Setup Guide

2 Pages

PiCCO2® Setup Guide Tips & Tricks •  Do you have a good arterial waveform trace? If the trace is absent or dampened check patient position and/or catheter. Ensure there are no kinks in the tubing. Check pressure bag is inflated. Flush arterial line. Check cables are attached correctly.  •  Do you have an arterial pressure appropriate for the patient? If a false low/high measurement is suspected, check transducer level is correct. Check the monitor was zeroed correctly.  •  Do you have an adequate/appropriate Cardiac Output/Index for the patient’s clinical situation? If Cardiac Index appears inappropriate (too low/high) check the patient’s height and weight have been entered correctly (press patient info bar at top of monitor). Check correct injectate volume was used.  •  Did you inject the correct amount and temperature of injectate? The amount of injectate can be changed in the TD screen depending on the patient weight. 15cc is recommended for adult patients. Injectate temperature must be < 8° C / 46° F in all cases.  •  Did the monitor display “injection” while you were injecting? If not the injection has not been detected. Check the monitor displayed <READY> before pressing <START>, check amount and temperature of injectate. Check injectate sensor cable connection. Is there the correct distance between the injectate sensor housing and the distal lumen hub? The point of injection must be directly through the injectate sensor housing, which must be as close to the patient as possible. Remove additional stop cocks and lines causing excess dead space.  •  Were all other high volume infusions switched off during the injection? Suspend any infusions that are running through the same lumen you are injecting through. Wait until the TD curve is complete before turning back the 3 way stop cock.  •  Is the temperature sensor cable connected correctly? Ensure cables are connected correctly at the injectate sensor and PiCCO catheter. Pins must not be bent.  •  Thermodilution  Injectate sensor cable  Injectate sensor housing  • CVP line  Distal lumen of CVC  Injectate sensor cable  2  3  Arterial connection cable  1  A. • •  Pressure output adapter  Pressure connection cable  PiCCO® Catheter  Connect the arterial connection cable 1 to the orange socket on the right side of the PiCCO2® device. Connect the injectate sensor cable 2 to the brown socket of the arterial connection cable 3 . 2  1  For further information on troubleshooting, please refer to the Train the Trainer manual. If you still are experiencing problems please contact your local clinical sales executive. See instructions for use and package insert for full prescribing information. Technical specifications are subject to © 2010 PULSION Medical Systems AG all rights reserved. change without further notice. PULSION Medical Inc. • 2445 Gateway • Drive Suite 110 • Irving, Texas 75063 Toll free 877.655.8844 • Office 214.446.8500 • Fax 214.446.6702 info@pulsion.com • www.PULSION.com  Prime the injectate sensor housing with saline before attaching to the central venous catheter Connect the blue end of the injectate sensor cable to the injectate sensor housing which should be preferably on the distal lumen of the central venous catheter.  Flush bag  Were you in the TD screen when you performed the injection? When in this screen wait for „Inject XXml“ to appear, then deliver a quick smooth injection (under 7 secs). Check the monitor displays <READY> before pressing <START>.  •  •  Setup PiCCO2 Hemodynamic Monitor  MPI851505US_R00  © PULSION 01/2010  3
File Type: PDF
File Size: 1.55 MB
File Name: picco2_setup_guide.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions