Instruction Manual
104 Pages

Preview
Page 1
ENGLISH
INSTRUCTION MANUAL
NON-MYDRIATIC RETINAL CAMERA
TRC-NW6SF/ TRC-NW6S
Copyrights and trademarks Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S is a trademark of the TOPCON Corporation. © TOPCON 2000
INTRODUCTION
Thank you for purchasing the TOPCON Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S. This manual contains instructions for the TOPCON Non-Mydriatic Retinal Camera. This includes the main features, basic operation, troubleshooting, and the maintenance and cleaning of this instrument. Please read the section Warning labels on page 5 and the section Safety precautions on page 6 carefully, before operating the instrument. To ensure the best performance of this instrument, read all instructions carefully and place this manual in a convenient location for future reference.
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
This machine offers the following innovative features
Observe the fundus of the eye and take images without using mydriatics. Improved picture quality and ease of use. Allows images of the periphery, using the alignment point and the external fixation target for peripheral photography. Allows Fluoro-photography by changing the photography mode (TRC-NW6SF only). Allows Fluoro-observation by using the built-in color video monitor (TRC-NW6SF only).
Precautions
4
This machine is a precision instrument. Be sure to install and use it in a controlled environment under normal temperature, humidity and atmospheric pressure conditions (10~40°C, 30-85%, 70~106kPa) and avoid direct exposure to sunlight. Never install the machine on a slope or an unstable surface. Before using the machine, ensure that all cables are correctly connected. Use a power supply within the range of ±10% of the rated voltage (50/60Hz). Ensure the surrounding area is always clean. When the instrument is not in use, turn off the Power switch, cap the objective lens, and install the dust cover. To ensure the best imaging results, handle the objective lens with particular care and keep it free of dust, fingerprints, or other stains.
Selecting externally connected equipment The TOPCON TRC-NW6SF/ TRC-NW6S complies with the CE standard. Before connecting a personal computer, image recorder, image printer, TV camera, or monitor to the TOPCON product, ensure that such external equipment is in compliance with the CE marking. Note:
The illustrations in this manual pertain to the model TRC-NW6SF only. Please note that the TRC-NW6S differs from the SF model.
Warning labels Important warnings are placed on the products and inserted in the instruction manuals, in order to encourage the safe use of products and prevent any danger to the operator and others or damage to existing facilities. We recommend that the following information about labels and icons be read before going on to read "Safety precautions" and the rest of the manual.
Meaning of labels
Label
Meaning
Ignoring or disregarding this label may lead to death or serious injury. WARNING
Ignoring or disregarding this label may result in personal injury or severe damage to the instrument or other facilities. PRECAUTIONS
Possible injuries include wounds, burns, electric shocks, etc. Damage to other facilities means extensive damage to buildings, equipment and furniture.
INSTRUCTION MANUAL
5
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
Meaning of icons
Icons
Meaning This icon indicates Caution. A further explanation or picture is placed close to the icon. This icon indicates Prohibition (DO NOT!). A further explanation or picture is placed close to the icon. This indicates Mandatory Action (DO!). A further explanation or picture is placed close to the icon.
Safety precautions This instruction manual details the safety precautions that need to be observed to prevent accidents. Always observe these precautions and use the instrument correctly. WARNING Icons
6
Precautions
Page
To avoid the risk of electric shock, disconnect the cable from the power supply before assembling the instrument. Always ensure the power supply is disconnected before assembly.
29
Ensure the power supply is equipped with a 3-plug AC receptacle to safeguard against power leakage. Power leakage could cause fire or electric shock.
36
To avoid the risk of electric shock, do not attempt disassembling, rebuilding and/or repairs. Always consult your dealer in case of repairs.
95
WARNING Icons
Precautions
Page
To avoid the risk of electric shock do not remove any instrument components, TV relay lens covers, chinrest or power supply components, other than the lamp house cover.
95
To avoid the risk of electric shock, always disconnect the power cable from the instrument before removing the fuse cover. Always replace the fuse cover before reconnecting the instrument to the power supply.
89
Install the instrument in a clean, dry area. -
To avoid the risk of fire or electric shock, do not place any liquids near the instrument.
-
To avoid the risk of electric shock, do not insert metal items into any vents and/or slots.
-
To avoid the risk of fire, always use the correctly rated fuse as specified on the fuse holder.
