Elisée 350
© ResMed
Waking people up to sleep
Field of application
• ICU / CCU ventilator • Pressure and volumetric ventilator • Invasive & Non Invasive ventilation • Single & Double circuit system • Adult & Pediatric ventilation
© ResMed
Suitable for paediatric use
• From 6 kg
• Circuit Ø 15 mm
• Minimum tidal volume : from 50 ml
• Maximum frequency : up to 80 cycles / min
© ResMed
Specifications of the ventilator
• Size : 290 x 250 x 130 mm • Weight : – – –
4,5 kg, ventilator only with internal battery 0,4 kg, mains supply pack 1,1 kg, external battery pack
• Autonomy : – – –
NimH batteries Internal autonomy : 2H30 External autonomy : 3H (removable pack)
Ventilation performances are preserved during the whole time of discharge © ResMed
In the forefront of technology • 2 patents : – Geometry of proportional valve – Flow-by structure : Combination of a high performance micro blower and a proportional 3 ways valve
• Thanks to this technology : – – – – –
© ResMed
Better control of flows Reduction of valve response time Maximum safety in case of HP no barautroma Maximum precision of flow-by Easy ventilation with high frequencies and little volumes
CONTENTS
Description of the ventilator
FiO2
Circuit assembly
Observation of ventilation
Turning the ventilator on / off
Respiratory mechanics option
Ventilation selection
Nebulization
Ventilation modes
Technical menu
Ventilation parameters Alarms management
© ResMed
Description of the ventilator Touch screen Connection of expiratory supports (double or single circuit)
Insufflation outlet Indicator for mains / external supply Indicator for internal / external battery « Silence Alarm » button
© ResMed
Luminosity sensors
Description of the ventilator Remote alarm socket
Nebulizer O2 outlet connector
« On / Off » button
Air outlet
External DC supply 12-28V
O2 low pressure connector Serial connection socket
Mains supply socket © ResMed
Access to the O2 sensor
Air inlet
O2 high pressure connector
Circuit assembly • Connection of a single or double patient circuit • Simplification of assembly / disassembly : system of clips No tools !
Clips
Center points
• Automatic detection of the circuit characteristics during the manual test © ResMed
Circuit assembly • 2 interchangeable supports : – « Single circuit » support allows to connect the exhalation valve control connector and the proximal pressure line
Proximal pressure line
Exhalation valve control © ResMed
Inspiratory limb of patient circuit (Ø 22 mm)
Circuit assembly • 2 interchangeable supports : – The « double circuit » support is composed of a support valve and a built-up spirometry. It allows to connect the exhalation limb.
Air outlet
Expiratory limb patient circuit (Ø 22 mm standard)
Inspiratory limb of patient circuit (Ø 22 mm standard) © ResMed
Turning the ventilator on • Start the ventilator by pressing the button on the right side of the device
• Then the following welcome screen appears : To display the results of the last manual test
To choose a preset ventilation mode To choose the previous ventilation mode
© ResMed
To set a new ventilation mode
Turning the ventilator off • Switch off the device by pressing the button on the right side of the device
• Depending on the ventilation status, current or not, one of the two following screens appear :
No ventilation
• Confirm you want to stop the ventilator © ResMed
While ventilating
Manual test • Doing a manual test allows : – –
To detect what type of patient circuit is used To determine the resistance / compliance of the circuit
• When doing a manual test ? – – – –
Each new patient Each time the ventilation program is changed When changing the circuit After adding / removing an accessory (humidifier …)
• Precautionary measures : – –
Connect all the accessories used for the ventilation Disconnect the ventilator from the patient
If the test fails, call your technical services © ResMed
Manual test • 3 measure stages : – – –
© ResMed
Sensors offset (O2, pressure …) Circuit inspiratory resistance Expiratory resistance (if double patient circuit), compliance and O2 sensor gain
Ventilation selection
5 preset programs Memorization of the whole ventilation configuration : - Mode - Type of patient - Ventilation parameters - Alarms settings
Reduction in training time for hospital staff ! Ideal to differentiate day / night ventilation !
© ResMed
Ventilation selection
© ResMed
New setting • To access to new ventilation programming, press the button “New patient” • The following screens allows to choose the patient configuration :
Each element of the configuration is reminded on the screen thanks to an icon © ResMed
New setting •
Once the patient configuration has been chosen, the following screen displays in order to choose the ventilation mode : Pressure modes
Volumetric modes
Dual mode
•
Once the mode has been selected, the screen of parameters setting appears
© ResMed
Ventilation modes : P(A)CV
• P(A)CV = Assisted Pressure-Controlled Ventilation The ventilator delivers cycles characterized by a fixed insufflation's time and pressure parameter –
PCV = cycles are released by the ventilator, insufflation does not depend on patient’s effort. A frequency should be adjusted.
–
PACV = inspiration can be released by the patient (trigger)
–
The change from PCV to PACV mode depends on inspiratory triggers adjustment
• In P(A)CV mode, the user can set recruitment cycles More about recruitment :
© ResMed