WOLF Lone Bone Plate System Cleaning and Sterilization Instructions

Revision Not Listed

TOBY Plate System

3 Pages

WOLF® Cleaning and Sterilization Instructions Toby Orthopaedics, Inc... 01 SW 42  nd  Ave, Suite 200  Miami, Florida 33134 USA Domestic Toll-free: 866... 9... BY (8629) Office: 406... 6... 60 Email: sales@TobyOrtho... m Atlantico Systems LTD 34 Oldfield Kingston Galway Ireland Phone: +35 391443609 Product  WOLF® Long Bone Plate System  Warnings  1... Reprocessing Limitations  1... Intended Use  1... Point of Use  1... Containment / Transportation  1... Preparation for Decontamination  Implantable devices (plates, screws, etc... are single use only... eusable instruments that are provided clean and ready to use are non-sterile unless the packaging indicates otherwise... lean and sterilize prior to each use, unless device is already provided in clean / sterile packaging... here applicable, disassemble instruments prior to cleaning... or additional warnings, reference the WOLF® Instructions For Use (Doc 60000024)... mplantable devices are single use only and may not be reused once in contact with bodily fluid, soft tissue, bone, or other sources of contamination... epeated processing has minimal effect on reusable WOLF® surgical instruments... nd of life is normally determined by wear and damage due to the use of the instruments... he WOLF® Plate system is indicated for fractures, osteotomies, and non-unions of upper extremity diaphyses... he plates and screws contained within the system are single use devices, which are designed for implantation per the relevant surgical procedure... he drill bits, drill guides, depth gauges, k-wire, drivers, and other instrumentation are designed for use during the surgical procedure and require decontamination and cleaning prior to reuse... he instruments typically contact blood, bone, tissue and bodily fluids during normal use, and therefore, require that the cleaning and processing steps are completed for each usage... emove gross soiling by submerging the instrument into cold water (<40ºC) immediately after use... on't use a fixating detergent or hot water (>40ºC) as this can cause the fixation of residue, which may influence the result of the reprocessing process... emove surface contamination with paper tissue... ospital precautions for the handling of contaminated/bio-hazardous materials should be followed... nstruments should be cleaned within 30 minutes after use to minimize the potential of staining, damage, and drying... If possible, the devices must be reprocessed in a disassembled or opened state... Disassemble the Depth Gauge to clean each component of the assembly: Depth Gauge Slide and Housing... ully retract the Slide in the Housing... ully separate the Slide from the Housing while taking care to not bend the Tip... ®  TOBY ORTHOPAEDICS 50000133-03 Page 1 of 3   c... Inspect the O-ring and remove / clean / replace / reinstall as necessary... Pre-Cleaning  Rinse with cold water for approximately 3 minutes... Manual Cleaning Procedure  Applicable towards reusable devices and instrumentation... ll reusable instrumentation shall be considered critical, especially the items with small lumens / cannula or silicone surfaces... 2... 4... 6... Immerse the device(s) in an enzymatic detergent solution prepared in warm (30-35ºC) tap water (8 mL EcoLab Neutral Enzymatic Detergent per liter or equivalent detergent solution) and allow device(s) to soak for not less than five (5) minutes... spirate not less than sixty (60) mL of the detergent solution through any lumens present, as applicable, using an appropriate sized syringe... rush all lumens present on device(s), as applicable, using an appropriately sized nylon bristled channel brush... et the brush in the detergent solution and run the brush down the entire length of each lumen and back, not less than five (5) times... rush any silicone surface with a nylon bristled instrument cleaning brush for not less than two (2) minutes per device where silicone material is present... repare an enzymatic detergent solution in a sonicating water bath using purified water (2 mL EcoLab Neutral Enzymatic Detergent per liter of water or equivalent detergent solution)... onicate the device(s) for not less than ten (10) minutes in the enzymatic detergent solution... pon completion of the manual brushing and sonication steps, process the device(s) through a standard washer / disinfector surgical instruments cycle with the following parameters (or equivalent):  Program Step Pre-rinse Wash Rinse 1 Rinse 2 Thermal Disinfection Drying 8... Automated Cleaning Cycle Parameters Medium Temperature (ºC) Time Cold Water N/A 3 minutes 4 minutes heating from Cold and 35-60ºC, 60+ºC 35 to 60ºC, Warm Water 4 minutes > 60ºC Warm Water N/A 3 minutes Warm Water N/A 1 minute 5 minutes heating from Purified Water 30-80ºC, 80+ºC 30 to 80ºC, 5 minutes > 80ºC Air 120ºC 11 minutes  Perform a final rinse of the device(s) in running purified water, not less than ~250 mL per device, and allow the device(s) to air dry... erform a visual inspection of the device(s) to ensure that all contaminants have been removed... esidual bodily fluids, excessive discoloration, unacceptable corrosion, etc... are typical reasons to further process the device(s)... f the device(s) are determined to not be visually clean, the above relevant steps shall be repeated... f the device cannot be cleaned effectively, then it shall be disposed of safely or Toby Orthopaedics’ Customer Service shall be contacted at the phone number listed above... Automated Cleaning Procedure  Automated systems are only recommended for the WOLF® surgical instruments when combined with the manual cleaning procedure detailed in the section above... Disinfection  1... Drying  Dry the instruments with a lint-free towel... he instruments may never be heated >140ºC... o avoid water residues, insufflate cavities of instruments by using sterile compressed air... Inspection / Function Testing  1... Submerge instruments in a disinfection detergent according to the detergent manufacturer's instructions... inse the instruments with sterile water to remove the detergent... ote: Disinfection is only a supplementary step and does NOT replace steam sterilization... Carefully inspect each device to ensure that all visible blood and soil has been removed... isually inspect for damage and/or wear... heck the action of moving parts to ensure smooth operation throughout the intended  ®  TOBY ORTHOPAEDICS 50000133-03 Page 2 of 3
File Type: PDF
File Size: 219KB
File Name: wolf_lone_bone_plate_system_cleaning_and_sterilizat.pdf
Available for download

Available for download in the BioClinical Services library

If you already belong to a hospital with access, please login via your intranet.

Contact Us to Register Your InterestAlready a member? Log in hereAbout Us

wolf_lone_bone_plate_system_cleaning_and_sterilizat.pdf

Page 1

WOLF®

Cleaning and Sterilization Instructions

Toby Orthopaedics, Inc.

401 SW 42

nd

Ave, Suite 200

Miami, Florida 33134 USA

Domestic Toll-free: 866.979.TOBY (8629)

Office: 406.556.3260

Email: sales@TobyOrtho.com

Atlantico Systems LTD

34 Oldfield

Kingston

Galway

Ireland

Phone: +35 391443609

Product

WOLF® Long Bone Plate System

Warnings

1.

2.

3.

4.

5.

Reprocessing

Limitations

1.

2.

Intended Use

1.

2.

3.

Point of Use

1.

2.

3.

Containment /

Transportation

1.

2.

Preparation for

Decontamination

Implantable devices (plates, screws, etc.) are single use only.

Reusable instruments that are provided clean and ready to use are non-sterile unless the

packaging indicates otherwise.

Clean and sterilize prior to each use, unless device is already provided in clean / sterile

packaging.

Where applicable, disassemble instruments prior to cleaning.

For additional warnings, reference the WOLF® Instructions For Use (Doc 60000024).

Implantable devices are single use only and may not be reused once in contact with

bodily fluid, soft tissue, bone, or other sources of contamination.

Repeated processing has minimal effect on reusable WOLF® surgical instruments. End

of life is normally determined by wear and damage due to the use of the instruments.

The WOLF® Plate system is indicated for fractures, osteotomies, and non-unions of

upper extremity diaphyses.

The plates and screws contained within the system are single use devices, which are

designed for implantation per the relevant surgical procedure.

The drill bits, drill guides, depth gauges, k-wire, drivers, and other instrumentation are

designed for use during the surgical procedure and require decontamination and cleaning

prior to reuse. The instruments typically contact blood, bone, tissue and bodily fluids

during normal use, and therefore, require that the cleaning and processing steps are

completed for each usage.

Remove gross soiling by submerging the instrument into cold water (<40ºC) immediately

after use.

Don't use a fixating detergent or hot water (>40ºC) as this can cause the fixation of

residue, which may influence the result of the reprocessing process.

Remove surface contamination with paper tissue.

Hospital precautions for the handling of contaminated/bio-hazardous materials should be

followed.

Instruments should be cleaned within 30 minutes after use to minimize the potential of

staining, damage, and drying.

If possible, the devices must be reprocessed in a disassembled or opened state.

1.

Disassemble the Depth Gauge to clean each component of the assembly: Depth Gauge

Slide and Housing.

a. Fully retract the Slide in the Housing.

b. Fully separate the Slide from the Housing while taking care to not bend the Tip.

®

TOBY ORTHOPAEDICS

50000133-03

Page 1 of 3

Page 2

c.

Inspect the O-ring and remove / clean / replace / reinstall as necessary.

Pre-Cleaning

Rinse with cold water for approximately 3 minutes.

Manual

Cleaning

Procedure

Applicable towards reusable devices and instrumentation. All reusable instrumentation shall

be considered critical, especially the items with small lumens / cannula or silicone surfaces.

1.

2.

3.

4.

5.

6.

7.

Immerse the device(s) in an enzymatic detergent solution prepared in warm (30-35ºC)

tap water (8 mL EcoLab Neutral Enzymatic Detergent per liter or equivalent detergent

solution) and allow device(s) to soak for not less than five (5) minutes.

Aspirate not less than sixty (60) mL of the detergent solution through any lumens

present, as applicable, using an appropriate sized syringe.

Brush all lumens present on device(s), as applicable, using an appropriately sized nylon

bristled channel brush. Wet the brush in the detergent solution and run the brush down

the entire length of each lumen and back, not less than five (5) times.

Brush any silicone surface with a nylon bristled instrument cleaning brush for not less

than two (2) minutes per device where silicone material is present.

Prepare an enzymatic detergent solution in a sonicating water bath using purified water

(2 mL EcoLab Neutral Enzymatic Detergent per liter of water or equivalent detergent

solution).