-
To avoid the risk of fire, always use the correctly rated fuse as specified on the fuse holder.
89
INSTRUCTION MANUAL
7
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
CAUTION Icons
8
Precautions
Page
Ensure that the monitor lamp is no brighter than necessary, to avoid any discomfort or injury to the patient.
-
Ensure that the photography light is no brighter than necessary, to avoid any discomfort or injury to the patient.
65, 66
Do not continue visible fluoro-observation any longer than necessary (TRC-NW6SF only), to avoid any discomfort or injury to the patient.
77
Always install the instrument on a fixed, stable surface, to prevent damage or injury.
To avoid potential injury, do not place your hand or fingers into the area between the instrument body and the power supply unit.
29, 34, 85
68
To avoid potential burns, do not touch the lamp during, or immediately after use.
86
To avoid the risk of electric shock, never handle a plug with wet hands.
36
To avoid the risk of electric shock, never touch the xenon lamp immediately after a flash or burn out.
87
To avoid potential injury to the patient, always watch the patient carefully while adjusting the chinrest.
61
CAUTION Icons
Precautions
Page
To avoid potential injury to the patient while moving the instrument, always keep the instrument at a safe distance from the patient.
68
Always ensure the TV relay lens and TV camera are securely mounted, to avoid potential injury during adjustment and operation.
32, 85
Use an approved automatic instrument table to avoid potential injury while moving the instrument.
34
Ensure the fixing knob at the bottom of the instrument is secure, before moving the instrument.
34
The instrument should always be carried by two people, both holding the bottom of the instrument. Attempting to carry the instrument alone could result in backache or other injury. Attempting to carry the instrument without holding it at the bottom could result in injury or damage.
29, 34
Ensure that the camera mount lever is firmly tightened to prevent the TV camera from falling and causing injury or damage.
85
Ensure that the lens mount lever is firmly tightened to prevent the TV relay lens from falling and causing injury or damage.
85
To avoid the risk of electric shock, always disconnect the power supply and unplug the power cable before replacing the lamp.
86, 87
To prevent the instrument from falling and to avoid injury during movement, use the base locking knob to secure the power supply unit to the instrument.
-
INSTRUCTION MANUAL
9
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
Operation and maintenance Purpose This Non-Mydriatic Retinal Camera is a precision electronic medical instrument, and should only be used under the direction of a medical doctor.
User maintenance To ensure the safe and proper operation of the instrument, unless otherwise specified in this manual, all maintenance should be carried out by a trained service technician. The following maintenance tasks, however, can be performed by the user. Please refer to the relevant text in this manual for the maintenance method.
10
Replacing lamps
The illumination lamp and xenon lamp may be replaced by the user. For details, see Replacing the illumination lamp on page 86 and Replacing the xenon lamp on page 87.
Replacing fuses
Fuses used with the instrument may be replaced by the user. For details, see Changing the fuses on page 89.
Cleaning the objective lens
The objective lens may be cleaned by the user. For details, see Cleaning the objective lens on page 91.
TOPCON Instruments general statements
TOPCON shall not bear any responsibility for damage due to fire, earthquakes, actions by third persons and other accidents, or the negligence and misuse by the user and use under unusual conditions. TOPCON shall not bear any responsibility for damage resulting from the inability to use this equipment, such as a loss of business profit and suspension of business. TOPCON shall not bear any responsibility for damage caused by operations other than those described in this Instruction Manual. The diagnoses that are made are the responsibility of relevant doctors, and TOPCON shall not take any responsibility for the results of such diagnoses.
Warning labels and positions To ensure safe operation of this equipment, warning labels are provided. Please adhere to the following warning instructions.
INSTRUCTION MANUAL
11
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
CAUTION • To avoid potential injury to the patient, always watch the patient carefully while adjusting the chinrest.
CAUTION • Never touch the patient's eyes or nose with the instrument during operation as this may result in personal injury.
WARNING • Electrical shock may cause burns or fire. Turn the main power switch OFF and UNPLUG the power cable before replacing fuses. Only replace with fuses of the correct rating.
CAUTION • To avoid the risk of electric shock, always disconnect the power supply and unplug the power cable before replacing the lamp. • To avoid potential burns, do not touch the lamp during, or immediately after use.
WARNING • To avoid the risk of electric shock, do not attempt to remove the instrument cover.
