Instruction Manual
RIWO Objective Lenses autoclavable
8526x.xxx / 85261xxx
GA-S 022 / USA / 2013-04 V3.0 / ECO 2013-0154
(RW: 2012-03 V2.0 / PDB 11-5134)
Important general notes and instructions for use Make sure that this product is used only as intended and described in this instruction manual, by adequately trained and qualified medical personnel, and that maintenance and repair are only carried out by authorized experts. Use the product only in the combinations and with the accessories and spare parts specified in this instruction manual. Use other combinations, accessories and replacement parts only if they are expressly intended for the planned applic ation and if the performance characteristics and safety requirements are not impaired. Do not alter the product in any way. Reprocess the products before every application and before returning them for repairs as required by the instruction manual in order to protect the patient, user and others. This manual is an integral part of the product and must be stored in such a way that it is accessible at any time during its entire life cycle. This manual must be passed on to any subsequent owner. Immediately upon receipt, check the product and its accessories for completeness and possible damage. Should the shipment give right to complaints, please inform the manufacturer or supplier immediately. Subject to technical changes! Due to ongoing developments, the illustrations and technical data may deviate slightly. CAUTION : Federal law restricts this device to sale by or on the order of a physician.
Safety instructions and levels of danger Symbols
Level of danger
WARNING! Failure to observe can result in death or serious injury. CAUTION! Failure to observe can result in slight injury or damage to the product.
.
IMPORTANT! Failure to observe can result in damage to the product or surroundings.
.
NOTE! Tips for optimum use and other useful information. USA RICHARD WOLF Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Toll Free: 001 (800) 323 - 9653 Phone: 001 (847) 913 - 1113 Fax: 001 (847) 913 - 1488 MANUFACTURER [email protected] www.richardwolfusa.com
GERMANY RICHARD WOLF GmbH 75438 Knittlingen Pforzheimerstr. 32 Telephone: +49 70 43 35-0 Telefax: +49 70 43 35-300 [email protected] www.richard-wolf.com
UK RICHARD WOLF UK Ltd. Waterside Way Wimbledon SW17 0HB Telephone: + 44 20 89 44 74 47 Telefax: + 44 20 89 44 13 11 [email protected] www.richardwolf.uk.com
BELGIUM / NETHERLANDS N.V. Endoscopie RICHARD WOLF Belgium S.A. Industriezone Drongen Landegemstraat 6 9031 Gent Drongen Telephone: +32 92 80 81 00 Telefax: +32 92 82 92 16
FRANCE RICHARD WOLF France S.A.R.L. Rue Daniel Berger Z.A.C. La Neuvillette F-51100 Reims Telephone: +33 3 26 87 02 89 Telefax: +33 3 26 87 60 33
AUSTRIA RICHARD WOLF Austria Ges.m.b.H. Wilhelminenstraße 93 a A-1160 Vienna Telephone: +43 14 05 51 51 Telefax: +43 14 05 51 51 45
[email protected] www.richard-wolf.com
[email protected] www.richard-wolf.be
0
Marketing Office U.A.E RICHARD WOLF Middle East P.O. Box 500283 AL Thuraya Tower 1 9th Floor, Room 904, Dubai Telephone: + 9 71 43 68 19 20 Telefax: + 9 71 43 68 61 12
INDIA RICHARD WOLF India Private Ltd. JMD Pacific Square No. 211 A, Second Floor Behind 32nd Milestone Gurgaon - 122 001 National Capitol Region Telephone: + 91 12 44 31 57 00 Telefax: + 91 12 44 31 57 05
[email protected] www.richard-wolf.com
[email protected] www.richard-wolf.com
GA-S 022
Contents 1
Technical description...
1
2
Intended use...
1
3
Indications and field of application...
1
4 4.1 4.2
Contraindications and side effects... Contraindications... Side effects...
1 1 1
5
Combinations...
2
6 6.1 6.2 6.2.1 6.2.2 6.2.3 6.3 6.3.1
Illustration... Legend and identification... System overview of RIWO objective lenses... Camera heads with ¼“ - sensors... Camera heads with ⅓“ - sensors... Camera heads with ½“ - sensors... RIWO objective lenses and adapters... RIWO objective lenses with C-mount thread and snap-on mechanism / clamping mechanism... RIWO objective lenses with RW-mount thread and clamping mechanism... RIWO-objective lenses with RW-mount thread and plug-on connection... Adapters for RIWO objective lenses...
