stryker_core_electric_handpieces_and_attachments_st.pdf
Page 1
Instruments
50-minute minimum exposure time
•
8-minute minimum drying time
270°F Gravity:
•
Gravity displacement sterilizer
•
270-275°F (132-135°C)
•
Double wrapped in an instrument tray
or fully perforated sterilization case
•
35-minute minimum exposure time
•
8-minute minimum drying time
Dynamic Air Removal (PreVac
or HiVac):
•
Dynamic Air Removal sterilizer
•
270-275°F (132-135°C)
•
Double wrapped in an instrument tray
or fully perforated sterilization case
•
4-minute minimum exposure time
•
8-minute minimum drying time
NOTE: Stryker does not recommend “Flash”
sterilization for routine sterilization of surgical
instruments. “Flash” sterilization should only
be used when individual surgical instruments
require immediate sterilization and use.
“Flash” Autoclave:
•
Gravity displacement sterilizer
•
270-275°F (132-135°C)
•
Unwrapped in a an instrument tray
•
10-minute minimum exposure time
AD
LO
! CAUTION! DO NOT run water directly into
open ends of handpiece.
2. Hold handpiece on an incline, distal tip pointing
down, flush with running water to rinse away
cleaning solutions.
T
EC
EJ
3. Use only Stryker recommended cleaning brush to
remove debris around bur and blade mounts.
4. Wipe dry with lint free cloth or force air dry.
To Clean a Handswitch:
1. Wipe clean with a mild detergent and water or alcohol.
2. Extend the finger rest to clean all surfaces
and rinse with water.
3. Wipe dry with a lint free towel or force air dry.
To Clean an Attachment:
1. Clean an attachment with a brush or sponge,
a hospital enzymatic cleaning solution and water.
2. Use Stryker recommended cleaning brush to clean
cannula area. DO NOT use a brush in the nose or
cannula of an Elite or Saber SD series attachment.
N
•
RU
Double wrapped in an instrument tray
or fully perforated sterilization case
AD
•
1. Wash exterior of all CORE components with
Medical Grade Cleaner or soap solution.
LO
250-254°F (121-123°C)
N
•
of the handpiece or into electrical connections.
Failure to comply may result in handpiece damage.
RU
Gravity displacement sterilizer
! DO NOT allow water to run into the distal end
N
•
in a washer/sterilizer.
RU
250°F Gravity:
! DO NOT immerse or place any component
AD
To obtain optimal performance and help
prevent damage, perform one of the
following sterilization procedures:
To Clean the Handpiece:
LO
Sterilization
RU
N
Electric Handpieces
and Attachments
Sterilization & Cleaning
3. The distal tips of attachments may soak in a cup
or beaker of detergent for 5-10 minutes.
4. Hold the attachment on an incline and flush with
running water to rinse away the cleaning solution.
“Flash” Dynamic Air Removal
(PreVac or HiVac):
5. Use a lint free towel to dry or force air dry.
•
Dynamic Air Removal sterilizer
•
270-275°F (132-135°C)
Lubrication Recommendations
•
Unwrapped in an instrument tray
•
4-minute minimum exposure time
To lubricate an Elite attachment, use Stryker TPS lubricant REF 5110-901-900.
•
1-minute or no dry time
To lubricate a Saber SD Series attachment use Stryker ContraClean REF 5110-901-920.
The information presented in this brochure is intended to
demonstrate the breadth of Stryker product offerings. Always
refer to the package insert, product label and/or user instructions
before using any Stryker product. Products may not be available
in all markets. Product availability is subject to the regulatory or
medical practices that govern individual markets. Please contact
your Stryker representative if you have questions about the
availability of Stryker products in your area.
N
RU
AD
! DO NOT lubricate after every use. Lubricate every 2 months or 40 sterilization cycles.
LO
*Sterilization Recommendations Validation based on
AAMI protocol.
! Most attachments do not require lubrication.
T
EC
EJ
Products referenced with TM designation are trademarks of Stryker.
Products referenced with ® designation are registered trademarks of Stryker.
Literature Number: 9100-001-009 Rev. A