCITATION)
In the event of cardiac arrest, disconnect both red airpipes: 1.
While system is operating normally, disconnect from the power unit.
2.
While moving the bed, disconnect from each other.
3.5
DEFLATION, REMOVAL, CARRYING AND STORAGE
POWER UNIT 1.
Switch off.
2.
Silence alarm.
3.
Disconnect from mains.
4.
Disconnect all airpipes.
5.
Remove from bed.
OVERLAY 1.
Undo all securing straps.
2.
Roll up the Overlay.
LFT5508 VIACLIN MODEL 2502
3.5
Issue 3 Oct 02
3.6
ALARMS AND FAULT FINDING
ALARMS The VIACLIN Alternating Pressure Mattress Overlay is equipped with a sophisticated set of alarm functions, these alert the user to the status of the available mains supply and any mattress overlay defect. In normal conditions with mains power available, the power unit turned on, the mattress overlay fully inflated and alternating, the status light on the power unit front panel will be green. If at any time the mains power should be removed from the power unit, an audible alarm will be heard. This audible alarm may be cancelled by pressing the alarm reset button for approximately 2 seconds. If the product develops a defect whilst in use, the status light will turn red and an audible alarm will be heard. This alarm may be cancelled by turning the mains power switch on the side of the power unit to the OFF (O) position, then pressing the alarm reset button for approximately 2 seconds. FAULT FINDING.
SYMPTOM
ACTION
No lights on power unit
Check power unit is connected to the mains power supply and that the power unit switch is turned to the ON (1) position.
Status light red and alarm sounding, or alarm sounding but no lights
Turn mains switch on power unit OFF, press alarm reset button for approximately 2 seconds until alarm cancels, then switch on again: a.
If the status light does not illuminate, mains power to the power unit has been lost. Check that the power unit is connected to the mains supply and that the power unit switch is turned to the ON (1) position
b.
If the alarm clears then returns after approximately 12 minutes, check air connection between power unit and mattress overlay, ensuring that all air connectors are pushed fully home and that the airpipe is not folded or bent causing air flow to be reduced. Turn mains switch on power unit OFF, press the alarm reset button for approximately 2 seconds until alarm cancels, then switch on again
Mattress overlay not inflating, status light green
Check all air pipes as in (b) above
LFT5508 VIACLIN MODEL 2502
3.6
Issue 3 Oct 02
SECTION 4 – CONSTRUCTION OF PRODUCT
4.1
MAJOR ELEMENTS OF SYSTEM
The product comprises two major elements:
4.2
1.
Mattress Overlay
2.
Power Unit
MATTRESS OVERLAY
The mattress has 18 transverse cells interconnected, such that there are 2 sets of cells - each being fully inflated then fully deflated. The cells are chemically bonded to pneumatic tubing, which is formed into a harness terminated by the CPR connector. The Mattress has a cover which protects the mattress and the bed base from soiling by body fluids, and has a smooth surface finish to facilitate cleaning.
4.3.
POWER UNIT
The Power Unit comprises a housing that has a moulded lid, a base and chassis assembly that is fixed together by four socket-headed screws that also secure the hook assembly. When closed the casing is sealed to IP44. The lid incorporates the following:
1.
Mains switch - A double-pole, switch mounted the side wall of the casing.
2.
Alarm reset button - Fixed adjacent to the mains switch.
3.
Fuses - Two 1 Amp, anti surge fuses are fitted adjacent to the mains switch.
4.
Air connectors - There are three connectors to enable connection of the airpipe fitted to the opposite side wall of the casing to that used for the mains switch.
5.
Front panel label - The label has a hole for the tri-coloured LED that indicates power supply on (amber), normal running mode (green) and fault (red). The LED is connected to the integral electronic control unit (ECU) via an IDT connector.
LFT5508 VIACLIN MODEL 2502
4.1
Issue 3 Oct 02
The base incorporates the following: 1.
Suspension hooks - To enable the unit to fit over the footboard of a bed.
2.
Air intake holes - Covered on the inside by a small block of foam that serves as a primary air filter and protects the unit from ingress of dust.
