Page 1
Instructions for use
Fiber Light Cable Sets Fusion fiber light cable sets - high temperature resistant Fluid light cable
GA-A287 / en-EU / 2021-02 V1.0 /
Richard Wolf reserves all rights in this document, in particular relating to reproduction, distribution, public communication and public dissemination. This document may not be modified, and texts and images must not be reproduced or translated without the express prior permission of Richard Wolf GmbH.
Copyright © RICHARD WOLF GmbH
Subject to technical changes! Due to ongoing development, the product descriptions, illustrations and technical data may deviate slightly from the current state. For further information about our products please contact Richard Wolf GmbH or your representative.
Manufacturer RICHARD WOLF GmbH Pforzheimer Straße 32 75438 Knittlingen GERMANY Telephone: +49 70 43 35-0 Telefax: +49 70 43 35-4300
www.richard-wolf.com
2
GA-A287
Table of contents 1
General information ...
5
1.1
Safety instructions and levels of danger...
5
1.2
Symbol...
5
2
General safety instructions and guidance for use ...
7
3
Product description...
8
4
Intended purpose ... 10
5
Indications ... 12
6
Contraindications and side effects ... 13
6.1
Contraindications... 13
6.2
Side effects... 13
7
Combinations ... 14
7.1 7.1.1
Overview of permissible combinations and requirements ... 15 Connection with fiber light cable (RICHARD WOLF) to light sources and endoscopes (RICHARD WOLF) ... 15 Connection with fluid light cable (RICHARD WOLF) to light sources and endoscopes (RICHARD WOLF) ... 16 Connection with special fiber light cable (RICHARD WOLF) to light sources (RICHARD WOLF) and endoscopes (OTHER MANUFACTURERS)... 17 Connection with fiber light cable (RICHARD WOLF) to light sources and endoscopes (OTHER MANUFACTURERS) ... 18 Connection with fluid light cable (RICHARD WOLF) to light sources and endoscopes (OTHER MANUFACTURER)... 19 Connection with fiber light cable (OTHER MANUFACTURER) to light sources and endoscopes (RICHARD WOLF) ... 20
7.1.2 7.1.3 7.1.4 7.1.5 7.1.6 8
Illustration ... 21
9
Checks... 23
9.1
Visual check ... 23
9.2
Function check ... 23
9.3
Service life ... 24
10
Application ... 25
10.1 10.1.1 10.1.2
Preparation ... 25 Attaching the adapter (1) (6) to the light cable (3) (7) ... 25 Identification of fiber bundle diameter on fiber light cable (1) (7) ... 26
10.2 10.2.1 10.2.2
Indications for Use... 27 Replacing the color coding ring (2) on the light cable connector facing the light source (only if required). 28 Light... 29
11
Reprocessing and maintenance ... 30
11.1
Reprocessing sequence – Fiber light cable set [ A ], Fusion fiber light cable set [ B ], Special fiber light cable set [ C ] ... 32 Machine reprocessing ... 33 Manual reprocessing ... 33 Checks and maintenance ... 35 Packaging... 35 sterilization... 35 Storage ... 36 Validation... 36
11.1.1 11.1.2 11.1.3 11.1.4 11.1.5 11.1.6 11.1.7
GA-A287
3
11.2 11.2.1 11.2.2 11.2.3 11.2.4 11.2.5 11.2.6 11.2.7
Reprocessing sequence for fluid light cable [ D ] ... 36 Machine reprocessing ... 37 Manual reprocessing ... 37 Checks and maintenance ... 38 Packaging... 39 sterilization... 39 Storage ... 40 Validation... 40
12
Technical description ... 41
12.1 12.1.1 12.1.2 12.1.3
Technical data ... 41 Endoscopes without identification of the fiber bundle diameter ... 41 Fiber Light Cable Sets ... 42 Fluid light cable [ D ] ... 45
12.2
Operating, storage, transport and shipping conditions ... 45
13
Spare parts ... 46
14
Disposal of product, packaging material and accessories ... 46
4
GA-A287
1
General information
1.1
Safety instructions and levels of danger WARNING This signal word is used to indicate a potentially dangerous situation. Not avoiding this situation can result in death or extremely serious injuries.