Sonicate the device(s) for not less than ten (10) minutes in the enzymatic detergent

solution.

Upon completion of the manual brushing and sonication steps, process the device(s)

through a standard washer / disinfector surgical instruments cycle with the following

parameters (or equivalent):

Program Step

Pre-rinse

Wash

Rinse 1

Rinse 2

Thermal

Disinfection

Drying

8.

9.

Automated Cleaning Cycle Parameters

Medium

Temperature (ºC)

Time

Cold Water

N/A

3 minutes

4 minutes heating from

Cold and

35-60ºC, 60+ºC

35 to 60ºC,

Warm Water

4 minutes > 60ºC

Warm Water

N/A

3 minutes

Warm Water

N/A

1 minute

5 minutes heating from

Purified Water

30-80ºC, 80+ºC

30 to 80ºC,

5 minutes > 80ºC

Air

120ºC

11 minutes

Perform a final rinse of the device(s) in running purified water, not less than ~250 mL per

device, and allow the device(s) to air dry.

Perform a visual inspection of the device(s) to ensure that all contaminants have been

removed. Residual bodily fluids, excessive discoloration, unacceptable corrosion, etc.,

are typical reasons to further process the device(s). If the device(s) are determined to

not be visually clean, the above relevant steps shall be repeated. If the device cannot be

cleaned effectively, then it shall be disposed of safely or Toby Orthopaedics’ Customer

Service shall be contacted at the phone number listed above.

Automated

Cleaning

Procedure

Automated systems are only recommended for the WOLF® surgical instruments when

combined with the manual cleaning procedure detailed in the section above.

Disinfection

1.

Drying

Dry the instruments with a lint-free towel. The instruments may never be heated >140ºC. To

avoid water residues, insufflate cavities of instruments by using sterile compressed air.

Inspection /

Function

Testing

1.

2.

3.

Submerge instruments in a disinfection detergent according to the detergent

manufacturer's instructions.

2. Rinse the instruments with sterile water to remove the detergent.

Note: Disinfection is only a supplementary step and does NOT replace steam sterilization.

Carefully inspect each device to ensure that all visible blood and soil has been removed.

Visually inspect for damage and/or wear.

Check the action of moving parts to ensure smooth operation throughout the intended

®

TOBY ORTHOPAEDICS

50000133-03

Page 2 of 3

Page 3

range of motion.

Check instruments with long slender features (particularly rotating instruments) for

distortion.

5. Where instruments form part of a larger assembly, check that the devices assemble

readily with mating components.

Note: If damage or wear is noted that may compromise the function of the instrument,

contact your Toby Orthopaedics representative for a replacement.

4.

Maintenance

Lubricate hinges, threads and other moving parts with a commercial water-based surgical

grade instrument lubricant (such as instrument milk) to reduce friction and wear.

Reuse Life

1.

2.

3.

4.

Implantable components may not be reused, but may be resterilized, as required,

through the moist heat method described herein. The implantable components have

been validated for undergoing at least 15 sterilization cycles without loss of performance.

Drill guides, drill guide sleeves, drivers, and similar instruments may be reused as long

as the functionality of the device is sufficient to perform the intended use. Typically these

types of instruments have a lifetime of 50+ surgical cases unless the threads or driver tip

is damaged during use.

The depth gauge assembly and drill bits / k-wires may be reused as long as the

functionality of the device is sufficient and any marking used for measurements are

clearly legible. The depth gauge scale and other similar laser markings on these devices

begin to deteriorate after 10 to 15 sterilization / cleaning cycles, and should be visually

inspected to ensure accuracy is still guaranteed. The cutting edges of the drill bits / kwires deteriorate at various times based on use, and the drill bits should be safely

discarded once the edges become dull. The edges can become dull after one use or

may last 10 to 15 surgical cases. Dull drill bits should be safely discarded to prevent

localized burning of the bone during use. Similarly, it is not recommended that K-wire or

drill bits be reused for multiple cases due to the high velocity rotation of the devices

causing possible deformation of the item and deterioration of performance.

The sterilization trays are designed to last hundreds of sterilization / cleaning cycles and

only require replacement when functionality is lost or excessive contamination cannot be

remedied.

Packaging

Instruments shall be loaded into the dedicated WOLF® sterilization tray for sterilization. If

applicable, use standard medical grade, FDA-approved or equivalent, steam sterilization

wrap following the AAMI double wrap method (ANSI/AAMI ST79).

Sterilization

Steam sterilize using a pre-vacuum cycle for 3 minutes at a minimum temperature of 135ºC.

20 Minute minimum drying time in accordance with ANSI/AAMI ST79.

Note: Flash Sterilization is not recommended for WOLF®.

Storage

WOLF® surgical instruments should be stored in a cool dry place.

This document is in accordance with ISO 17664 current revision

CONFIDENTIAL: This document is property of Toby Orthopaedics, Inc., and shall not be reproduced or distributed without the

express written consent of Toby Orthopaedics, Inc..

®

TOBY ORTHOPAEDICS

50000133-03

Page 3 of 3