rocket_r54569_psu_service_manual_rev_01_may_2017.pdf
Page 1
Rocket Medical
R54569 Digital low vacuum/low flow Suction Pump
Service Manual
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 2
Rocket Digital low vacuum/low flow Suction Pump – Service Manual
Revision History
Rev. No.
01
Rev. Date
080517
Summary of Changes
Initial Release
Chapters Affected
-
2
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 3
Service Manual for Rocket Chest Drain Pump.
Purpose of Document
This document is to assist trained service personnel in the routine maintenance and service of
the Rocket Chest Drain (CDSU).
Table of Contents
REVISION HISTORY .................................................................................................................................................................. 2
SERVICE MANUAL FOR ROCKET CHEST DRAIN PUMP. ........................................................................ 3
PURPOSE OF DOCUMENT ........................................................................................................................................................ 3
1 Safety ................................................................................................................................................................................ 4
1.1 Special Precautions................................................................................................................................................ 4
1.2 Electrical Shock Hazard ........................................................................................................................................ 4
1.3 Explosion Hazard..................................................................................................................................................... 4
2 Operating manual ....................................................................................................................................................... 5
3 Basic Trouble shooting guide ................................................................................................................................. 5
4 Specification .................................................................................................................................................................. 6
5 Warnings ........................................................................................................................................................................ 7
6 Technical Description................................................................................................................................................ 8
6.1 Power Supply. ........................................................................................................................................................... 8
6.2 User Interface............................................................................................................................................................ 8
6.3 System Control ......................................................................................................................................................... 8
6.4 System Safety ............................................................................................................................................................ 8
7 Calibration checks....................................................................................................................................................... 9
8 Routine service tasks .............................................................................................................................................. 10
9 Disassembly / Reassembly................................................................................................................................... 11
9.1 Unit Top assembly................................................................................................................................................ 12
9.2 Unit Base assembly .............................................................................................................................................. 12
9.3 PCB Top .................................................................................................................................................................... 13
9.4 PCB Bottom ............................................................................................................................................................. 13
9.5 PCB Top Overlay ................................................................................................................................................... 14
9.6 PCB Bottom Overlay ............................................................................................................................................ 15
10 Circuit Diagram ...................................................................................................................................................... 16
11 Parts Listing ............................................................................................................................................................. 20
12 Contact Details ........................................................................................................................................................ 24
3
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 4
1
Safety
The CDSU is an external, mobile suction device for use with the Rocket Medical Chest
Drainage range. Rocket Medical chest drains are required to prevent the build-up of air
inside the body (pleura) whilst breathing and also remove any unwanted fluids after
surgery. This development is a new design intended to be an alternative to existing
hospital wall suction and will also allow the patient to become ambulatory in their
recovery
1.1
Special Precautions
The CDSU has been designed for electrical safety. All the safety and
operating instructions should be read before operating the unit. Failure to do
so could result in injury to the user, patient, or damage to the system and
accessories.
Always remove from the patient area when undertaking service on the unit.
Always take anti-static precautions when service this unit, failure to do so will
cause damage to the sensitive electronic circuits.
1.2 Electrical Shock Hazard
Do not defeat the grounding integrity of this system. Protection against
electrical shock, in the event of failure of insulation, is provided by the
connection of the psu unit to the safety ground. Safety grounding occurs only
when the 3-wire cable and plug provided with the system are connected to a
properly grounded receptacle.
Do not remove the system cover. The system should be serviced by trained
and qualified personnel only. Contacting the hazardous voltages within the
system could cause serious injury.
Never attempt to open the mains power supply unit, this unit is sealed and
must be returned to the factory if damaged or faulty.
Do not use the system if the power cord has any cuts or openings.
Do not use the unit or power supply if their enclosure is cracked.
Do not immerse the unit or power supply in any liquids.
1.3 Explosion Hazard
Do not operate or use this system in the presence of flammable anesthetics,
gases or oxygen rich environments as it could lead to explosion.
4
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 5
2
Operating manual
Before undertaking any service on this unit read the system operating manual in full. If it
is not available contact Rocket Medical who will provide an electronic copy
The Manual will provide guidance on the use of the controls, identification and meanings
of the symbols and guidance over the units use.