CAUTION This signal word is used to indicate a potentially dangerous situation. Not avoiding this situation can result in minor or moderate injury.
ATTENTION This signal word without warning sign is used to indicate a potential danger of material damage.
NOTE This signal word indicates additional useful information for the reader, such as hints for easier operation as well as cross references.
1.2
Symbol
Symbols
Designation Follow the manual Medical device Product number Lot code Manufacturer Manufacturing date Number, amount
Data Matrix Code
GA-A287
5
Symbols
Designation CE marking in conformity with Directive 93/42/EEC on medical products or EU Regulation 2017/745 on medical devices. Only valid, if the product and/or packaging is marked with this identification. Products as of category IIa, sterile products and products of class I, products with measuring function of class I, as well as reusable surchical instruments are additionally marked with the four-digit code number of the competent authority. The CE identification on the title page of these instructions for use relates exclusively to the Richard Wolf main product or, if serveal identical products are described, to the Richard Wolf product with the highest classification. The CE identification of the other Richard Wolf products and, if applicable, of products made by other manufacturers, described in these instructions for use results exclusively from the identification on the product and/or packaging.
6
GA-A287
2
General safety instructions and guidance for use The product must only be used as intended following the instruction manual by adequately trained and qualified medical personnel. Maintenance and repair must be carried out by authorized experts. Use the product only in the combinations and with the accessories and spare parts specified in this instruction manual. Use other combinations, accessories and replacement parts only if they are expressly intended for the planned application and if the performance characteristics and safety requirements are met. The product must not be altered in any way. Reprocess the products in accordance with the manual before every use and before return shipment to protect the patient, user and third parties. This manual is an integral part of the product and must be stored in such a way that it is accessible at any time during its entire life cycle. This manual must be passed on to any subsequent owner or user. Immediately on receipt, check the product and its accessories for completeness and possible damage. Should the shipment give rise to complaints, please inform the manufacturer or supplier immediately. Any severe incident occurring in conjunction with this product must be reported to the manufacturer and the competent authority of the country where the user and/ or patient resides.
GA-A287
7
3
Product description The fiber light cable sets are available in different product versions: n n n n
Fiber light cable set with color coding [ A ] Fusion fiber light cable set, high temperature resistant [ B ] Special fiber light cable set [ C ] Fluid light cable [ D ]
The products described differ in the following features: Fiber light cable set with color coding [ A ] n
n
n n
on projector side – with replaceable color coding ring (standard color black). This color coding ring is available in different colors. on endoscope side – with color coding ring and numeric code for fiber bundle diameter. Adapter, projector side, can be unscrewed Adapter, endoscope side, can be unscrewed – with "snap" quick-release coupling (ring actuation)
Fusion fiber light cable set, high temperature resistant [ B ] n
n
n n
on projector side – with identification: fusion – with replaceable color coding ring (standard color black). This color coding ring is available in different colors. on endoscope side – with color coding ring and numeric code for fiber bundle diameter Adapter, projector side, can be unscrewed Adapter, endoscope side, can be unscrewed – with "snap" quick-release coupling (ring actuation)
Special fiber light cable set [ C ] n
n n
on endoscope side – with color coding ring and numeric code for fiber bundle diameter Adapter, projector side, fixed Adapter, endoscope side, can be unscrewed – with "snap" quick-release coupling (ring actuation)
Fluid light cable [ D ] n n n
8
projector side and endoscope side, straight Adapter, projector side, can be unscrewed Adapter, endoscope side, can be unscrewed – with tab, long version
GA-A287
NOTE The fiber light cables are also available without adapter. The color coding ring system on both the projector and the endoscope sides allows easy and quick allocation of the fiber light cable. n Color coding ring projector side (replaceable): Hospital-specific identification of fiber light cables (e.g., for different disciplines) n Color coding ring endoscope side: Fiber bundle diameter of fiber light cable matches the endoscope diameter See section Technical data and order data
GA-A287
9
4
Intended purpose NOTE For detailed information regarding fluorescence diagnostic applications, please refer to the corresponding system instruction manuals.
