User Manual
8 Pages
Preview
Page 1
USER MANUAL: ®
Rocket Craft™ Suction Pumps - All Models
Head Office & Export Customer Services Rocket Medical plc Imperial Way. WATFORD. WD24 4XX ENGLAND Tel: +44 (0) 1923 651400 Fax: +44 (0) 1923 230212 Email: [email protected]
UK Customer Services Rocket Medical plc. Sedling Road. WASHINGTON. NE38 9BZ. ENGLAND Tel: +44 (0) 191 419 6988 Fax: +44 (0) 191 419 6989 Email: [email protected]
US Office: Rocket Medical Unit 3. 150 Recreation Park Drive HINGHAM. MA 02043. USA Tel: 1 781 749 6223 Fax: 1 781 749 6235 Email: [email protected]
USERS MANUAL – ROCKET® CRAFT™ SUCTION PUMPS – ALL MODELS Page 2
1. ROCKET® CRAFT™ SUCTION PUMPS 1.1. General Description: The Rocket® Craft™ Suction Pumps have been developed to provide a smooth, low volume vacuum at a pre-determined negative pressure. Vacuum is activated by a foot operated switch controlled by the surgeon performing oocyte collection. The range of vacuum is variable from 0-400mm Hg and adjustable over this range by a pre-set control. Vacuum is provided by a quiet diaphragm pump, working from a mains voltage supply and manufactured to BS 5724 Class 1 and CE safety regulations. The Rocket® Craft™ is supplied with both a reusable water trap set as standard designed to help prevent aspirated fluids being drawn into the vacuum power unit. The trap also acts as a reservoir to provide an even vacuum pressure during procedures. A single patient use, disposable water trap set is also available. The foot control has an air-operated solenoid, switching the main pump unit. No electrical connections are made to the sealed foot operated control. A suitable collection test tube for use with oocyte needle sets is B.D. Falcon test tube No. 2001F, 17 x 100mm.
1.2. References: Craft I, McLeod F. Edmonds K, ‘Human embryo transfer technique’. Lancet 1961 ii 1104-5 Craft I, Diahanbakch O. McLeod F et al ‘Human pregnancy following oocyte and sperm transfer to the uterus.’ Lancet 1992 i 1031-3 Craft I, (1984) ‘Clinical Methodology’ British Journal of Hospital Medicine 90-102
ZDOCK105
Rev. 07
15/01/09
Copyright© ROCKET MEDICAL PLC 2001- 2009
USERS MANUAL – ROCKET® CRAFT™ SUCTION PUMPS – ALL MODELS Page 3
1.3. Warranty Rocket® Craft™ 220V Suction Pumps are sold by Rocket Medical plc under the warranties set forth in the following paragraphs. Such warranties are extended only with respect to the purchase of the Products directly from Rocket Medical plc as new merchandise and are extended to the first Buyer thereof, other than for resale. For a period of twelve (12) months from the date of shipment the Products are warranted to be free from functional defects in materials and workmanship and to conform to the description of the Products contained in the operating manual and accompanying labels, provided the same is properly operated under conditions of normal use, that annual maintenance and service is performed at an authorised Rocket Medical plc service facility Removal of any QC seal voids the warranty. The foregoing warranties shall not apply if the Products have been repaired other than by Rocket Medical plc or other than in accordance with written instructions provided by Rocket Medical plc , or altered by anyone other than Rocket Medical plc, or if the Products have been subject to misuse, negligence, or accident. Rocket Medical plc 's sole and exclusive obligation and Buyer's sole and exclusive remedy under the above warranties is limited to repairing or replacing, free of charge, at Rocket Medical plc 's option, Products, which are telephonically reported to Rocket Medical plc and which, if so advised by Rocket Medical plc, is thereafter returned with a statement of the observed deficiency, not later than seven (7) days after the expiration date of the warranty, to Rocket Medical plc during normal business hours, transporting charges prepaid and which, upon Rocket Medical plc's examination, is not found to conform with the above warranties. Rocket Medical plc shall not be otherwise liable for any damages including but not limited to incidental damages, consequential damages or special damages. There are no express or implied warranties which extend beyond the warranties herein above set forth. Rocket Medical plc makes no warranty of merchantability or fitness for a particular purpose with respect to the Products or parts thereof.
1.4. Service and Repairs: Craft™ Pumps normally require little routine maintenance, however they must be serviced and calibrated annually at an Rocket Medical plc service facility. Please contact any of our Customer Services Departments for details of your nearest service centre
ZDOCK105
Rev. 07
15/01/09
Copyright© ROCKET MEDICAL PLC 2001- 2009
USERS MANUAL – ROCKET® CRAFT™ SUCTION PUMPS – ALL MODELS Page 4
2. OPERATING INSTRUCTIONS ROCKET® CRAFT™ VACUUM UNIT R29.654 & R29.655 9
Connect main lead to electrical supply 230/240 volt 50 Hz (R29.655) or 110/120 volt 60 Hz. (R29.654)
9
The reusable water trap set should be packed and sterilised by autoclaving at 135°C @ 3 bar for 3 mins. prior to each use.