3 3 5 5 6 6 7
6.3.2 6.3.3 6.3.4 7 7.1 7.1.1 7.2 7.3 7.4
7 7 8 8
7.6 7.7 7.7.1 7.7.2
Use... Preparation... Removing the caps / objective lens protection caps... Connecting the objective lens to the camera head... Connection C-mount / RW-mount adapter 8885.952... Connecting the RIWO objective lens, RIWO angular objective lens to an endoscope with eyecup... RIWO objective lens with snap-on mechanism (1)... RIWO objective lens, RIWO angular objective lens with clamping mechanism (2) . . . Connecting the objective lens to an endoscope with plug-on eyepiece adapter... Connecting the objective lens to an endoscope with plug-on eyecup 8885.901 / 8885.902... Sterilizable adapter 8580.91 (option)... Additional notes and instructions for use... Zoom objective lenses... RIWO objective lenses with snap-on mechanism (1)...
8 8.1 8.2
Checks... 15 Visual check... 15 Function check... 15
9 9.1 9.2 9.2.1 9.3 9.4
Reprocessing and maintenance... Disassembly before cleaning... Illustrations on reprocessing... Manual cleaning... Reprocessing procedures... Alternate Sterilization Methods...
10
Technical data and order data... 18
11
Spare parts and accessories... 18
12 12.1
Operating, storage, transport and shipping... 18 Disposal of product, packaging material and accessories... 18
13
Warranty and Customer Service... 19
7.4.1 7.4.2 7.5 7.5.1
GA-S 022
9 9 9 10 10 11 11 11 12 12 13 14 14 14
16 16 16 16 16 17
I
1
Technical description
RIWO objective lenses and adapters are available in different product versions and are distinguished as follows: Z RIWO objective lenses with C-mount thread and snap-on mechanism / clamping mechanism RIWO angular objective lens ' with snap-on mechanism / clamping mechanism ' with focusing ring / focusing screw and zoom ring ' with cap for snap-on mechanism / clamping mechanism ' with protection cap for objective lens / cap with thread for C-mount thread ' with indication of focal length Z RIWO objective lenses with RW-mount thread and clamping mechanism ' with clamping mechanism ' with focusing ring and zoom ring ' with cap for clamping mechanism ' with plug-on cap for camera heads ' with indication of focal length Z RIWO objective lenses with RW-mount thread and plug-on connection ' with plug-on eyepiece adapter for endoscope with plug-on eyecup ' with clamping screw ' with indication of focal length Additional feature of miniature RIWO objective lens with pendulum: ' for image alignment Z Adapter for RIWO objective lenses Plug-on eyecup / plug-on eyecup, rotatable ' Adapter adapts endoscope with plug-on eyepiece and C-mount objective lenses ' with clamping screw C-mount / RW-mount adapter ' Adapts endoscope and CCD ENDOCAM with RW-mount thread C-mount ob jective lenses Sterilizable adapter ' Adapts endoscope and objective lenses with snap-on mechanism / clamping mechanism
2
Intended use
3
Indications and field of application
4
Contraindications and side effects
4.1
Contraindications
4.2
Side effects
The objective lens serves to connect endoscope to video cameras. For examination, diagnosis and therapy in conjunction with endoscopes in differ ent medical disciplines such as surgery, urology (transurethral resection), gyneco logy (myoma resection) and ENT by adequately trained and qualified medical personnel.
1
Contraindications directly related to the product are presently unknown. On the basis of the patient's general condition the doctor in charge must decide whether the planned use is possible or not. Side effects are not known if instruments are applied as defined in the intended use.
GA-S 022
5
Combinations
RIWO objective lenses are used in conjunction with endoscopes and CCD cam eras.
CAUTION! Do not combine products incorrectly! Injuries of the patient, user or others as well as damage to the product is possible. Combine the different products only if their intended uses and relevant technical data (working length, diameter, etc.) are the same. Follow the instruction manual of the products used in combination with this product.
.
NOTE! We recommend using a sterile objective lens. When using an unsterile objective lens, the sterilizable adapter 8580.91 can be connected between the endoscope and objective lens as an option to fasten the drape. This will allow subsequent endoscope changes without contaminating the sterile field (see section 7.6). Fig. 1 Adapter 8580.91 option
Endoscope Fig. 1
GA-S 022
Drape Objective lens and camera head
2
6
Illustration
3
1
4
3
5
1
5
# 6
2.1
#
3
#
2
2.2 #
2
#
# 7
*
*
2
2 8
8
Fig. 2
6.1
Legend and identification
3
Item
Designation
Item
Designation
1
Snap-on mechanism
6
Zoom ring
2
Clamping mechanism
7
Clamping screw
2.1
Grip
8
Focal length (f)
2.2
Tab
3
Focusing ring / focusing screw
#
Product no.