A manufacturer's label is affixed to the rear of the base which bears electrical rating information and the serial number of the unit. The interior of the base carries a mild steel chassis that is secured to the casing front by two screws. The chassis incorporates the following: Compressor Rotorvalve Pressure Switch Alarm Buzzer
LFT5508 VIACLIN MODEL 2502
ECU Silencer / Accumulator Restrictor Block
4.2
Issue 3 Oct 02
4.4
TECHNICAL DETAILS
Cycle Control
Purpose designed distributor valve supplying operating air to the inflatable cells
Cycle Time
12 minutes
Supply Voltage
220/240Vac 50Hz
Power Rating
40VA
Fuse Rating
Power Unit:
1A Antisurge (x2)
Mains Plug:
3A
Noise Level
NC30
Nominal Dimensions
Mattress (Inflated)
Mattress (narrow)
Length:
1900mm
Length
1900mm
Width:
880mm
Width
800mm
Height:
130mm
Height
130mm
Weight:
3.8kg
Weight
3.5kg
Power Unit Height:
203mm
Width:
220mm
Depth:
115mm (max)
Weight:
2.3kg
Electrical Safety
Conforms to BSEN 60601-1
Classification
Class 1, Type B
Mode of Operation
Continuous
Symbols
Alternating Current Mains Switch - 1 (On) Power connected to the mains supply 0 (Off) Power disconnected from mains supply Type B
Environment (Usage and Storage)
Air Humidity:
30% to 75%
Ambient Temperature: 0ºC (32ºF) to 40ºC (104ºF) EMC
LFT5508 VIACLIN MODEL 2502
This equipment complies with EMC requirements. If effects are noticed the affected equipment should be moved apart.
4.3
Issue 3 Oct 02
SECTION 5 – INFECTION CONTROL/CLEANING 5.1
INFECTION CONTROL
Infection Control and routine cleaning must be carried out in accordance with your local Infection Control Policy. 5.2
CLEANING GUIDELINES
WARNING:
ENSURE THE POWER UNIT IS DISCONNECTED FROM THE MAINS ELECTRICITY SUPPLY BEFORE CLEANING.
CAUTION
Do not immerse the power unit in water. Do not high temperature autoclave, or use Phenolic based products for cleaning.
It is recommended the system is cleaned between users or approximately every 2 weeks if in constant use. Overlay and Cover Wipe down with warm water containing detergent, dry thoroughly before use. The cover may also be machine washed (80º C). Swab with a solution of Sodium Hypochlorite or similar (1,000 to 10,000ppm available chlorine), dry thoroughly before use. It is recommended that the system is cleaned regularly and after each patient use. In many cases it will only be necessary to remove the overlay cover for cleaning. Where there is obvious soiling, a full cleaning or sterilization procedure will have to be carried out.
Ensure overlay and cover are fully dry before use
CAUTION: SODIUM HYPOCHLORITE (1,000 - 10,000 PARTS PER PARTS PER MILLION AVAILABLE
DECONTAMINATE USING ETO (ETHYLENE OXIDE) OR DRAEGER SYSTEM
COVERS ONLY
80 C
PHENOL
CAUTION
LFT5508 VIACLIN MODEL 2502
5.1
Issue 3 Oct 02
Power Unit Wipe down with warm water containing detergent, dry thoroughly before use. Wipe down with a solution of Sodium Hypochlorite or similar (up to 10,000ppm available chlorine), dry thoroughly before use
CAUTION
Cleaning procedure
+
SODIUM HYPOCHLORITE (1,000 - 10,000 PARTS PER MILLION AVAILABLE CHLORINE (Cl))
DECONTAMINATE USING ETO (ETHYENE OXIDE) OR DRAEGER SYSTEM
CAUTION:
PHENOL
CAUTION
The overlay mattress, overlay cover and power unit may also be decontaminated by using ETO (Ethylene Oxide) or the Draeger method.
LFT5508 VIACLIN MODEL 2502
5.2
Issue 3 Oct 02
SECTION 6 – EQUIPMENT MAINTENANCE
Overlay
To Atmosphere
'B' Cells
'A' Cells Rotorvalve
Motor Gearbox
Restrictor Block
Supply Voltage PL5
Compressor PL4
PL6
PL3
Supply Voltage
2
Mains On/Running Normally PL1
Alarm PL2
A.C. Supply
1
ECU
From Atmosphere
3
4
5
Pressure Switch
6
Side Formers
7 Air Connector
8
Tri-Colour LED
Alarm Reset
Buzzer
The item numbers shown at the top left hand corner of the box drawn around each item are used in the Maintenance and Parts sections for identification purposes, i.e. Item 1 in each section is the ECU, Item 2 is the compressor, etc.
LFT5508 VIACLIN MODEL 2502
6.1
Issue 3 Oct 02
Notes: i. ii. iii. iv.
It is recommended that the Viaclin system is part of a planned preventative maintenance scheme. To prevent damage, normal electrostatic control procedures must be observed when working on the Power Unit. Electrical safety checks must be carried out on completion of any routine maintenance or workshop repairs. Details of tubing in the power unit is shown in Section 8 ‘TUBING DETAILS – SCHEMATIC’
Power Unit – General ECU Conn Maint/Repair
Test/Set-Up Test Eqpt
N/A 1.
Check the security of the wiring and tightness of screws and integrity of component mountings rectify as necessary.
2.
Check pneumatic tubing for hardening or cracking and replace as necessary.
3.
Switch on the electrical power supply.
4.
Check all pneumatic connections for air leaks and rectify as necessary.