3
Basic Trouble shooting guide
The Mobi distinguishes between warnings, alarms and internal errors. If Mobi detects
any of these situations, an acoustic warning signal sounds and an error number will be
displayed.
The error number corresponds to the problem as shown below.
Error
Problem Description
Trouble Shooting
Source of Error
501
Self test failed
Sensor defect
Switch off unit and restart.
502
Battery low
Connect power supply
Plug in Power Supply to recharge battery
503
Self test failed
Battery defect
Do not disconnect
Power supply
Battery defect, do not
disconnect power supply
Filter may not have
been recognised,
problem with sensor.
Battery low on charge,
approximately 30 mins
of power remaining
Battery has
malfunctioned,
possible battery
replacement required.
504
No filter
Please insert filter
505
Battery empty
Please charge battery
Check to see if filter is in
place. Remove pump from
bottle to check filter in in
position.
Connect Power Supply to
charge battery
506
Data storage error
507
Battery empty
Power Off
508
Battery life cycle warning
Turn unit off and back on. If
error occurs please contact
manufacturer for service.
Battery has no charge,
connect power supply to
charge or use connected to
power supply.
The battery in the unit has
almost reached the
maximum number of
charges allowed for the
battery to run efficiently.
This warning is displayed
with 50 charges remaining.
Unit may have been
turned on without filter
in place.
Battery has been
allowed to discharge
fully
Memory malfunction,
failure in storage
access
Battery charge has
been left to discharge
fully. The unit does not
have enough power to
function
The battery in the unit
has a 500 charge life
cycle. Please arrange
for the unit to be
serviced and the
battery replaced.
5
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 6
4
Specification
Power Input to Pump:
Power Supply:
Consumption:
Frequency:
Maximum current:
Environmental conditions:
Service interval:
Dimensions:
Weight:
Suction Ranges:
Suction Range Accuracy:
18VDC
100 - 240VAC 50/60Hz
0.096 KW/h
50/60 Hz
2.5A @ 18V
+5°C to +35°C
500 Battery charges
W - 190mm x H - 118mm x D - 140mm
Unit - 0.96Kg
User Set: -0.5Kpa to -5Kpa
±10% at 2.5Kpa.
6
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 7
5
Warnings
Not intended for use where a suction (vacuum) value of over -5 Kpa or a flow of over
5L/min is required.
The Mobi device is not approved for connection to any other device except the Rocket
Medical R54571 single use filter and R54500 chest drain unit and tubing set (R54502,
R54539).
The R54500 chest drain unit and tubing set (R54502, R54539) are single use parts and
their instructions for use must be followed at all times.
Do not use in an area where flammable gases are present.
If a persistent air leak is shown, check the filter is positioned correctly on the chest
drain unit and that the suction unit is securely attached to the chest drain unit.
Precautions for EMC safety should be observed. The Mobi complies with IEC606011-2:2007, however:
- Electronic equipment in the vicinity of the Mobi device may affect its
operation and potentially cause unpredictable operation of the device.
- Wherever possible the Mobi device should be distanced from surrounding
electromagnetic equipment and cables to this equipment in order to reduce
possible electromagnetic interference.
- The Mobi device power cable should only be connected to a correctly wired
grounding receptacle in order to avoid the risk of electrical shock.
When positioning the unit, ensure that access is available to the DC power cable
located on the rear of the unit.
To avoid damage to the unit only use the approved mains power adaptor supplied with
the unit, Model Number: R54569-PSU-UK.
The Mobi contains a Li-ion battery and that is not user replaceable and that
unauthorised attempts to repair could lead to damage to the unit, environment and/or
human health if incorrectly fitted, handled or disposed of improperly.
The Li-ion battery must only be replaced by a suitably trained and Rocket Medical
approved service centre.
Fluids should not be allowed to enter the device as this may result in damage to the
system.
Only qualified personnel should service the device and the unit cabinet should not be
opened except by these personnel due to the risk of hazardous electrical shock. All
service requirements should be referred to a Mobi authorised representative.
7
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 8
6
Technical Description
The CDSU is an external, mobile suction device for use with the Rocket Medical Chest
Drainage range. Rocket Medical chest drains are required to prevent the build-up of air
inside the body (pleura) whilst breathing and also remove any unwanted fluids after
surgery. This development is a new design intended to be an alternative to existing
hospital wall suction and will also allow the patient to become ambulatory in their
recovery. The unit is both mains and battery powered (rechargeable) for limited time use
while mobile.
The unit is intended to function only with a disposable filter set. The filter set will be a
sterile, single use and is a disposable component. The filter will act as protection for the
CDSU but also as a bacterial barrier for the patient.
The CDSU consists of a vacuum pump, membrane keypad, LCD display (with
backlight), monitoring electronics (pressure sensor & filter detection) and battery.
The system will be mounted on top of the chest drainage bottle in which the vacuum is
generated. Here it will control the vacuum at predetermined levels. The vacuum can be
set to a level of suction between 0 – 50cmH20 at 5cmH20 intervals.
6.1 Power Supply.
The unit operates from either an 18VDC input or from the built in 14.8V Lithium-ion
battery. The voltages are switched internally to provide the main operating voltages of
12VDC and 3V3DC. The Lithium –ion battery is charged from the DC input supply
using a specialised Charge IC BQ24610RGET which controls the charge current and
monitors the temperature of the battery.
6.2 User Interface
A back lit LCD provides the user messages and operating values of the device with a
tactile 3 button keyboard allowing user interaction with the device to set pressure etc.
6.3 System Control
The entire system is controlled by software within an Atmel ATSAM3S4CA-AU
processor (32-Bit) with 256KB of program memory This provides the display
information for the LCD display, the action from the key presses, the drive of the pump
motor and monitors the feedback from the pump to provide volume flow and a pressure
sensor for the pressure generated.
Software within the device controls the system to the predetermined set values and
stores these values for later USB transmission to data logging software.
6.4 System Safety
Pressure safety is controlled by an electronic over pressure valve opening when the
values are outside the prescribed range with a backup mechanical discharge valve to
protect from system failures.
Electronic system failures are protected from short circuits by built in protection in the
Lithium – ion battery pack.
The unit is reverse voltage protected via an input protection diode when connected to an
external DC power supply.
8
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 9
7
Calibration checks
System calibration checks can be made by connecting a suitable flow meter to the input tube
of a collection bottle and the pressure can be checked by connected a pressure meter to the top
of the bottle. The following photograph is an example of a calibration check system.
Unit Under
Test
Digital
Pressure
Meter
Digital Flow
Meter
Flow control
Collection
bottle with
800mL of water
Set the pressure to a suitable level and control the flow by restricting the input tube. The
measure values on the pressure meter and flow rate of the device should be within 10%.
There are no PCB adjustments for pressure or flow, the system is entirely digital so if the unit
is not correct then the control pcb will need to be exchanged.
9
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 10
8
Routine service tasks
The following parts should be part of a regular service maintenance program.
Replace mechanical safety valve.
Replace one direction flow valve
Replace lithium ion battery
Perform and document system test
Reset system use parameters
Electrical safety test power supply unit.
Software upgrade if available. (only available from a factory service)
WARNING: Take care when replacing the lithium battery, ensure correct connection and
that the new battery is correctly fitted into the moulding recess. Dispose of used lithium
battery correctly and with care, observing current legislation.
Never use a replacement part that is not supplied by Rocket Medical.
It is recommended that these takes are undertaken every two years or sooner if indicated by
the system notifying the user of charge 500 cycles.
10
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 11
9
Disassembly / Reassembly
Parts identification:
1) Carry handle
2) Unit On/Off button
3) User selection buttons
4) LCD display
5) Rotating locking bar
6) Position for single use filter
To disassemble remove locking bar
retention block.
Rotate Locking bar (5) so that it is
removed from the unit base.
This reveals fixing screws, remove all
the screws.
The unit will then separate to reveal the PCB in the top of the unit and the battery / pump in
the base.