Light cable 80655023
FIBER LIGHT CABLE Ø 5 MM TL 2.3 M
80655030
FIBER LIGHT CABLE Ø 5 MM TL 3 M
80655035
FIBER LIGHT CABLE Ø 5 MM TL 3.5 M
80661618
FIBER LIGHT CABLE Ø 1.6 MM TL 1.8 M
80661623
FIBER LIGHT CABLE Ø 1.6 MM TL 2.3 M
80661630
FIBER LIGHT CABLE Ø 1.6 MM TL 3 M
80662518
FIBER LIGHT CABLE Ø 2.5 MM TL 1.8 M
80662523
FIBER LIGHT CABLE Ø 2.5 MM TL 2.3 M
80662530
FIBER LIGHT CABLE Ø 2.5 MM TL 3 M
80663518
FIBER LIGHT CABLE Ø 3.5 MM TL 1.8 M
80663523
FIBER LIGHT CABLE Ø 3.5 MM TL 2.3 M
80663530
FIBER LIGHT CABLE Ø 3.5 MM TL 3 M
80663535
FIBER LIGHT CABLE Ø 3.5 MM TL 3.5 M
80665023
FIBER LIGHT CABLE Ø 5 MM TL 2.3 M
80665030
FIBER LIGHT CABLE Ø 5 MM TL 3 M
80665035
FIBER LIGHT CABLE Ø 5 MM TL 3.5 M
80673523
FIBER LIGHT CABLE BLUE Ø 3.5MM
4070.253
FLUID LIGHT CABLE Ø 3.0MM TL 2.3MM
The products are used to transmit light in the visible spectral range from the light source to the application-specific instrument set. Adapter for light cable
10
807001
ADAPTER, ENDOSCOPE SIDE
807002
ADAPTER, ENDOSCOPE SIDE
808801
ADAPTER, ENDOSCOPE SIDE
8070.00
ADAPTER, ENDOSCOPE SIDE
8087.00
ADAPTER, ENDOSCOPE SIDE
8088.00
ADAPTER, ENDOSCOPE SIDE
8088.80
ADAPTER, ENDOSCOPE SIDE
GA-A287
8088.81
ADAPTER, ENDOSCOPE SIDE
8091.00
ADAPTER, ENDOSCOPE SIDE
8095.05
ADAPTER, ENDOSCOPE SIDE
809506
ADAPTER, ENDOSCOPE SIDE
8095.07
ADAPTER, PROJECTOR TO LIGHT CABLE
809509
ADAPTER, ENDOSCOPE SIDE
8096.811
ADAPTER, PROJECTOR TO LIGHT CABLE
The products are used to connect light guides to light sources or endoscopes. These products are exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons.
GA-A287
11
5
Indications The products are used in endoscopic procedures in various medical disciplines.
NOTE For detailed information regarding the blue light mode, please refer to the corresponding system instruction manuals.
Patient population The intended patient population is not restricted to ethnics, age, gender, body height and weight. Before use the attending physician must ensure that the product can be safely used in terms of its dimensions or settings.
12
GA-A287
6
Contraindications and side effects
6.1
Contraindications There are currently no known contraindications directly related to the product. Medical contraindications for the endoscopic procedures to be performed exclude the use of the products concerned. The responsible physician must decide on the basis of the patient's general condition whether the intended use is possible or not.
NOTE For any limitations regarding the blue light mode, please refer to the corresponding system instruction manuals.
6.2
Side effects No side effects related to the products are known if the products are used as intended.
GA-A287
13
7
Combinations CAUTION Incorrect combination of products! Injuries to the patient, user, or others, as well as damage to the product are possible. n The different products must only be used jointly if the intended use and the relevant technical data, such as working length, diameter, etc., are the same. n Follow the instruction manuals and the system instruction manuals of the products used in combination with this product.
