TEM

Thromboelastometry Systems

ROTEM sigma How to Run a Test Guide

Guide

1 Page

ROTEM® sigma – HOW TO RUN A TEST Have patient sample (blue top tube filled to line) and sealed ROTEM cartridge ready  Always follow the on screen instructions: STEP 1:  ü To LOGIN, select username icon: xxxx and input password: xxx (lowercase)  ü Press START ü Open & remove cartridge from packet PLEASE NOTE: If a message comes up requesting maintenance or QC, please select CANCEL  STEP 2: ü Insert the cartridge into the ROTEM® sigma Cartridge must be used within 10 minutes after opening  STEP 3: ü Enter patient name and URN ü Press  to confirm  STEP 4: ü Gently mix the citrate sample (blue top tube) – DO NOT SHAKE  ü WAIT while the cartridge is being prepared ü When PROMPTED, insert the sample into the cartridge The measurements will start automatically after 5 minutes  STEP 5: ü Let the test run until desired results are available (A5 / A10 or longer) You can print at ANY TIME - please ALSO print when test has finished  ü To PRINT: select SIGMA on print switch, press PRINT and select 4 ü Stop all channels by pressing STOP ALL when required ü To save the measurement(s) in the database and clear the channel(s), then press SAVE/CLEAR ALL  STEP 6: IMPORTANT STEP ü WAIT until prompted to remove cartridge from device ü When prompted, remove and dispose of the cartridge into the  SHARPS BIN (DO NOT remove sample from cartridge – discard together) ü Device is now ready for the next sample…
File Type: PDF
File Size: 1.75 MB
File Name: rotem_sigma_how_to_run_a_test_guide.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions