BCS
  • Home
  • Contact Us
  • Sign In

TORNIER SAS

TORNIER Instrumentation Instructions for Use

Revision Not Listed

TORNIER Instrumentation

3 Pages

UTBI2013...  TORNIER INSTRUMENTATION INSTRUCTIONS FOR USE  The manufacturer recommends that all personnel responsible for handling and implanting the devices read and understand this information before use... he use of surgical instrumentation requires knowledge of anatomy, biomechanics, and reconstructive surgery of the musculo-skeletal system... urgical instrumentation must be used only by a qualified surgeon operating in accordance with current information on the state of scientific progress and the art of surgery... he user must ensure the adequate condition and function of surgical instrumentation before use... MPORTANT: When the hospital does not own the surgical instrumentation, it accepts invoicing and delivers payment in the following cases: - when the instrumentation must be destroyed according to sanitary rules - when the instrumentation has been damaged due to non-respect of Tornier instructions mentioned herein... ESCRIPTION The surgical instrumentation consists of ancillary instruments, packaging trays as well as containers... he instrumentation type is inscribed on the metal container or, if the instrument is delivered individually, on the packaging... he exact designation of each instrument is given on the instrumentation list supplied or, if the instrument is delivered individually, on the package label... ymbols are sometimes used to identify instruments (labeling or marking) and they have the following meanings: XS = Extra-small; S or SM = Small; S+ or SM+ = Small+; M or ME = Medium; M+ or ME+ = Medium+; L or LA = Large; L+ or LA+ = Large+;  XL = Extra-large; 2XL = Extra-extra-large; 3XL = Extra-extra-extra-large; L = Left; R = Right; "-" = Short neck; "0" = Medium neck; "+" = Long neck... ornier surgical instrumentation has been specially designed to facilitate the implantation of Tornier implants and must be used solely for this purpose... t is important to refer to the technical documentation prior to the operation, or contact your Tornier representative for a more detailed description of how to use the instrumentation... nder no circumstance should an instrument be implanted... NTENDED USE: The surgical instrumentation is used by surgeons for the implantation of orthopedic implants... o not modify the instruments... NSTRUMENT DELIVERED NON-STERILE When surgical instrumentation is delivered non-sterile, the hospital is responsible for its cleaning and sterilization prior to use, in accordance with validated methods... he following recommendations do not substitute for the sanitary rules in force: standards, guides, government notices,  ministerial  texts,  etc…   Before any operation, it is necessary to remove wedging foam in the metal containers as well as plastic bags if the instrument is delivered individually... nstruments made up of removable components must be dismantled before pre-disinfection and cleaning, in accordance with instructions provided in the inventory list... rticulated instruments must be opened in order to allow the cleaning of all interstices... he instructions hereafter must be followed in order to maintain optimal efficiency and safety of instruments: - The use of metallic brushes, scrub pads and other articles likely to damage the instruments must be avoided...  Chemicals such as chlorine or soda as well as organic or ammoniated acids or solvents (e...  acetone) which are likely to damage the instruments must not be used...  Chemicals including soda must not be used for metal containers... - Phosphoric acid must not be used for the neutralization of alkaline residues after the cycle of automated machine cleaning on instrumentation packaging trays and on instruments made up of polymere pieces (example : polymer handle)... ote: Orthopedic procedures are not considered at risk in relation to NCTA (Non-Conventional Transmissible Agents)...  complete cleaning using molar sodium (1N) or sodium hypochlorite with a concentration of 2% active chlorine should only be reserved for instruments that have been used on a patient with suspected or confirmed TSE (Transmissible Spongiform Encephalopaties) before the invasive procedure...  STORAGE AND HANDLING: Surgical instrumentation must be handled with care and stored in an appropriate, clean and dry location... t is recommended to remove instruments from plastic bags before storing them to avoid condensation... nstruments must not be stored in contact with or near products that may have a corrosive effect...  PRE-DISINFECTION : Pre-disinfection aims to reduce the population of micro-organisms and to make subsequent cleaning easier... t is also intended to protect staff while handling instruments and avoid contamination of the environment... ll reusable devices must undergo immediate pre-disinfection or be immediately treated in a washerdisinfector... re-disinfection is achieved by dipping instruments, for a minimum of 15 minutes, in a neutral or alkaline decontaminant/disinfectant bactericidal, fungicidal and possibly virucidal solution that does not contain aldehyde nor ethanol... he use of brushes is authorized to clean the parts from all soils that can potentially alter the action of detergents and decontaminants... he instruments should then be carefully rinsed in a controlled water to avoid interference between the decontaminant/disinfectant and cleaning solutions... t is important to refer to the instructions supplied by the manufacturer of these products... AUTION : Packaging trays and baskets must not be in contact with decontaminating solutions for a long time... lean dirty areas and rinse immediately...  CLEANING: The equipment is then thoroughly cleaned out of the container (the efficiency of parts cleaned inside their loading container is not ensured), after disassembly if necessary...  