No user serviceable parts inside, refer servicing to qualified personnel.
WARNING • Electrical shock may cause burns or fire. Turn the main power switch OFF and UNPLUG the power cable before replacing fuses. Only replace with fuses of the correct rating.
12
Contents
INTRODUCTION This machine offers the following innovative features
4
Precautions
4
Warning labels
5
Meaning of labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Meaning of icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety precautions Operation and maintenance Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . User maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1
5 6
6 10 10 10
TOPCON Instruments general statements
11
Warning labels and positions
11
COMPONENTS 1.1
Component names
17
1.2
Control panel components
21
1.3
Monitor screen
22
1.4
Standard accessories
24
1.5
Assembly components
26
Non-Mydriatic Retinal Camera TRC-NW6SF/TRC-NW6S
2
ASSEMBLY PROCEDURE 2.1
Assembling the instrument body
29
2.2
Connecting the tv relay lens with the instrument body
32
2.3
Confirmation after assembly
33
2.4
Installing the instrument
34
2.5
Connecting the power cable
36
2.6
Connecting the external device
36
2.6.1
Connecting the TV camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
2.6.2
Connecting the external recording device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2.6.3
Outputting images to an external monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
2.6.4
VIDEO OUT terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
2.7
39
2.7.1
Preparation for menu setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
2.7.2
Displaying the Menu screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
2.7.3
Returning to the Monitor screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
2.7.4
Setting the Record/Playback mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
2.7.5
Switching of fixation target on/flicker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
2.7.6
Switching of internal/external fixation targets . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
2.7.7
Fixation target pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
2.7.8
Flash level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
2.8
3
Reset from power save state
43
INITIAL SETTING 3.1
Preparation for initial set-up
45
3.2
Setting the record/playback mode
46
3.3
Setting the fixation target
47
3.3.1
Switching the fixation target on/flicker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3.3.2
Switching of internal/external fixation targets . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
3.4
Setting peripheral photography
48
3.4.1
Fixation target pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
3.4.2
R/L reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.4.3
Display mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.5
xiv
Menu setting
Setting the screen display
51
3.5.1
Flash level compensation display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
3.5.2
Flash level display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.5.3
Illumination level display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.5.4
Picture angle display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
3.5.5
Illumination aperture display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
3.5.6
Peripheral fixation position display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
3.6
System setting
55
3.7
Setting the initial state
55
3.7.1
Flash level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
3.7.2
Commercial power supply frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
3.7.3
LED display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
3.7.4
Timer sound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
3.7.5
Operation sound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
3.7.6
Power save time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
3.7.7
Quick mirror up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
3.8
4
Setting the language
60
BASIC OPERATION 4.1
Preparation for photography
61
4.1.1
Connecting to the power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4.1.2
Positioning the patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4.2
Color photography (center)
64
4.2.1
Setting the photography mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
4.2.2
Setting the picture position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
4.2.3
Setting the picture angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
4.2.4
Setting illumination level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
4.2.5
Setting the flash level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4.2.6
Changing the diopter compensation lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
4.2.7
Collimation and photography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
4.3
Fluoro-photography (center)
75
4.3.1
Setting the photography mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.3.2
Setting the picture position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.3.3
Setting the picture angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.3.4
Setting illumination level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.3.5
