SONOPET 5450-800-039 Torque Wrench Quick Setup Guide.pdf
Page 1
SONOPET
®
Ultrasonic Aspirator
Quick setup guide
To install ultrasonic tip
3
Do not over tighten the ultrasonic tip during installation.
Stop tightening immediately when the wrench handle
pivots out of alignment with the wrench head. Failure to
comply may result in product damage, loss of power and/or
overheating of the handpiece during use.
Place the torque wrench on the ultrasonic tip as
shown. The side of the torque wrench with the
symbols must face the distal end of the tip (figure 2).
Note:
When used correctly,
the wrench handle will
pivot out of alignment
with the wrench head
during installation.
Figure 2
Position
torque wrench.
1
Thread the ultrasonic tip onto the handpiece.
Note:
There’s no need to remove the handpiece cover.
2
4
Place the handpiece tip joint into the notched bracket
on the sterilization and assembly tray (figure 1).
Hold the handpiece in place and rotate the torque
wrench in the direction of the arrow to tighten the
ultrasonic tip. Stop tightening immediately when
the wrench handle pivots out of alignment with the
wrench head, indicating that adequate force has
been applied (figure 3).
Figure 1
Place handpiece in
sterilization tray.
Instructions for torque wrench part number 5450-800-039 only.
Please refer to the product IFU for additional detailed information.
Figure 3 Install ultrasonic tip.
Page 2
SONOPET
To remove ultrasonic tip
Ultrasonic Aspirator
Caution:
When removing the ultrasonic tip,
DO NOT remove the tip joint.
Note:
There is no need to remove the handpiece
cover. Removing the tip cover alone will
expose the tip joint.
1
Place the handpiece tip joint into the notched bracket
on the sterilization and assembly tray (figure 4).
2
Place the torque wrench on the ultrasonic tip as
shown. The side of the torque wrench with the
symbols must face the distal end of the tip (figure 5).
3
Figure 4
Place handpiece in
sterilization tray.
If necessary, reset the head of the torque wrench to
its proper position by rotating the handle in the
opposite direction of the arrow until it clicks into
alignment with the head (figure 6).
4
Hold the handpiece in place and rotate the torque
wrench in the opposite direction of the arrow to
loosen and remove the ultrasonic tip (figure 6).
5
Discard the ultrasonic tip. See the Disposal/Recycle section
in Instructions for Use.
Figure 5
Position torque
wrench.
Figure 6 Remove ultrasonic tip.
Instructions for torque wrench part number 5450-800-039 only.
Please refer to the product IFU for additional detailed information.
Neuro Spine ENT
This document is intended solely for the use of healthcare professionals. A surgeon must always rely on his or her own professional clinical
judgment when deciding whether to use a particular product when treating a particular patient. Stryker does not dispense medical advice
and recommends that surgeons be trained in the use of any particular product before using it in surgery.
The information presented is intended to demonstrate a Stryker product. A surgeon must always refer to the package insert, product label
and/or instructions for use, including the instructions for cleaning and sterilization (if applicable), before using any Stryker product. Products
may not be available in all markets because product availability is subject to the regulatory and/or medical practices in individual markets.
Please contact your Stryker representative if you have questions about the availability of Stryker products in your area.
Stryker Corporation or its affiliates own, use, or have applied for the following trademarks or service marks: SONOPET, Stryker.
All other trademarks are trademarks of their respective owners or holders.
Literature Number: 9100-004-059 Rev. None
Copyright © 2016 Stryker
Manufactured by:
Stryker Neuro Spine ENT
4100 East Milham Avenue
Kalamazoo, MI 49001 USA
Tel: 269 323 7700
Fax: 269 353 3026
Toll free: 800 253 3210
stryker.com/nse