Radi Medical Systems

Coronary Pressure Measurement Monitors

RadiAnalyzer Xpress Quick Guide Rev 04 July 2008

Quick Guide

2 Pages

QUICK GUIDE 1.  CONNECT CABLES Press and hold F1 for Auto Setup  ON/OFF switch  Mains power PressureWire®  AO adapter AO transducer  Cathlab AO input* Cathlab PressureWire® input*  * Note: These connections are optional, FFR calculation is independent of the cathlab monitor system.  2.  ZERO CATHLAB PRESSURE CHANNELS  Select ZERO CATHLAB and press ENTER  3.  CALIBRATE AO TRANSDUCER  Select CAL AO and press ENTER  4.  Open AO to air and press ENTER. Close valve to resume pressure reading.  CALIBRATE PRESSUREWIRE®  Select CAL WIRE and press ENTER.  5.  Zero both cathlab pressure channels, press ENTER when done. (Optional: Press LEFT/RIGHT to test AO/PressureWire® OUT.)  Flush the PressureWire® coil, place flat on table and press ENTER.  VERIFY EQUAL AO AND PRESSUREWIRE® PRESSURES 1. Advance sensor element just outside guiding catheter opening. 2. Verify that AO (Pa) and PressureWire® (Pd) pressures are equal at that position. 3. If not equal: Make sure that the AO transducer is fixed at heart level and remove any residual difference by pressing EQUALIZE for 3 seconds.
File Type: PDF
File Size: 198 KB
File Name: radianalyzer_xpress_quick_guide.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions