Operating Instructions & Maintenance Manual
Model DM10-001
DCell Suction
Clearing The Airway Is Our #1 Priority
, INC. 11064 Randall Street Sun Valley, CA 91352 USA Tel 1+818-504-4054 www.sscor.com shop.sscor.com Email: [email protected] [email protected]
The world’s 1st and only full size portable suction device powered by ALKALINE BATTERIES.
Clearing The Airway Is Our #1 Priority
Caution-Notice 1. Federal law restricts this device to the sale, distri- 10. Make sure you insert batteries into your device bution, and use by, or on the order of a physician, properly, with the + (plus) and – (minus) terminals emergency medical technician, or other medical aligned correctly. Always install batteries per the practitioner. For use by medical personnel trained labeling on the battery holder. in suctioning techniques and in the use of medical 11. Store batteries in a dry place at normal room temsuction equipment. perature. Extreme temperatures reduce battery 2. SSCOR suction units are not designed or intendperformance. ed for use in extended procedures that require 12. Any battery may leak harmful chemicals which prolonged high vacuum/low airflow applications, may damage skin, clothing, or the inside of the as is the case in wound drainage or endoscopic battery holder. To avoid risk of injury, do not let use or in any other procedure that produces high any material leaked from a battery come in convacuum levels within an occluded system for an tact with eyes or skin. If batteries leak, replace the extended period of time. Turn the suction unit off battery holder. when it is not in use. 13. Any battery may rupture or explode if put in a fire 3. This manual is restricted to the discussion of the or otherwise exposed to excessive heat. To avoid use and maintenance of this device. It does not risk of injury, do not expose batteries to excessive attempt to discuss professional techniques in sucheat, or moisture. tioning procedures. 14. Use only D Cell alkaline batteries from a reputable 4. Operator should be thoroughly familiar with these manufacturer. operating instructions before this device is used. 15. Do not use different brands or chemistries of bat5. Do not use in the presence of flammable agents teries in the same battery holder. or anesthetics. 16. Do not use Lithium batteries in the battery holder. 6. Before testing for vacuum over -300mmHg look for an expiration date on the canister (where appli- 17. When replacing batteries, always replace the whole set. cable) and change the canister if the canister has passed the expiration date to minimize the possi- 18. Batteries will not recharge when installed in the bility of implosion, which can occur when a canissuction device. ter is aged or damaged. 19. Follow battery manufacturer’s instructions as to 7. Do not use the batteries after the “Use By” date proper handling, storage and disposal of batteries. on the battery label. 20. When using the SSCOR DCell Suction Device, 8. Do not store batteries in the SSCOR DCell Sucensure that the canister is in an upright position. tion device if it is not expected to be used for sev21. Connect the suction tubing from the device to the eral months. vacuum port of the canister prior to use. 9. Keep battery contact surfaces and battery compartment contacts clean by rubbing them with a rough cloth each time you replace batteries.
2
SSCOR DCell Suction – Model DM10-001 Rev. F (10/19)
Clearing The Airway Is Our #1 Priority
Table of Contents Caution Notice
Page 2
General Description and Warranty
Page 3
Description of Symbols
Page 4
Operating Instructions
Pages 5 & 6
Trouble Shooting
Page 7
General Specifications
Page 8
Maintenance
Page 9
Internal Access
Page 10
General Description A battery pack containing 12 fully charged alkaline batteries at full capacity will power the device for approximately 6-9 hours. This run time is defined as the period of time the device generates more than 300mmHg negative pressure. This run time specification is based upon tests conducted using new Duracell® CopperTop batteries. Run times are affected by many variables which include the brand of battery, the age of the battery, and ambient temperature. The red LED on the control panel will light indicating the batteries are losing capacity and new batteries should be installed in the 12 X D battery holder. See page 5 for instructions on battery replacement. For replacement parts and accessories, please visit shop.sscor.com. The SSCOR DCell Suction device is designed for hospital crash carts, patient transport and emergency medical service. Suction power can be regulated when full power may be considered harmful to the patient. SSCOR suction units are designed to provide instant, effective suctioning, independent of external sources of power and can be pre-set to be activated immediately upon reaching the distressed patient. All controls are clearly labeled and easily accessible.
