GA-SG-500-01-E Version 1.1A Stand 01.03.2005
Instructions of use for Oszillating Plastersaw SG-500-01
Content
Validity ______________________________________________________________________________________ 1 Certificate of compliance _______________________________________________________________________ 1 Purpose of use________________________________________________________________________________ 2 Inspection____________________________________________________________________________________ 2 First use _____________________________________________________________________________________ 2 Staring the machine ___________________________________________________________________________ 2 Changing saw blade ___________________________________________________________________________ 2 Saw blades ___________________________________________________________________________________ 2 Safety Instructions ____________________________________________________________________________ 2 Scope _______________________________________________________________________________________ 2 Opening of bandages __________________________________________________________________________ 2 Special instructions ___________________________________________________________________________ 2 Additional Tips ________________________________________________________________________________ 2 Troubleshooting_______________________________________________________________________________ 2 Care_________________________________________________________________________________________ 2 Maintenance __________________________________________________________________________________ 2 Guarantie:____________________________________________________________________________________ 3 Technical data ________________________________________________________________________________ 3 Detail drawing ________________________________________________________________________________ 4
Validity These instructions are valid for all SG-500-01 Plaster Saws and accessoires, sold by SCHREIBER Instruments. Certificate of compliance Manufacturer´s name and address: SCHREIBER GmbH Chirurgische Instrumente Unterer Damm 15 D-78567 Fridingen Product name: SG-500-01, oscillating one-hand plaster saw and accesoires (saw blades) in accordance to the valid catalogue and price-list. The Schreiber GmbH confirms herewith in sole responsibility, that the above mentioned products comply with the requirements according to annex I of the council directive 93/42/EEC and EN 606011-2. These products are permitted to carry the CE-mark We have established a certified quality management system according to DIN/EN/ISO 13485:2003 1
Signature 2 4
3
Armin Zepf Quality Manager
SCHREIBER GmbH, Unterer Damm 15, D-78567 Fridingen, Tel.: +49-7463-307, Fax: +49-7463-8870, http://www.schreibergmbh.de, [email protected]
1
Instructions of use for Oszillating Plastersaw SG-500-01 With purchasing this SCHREIBER device you have decided to use a high quality product. Proper function an safety are only guaranteed for a long time, if the following points are strictly observed. Purpose of use Cutting of plaster and synthetic bandages. The safety rules of MDD (medical device directive) and the national law for safe devices (Arbeitsmittel-Gerätesicherheitsgesetz) are valid. Inspection SCHREIBER oscillating plaster saws must be inspected for damages or faults immediately after transportation. Claims for compensation are only possible, if shipper or dealer are informed immediately.
First use SCHREIBER oscillating plaster saws must not be used, before these instructions are worked through. • For use and maintenance the valid (national) laws for safety and maintenance must be observed. • make sure, that the voltage on the machine is similar to your socket. • Check cables and extensions regularly. • do not screw or rivet external labels on the body of the machine. This may cause electrical danger. Staring the machine • Make sure, that the plug is connected to the socket properly. • Plaster saw is equipped with an electronic speed control button. The optimal speed must be adjusted, depending on the sawblade and use. Speeed can be adjustaed by turning the button (2) (pre-selection= 4). • Turn switch (1) forwards to "ON"position for turning on and backwards for turning off. Changing saw blade -unplug cable to prevent unintentional turning on. 2
-Untighten locking screw (4) with the enclosed hexagonal key. -remove saw blade (3) -clean thread and bearing surface -insert new saw blade and center on the shaft. -screw down locking screw by hand and tighten with hexagonal key (6-7 Nm) Make sure, that the shaft is not turned during untightening and tightening the screw. Turning the excentric axis inside the machine may damage the transmission. Saw blades Attention: blunt saw blades must be changed immediately. Hot saw blades may burn the patients skin and overload the machine. Available are saw blades for plaster bandages (SG-501-02) and synthetic bandages (SG-501-04). Safety Instructions • Only use original spare parts. • Improper use may cause overheating and damage of the machine. • Never bring the machine in contact with water or other solutions. • Maschine is not explosion protected. Keep away from flammable liquid or gas. Scope Generally it is not recommended to misuse the saw for medical or other purposes, that means the machine must only be used as saw for plaster or sysnthetic bandages. It is not suitable for surgical operations. The patient must not have direct contact to the sawblade. Opening of bandages ATTENTION: Unpadded bandages must be opened with special care, because sticked skin might be burned! Place the saw with the fixed saw blade on the bandage with slightly pressure. The thumb can serve as guide and depth gauge and prevents abruptly penetration into the bandage. As long as the saw blade is cutting, there is a resisatnce, that decreases, when the saw blade goes through the bandage. Then the saw must be lifted without beeing taken out of the groove and
GA-SG-500-01-E Version 1.1A Stand 01.03.2005
mooved about 15 mm into cutting direction. Repeat these steps, until the complete bandage is opened by several linear cuts. Special instructions • Bandages with metal inserts should not be opened with this saw.. • Avoid permanent friction on the soft furnishing (heat). Additional Tips The cutting of unpadded bandages should be avoided as far as possible (see "opening of bandages"). If a close bandage is necessary, it is useful to integrate a gaze or cotton stripe on the patietns skin and mark the position on the bandage with a marker. Later the bandage can be opened along the line. After surgical operations, the position of the wound should also be marked on the bandage, so it is possible to open a window for wound inspection, extraction of wires or incisions. Troubleshooting REPAIR OF THE MACHINE MUST BE MADE BY EXPERTS ONLY! If the saw does not work:: • If the coals are use up on 7 mm, they must be replaced by an expert. . • Check the cable for damages • Check the fuse. If the saw is too loud: Turn off and send to the manufacturer for inspection. Care The machine does not need special service, but it should be cleaned with a moist duster regularly. The saw must not be immersed into liquids or rinsed with water. Take care, that no plaster debris is disturbing the oscillating shaft or the on/off switch. Keep the airing grids free of plaster debris and make sure that the cool air can circulate properly. The saw should be cleaned and checked all 6 months, depending on use and duration. Maintenance Electrical devices must only be checked,repaired and maintenanced by experts. Alway unplug the machine, when it is not used. Blunt saw blades
SCHREIBER GmbH, Unterer Damm 15, D-78567 Fridingen, Tel.: +49-7463-307, Fax: +49-7463-8870, http://www.schreibergmbh.de, [email protected]
GA-SG-500-01-E Version 1.1A Stand 01.03.2005
Instructions of use for Oszillating Plastersaw SG-500-01
must be removed, to avoid overheating of the machine. Saw must be stored under dry conditions. Device must not be disassembled. There are no components inside, that can be repaired by the user. Guarantie: We give 1 year guaranty on this machine and accessoires on material and manufacturing faults. Improper use, repair by third parties and modifications void guaranty and CE-marking.
Technical data
Cat.-N°
SG-500-01
Voltage
230V/50Hz
Power
180 Watts
Power Input
0,8A
Strokes
12000-21000 /min
Noise, A-tipical
70 dB(A)
Noise while sawing
>85dB(A)
Hand/Arm-Vibration
<2.5 m/s
Weight
1.1 Kg
Conformity
CE,DIN EN 60601-1,2
Safety Class
Class II safty isolated
Fuse
T 1,6A TR5