11
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 12
9.1 Unit Top assembly
LCD
Key Pad
Control PCB
9.2 Unit Base assembly
Unit Lithium
ion battery
Electronic
Discharge Valve
Mechanical
one way valve
Mechanical Over
pressure dump valve
Pump
Alarm
Speaker
Clock Battery
Backup
Filter present
detection
12
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 13
9.3 PCB Top
Display
backlight
connector
Keypad
connector
Display
connector
Alarm
speaker
connector
Lithium ion
battery
connector
Pump
Connector
System
pressure sensor
Unit DC
input
connector
Clock Battery
Backup connector
System USB
connector
9.4 PCB Bottom
Main CPU
13
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 14
9.5 PCB Top Overlay
14
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 15
9.6 PCB Bottom Overlay
15
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 16
10 Circuit Diagram
16
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 17
17
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 18
18
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 19
19
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 20
11 Parts Listing
Chest Drain R54569
Part Number
Ref
Description
Quantity
Per
R3
2R2 1% 1206
1
R23
10R 1% 1206
1
R45, 46
0.01R 1% 2010
2
R61
0.02R 1% 2010
1
R7
100R 1% 0603
1
R24
330R 1% 0603
1
R26
150R 1% 0603
1
R33, 60, 66
750R 1% 0603
3
R2, 4, 17, 58
1K 1% 0603
4
R30
1K5 1% 0603
1
R27
2K4 1% 0603
1
R41, 42
4K7 1% 0603
2
R57
9K1 1% 0603
1
R5
9K31 1% 0603
1
R11, 21, 32, 35, 37, 54, 63,
64, 65, 70, 72, 76
10K 1% 0603
12
R50
11K 1% 0603
1
R29
12K 1% 0603
1
R13
14K7 1% 0603
1
R51
15K 1% 0603
1
R62
30K 1% 0603
1
R12, 68, 69, 71
33K 1% 0603
4
R28
75K 1% 0603
1
R1, 8, 9, 10, 14, 15, 16, 18,
19, 22, 25, 31, 34, 39, 44,
47, 49, 53, 56, 67, 73, 74,
75, 77, 78, 79, 80
100K 1% 0603
27
R48, 52
240K 1% 0603
2
R59
270K 1% 0603
1
R6, 38
430K 1% 0603
2
R55
470K 1% 0603
1
R36
1M 1% 0603
1
R20, 22
5M1 1% 1206
2
R40
Resistor 0R 0603
1
22pF COG 50V 0603
1
27pF COG 50V 0603
2
C15
3n3F X7R Ceramic 0603
1
C26
10nF X7R Ceramic 0603
1
C43
47nF X7R Ceramic 0603
1
C16
C27, 30
20
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 21
C1, 3, 5, 6, 7, 11, 20, 24,
25, 28, 32, 33, 37, 40, 41,
42, 46, 47, 48, 49, 50, 52,
53, 54, 55, 56, 57, 58, 59,
60, 62, 63, 65
100nF X7R 50V Ceramic 0603
34
C4, 36
1uF X7R 25V Ceramic 0603
2
C12, 31
1uF X7R 25V Ceramic 1206
2
C13
1uF X7R 25V Ceramic 0805
1
C2
2u2 X5R 50V Ceramic 1206
1
C18, 21
2u2 X5R 50V Ceramic 0805
2
C8, 9, 10, 19
4u7 X5R 16V Ceramic 0805
4
C34, 35, 38, 39 44, 51, 66
10uF X7R 35V Ceramic 1210
7
C17, 23
22uF X5R 16V Ceramic 0805
2
C67
4u7F Tantalum 20V 10% case A
1
C14
33uF Tantalum 20V 10% case C
1
C45
100uF Electrolytic 16V SM FC
1
C64
100uF Electrolytic 6V3 SM FC
1
L4
10U-1A4-0R12-WE-PD2 74477310
1
L2
6U8-4A3-0R025-WE-PD 7447715006
1
L3