Setting the flash level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.3.6
Changing the diopter compensation lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.3.7
Preparing the patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.3.8
Collimation and photography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.3.9
Fluoro-observation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
4.4
Peripheral photography
79
4.4.1
Setting the picture position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
4.4.2
Other settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
4.4.3
Collimation and photography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
INSTRUCTION MANUAL
xv
Non-Mydriatic Retinal Camera TRC-NW6SF/TRC-NW6S
4.5
Setting the photography mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
4.5.2
Setting the picture position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
4.5.3
Setting the picture angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
4.5.4
Setting the illumination level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
4.5.5
Changing the diopter compensation lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
4.5.6
Collimation and photography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
5
Finishing
84
MAINTENANCE AND CHECKS 5.1
Daily care
85
5.2
Changing the lamp
86
5.2.1
Replacing the illumination lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
5.2.2
Replacing the xenon lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
5.3
Changing the fuses
89
5.4
Supplying the chinrest tissue
90
5.5
Adjusting the color video monitor
90
5.6
Cleaning the instrument
91
5.6.1
Cleaning the dust cover, control panel and monitor screen . . . . . . . . . . . . . . . . 91
5.6.2
Cleaning the objective lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
5.7
List of expendables and spare parts
92
5.8
Optional accessories
93
5.8.1
5.9
6
xvi
82
4.5.1
4.6
7
Anterior segment photography
Adjustable instrument table AIT-11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Contacting TOPCON
94
TROUBLESHOOTING 6.1
Before requesting service
95
6.2
Troubleshooting table
96
6.3
Error code list
99
6.4
Emergency chinrest operation
99
SPECIFICATIONS
1
COMPONENTS
1.1 Component names
(1)
(2) (9) (3)
(10) (11) (12) (13)
(4)
(14) (15) (16)
(5)
(17)
(6) (18) (7) (19) (8) (20)
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
(1)
TV camera connection terminal
(2)
Diopter compensation lens selector
(3)
Focusing knob
(4)
IR filter selector
(5)
Image quality adjustment knob
(6)
Base brake knob
(7)
Vertical position mark
(8)
Fuse holder
(9)
TV relay lens attaching lever
(10)
TV relay lens
(11)
Angle selector
(12)
Mode selector *
(13)
Fluoro-observation knob *
(14)
Name plate slot
(15)
Color video monitor
(16)
Photography switch
(17)
Omni-directional joystick
(18)
Power lamp
(19)
Control panel
(20)
Fixing knob (for use when carrying the instrument)
* TRC-NW6SF only
18
(21)
(22)
(26)
(27) (23) (28) (24)
(25)
(21)
Forehead rest
(22)
Canthus marker
(23)
Chin-rest tissue pin
(24)
Lamp house cover
(25)
Chinrest
(26)
Anterior segment fixation target
(27)
Objective lens
(28)
Lamp house cover screw
INSTRUCTION MANUAL
19
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
(34)
(29)
(30)
(31) (32) (33)
20
(29)
Base
(30)
Power switch
(31)
Fuse holder
(32)
Power supply unit
(33)
External output terminal:
(34)
External fixation target
1.2 Control panel components
(40) (35)
(36)
(37)
(41)
(42)
(38)
(39)
(35)
Menu switch Displays the Menu screen.
(36)
Split switch Turns split lines on/off.
(37)
Flash Level switch ([-], [reset], [+]) Adjusts the flash level for the patient’s eye condition.
(38)
Illumination Level switch ([-], [+]) Adjusts the illumination level for the patient’s eye condition.
(39)
Chinrest switch ([down], [up]) Positions the patient’s head by electronically moving the chinrest up/down.
(40)
Illumination Aperture switch * Changes the flash level for color photography. Also, changes the illumination aperture in fluoro-photography with 45°/ picture angle.
(41)
Timer switch * Displays the elapsed time on the monitor/photography image.
* TRC-NW6SF only
INSTRUCTION MANUAL
21
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
(42)
Periphery Fixation switch ([left turn], [reset], [right turn]) Switches the position of the internal fixation target, therefore guiding the patient’s eye to the periphery fixation point.
1.3 Monitor screen Monitor screen (43) (44) (45) (46)
(51) (52)
(53)
(47) (48) (49) (50)
(43)
Nameplate/timer image
(44)
Xenon charging display
(45)
Flash level compensation display
(46)
Flash level display
(47)
IR warning display
(48)
Picture angle display
(49)
Illumination aperture display *
(50)
Fixation target position display
(51)
Scale
(52)
Alignment bright spots
(53)
Split lines
(54)
Illumination level
* TRC-NW6SF only
22
(54)
Menu screen
(55)
(56)
Preview display (color) (57)
(58)
(55)
Cursor
(56)
Input guide display
(57)
Nameplate/timer image
(58)
Photography image
INSTRUCTION MANUAL
23
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
1.4 Standard accessories Upon unpacking, make sure that all the following standard accessories are delivered. Figures in parentheses, ( ), indicate the quantity.
(59)
(60)
(61)
(62)
(63)
(64)
(65) (66)
24
(59)
Power cable (1)
(60)
Nameplate (10)
(61)
Illumination lamp (1)
(62)
Fuse (9)
(63)
Chinrest tissue (1)
(64)
Chinrest tissue pin (2)
(65)
Emergency chinrest knob (1)
(66)
Spare parts case (1)
(67)
Instruction Manual
(68)
Warranty (1)
(69)
Dust cover (1)
(70)
External fixation target (1)
INSTRUCTION MANUAL
(67)
(68)
(69)
(70)
(71)
(72)
25
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
(71)
Power cable (1)
(72)
TV camera signal cable (1)
1.5 Assembly components
(6) (3)
(7) (4) (1) + (2)
(5)
(9)
(10)
(11)
(8)
(12)
(15)
(13)
(16)
(17)
(14)
(20) (18)
26
(19)
1.5.1
Components Description
Quantity
Container
(1) Instrument body
1
O
(2) TV relay lens
1
O
(3) Rail cover
2
O
(4) TV camera power cable
1
O
(5) TV camera signal cable
1
O
(6) Power cable
1
O
(7) External fixation target
1
O
(8) Star screwdriver
1
O
(9) Hexagon rod spanner
1
O
(10) Focusing tool
1
Accessories case
(11) Dust cover
1
O
(12) Nameplate
10
(13) Chin rest tissue pin (spare)
2
(14) Illumination lamp
1
(15) Chinrest tissue
1
(16) Fuse (spare)
9
(17) Emergency chinrest knob
1
(18) Accessories case
1
O
(19) TRC-NW6SF Instruction Manual
1
O
(20) Warranty
1
O
INSTRUCTION MANUAL
Accessories case Accessories case Accessories case Accessories case Accessories case Accessories case
27
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
28
2
ASSEMBLY PROCEDURE
2.1 Assembling the instrument body
To avoid the risk of electric shock, disconnect the cable from the power supply before assembling the instrument. Never plug in the power cable before assembly.
WARNING
Always install the instrument on a fixed, stable surface, to prevent damage or injury.
CAUTION
CAUTION
Note:
The instrument should always be carried by two people, both holding the bottom of the instrument. Attempting to carry the instrument alone could result in backache or other injury. Attempting to carry the instrument without holding it at the bottom could result in injury or damage.
Since the upper part and lower part of the instrument are connected only by the power cable, take them out together so they do not separate from each other.
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
1 Remove the instrument body (1) from the container and place it on the appropriate table. 2 Slightly raise the omni-directional joystick, and pull out the cushion from the lower part of the base in the direction of the arrow.
Omni-directional joystick
Styrofoam
Sponge Sponge Cushion
3 Wipe the sliding board with a cloth to remove any dirt. 4 Remove the styrofoam from the transportation bracket (A), on the left hand side of the patient, and unscrew the bracket (B).
transportation bracket (A)
transportation bracket (B)
5 Slide the base to the right and unscrew the transportation bracket (A).
30
6 Fasten the rail covers (3), using the small screws attached.
7 Set the external fixation target (7). Align the connector groove of the external fixation target with the connector of the instrument body and insert it. Secure it firmly and ensure that it cannot be moved.
8 Fix the chinrest tissue (15) with the chinrest tissue pin. Chinrest tissue pin
Chinrest tissue
INSTRUCTION MANUAL
31
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
2.2 Connecting the tv relay lens with the instrument body
Always ensure the TV relay lens and TV camera are securely mounted, to avoid potential injury during adjustment and operation.
CAUTION
This machine can be connected to an optional TV camera. If your TV camera is SONY DXC-970MD, the attached TV camera power cable (4) and TV camera signal cable (5) can be connected to the instrument body (1). 1 Turn the TV camera fixing ring in the direction of the arrow until it reaches the stopper, and remove the protective cap.
Lens fixing ring
Protective cap
2 Mount the TV camera so the lens positioning hole is aligned to the pin of the TV relay lens (2).
32
3 Turn the lens fixing ring in the direction of the arrow and fasten it.
2.3 Confirmation after assembly 1 Slide the base to the left of the operator, and ensure that the primary voltage is within the correct range of voltage as set by the voltage selector.
2 Ensure that the input voltage is within the range ±10% of the rated voltage. If the input voltage exceeds the range, use a constant-voltage power supply (marketed: 400VA up). 3 Loosen the base locking knob, and move the omni-directional joystick to confirm that it moves smoothly. - Right-left movement - Back-forth movement - Up-down movement
INSTRUCTION MANUAL
33
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
Note:
The right-left movement of the instrument may be uneven when it is first unpacked. If so, move the joystick to its limit in each direction.