4
Cap for snap-on mechanism
*
Serial no.
5
Cap for C-mount thread
GA-S 022
Symbol
Designation Follow the instruction manual
REF
Order number
LOT
Lot designation
SN
Serial no. Identification in conformity with EEC directive 93/42/EEC on medical products, is only valid if the pro duct and/or its packaging is provided with this identification. Products of category IIa and above, as well as sterile products or products with measuring function of category I, are additionally marked with the code no. of the notified body (0124).
GA-S 022
4
6.2
System overview of RIWO objective lenses
6.2.1
Camera heads with ¼“ - sensors
ENDOSCOPES
OBJECTIVE LENSES
CAMERA HEADS
85264.501 flexible
85264.121/ .171
85264.231/ .271/ .381
with fixed eyepiece with eyecup Adapter 8580.91 (option) for all clamping mechanisms
1 CCD camera head, RW-mount e.g. 5520.901 8580.91
Plug-on eyecup 8885.901
with plug-on connection 85266.231/ .271
Fig. 3
5
85267.231/ .271
Direct connection optional
GA-S 022
6.2.2
Camera heads with ⅓“ - sensors
ENDOSCOPES
OBJECTIVE LENSES
Adapter 8580.91 (option) for all clamping mechanisms
flexible
CAMERA HEADS
8580.91
85261504
3-chip HD camera head, C-mount e.g. 5509xxx
with fixed eyepiece with eyecup 1 CCD HD camera head, C-mount e.g. 85509xxx
85261144 / 244 with plug-on connection
Fig. 4
6.2.3
Plug-on eyecup 8885.901
Direct connection optional
Camera heads with ½“ - sensors
ENDOSCOPES
OBJECTIVE LENSES
85261.502 flexible
Adapter 8580.91 (option) for all clamping mechanisms
CAMERA HEADS
3 CCD camera head, C-mount 5506.xxx, 5516.xxx
8580.91
with fixed eyepiece with eyecup 85261.501
1 CCD camera head, C-mount e.g. 5512.xxx 85261.172/ .212/ .242/ .272/ .322/ .382 with plug-on connection
Fig. 5
GA-S 022
Plug-on eyecup 8885.901
Direct connection optional
6
6.3
RIWO objective lenses and adapters
6.3.1
RIWO objective lenses with C-mount thread and snap-on mechanism / clamping mechanism autoclavable
Illustration
134_C / 272_F
Designation
Focal length (f)
Product no.
RIWO objective lens with clamping mechanism C-mount thread, immersible and autoclavable
f = 17 mm f = 21 mm f = 24 mm f = 27 mm f = 32 mm f = 38 mm
85261.172 85261.212 85261.242 85261.272 85261.322 85261.382
f = 14 mm
85261144
f = 24 mm
85261244
f = 13 - 29 mm
85261504
f = 21 - 36 mm
85261.501 85261.502
Designation
Focal length (f)
Product no.
Miniature RIWO objective lens with clamping mechanism RW-mount thread, immersible and autoclavable
f = 23 mm
85264.231
f = 27 mm
85264.271
f = 38 mm
85264.381
f = 12 mm
85264.121
f = 17 mm
85264.171
f = 12 - 26 mm
85264.501
for ENDOCAM 5501, 5503, 5506, 5507, 5509, 5511, 5512, 5516, 5508 (only for 85261.172)
RIWO objective lens with snap-on mechanism C-mount thread, immersible and autoclavable
134_C / 272_F
for ENDOCAM 5509, 5525
RIWO zoom objective lens with snap-on mechanism C-mount thread, immersible and autoclavable
134_C / 272_F
for ENDOCAM 5509, 5525
RIWO objective lens with clamping mechanism C-mount thread, immersible and autoclavable
134_C / 272_F
for ENDOCAM 5501, 5503, 5506, 5507, 5509, 5511, 5512, 5516
6.3.2
RIWO objective lenses with RW-mount thread and clamping mechanism autoclavable
134_C / 272_F
Illustration
for ENDOCAM 5501, 5502, 5511
134_C / 272_F
Miniature RIWO objective lens with clamping mechanism RW-mount thread, immersible and autoclavable for ENDOCAM 5520
134_C / 272_F
Miniature zoom objective lens with clamping mechanism RW-mount thread, immersible and autoclavable for ENDOCAM 5520
7
GA-S 022
6.3.3
RIWO-objective lenses with RW-mount thread and plug-on connection autoclavable
Illustration
Designation
Focal length (f)
Product no.
f = 23 mm
85266.231
f = 27 mm
85266.271
f = 23 mm
85267.231
f = 27 mm
85267.271
Miniature RIWO objective lens with plug-on connection RW-mount thread, immersible and autoclavable
134_C / 272_F
for ENDOCAM 5501, 5502, 5511
Miniature RIWO objective lens with pendulum and plug-on connection RW-mount thread, for image alignment, immersible and autoclavable
134_C / 272_F
for ENDOCAM 5501, 5502, 5511
6.3.4
Adapters for RIWO objective lenses autoclavable
Illustration
Designation Plug-on eyecup immersible and autoclavable for the connection of C-mount objective lenses to endoscopes (telescopes) with plug-on connection
134_C / 272_F
Plug-on eyecup, rotatable specially designed for VIDEO-TUR immersible and autoclavable
Product no.
8885.901
8885.902
for the connection of C-mount objective lenses to endoscopes (telescopes) with plug-on connection
C-mount / RW-mount adapter immersible and autoclavable 134_C / 272_F
134_C / 272_F
GA-S 022
for the connection of C-mount objective lenses to RW-mount camera heads for CCD ENDOCAM 5502
Adapter, sterilizable immersible and autoclavable
8885.952
8580.91
allowing a telescope change
8
7
Use CAUTION! The products have only limited strength! Applying excessive force will lead to damage, impair the function and endanger the patient. Immediately before and after each use, check the products for damage, loose parts and completeness. Do not use the products if they are damaged or incomplete or have loose parts.
7.1
Preparation
Z Carry out a check: see section 8
.
7.1.1
IMPORTANT! Before assembling the objective lens, make sure that the glass surfaces of the objective lens and camera head are dry and dust-free. Remove any soiling with a cotton swab soaked with alcohol (wooden or plastic stick, not metal).
Removing the caps / objective lens protection caps RIWO objective lenses with snap-on mechanism (1) Fig. 6 Unlocking: Z Push the snap-on mechanism (1) in direction of arrow (Fig. 6a). ' The snap-on mechanism (1) locks in open position. Z Remove the cap (4). Z Unscrew the cap (5) from the C-mount thread (Fig. 6b)
1 Fig. 6a
Fig. 6b
4
Fig. 6
85261504
5
RIWO objective lenses with clamping mechanism (2) Fig. 7 Z Push together the tab (2.2) and the nose (2.1) as indicated by the arrows "open" and remove the cap.
2.1
2 2.2
Fig. 7
9
GA-S 022
7.2
Connecting the objective lens to the camera head Fig. 8 Z Screw the objective lens in clockwise direction onto the camera head. ' Check for correct connection.
Fig. 8
7.3
Connection C-mount / RW-mount adapter 8885.952 The C-mount / RW-mount adapter allows the connection of C-mount objective lenses to RW-mount camera heads. Fig. 9 Z Screw the C-mount adapter in clockwise direction onto the camera head.
8885.952 Fig. 9
GA-S 022
10
7.4
Connecting the RIWO objective lens, RIWO angular objective lens to an endoscope with eyecup The endoscope can be fastened to the clamping mechanism (2) in two ways either a rotatable or non-rotatable connection.
7.4.1
RIWO objective lens with snap-on mechanism (1) Usually the snap-on mechansim (1) is unlocked (open) and preloaded. Fig. 10 Locking: Z Place the endoscope onto the objective lens. ' The snap-on mechanism (1) lock automatically. The endoscope is locked in place. Z Check for secure connection. Unlocking: Z Open the snap-on mechanism (1). ' see section 7.1.1, Fig. 6.
1
Fig. 10
7.4.2
RIWO objective lens, RIWO angular objective lens with clamping mechanism (2) Fig. 11 Locking the endoscope Z Push together the tab (2.2) and the nose (2.1) as indicated by the arrow "open", insert the endoscope, and let go of the tab (2.2). Z Check for secure connection. Locking the rotatable clamping mechanism (2) Z Push the tab (2.2) in direction of arrow "close" while holding the nose (2.1). Z Check for secure connection. 2.1 2.1
2.2
2.2
2
Fig. 11
11
GA-S 022
7.5
Connecting the objective lens to an endoscope with plug-on eyepiece adapter Fig. 12 Z Before the connection, open the clamping screw (7) in direction of arrow "open" as far as it will go. Z Insert the endoscope into the objective lens as far as it will go. ' The endoscope is rotatable. Z To lock the endoscope in place, turn the clamping screw (7) in direction of arrow "close".
7 open close
Fig. 12
7.5.1
Connecting the objective lens to an endoscope with plug-on eyecup 8885.901 / 8885.902 The plug-on eyecup 8885.901 / 8885.902 allows the connection of all objective lenses with snap-on mechanism (2) / clamping mechanism (2). The rotatable plug-on eyecup 8885.902 has been specially designed for VIDEOTUR. Fig. 13 Z Connection of plug-on eyecup to endoscope: see section 7.5. Z Connection of objective lens to the plug-on eyecup: see section 7.4.
7 open close
Fig. 13
GA-S 022
12
7.6
Sterilizable adapter 8580.91 (option) Fig. 14 For fastening to camera drape Connecting or changing the endoscope during use without interrupting the sterile chain.
Fig. 14
.
IMPORTANT! The use of the adapter is only possible for objective lenses with snap-on me chanism (1) / clamping mechanism (2)! Fig. 15 Z Insert the adapter into the drape from the camera side. Z Enlarge the opening of the adhesive tape and fasten the drape to the adapter (Fig. 15a). Z Connect the objective lens with camera head and endoscope to the adapter (Fig. 15b). Adapter
Fig. 15b
Fig. 15a
Adapter
Fig. 15
13
GA-S 022
7.7
Additional notes and instructions for use
7.7.1
Zoom objective lenses
3
4
Image magnification Fig. 16 Z Adjusting to long focal lengths: Turn the zoom ring (4) in direction of arrow ' For full format image see Fig. 16a.
.
CAUTION! Loss of image information when using full format setting! When using the full format setting (Fig. 16a), parts of the endoscopic image (hatched surface) are not displayed on the monitor.
Fig. 16
Image demagnification Z Adjusting to short focal lengths: Turn the zoom ring (4) in direction of arrow . ' Inscribed circle, see Fig. 16b.
.
Fig. 16a
NOTE! The complete endoscopic image (gray surface) is displayed on the monitor (Fig. 16b).
Focusing Z Turn the focusing ring (3) until the image on the monitor is in focus. . 'For the close-up range turn in direction of arrow . 'For distant focusing turn in direction of arrow
Fig. 16b
7.7.2
RIWO objective lenses with snap-on mechanism (1)
.
NOTE! In the case of a defect (repair case), the snap-on mechanism can be re placed.
Fig. 17
(x)
1
Fig. 17
GA-S 022
(y)
Z Unscrew the socket head screws (x) using the supplied Allen wrench size 1.5 (y) and remove the snap-on mechanism (1). Z Mount the new snap-on mechanism (1) to the RIWO objective lens using the supplied socket head screws (x). Z Check for secure and correct connection. Z Carry out a function check: see section 8.2
14
8
Checks CAUTION! Be careful if products are damaged or incomplete! Injuries of the patient, user or others are possible. Run through the checks before and after each use. Do not use the products if they are damaged, incomplete or have loose parts. Return damaged products together with any loose parts for repair. Do not attempt to do any repairs yourself.
8.1
Visual check
Z Check the RIWO objective lenses and accessories for ' damage ' sharp edges ' loose or missing parts ' rough surfaces Z Any lettering, labeling or identification necessary for the safe intended use must be legible. ' Missing or illegible inscriptions and labels that can lead to wrong handling and reprocessing must be reinstated. Z When screwing the objective lens onto the camera head or when connecting the endoscope, the glass surfaces of the connection must be dry and dustfree.
8.2
15
Function check
Z Check the individual components for compatibility. Z Check for secure connections. Z Check the individual instruments for easy assembly and proper functioning of the locking mechanism(s). Replace the instruments if the connection ' although locked is not secure. ' cannot be locked or is difficult to lock. Z Check the function of the snap-on mechanism (1) / clamping mechanism (2) or plug-on connection.
GA-S 022
9
Reprocessing and maintenance
9.1
Disassembly before cleaning Z Remove the caps from the RIWO objective lens: see section 7.1.1
9.2
Illustrations on reprocessing
9.2.1
Manual cleaning
.