9
Rocket Medical plc recommend the use of Rocket® Craft™ Single Use Water Trap sets which are available from your local agent or Customer Services.
9
Using aseptic technique, unpack the water trap and insert the glass water trap bottle into the clip at the left hand side of the casing.
9
Connect the short tube to the port marked SUCTION and secure with a light push fit.
9
Pass the longer patient connection tube into the operative field.
9
Connect foot switch tube to connector on front panel.
9
Activate vacuum by depressing foot switch. The pump will stop when the foot is lifted from the pad.
9
Set the required vacuum required by occluding the vacuum line distal (patient side) to the water trap and adjusting the control to the desired level as shown by the needle on the gauge
Craft™ Single Use Water Trap Set
9
During the aspiration of the fluid the control knob may be adjusted to the precise vacuum require under load, approximately 100-200mm Hg.
9
NOTE: Thinner gauge needles (17G) and those with longer tube sets, 7090cm may require vacuums in the region of 150-200mmHg to achieve satisfactory flow rates.
9
The unit is then ready for use.
9
If vacuum lines become blocked due to debris or viscous fluids, a temporary high vacuum can be obtained to unblock the needle and tubing by occluding the finger bleed hole on the front panel, marked “OCCLUDE FOR MAXIMUM VACUUM”. High vacuum should NOT be used to aspirate oocytes.
9
Placing a finger on the red disc for a few seconds will clear the tubing lines at 300-400mm Hg temporary high vacuum.
9
On completion depress front panel switch at “O” to switch the unit off.
Craft™ Suction Pump with single foot control, shown with re-us
ZDOCK105
Rev. 07
15/01/09
Copyright© ROCKET MEDICAL PLC 2001- 2009
USERS MANUAL – ROCKET® CRAFT™ SUCTION PUMPS – ALL MODELS Page 5
3. OPERATING INSTRUCTIONS ROCKET® CRAFT™ VACUUM UNIT R29.660 & R29.661 9 9
Connect main lead to electrical supply 230/240 volt 50 Hz (R29.660) or 110/120 volt 60 Hz. (R29.661) Connect the dual foot switch tubes to the push connectors on front panel.
9
The reusable water trap set should be packed and sterilised by autoclaving at 135°C @ 3 bar for 3 mins prior to each use.
9
Rocket Medical plc recommend the use of Rocket® Craft™ Single Use Water Trap sets which are available from your local agent or Customer Services.
9
Using aseptic technique, unpack the water trap and insert the glass water trap bottle into the clip at the left hand side of the casing. Reusable water trap sets are available, contact your local agent or Customer Services.
9
Connect the short tube to the port marked SUCTION and secure with a light push fit.
9
Pass the longer patient connection tube into the operative field.
9 9
Connect foot switch tube to connector on front panel. Activate vacuum by depressing LOW VACUUM (Larger) foot switch. The pump will stop when the foot is lifted from the pad.
9
Set the required vacuum required by occluding the vacuum line distal (patient side) to the water trap and adjusting the control to the desired level as shown by the needle on the gauge
9
During the aspiration of the fluid, the control knob may be adjusted to the precise vacuum required under load, approximately 100-200mmHg.
9
NOTE: Thinner gauge needles (17G) and those with longer tube sets, 70-90cm may require vacuums in the region of 150-200mmHg to achieve satisfactory flow rates.
9
The unit is now ready for use.
9
If vacuum lines become blocked due to debris or viscous fluids, a temporary high vacuum can be obtained to unblock the needle and tubing by depressing the smaller HIGH Vacuum foot switch this provides a temporary high vacuum up to 440mmHg.
9
High vacuum should NOT be used to aspirate oocytes.
9
On completion depress font panel switch at “O” to switch the unit off.
®
Rocket Craft™ Single Use Water Trap Set
Craft™ DuoVac with twin foot control and reusab
ZDOCK105
Rev. 07
15/01/09
Copyright© ROCKET MEDICAL PLC 2001- 2009
USERS MANUAL – ROCKET® CRAFT™ SUCTION PUMPS – ALL MODELS Page 6
Rocket® Single Use Water Trap Set for use with all Rocket® Craft™ Suction Pumps The Rocket® Water Trap Set for use with Rocket® Craft™ Suction Pumps offers a simple, cost effective solution to contamination risks and reuse problems. • COMPLETE SET: Provides the complete bottle and tubing set in single, sterile pack for one time patient use. • GLASS TRAP BOTTLE: To maintain clear view in the event of potential pump contamination with aspirate. Designed to withstand >450mmHg vacuum. • REMOVES CONTAMINATION RISK from irregular or uncontrolled decontamination of reusable bottle, bung and angle tube set. • PATIENT TUBE SET: 2m patient tube set with integral luer lock connector to ensure secure, reliable connection to the aspiration needle set. • SUITABLE FOR USE: Compatible with all versions of Craft™ Suction and Duo Vac. • Also compatible with the genX and Pioneer Pro Pumps.