2.4 Installing the instrument
Ensure the fixing knob at the bottom of the instrument is secure, before moving the instrument.
CAUTION
CAUTION
The instrument should always be carried by two people, both holding the bottom of the instrument. Attempting to carry the instrument alone could result in backache or other injury. Attempting to carry the instrument without holding it at the bottom could result in injury or damage.
Use an approved automatic instrument table to avoid potential injury while moving the instrument.
CAUTION
CAUTION
34
Always install the instrument on a fixed, stable surface, to prevent damage or injury.
1 Fasten the fixing knob. 2 Hold the instrument firmly at the specified positions and place it on the adjustable instrument table. For details about the adjustable instrument table, see Optional accessories on page 93.
Holding positions
Fixing knob
Holding the instrument body
3 After installing the instrument, loosen the fixing knob fully. The instrument body can then be moved. 4 If the instrument body is slightly off level, fine-adjust the height using the four adjusters. Do not extend the adjuster past 1 cm.
Adjuster
INSTRUCTION MANUAL
35
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
2.5 Connecting the power cable
Ensure the power supply is equipped with a 3-plug AC receptacle to safeguard against power leakage. Power leakage could cause fire or electric shock.
WARNING
To avoid the risk of electric shock, never handle a plug with wet hands.
CAUTION
1 Make sure that the power switch (30) of the instrument body is OFF. 2 Connect the power cable to the instrument body.
3 Plug the power cable into an 3-p AC outlet that is properly grounded.
2.6 Connecting the external device 2.6.1
Connecting the TV camera This machine can be connected and used with an optional TV camera. If your TV camera is SONY DXC-970MD, the attached TV camera power cable and TV camera signal cable can be connected to the instrument body.
36
1 Connect the TV camera connection terminal of the instrument body to the DC IN/REMOTE terminal of the TV camera with the TV camera power cable. 2 Connect the TV camera connection terminal of the instrument body to the RGB/SYNC terminal of the TV camera with the TV camera signal cable. DC IN/REMOTE terminal
RGB/SYNC terminal
TV camera (SONY DXC-970MD)
Note:
If your TV camera is not a SONY DXC-970MD, contact your dealer or TOPCON (see the back cover) for details about connection.
2.6.2
Connecting the external recording device This machine can be connected to an external recording device via the external connection terminal. 1 Connect the external recording device to the image connection terminal of the instrument with an image cable (optional). 2 Connect the external recording device to the control terminal of the instrument with a control cable (optional).
INSTRUCTION MANUAL
37
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
3 When using a TOPCON IMAGEnet system, connect the photography data cable (optional) to the photography data output terminal of the instrument.
Image connection terminal
Control terminal
Photography data output terminal
Note:
Various types of connection cables are available for connectiing the instrument with external recording devices. For details about connecting external recording devices, contact your dealer or TOPCON (see the back cover).
2.6.3
Outputting images to an external monitor The images displayed on the color video monitor may be outputted to an external monitor via the VIDEO OUT terminal of the external connection terminal. 1 Connect the BNC cable (optional) to the VIDEO OUT terminal of the instrument. 2 Connect the other end of the BNC cable to the input terminal of the external monitor.
Control terminal
38
2.6.4
VIDEO OUT terminal
Note:
For details about video output to external recording devices, contact your dealer or TOPCON (see the back cover).
2.7 Menu setting The following items can be set within Menu setting: record/playback, fixation target on/flicker, internal/ external fixation targets, fixation target patterns and flash level.
2.7.1
Preparation for menu setting 1 Check the power cable connection. For details about the connection, see Connecting the power cable on page 36. 2 Turn the power switch (30) on.
2.7.2
Displaying the Menu screen 1 Check the monitor screen. 2 Press the Menu switch (35) on the control panel. Check the MAIN MENU screen.
Cursor
3 Press the flash level switch (37) (+) or (-). The cursor then moves.
INSTRUCTION MANUAL
39
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
2.7.3
Returning to the Monitor screen 1 Press the menu switch (35).