15U-5A0-0R027-WE-PD 744770115
1
L7
90Z-0A37-0R3-WE-CNSW-1206 744232090
1
L1, 5, 6
300Z-2A0-0R15-WE-CBF-742792641
3
T2, 4, 5, 7, 8,9,11, 12,13,15
Si7463ADP-T1-GE3
10
T1,3
CSD18504Q5A
2
T6,10, 14
BSS123
3
Q1,2
BSS84
2
T18
BC869
1
T19
BC869
1
D2,4, 6
SK56
3
D12,11
MMSZ4684T1G
2
D3,6,13
PMEG4005AEA
3
D1
SSB43L
1
D7,9
BAT54W
2
D14
BAS16WS
1
D11
1N4148W SOD123 package
1
LD1, 2, 3
SML-311UTT86K
3
D5
VLMT3100-GS08
1
D15
WE-TCS-824015
1
IC1
BQ24610RGET
1
IC2
TPS62177DQC
1
IC3
LM25085MY
1
IC4
FT232RL
1
21
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 22
IC5
ATSAM3S4CA-AU
1
IC6
AT25SF041-SSHD
1
IC7
PCF2129AT/2
1
IC8, 9
LM258DG
2
IC11
LM7301IM5/NOPB
1
SEN2
SQ277-14015 (pressure sensor)
1
BU1
DC Power Socket RASM712X
1
VAR1
VC080505C150DP
1
BU4
Connector FPC 32 way Wurth 68713214022
1
X4
USB connector Molex 67503-1020
1
ST4
Keypad 6 way TE 5-102523-4
1
ST1, 2
4 way connector B4B-EH-A
2
JP3
3 way connector B3B-EH-A
1
JP1, 2, 4, 6, 7
2 way connector JST B2B-EH-A
5
BU3
JTAG connector 20way FCI20021321-00020C4LF
1
-
4 way connector housing EHR4
2
-
3 way connector housing EHR3
1
-
2 way connector housing JST EHR2
5
Q3
Crystal 12.00MHZ-FUN-CM7050 (RS 478-9729)
1
BZ2
Sounder PKLCS1212E4001-R1
1
Clock Backup
Saft lithium
1
Battery
Lithium 4S1P BM18650Z3 Block
1
LCD
Display CG240160B-LB-FT-N6-CLO
1
Pump
Thomas Pump 14200500
1
Detection switch assy
Switch B3WN6002 + PCB
1
Speaker
Visaton K1650 OHM art 2816
1
Discharge Valve
Parker Valves 991-003063-014015
1
Tubing
Silicone 3mmID x 5mmOD
0.03
Tubing
Silicone 3mmID x 5mmOD
0.035
Tubing
Silicone 3mmID x 5mmOD
0.02
Tubing
Silicone 3mmID x 5mmOD
0.09
Tubing
Silicone 3mmID x 5mmOD
0.125
Tubing
Silicone 3mmID x 5mmOD
0.035
Tubing
Silicone 3mmID x 5mmOD
0.0245
Tubing
Silicone 3mmID x 5mmOD
0.1
Tubing
Silicone 3mmID x 5mmOD
0.07
Tubing
Silicone 3mmID x 5mmOD
0.04
Tubing
Silicone 3mmID x 5mmOD
0.04
Tubing
Silicone 2mmID x 3mmOD
0.16
22
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 23
Safety Valve
E-C-PP-1C1C-E-001 eoxshop.com
1
One Direction Valve
SCV67230 (Value Plastics)
1
Tube connections
right angle 10 -32 UNF 2mm tube KL220-6005
1
Tube connections
1
Tube connections
right angle 10 -32 UNF 3mm tube KL230-6005
streight connector M5 x 8mm 3 mm tube M52306005
Tube connections
T junction 3 mm tube 1/8 T230-6005
4
Membrain switch
1
1
Reset switch
Omron B3F-3100
1
Power Supply
XP Power AFM60US18
1
Mains Lead Set (UK)
Power Cord - Mededical Grade 13A to IEC socket
1
USB Lead
USB A plug- USB B mini plug 1M long
1
23
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517
Page 24
12 Contact Details
Customer Services:
Rocket Medical plc.
Sedling Road. WASHINGTON. NE38 9BZ. ENGLAND
Tel: +44(0)191 419 6988. Fax: +44(0)191 419 6989
Email: customerservices@rocketmedical.com
24
Copyright© 2017
Rocket Medical plc .
All rights reserved.
ZDOCK297
Rev:01
080517