IMPORTANT! Manual cleaning is possible with the camera head in place. In the case of frequent use, unscrew the objective lens from the camera head once a week, clean the thread from any residues and apply 1 - 2 drops of oil. Remove any excess oil.
.
IMPORTANT! Do not clean the plastic parts with metal or sharp-edged tools (e.g. brushes). Fig. 18
4
5
Fig. 18
9.3
Reprocessing procedures
The following tables contain a description of the reprocessing procedures for the indi vidual products validated by Richard Wolf.
GA-S 022
.
IMPORTANT! ' When used as intended and when the manual is observed it is not necessary to limit the number of possible reprocessing cycles. ' Careful and gentle handling during the entire reprocessing process has an es sential influence on the service life of medical products. ' Before being returned for repair, defective products must have gone through the entire reprocessing cycle. ' The user is obliged to make sure that the reprocessing process including the resources, materials and personnel is suited to yield the required results. ' Follow the national and international validation requirements.
.
IMPORTANT! Reprocess the products before every application and before returning for repairs as required by the instruction manual in order to protect the patient, user, and others.
.
IMPORTANT ! Do not sterilize RIWO objective lenses in hot-air sterilizers.
16
Product:
RIWO objective lenses, adapters and caps
Reprocessing guide: Preparation at the point of use:
Directly after use, remove any coarse soiling from the RIWO objective lens. Do not use any fixing or setting agents or hot water (>40C) as this will bake the residues to the surfaces and may have a negative impact on cleaning.
Transport:
Safe storage and transport of the instruments to the reprocessing room in a closed container to prevent damage to the instruments and contamination of the environment.
Precleaning:
No special requirements.
Cleaning:
Manual
Machine
See section 9.2.1, Fig. 18
Place the RIWO angular objective lens and caps in sieve tray of the washer-disinfector.
1.Immerse in a cleaning solution for at least 5 minutes, then brush inside. 2.Brush with a soft brush for at least 5 seconds to remove any residues. 3.Finally, rinse the RIWO objective lens, adapters and caps thoroughly with processed water and dry with a lintfree disposably cloth. 4.Clean the glass surfaces on the RIWO objective lens with a cotton swab soaked with alcohol (wooden or plastic swab carrier, not metal).
' Prewash for >4 min. with cold water ' Empty ' Wash for >6 min with a cleaning agent at 55C ' Empty ' Neutralize*) for >3 min with hot tap water (>40C) ' Empty ' >2 min of intermediate rinsing with hot tap water (>40C) ' Empty
Disinfection:
1.Immerse the RIWO objective lens, adapters and the Carry out thermal machine disinfection following the na caps in an approved disinfectant solution, exposure tional requirements regarding the A0 value (see DIN EN time as specified by the manufacturer. ISO 15883). 2. Finally, rinse the RIWO objective lens and caps thor oughly for at least 20 seconds with processed water**). 3.Clean the glass surfaces on the RIWO objective lens with a cotton swab soaked with alcohol.
High Level disinfection:
Cidex OPA, Cidex Glutaraldehyde
Drying:
Manual drying: Dry outside of RIWO objective lens, adapters and caps with a lint-free disposable cloth or swab or alternatively in a drying cabinet.
Function test, visual check maintenance:
Check visually for cleanliness. If necessary, repeat the reprocessing process until the RIWO objective lens, ad apters and caps are visually clean. ' Carry out a visual check: see section 8.1
Packaging:
Package the RIWO objective lens, adapters and caps for sterilization as required by the standards.
Sterilization:
Pre-Vac steam cycle at 270°F (132C) with 4 minutes exposure and 20-30 minutes dry time.
Storage:
Store the sterilized instruments in a restricted area at approximately 24°C / 75°F, with at least 4 air exchanges per hour and a relative humidity that does not exceed 70%, in accordance with ANSI / AAMI ST79.
Drying of instruments using the washer-disinfector drying cycle. If necessary, additional manual drying can be achieved with a lint-free disposable cloth or swab or alternatively in a drying cabinet. Dry hollow spaces with filtered com pressed air.
*) depending on the quality of the neutralization solution, add an acid based on citric acid. **) use sterile water if manual disinfection is not followed by sterilization.
9.4
17
Alternate Sterilization Methods
Z For possible alternate sterilization methods, refer to our website, www.richardwolfusa.com under „Reprocessing & Sterilization“ for listings of Richard Wolf instruments that have been approved for various reprocessing methods.