2.7.4
Setting the Record/Playback mode It is possible to alternate between Record and Playback modes. In the Record mode, ordinary photography is available. Use the Playback mode to display the recorded images of the recording device to the color video monitor. The record mode RECORD is set automatically once the power supply is turned on. 1 On the MAIN MENU screen, make sure that the cursor is at RECORD/PLAYBACK, and press the flash level switch (37) (reset). The monitor screen changes to the RECORD/PLAYBACK screen.
2 Press the flash level switch (37) (+) or (-) and select RECORD (record mode) or PLAYBACK (playback mode). 3 Press the flash level switch (37) (reset); the setting is complete and the MAIN MENU screen returns. Note:
When the playback mode is on, photography is prohibited.
2.7.5
Switching of fixation target on/flicker The fixation target can be switched between on and flicker states. When the instrument is shipped, FLICKERING is set.
40
1 Move the cursor to FIXATION TYPE on the MAIN MENU screen, and press the flash level switch (37) (reset) to display the FIXATION TYPE screen.
2 Press the flash level switch (37) (+) or (-) and select CONSTANT (on) or FLICKERING (flicker). 3 Press the flash level switch (37) (reset); the setting is complete and the MAIN MENU screen returns.
2.7.6
Switching of internal/external fixation targets It is possible to alternate between Internal and External fixation targets. When the instrument is shipped, INTERNAL (internal fixation) is set.. 1 Move the cursor to INTERNAL/EXTERNAL FIX on the MAIN MENU screen, and press the flash level switch (37) (reset) to display the INTERNAL/EXTERNAL FIX screen.
2 Press the flash level switch (37) (+) or (-) and select INTERNAL (internal fixation) or EXTERNAL (external fixation). 3 Press the flash level switch (37) (reset); setting is complete and the MAIN MENU screen returns.
2.7.7
Fixation target pattern It is possible to select a fixation target pattern.
INSTRUCTION MANUAL
41
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
When the instrument is shipped, 8 is set. 1 Move the cursor to FIXATION PATTERNS on the MAIN MENU screen, and press the flash level switch (37) (reset) to display the FIXATION PATTERNS screen.
2 Press the flash level switch (37) (+) or (-) and select the desired pattern from the following:
Note:
4A
(diagonal 4 positions),
4B
(vertical-horizontal 4 positions),
5A
(diagonal 4 positions + center),
5B
(vertical-horizontal 4 positions + center),
8
(8 positions), or
9
(8 positions + center).
The center position is assigned to the optical axis center and is used for photographing around the macular segment.
3 Press the flash level switch (37) (reset); setting is complete and the MAIN MENU screen returns.
2.7.8
Flash level It is possible to change the reference value of the flash level. When the instrument is shipped, 0 (no change) is set.
42
1 Select FLASH LEVEL on the MAIN MENU screen, and press the flash level switch (37) (+) or (-) to display the FLASH LEVEL screen.
2 Press the flash level switch (37) (+) or (-) and select the desired level from the following: +2
(2 steps up),
+1
(1 step up),
0
(no change),
-1
(1 step down), or
-2
(2 steps down).
3 Press the flash level switch (37) (+) or (-); setting is complete and the MAIN MENU screen returns. Note:
A step up/down changes the flash level reference value by about 20%.
2.8 Reset from power save state This machine features a power save function. When the instrument body has not been operated for a set time, the power save function stops power supply to the monitor, CCD camera, illumination light source and photography light source. During power saving, the power lamp of the control panel flickers and the monitor screen switches off. 1 Press the photography switch (16). In a few seconds, the video monitor displays and is ready for photography. Note:
When the instrument is shipped, the power save set time is 10 minutes. To change the set time, contact your dealer or TOPCON (see the back cover).
INSTRUCTION MANUAL
43
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
44
3
INITIAL SETTING
On the INITIAL MENU screen, the following items can be programmed; record/playback, internal/external fixation targets, peripheral photography, system setting, initial setting and language.
3.1 Preparation for initial set-up 1 Ensure that the power cable is connected. 2 Press the menu switch (35) and photography switch (16) of the control panel and turn the power switch (30) on. Press the menu switch (35) and photography switch (16) until the buzzer sounds. The Title screen is displayed, and a few seconds later, the Monitor screen is displayed.
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
3 Press the menu switch (35) of the control panel to display the SETTING MENU screen.
4 To finish with the INITIAL MENU screen, press the menu switch (35). Note:
Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.