Warranty SSCOR warrants that each new product is free from defects in material and workmanship under normal use and service for a period of three years from the date of purchase. This warranty gives you specific legal rights and you may also have other rights that vary from jurisdiction to jurisdiction. For countries where minimum warranty terms are determined by statute, the warranty term is the longer of the statutory period or the term listed above. Batteries, disposable items including collection canisters, patient tubing and catheters are excluded from this warranty. See the SSCOR Warranty for terms and conditions, available on www.sscor.com.
3
SSCOR DCell Suction – Model DM10-001 Rev. F (10/19)
Clearing The Airway Is Our #1 Priority
Description of Symbols
Symbol
4
Location
Meaning
Control Panel Label
12V DC Power Input
On/Off Switch
Turns Device On/Off
Control Panel Label
Battery Voltage Indicator
Serialized Label
Separate Collection for Electronic Equipment
Serialized Label
Type BF Equipment
Serialized Label
Attention-Consult Accompanying Documents
Shipping Carton
Authorized Representative in the European Community
Shipping Carton
Date of Manufacture
Shipping Carton
Manufacturer
Shipping Carton
Transport Storage Range
Shipping Carton
Humidity
Shipping Carton
Atmospheric Pressure
SSCOR DCell Suction – Model DM10-001 Rev. F (10/19)
Clearing The Airway Is Our #1 Priority
Operating Instructions
Running the unit from the 12xD Cell Alkaline battery holder 1. Press the ON/OFF switch 1 on the control panel to start the vacuum pump.
5
4 1 3 2
Adjusting the vacuum level • Occlude the patient tube and turn the vacuum regulator 2 clockwise to the stop. • Observe gauge (located above the vacuum regulator). The vacuum reading should rise to 300mmHg from zero in 3 seconds. It should max out at approximately >= -525mmHg. Lower levels of negative pressure will be observed at altitude. 2. Press the on/off switch on the control panel to stop the vacuum pump.
WARNING: The tandem port on the canister is not an inlet port. Do not connect tubing to the tandem port of the canister. Battery Level of Charge Verification Once the unit is running from the internal battery pack, check the control panel and observe the battery voltage indicator to gauge the level of battery charge. The green LED when lit, indicates a high level of charge. WARNING: The red LED when lit, indicates the battery voltage is nearly depleted and the negative pressure level of the device will soon be below 300mmHg. Immediately install fresh batteries into the battery holder or connect device to an appropriate 12V DC power source. Replacing the 12xD Cell Batteries When the red Battery Voltage LED is lit indicating low battery voltage, the batteries in the 12xD Cell battery holder should be replaced with fresh D Cell Alkaline batteries. To access the battery holder, open the top of the chassis 5 exposing the battery holder: 1. Disconnect the Molex on the 12xD Cell battery holder from the wiring harness; 2. Lift the 12xD Cell battery holder out of the chassis, noting the orientation of the battery holder. The battery holder only fits in this orientation. Do not lift the battery holder by the exposed wires, use the fabric pull tab to lift the 12xD Cell battery holder out of the chassis. Replacement battery holder may be purchased at shop.sscor.com.
5
SSCOR DCell Suction – Model DM10-001 Rev. F (10/19)
Clearing The Airway Is Our #1 Priority
Operating Instructions (Continued) Remove the discharged batteries from the 12xD Cell battery holder and dispose of the batteries according to local / regional / national requirements for the disposal of alkaline batteries. Install 12 new fresh D Cell alkaline batteries into the 12xD Cell battery holder - being careful to assemble the batteries in the correct orientation as indicated on the battery holder. When the battery holder is loaded with 12 new D Cell alkaline batteries, set the 12xD Cell battery holder back into the chassis. The battery holder only fits in one orientation. Re-connect the Molex on the battery holder to the wiring harness. Turn the device on to verify the batteries are correctly installed and the battery holder is connected to the wiring harness. Observe the battery voltage indicator on the control panel. The green LED should be lit indicating the batteries have a high level of charge. Running the device using the #80665 DC power cord (Available for purchase on shop.sscor.com) Connect the DC power cord to the 12V DC cigarette lighter receptacle in the vehicle. Connect 3 the DC power cord to the suction unit by securely attaching the charging plug into the receptacle. If the suction unit is operated while it is hooked up to the vehicle it will utilize the vehicle power and save its own battery for emergency use. The DC power cord will not charge the 12xD Cell batteries when connected to the suction device. CAUTION: The #80529 power adapter will not charge the 12xD Cell batteries when connected to the suction device. If the 80529 power adapter is connected to the SSCOR DCell Suction device and the battery voltage indicator shows a red LED, the output from the power adapter is not sufficient to power the SSCOR DCell Suction device. The device is running from its internal batteries. The 80529 power adapter may be purchased at shop.sscor.com. WARNING: To avoid risk of shock, this equipment must only be connected to a supply mains with protective earth. The 80529 Power Adapter is for Indoor Use Only. WARNING: Periodic checking or replacement of “D” Cell Alkaline batteries is required. If for any reason poor battery quality is suspected, perform a battery test as described below. BATTERY TEST: Turn the SSCOR DCell Suction device on and run the unit for 30 seconds. Occlude the patient port of the canister and observe the negative pressure gauge. The negative pressure gauge should rise above 300mmHg in 3 seconds and reach 525mmHg. Observe the battery condition indicator LED 4 . If it lights red, it is time to install 12 fresh D Cell batteries into the SSCOR DCell Suction battery holder. Dispose of the batteries according to local /regional / national requirements for the disposal of electronic waste. Do not store batteries in the SSCOR DCell Suction device if it is not expected to be used for several months.
6
SSCOR DCell Suction – Model DM10-001 Rev. F (10/19)
Clearing The Airway Is Our #1 Priority
Trouble Shooting
Problem
Possible Cause
Corrective Action
Does not function when switch is on (DC Power)
Battery discharge
Replace Batteries
Molex connections disconnected
Lift battery cover and check to see Molex connections are secure
Defective PC Board
Replace PC Board
Fuse Blown
Replace Fuse
Red LED
Low battery capacity
Replace Batteries
Battery condition indicator lights not lit
Defective PC Board
Replace PC Board
No suction when pump is running
Vacuum line is disconnected
Re-connect vacuum line
Canister Defective
Replace canister
Canister lid loose
Tighten canister lid
Catheter thumb vent is open
Occlude thumb vent
Defective pump(s)
Replace pump(s)
7
SSCOR DCell Suction – Model DM10-001 Rev. F (10/19)
Clearing The Airway Is Our #1 Priority
General Specifications
Characteristics
Specifications
Size
12”H x 9”W x 11”D 30.48cm H x 22.86cm W x 27.94 cm D
Weight
9.5 LBS. / 4.3Kg
Vacuum Pump
12V DC Exceeds 525mmHg. Clinical Airflow >=30 LPM with 12 fully charged D Cell alkaline batteries at full capacity
Variable Regulator
Controls negative pressure
Battery
12 x D Cell alkaline batteries 12 x D Cell battery holder part number: 040-00001
Wiring Harness
Uses (1) 2A 250V slow acting fuse UL Listed and CSA Certified Type (1¼ x ¼)
Collection Canister
1200cc/ml SSCOR Part #48041
Patient Tubing
Vinyl tubing 9/32” ID, 72”L SSCOR Part #43200
Suction Tip
HI-D® “Big Stick®“ Large Bore Suction Tip with thumb control. SSCOR Part #44241
Environmental Conditions Operating Temperature Range: Operating Relative Humidity: Operating Atmospheric Pressure: Operating Altitude: Storage & Transport Temperature: Storage & Transport Relative Humidity: Storage & Transport Atmospheric Pressure:
8
0 °C (+32 °F) – +40 °C (+104 °F) 0 - 93% (non condensing) 10.2 Psi (70 kPA) – 15.4 Psi (106 kPA) <2000m -40 °C (-40 °F ) – +50 °C (+122 °F) 0 - 95% (non condensing) 7.3 Psi (50 kPA) – 15.4 Psi (106 kPA)
SSCOR DCell Suction – Model DM10-001 Rev. F (10/19)
Clearing The Airway Is Our #1 Priority
Maintenance Preventive Care Observe the following maintenance routine to ensure readiness at any time: 1. Do not store batteries in the SSCOR DCell Suction device if it is not expected to be used for several months. 2. Test the SSCOR aspirator at regular intervals; See page 6. 3. Make sure the SSCOR aspirator is always clean and ready for use. 4. The canister contains a float valve to prevent fluids from reaching the vacuum pump, however, if the procedure produced an excessive quantity of fluids, check the vacuum line (#) for evidence of moisture. If the vacuum line between the device and canister is moist, it is possible that fluids have reached the vacuum pumps. See Disinfection Instructions below. 5. For technical assistance, call +1 818-504-4054. Note: No part requires lubrication and lubricants should not be used. Sanitation As soon as possible after use, the single use disposable canister, patient tubing and catheter should be discarded according to local / regional / national requirements for the disposal of hazardous waste materials. Clean using a mild detergent and if necessary disinfect with a mild disinfectant such as 10 to 1 bleach and rinse using clear water to remove any residue. Do not reuse any single use disposable parts; do not submerge the device into any liquid, this will void the warranty and cause the device to malfunction. Disinfection Use personal protective equipment such as gloves, a smock, and face and eye protection when handling units that are suspected to be contaminated. Caution: Disconnect the unit from any power source prior to cleaning the unit. When cleaning the interior of the chassis, disconnect the battery from the PC Board to prevent damaging the PC Board.
Part
Cleaning & Disinfecting
Collection Canister
SSCOR #48041. Disposable item, re-use not permitted. Use new canister for each patient.
Patient Tubing (Type BF Applied Part)
SSCOR #43200. Disposable item, re-use not permitted. Use new patient tubing for each patient.
HI-D® Stick (Type BF Applied Part)
SSCOR #44241. Disposable item, re-use not permitted. Use new HI-D Stick for each patient.
Vacuum Pump
SSCOR #90207. Wipe with damp cloth or disinfectant wipe. Sterilization not permitted. Vacuum pump should be replaced if contaminated.
PC Board
SSCOR #050-00001. Sterilization not permitted. PC Board should be replaced if contaminated.
Plastic Chassis
Wipe with damp cloth or disinfectant wipe. Sterilization not permitted.
9
SSCOR DCell Suction– Model DM10-001 Rev. F (10/19)
Clearing The Airway Is Our #1 Priority
Internal Access WARNING DO NOT ATTEMPT TO SERVICE THIS EQUIPMENT IF YOU ARE NOT A BIOMEDICAL TECHNICIAN. NO MODIFICATION OF THE EQUIPMENT IS ALLOWED. UNAUTHORIZED SERVICE ATTEMPTS, OPENING OR TAMPERING WITH THE COMPONENTS OF THE SSCOR DCELL SUCTION™ CAN DAMAGE OR DISABLE THE DEVICE, AND WILL VOID THE WARRANTY. TAKE APPROPRIATE ESD ABATEMENT MEASURES WHEN HANDLING ALL INTERNAL COMPONENTS.
To open the device, first remove the 5 screws around the perimeter of the back chassis. Next, remove the two screws holding the D-Ring of the carry handle on each side of the device. The device will then open. To close the device, replace the screws and tighten. Make sure all pneumatic and wiring connections are secure before closing the chassis.
C
D
10
A B
5 1
4
1
3
2
2
(A) PC BOARD-ELECTRICAL CIRCUITS Do not attempt to service the PC Board. Return to SSCOR for service (B) Regulator Do not attempt to service the regulator. Return to SSCOR for service (C) Vacuum Pumps Do not attempt to service the vacuum pumps. Return to SSCOR for service (D) Battery Holder Do not attempt to service the battery holder. Return to SSCOR for service