Ordering Information Rocket® Water Trap Set suitable for all Rocket® Craft™ Suction Pumps
Sterile each
R57.685
Packed individually, 10 per carton
ZDOCK105
Rev. 07
15/01/09
Copyright© ROCKET MEDICAL PLC 2001- 2009
USERS MANUAL – ROCKET® CRAFT™ SUCTION PUMPS – ALL MODELS Page 7
4. SPARE PARTS: CRAFT SUCTION PUMPS: Item
1. 3. 10. 11. 12. 14. 15. 18. 19. 20. 21/22 22. 23. 25. 28 104.
Part No. R29.655 R29.654 B14-011 B02-009 B02-140 F06-006 R06-027 C12-030 F04-027 F04-023 E05-000 N05-016 K01-001 O04-000 G03-005 C01-035 I08-004 S10-027 C01-034 B05-005
Description CRAFT SUCTION PUMP, 240V. CRAFT SUCTION PUMP, 110V. Box, lower half Bottle Bracket Bottle Bracket assembly (bracket & white plastic coated clip) Front Panel Fin Rear Panel 240v Chassis bare Filter assembly (Silencer) Filter housing assembly Filter Element (part of item 15) Needle Valve assembly Knob Pointer Vacuum Control Orifice assembly Vacuum Gauge 0-600mmHG and connector assembly (items 21 & 22) Connector IEC Connector Switch 240v Flanged Connector Footswitch (Cat.No. R29.656)
59. 11. 19. 44 13. 34. 30. 42. 56. 55. 48. 58. 51. 50. 45. 52. 53 104. 22. -
R29.660 R29.661 B14-009 B14-011 C12-044 F06-022 B02-009 S10-007 S18-004 I08-004 S10-001 N05-014 K01-001 G03-004 B20-010 N06-002 N06-003 C01-034 C01-179 C01-178 B05-004 F04-004 C01-176 C01-177
CRAFT DUO-VAC SUCTION PUMP 240V. CRAFT DUO-VAC SUCTION PUMP 110V. Case Upper Half Case Lower Half Chassis Front Panel Bottle Bracket Pressure Switch Solenoid Valve IEC Connector Switch/Indicator Stop Pin/Nut (Part of N05-016 Needle Valve Assembly) Knob Vacuum Gauge 0-600mmHG Ventilator blanks Neon Indicator Green Neon Indicator Red Flanged connector Plastic Connector MALE (on front of Pump under standard vacuum green indicator) Plastic Connector FEMALE (on front of Pump under high red vacuum indicator) Footswitch, Dual Vacuum (Cat.No.R29.662) Filter Assembly (silencer) Set of 2 Plastic connectors for Footswitch, high vacuum Green Connector (Female) for lead on Dual Footswitch Red Connector (male) for lead on Dual Footswitch
ZDOCK105
Rev. 07
15/01/09
Copyright© ROCKET MEDICAL PLC 2001- 2009
USERS MANUAL – ROCKET® CRAFT™ SUCTION PUMPS – ALL MODELS Page 8
4.1.MOTOR SPARES & ACCESSORIES Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Part No. SO1-214 WO1-000 SO1-138 HO1-001 VO1-002 VO2-001 SO1-203 DO2-001 GO1-001 DO1-053 CO2-031 BO2-035 SO1-104 WO1-051 SO1-018 C03-084 MO1-311
15
MO1-312
-
S02-012
ZDOCK105
Rev. 07
Description Screw 4BA x 3/4" pan pozi b'zinc Washer for the above (flat) 6BA b'zinc Head Screw (centre) no. 4 x 1/4" pan pozi Head 800111 Valve diaphragm 800109B EPDM Valve plate 800110 Screw 6BA X 5/16" csk S/S Diaphragm plate upper 700172 Gasket 690260 Grey Poly Pump Diaphragm 700190C EPDM Crankcase 770137 Bracket 820170 blk. anodis Screw no 4 x 3/16" pan pozi 'B' b'zinc Washers S/P int. M4 bzn Screws M4 x 10 pan pozi bzn Complete Conrod Assembly, including top diaphragm plate Motor 770139 220/240v (includes motor, crank case and valves) Motor 730190 110/120v 50 Hz. (includes motor, crank case and valves) Service Kits (EPDM) includes items 3 and 8 plus a washer
15/01/09
Copyright© ROCKET MEDICAL PLC 2001- 2009
Qty. 4 4 1 1 1 1 1 1 1 1 1 1 4 2 2 1 1 1 1