3.2 Setting the record/playback mode It is possible to alternate between Record and Playback modes. Use the Playback mode to display the recorded images of the recording device to the color video monitor. The record mode RECORD is set automatically once the power supply is turned on. 1 Move the cursor to the RECORD/PLAYBACK of the INITIAL MENU screen, and press the flash level switch (37) (reset). The monitor screen changes to the CSTM SCREEN.
2 Press the flash level switch (37) (+) or (-) and select the RECORD or PLAYBACK mode. 3 Press the flash level switch (37) (reset) and the INITIAL MENU screen displays.
46
Note:
Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.
3.3 Setting the fixation target It is possible to set the fixation target on/flicker and internal/external fixation target switching on the FIXATION screen. 1 On the INITIAL MENU screen, move the cursor to FIXATION and press the flash level switch (37) (reset). The monitor screen changes to the FIXATION screen.
2 To finish with the FIXATION screen, press the menu switch (35). Note:
Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.
3.3.1
Switching the fixation target on/flicker The fixation target can be switched between on and flicker states. When the instrument is shipped, FLICKERING (flicker) is set. 1 Select FIXATION TYPE on the FIXATION screen to display the FIXATION TYPE screen.
INSTRUCTION MANUAL
47
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
2 Press the flash level switch (37) (+) or (-) and select CONSTANT (on) or FLICKERING (flicker). 3 Press the flash level switch (37) (reset); setting is complete and the FIXATION screen returns.
3.3.2
Switching of internal/external fixation targets Internal/external fixation targets can be switched. When the instrument is shipped, INTERNAL (internal fixation) is set. 1 Select INTERNAL/EXTERNAL FIX on the FIXATION screen, to display the INTERNAL/EXTERNAL FIX screen.
2 Press the flash level switch (37) (+) or (-) and select INTERNAL (internal fixation) or EXTERNAL (external fixation). 3 Press the flash level switch (37) (reset); setting is complete and the FIXATION screen returns.
3.4 Setting peripheral photography The following items can be programmed on the PERIPHERAL FIXATION screen: fixation target pattern, R/L reset and display mode.
48
1 Move the cursor to the PERIPHERAL FIXATION of the INITIAL MENU screen, and press the flash level switch (37) (reset). The monitor screen changes to the PERIPHERAL FIXATION screen.
2 To finish with the PERIPHERAL FIXATION screen, press the menu switch (35). Note:
Settings will not be changed if the power switch (30) is turned off without first finishing with the INITIAL MENU screen as outlined above.
3.4.1
Fixation target pattern It is possible to select a fixation target pattern. When the instrument is shipped, 8 is set. 1 Move the cursor to FIXATION PATTERNS on the PERIPHERAL FIXATION screen, to display the FIXATION PATTERNS screen.
2 Press the flash level switch (37) (+) or (-) and select the desired pattern from the following: 4A
(diagonal 4 positions),
4B
(vertical-horizontal 4 positions),
5A
(diagonal 4 positions + center),
5B
(vertical-horizontal 4 positions + center),
8
(8 positions), or
9
(8 positions + center).
3 Press the flash level switch (37) (reset); the setting is complete and the PERIPHERAL FIXATION screen returns.
INSTRUCTION MANUAL
49
Non-Mydriatic Retinal Camera TRC-NW6SF/ TRC-NW6S
Note:
The center position is assigned to the optical axis center and is used for photographing around the macular segment.
3.4.2
R/L reset If the base is moved during peripheral photography, right/left eye reset can be set. (When OFF is set, peripheral photography cannot be reset by right/left eye switching.) When the instrument is shipped, OFF (no reset) is set. 1 Select R/L RESET on the PERIPHERAL FIXATION screen, to display the R/L RESET screen.
2 Press the flash level switch (37) (+) or (-) and select ON (reset) or OFF (no reset). 3 Press the flash level switch (37) (reset); the setting is complete and the PERIPHERAL FIXATION screen returns.
3.4.3
Display mode It is possible to set the display mode of the peripheral fixation target. When FIXATION POSITION (fixation position display) is set, the monitor screen display responds to the fixation target, and when PHOTOGRAPHIC POSITION (picture position display) is set the monitor screen display responds to the photography result. When the instrument is shipped, FIXATION POSITION (fixation position display) is set. 1 Select FIXATION INDICATOR on the PERIPHERAL FIXATION screen, to display the FIXATION INDICATOR screen.
50