1. Important information – read prior to start-up You have acquired a high-quality RIESTER examination light manufactured in compliance with Directive 93/42/EEC for medical products and subject to continuous stringent quality control. Should you have any queries, please contact the Company or your RIESTER Agent who will be pleased to assist you. For addresses see last page of these Instructions for Use. The address of your authorised RIESTER Agent will be supplied to you on request. Please note that correct and safe operation of instruments will only be guaranteed when RIESTER instruments and accessories are used throughout.
2. Purpose The universally applicable ri-magic® halogen light by Riester has been manufactured for lighting and therefore diagnosis and minor surgery in many medical fields, offering a light intensity of approx. 100.000 lux at a distance of 100 mm.
3. Safety and electromagnetic compatibility Meaning of symbols on power pack and light plinth: Attention: Comply with Instructions for Use! Protection class II instrument The light complies with the requirements relating to stability and emission of electromagnetic interference. Please note that interference may not be completely ruled out under the increased influence of detrimental field strengths, i.e. when operating mobile telephones and radiological equipment. Note • There may be a risk of gas ignition when the light is used in the presence of inflammable drug mixtures such as air and/or oxygen or laughing gas! • The light should only be opened by skilled persons to avoid a risk of lethal electrocution. • Touch light head after extended use only at the controls provided (9 = rheostat incl. an On/Off switch, 10 = focusing ring). Sections of the light socket in particular will become very hot after extended use.
4. Assembly instructions prior to start-up a) ri-magic® desk model Attach universal clamp (1) and set screw (2) to a desk or wall rail. Insert universal light (4) from top into the bore of the screw clamp and fix with the knurled bolt (5). Connect cable plug (6) of the power pack (7) to the socket (8) from below until engaging with a click. Connect a transformer (7) to the socket. The light is now ready for operation. b) ri-magic® wall model Hold wall bracket (1a) to the desired point on the wall and mark 4 holes for dowels. Remove wall bracket and drill the 4 marked holes. Insert dowels into the wall. Apply wall bracket and fix with screws. Insert light (4) into bracket from above and fix by the locking screw (3a) on the side. Connect cable plug (6a) of power pack (7a) to socket (8a) from below until it engages with a click. Connect transformer (7a) to the socket. The light is now ready for operation. c) ri-magic® stand model Push bracket (17) into the tripod tube (11) and insert tripod tube (11) and bracket (17) into the mobile bottom section (12) by fixing with the knurled bolt (18) on the side and fix with the enclosed bolt (13) and locking ring (14). The tripod (base) is provided with 5 castors, two of which may be locked (15). Tilt lever (16) down for locking. Attach universal light (4) into bracket from above and fix with the knurled bolt (5b). Connect cable plug (6b) of power pack (7b) to the socket (8b) on the side until an audible click is heard. The light is now ready for operation.
5. Start-up and operation Initially connect the cable plug (6,a,b) to the socket (8,a,b). Connect power pack (7,a,b) to the mains socket. Voltages on the rating plate of the mains adapter must be noted at all times. There are two optional voltages: 230 V and 120 V.
7
On-Off Turn the rheostat (9) in clockwise direction. When a click is heard, the light is switched on. Turn the rheostat (9) in counter-clockwise direction until a click is heard. The light is now switched off. Rheostat The desired light intensity may be set by turning the rheostat (9). Increase light intensity by turning in clockwise direction and decrease by turning in counter-clockwise direction. Focusing A focusing ring (10) is provided at the front of the light head for controlling the light cone of the light. Note: • The flexible arm of the light should not be bent in too small a radius in order to prevent premature material fatigue. • For normal operation of the light, the bracket (4) and the universal clamp (1) and/or the tripod bracket (17) will heat up around the socket (8,a,b).
6. Light 6.1 Replacement of light Prior to replacing the light, switch light off and disconnect plug from mains, followed by letting the light head cool down for a few minutes. Turn the front section of the light head, together with the focusing ring in counterclockwise direction to remove. The light and mirror may now be carefully withdrawn from the socket. Insert a new light in reverse order. Ensure that the glass section is not touched. 6.2 Specification Light 8V
20W
2500 mA
average life 50 h
7. Cleaning and disinfection Prior to cleaning or disinfection of the light, disconnect plug from mains socket Cleaning and/or disinfection The ri-magic® examination light may be externally cleaned with a damp cloth. In addition, the following disinfectants may be used for disinfection: Aldehyde (formaldehyde, glutaraldeyhde, aldehyde derivatives), surfactants or alcohols. Means for cleaning or disinfection may be a soft, possibly lintfree cloth or Q-tips. Sterilisation According to the ruling teaching (Tübingen Medical Products Test Centre), sterilisation is only specified for surgery. There is no reason for sterilisation of a diagnostic light.
8. Spare parts Item No. 11311 Item No. 11341 Item No. 11342 Item No. 11345 Item No. 11346
Replacement light 8 V Transformer 230 V Transformer 120 V Wall bracket Universal clamp
9. Specification Model: Connection: Type of lamp: Operating temperature: Storage: Weight:
ri-magic® desk/wall/stand light For mains connection see details on power pack (230 V/120 V) Halogen light 8 V +10° to +40°C, relative humidity between 30 and 75% (non-condensing) -5° to +50°C, up to 85% relative humidity (non-condensing) Desk and wall model: approx. 3 kg Stand model: approx. 9 kg
10. Maintenance These instruments and their accessories do not require any specific maintenance. Should an instrument have to be examined for any specific reason whatsoever, please return it to the Company or an authorised RIESTER dealer in your area. Addresses to be supplied on request. 8