ATTENTION The special fiber light cable set is fitted with a fixed projector-side adapter and is only compatible with Richard Wolf light sources. The adapters (1) and (6) can be unscrewed and replaced by suitable adapters (a) and (b) for connection to the light source or endoscope (Fig. 1a). n
n
See section 7.1.1 Connection with fiber light cable (RICHARD WOLF) to light sources and endoscopes (RICHARD WOLF) See section 7.1.2 Connection with fluid light cable (RICHARD WOLF) to light sources and endoscopes (RICHARD WOLF)
Use of fiber light cable sets [ A ] [ B ] (RICHARD WOLF) (Fig. 1b) n
See section 7.1.4 Connection with fiber light cable (RICHARD WOLF) to light sources and endoscopes (OTHER MANUFACTURERS)
Use of special fiber light cable sets [ C ] (RICHARD WOLF) (Fig. 1c) n
See section 7.1.3 Connection with special fiber light cable (RICHARD WOLF) to light sources (RICHARD WOLF) and endoscopes (OTHER MANUFACTURERS)
Use of fluid light cables [ D ] (RICHARD WOLF) (Fig. 1d) n
See section 7.1.5 Connection with fluid light cable (RICHARD WOLF) to light sources and endoscopes (OTHER MANUFACTURER)
Use of fiber light cable (OTHER MANUFACTURER) n
14
See section 7.1.6 Connection with fiber light cable (OTHER MANUFACTURER) to light sources and endoscopes (RICHARD WOLF)
GA-A287
on projector side
on endoscope side
1
6
a
1
Fig. 1a
b
Fig. 1b
b
Fig. 1c
6
Fig. 1d
Fig. 1
7.1
Overview of permissible combinations and requirements
7.1.1
Connection with fiber light cable (RICHARD WOLF) to light sources and endoscopes (RICHARD WOLF)
Light source
Endoscope
Fiber light cable [ A ] [ B ]
Special fiber light cable [ C ]
Adapters
Adapters
projector side
endoscope side
8095.07
Adapter, fixed projector side
809509
Fig. 2
GA-A287
15
7.1.2
Connection with fluid light cable (RICHARD WOLF) to light sources and endoscopes (RICHARD WOLF)
Light source
Endoscope
Fluid light cable [ D ]
Adapter
Adapter
projector side
endoscope side
8095.07
809506
8095.05
Fig. 3
16
GA-A287
7.1.3
Connection with special fiber light cable (RICHARD WOLF) to light sources (RICHARD WOLF) and endoscopes (OTHER MANUFACTURERS)
Special fiber light cable [ C ]
Light source
Light source made by:
Adapter, fixed
Adapters
projector side
endoscope side
Other endoscope
Endoscope made by:
ACMI + Circon 807002
Henke-Sass-Wolf
Storz
Olympus
Stryker 807001
Fig. 4
GA-A287
17
7.1.4
Connection with fiber light cable (RICHARD WOLF) to light sources and endoscopes (OTHER MANUFACTURERS)
Light source from other manufacturer
Light source made by:
Storz
Fiber light cable [ A ]
Endoscope from other manufacturer
Adapter
Adapter
projector side
endoscope side
ACMI + Circon
fits directly 807002
Olympus
Henke-Sass-Wolf
Storz
8096.811
Stryker
Endoscope made by:
Wolf adapter fits Olympus
Stryker 807001
Fig. 5
18
GA-A287
7.1.5
Connection with fluid light cable (RICHARD WOLF) to light sources and endoscopes (OTHER MANUFACTURER)
Light source from other manufacturer
Light source made by:
Storz
Fluid light cable [ D ]
Endoscope from other manufacturer
Adapter
Adapter
projector side
endoscope side
fits directly
Endoscope made by:
Storz
Olympus 8096.811 Olympus Stryker
Wolf adapter fits
Stryker 8070.00
Fig. 6
GA-A287
19
7.1.6
Connection with fiber light cable (OTHER MANUFACTURER) to light sources and endoscopes (RICHARD WOLF)
Light source
Fiber light cable from other manufacturer
Endoscope
Color coding ring? Yes / No?
Adapter projector side
Fiber light cable made by:
Adapter endoscope side
Storz, 3.5 mm
Storz, 4.8 mm
Olympus Standard 808801
Olympus WA032...A 8088.81 fits directly
ACMI + Circon 8087.00
Fig. 7
20
GA-A287