5.
Lubricate spigot outlets sparingly with appropriate grease.
Carry out an overall operation test N/A
Item 1 – Electronic Control Unit The microprocessor controlled ECU is mounted on the chassis. The ECU has a rechargeable battery that provides power for the ECU and alarm buzzer in the event of power failure (it does not provide reserve power for the compressor or rotorvalve). The battery can only be recharged by the ECU. Control of the compressor is by the ECU, in response to signals provided by the pressure switch which continually monitors system pressures. The ECU also controls the status indicator on the front panel and the alarm buzzer and responds to activation of the alarm-reset pushbutton. ECU Conn Maint/Repair
N/A The battery is the only maintainer serviceable item; otherwise the ECU is repair by replacement. Note: To prevent damage, normal static control procedures must be observed when working on the PMU. Battery replacement (replacement only needed if battery is charged and power down alarm fails)
Test/Set-Up Test Eqpt
1.
Unsolder battery from ECU – note orientation of battery.
2.
Solder new battery onto ECU, ensuring it is in the same orientation as the old battery.
Testing of the ECU is carried out as part of the system test N/A
LFT5508 VIACLIN MODEL 2502
6.2
Issue 3 Oct 02
Item 2 – Compressor One Air Compressor is fitted, which takes air from atmosphere within the power unit and delivers air to the overlay bottom deck via the silencer, restrictor block, airpipe and to the top deck via the silencer, restrictor block rotorvalve and airpipe The Compressor has rubber-mounting feet attached by screws and nuts to an aluminium baseplate, a 4-way IDT connector for connection to the ECU, a spigot outlet over which 8mm-diameter silicone tubing is push-fitted to enable pneumatic connection to the check valve, from which tubing connects to the silencer. ECU Conn Maint/Repair
PL4
Test/Set-Up
Carry out compressor electrical tests: 1. Disconnect tubing from the compressor (if fitted). 2. Connect the compressor to the appropriate connector(s) on the test set (TE5315). 3. Connect the test set to mains and switch on.
Maintenance of the compressor consists of replacing the filter: 1. Remove the filter cover from the end of the compressor, remove the filter. 2. Clean the filter housing surfaces and the disk. 3. Fit a new filter and replace the cover.
No-Load Test Check compressor runs with no excess vibration or noise - if there is a defect the 1.6a fuse may blow. On-Load Test This test is carried out at the same time as the flow rate test. Carry out a compressor flow rate test with TE5315 connected as for the no-load test - if there is a Defect the 1.6a fuse may blow. Carry out a flow rate test on the compressor: Note: Ensure the compressor is in its ‘HOT’ state prior to carrying out the test: Start Service Mode and press the Alarm Reset button once, this will start the compressor. Let the compressor run for 30 minutes to get to the 'Hot' state. If the system has been running for at least 1 hour prior to starting Service Mode, the compressor is assumed to be in the 'Hot' state and do not need to be run for 30 minutes. 1.
Remove the tubing from the compressor and connect the test equipment (TE4540) in its place.
Note: Ensure Test Equipment is connected to the pressure outlet of the compressor, not the vacuum side. 2.
Switch the power unit On. Adjust the aneroid sphyg to read 100mmhg.
3.
Check the flow rate on the flow meter reads not less than 5.5 ltr/min. If using test equipment other than TE4540, the flow rate should read 7 ltr/min or greater.
4.
Switch off, remove test equipment; reconnect the tubing to the compressor.
Notes:
Test Eqpt
i. ii.
If using a 1-12 ltr glass tubed flow meter, read the value from the top of the float. If using a plastic bodied flow meter, read the value from the centre of the float.
TE4540, TE5315
LFT5508 VIACLIN MODEL 2502
6.3
Issue 3 Oct 02
Item 3 - Silencer The silencer smoothes pressure fluctuations in the pneumatic system and absorbs any noise being carried in the airflow from the compressor. It is fitted into a recess in the chassis and held in place by the casing. Spigot outlets from the silencer have 8mm-diameter silicone tubing push-fitted to enable pneumatic connection to the compressor and restrictor block. ECU Conn Maint/Repair
N/A 1.
There are no maintenance actions to be carried out on the silencer. Repair is by replacement
Test/Set-Up
2. It is recommended the silencer is replaced annually There are no tests to be carried out on the silencer.
Test Eqpt
N/A
Item 4 – Restrictor Block The restrictor block enables airflow to and from the following: Silencer Pressure Switch Air Connector (for side formers only) Rotorvalve ECU Conn Maint/Repair Test/Set-Up Test Eqpt
N/A Remove all tubing from restrictor block and ensure all spigots are clear of obstructions. There are no tests to be carried out on the restrictor block. N/A
LFT5508 VIACLIN MODEL 2502
6.4
Issue 3 Oct 02