cleaning process done out of qualification ranges can lead to sterility or toxicity issue... leaning eliminates contamination of the material... t must be performed in a washer-disinfector with a neutral or slightly alkaline detergent used at a maximum of 60°C... he detergent shall be selected for medical applications and present no known residual toxicity for the patient... he use of a mechanical action through manual of ultra-sonic means is recommended... n case the process cannot be done automatically, a manual process shall be used by reproducing the conditions described in the cleaning recommendations... he cleaning cycle must include a final rinse with a controlled water... ime, water flow and rinsing volumes shall be sufficient to produce a low level of cleaning agent residues left on the product surface... nstruments should be carefully dried to avoid recontamination...  thermal decontamination at 93°C is recommended... he correct operation of each instrument must be inspected thereafter per the specific product documentation supplied... he cleaning conditions validated by Tornier can be found in this document... n the case of patients with suspected or confirmed TSE, the cleaning procedure for the washer-disinfector shall be done after a decontamination process conform to the instruction DGS/RI3/2011/449 and 29CFR1910... 30... he use of lubricants to optimize the use of the instruments is authorized... he products used shall be selected specifically for a medical use... 1/3   Wrapped Immersion  Rinsing 1  Rinsing 2 Water  Drying  Solution  Time (min)  T°C  Water  1... re-disinfection / Decontamination (FR / UK conditions)  MEDICLEAN 0...  or ALKAZYME 0...  pH10-11  15  Ambient  Tap water 1’ in ambient temperature – 12...  minimum  2... anual precleaning on ultrasonic bath (FR / UK conditions)  MEDICLEAN FORTE 2% pH10... 10...  5  20-30°C  Tap water 1’ in ambient temperature – 12...  minimum  3... utomatic cleaning process (FR / UK conditions)  MEDICLEAN 0...  pH 8... 9...  10  93°C  Softened water in ambient temperature – 12...  minimum  1... re-disinfection / Decontamination (US conditions)  ALKAZYME 0...  pH10-11  15  Ambient  Tap water 1’ in ambient temperature – 12...  minimum  2... anual precleaning on ultrasonic bath (US conditions)  CIDEX 28 7... 8...  5  20-30°C  Tap water 1’ in ambient temperature – 12...  minimum  3... utomatic cleaning process (US condition)  Prolystica 0...  pH1...  10  93°C  Softened water in ambient temperature – 12...  minimum  STERILIZATION PARAMETERS RECOMMENDED IN THE UK6,7: The heaviest containers must be placed at the bottom of the autoclave... n order to avoid residual water in containers after sterilization we advise that a folded paper or non woven sheet is placed on the back of the container before sterilization, to improve vaporization during final drying... ornier recommends to sterilize through this method:  Purified water 1’ in ambient temperature – 12...  minimum  Y  T°C Incertainty Time Drying Wrapped  Purified water 1’ in ambient temperature – 12...  minimum  After sterilization, the wrapped trays shall be handled with care to the point of use to prevent any disrupture of the sterile barrier... he sterilization conditions validated by Tornier can be found in this document... TERILIZATION PARAMETERS RECOMMENDED IN THE USA5,7 The sterilization may be performed using the following method... ornier recommends to sterilize through this method :  Pre-vacuum method 132°C (269... F) +3°, -0° 4 min 20 min yes  STERILIZATION PARAMETERS RECOMMENDED IN FRANCE1,2,3,4,7 The heaviest containers must be placed at the bottom of the autoclave... n order to avoid residual water in containers after sterilization we advise that a folded paper or non woven sheet is placed on the back of the container before sterilization, to improve vaporization during final drying... ornier recommends to sterilize through this method :  T°C Incertainty Time Drying  Pre-vacuum method 134°C (273... F) +3°, -0° 18 min 20 min  Pre-vacuum method 134°C (273... F) +3°, -0° 3 min 20 min yes  Y  3...  STERILIZATION: Instruments, packaging trays and baskets are adapted for steam sterilization at a temperature not exceeding 140°C... terilization is mandatory for Tornier instruments, disinfection is not sufficient... t shall be done using the containers wrapped per the l’ANSI AAMI ST 79, Fig...  The instruments’ function shall be controlled before the sterilization per the specific product check-lists supplied... enerally speaking, the instruments must be sterilized when assembled, but certain instruments consisting of components that can be disassembled must be disassembled before predisinfection and cleaning if assembly/disassembly instructions are provided with the product check-list... T°C Incertainty Time Drying Wrapped  yes  3... AFTER USE: After each use and before returning to Tornier, the instrumentation (entire box or isolated instrument) must be pre-disinfected, cleaned and sterilized according to the aforementioned recommendations... nstruments that appear to be non-functional must immediately be sent to Tornier for maintenance or exchange... he nature of dysfunction must be clearly indicated... he instrumentation must be correctly packaged before being returned, and the original positioning of the components in corresponding containers should be respected... nstrumentation must be returned with the Count Sheet filled in and duly signed by authorized hospital personnel (with respect to position, qualification or authority)... INGLE USE INSTRUMENT DELIVERED STERILE 4...  PRECAUTIONS OF USE: - Never re-use an instrument designated for single-use, even if it appears to be in perfect condition, to prevent any risks of crosscontamination or a risk of reduced performances...  Never re-sterilize an instrument designated for single-use...  STORAGE AND HANDLING: The instrumentation must be stored in an appropriate, dry and clean location to prevent any loss of sterility... nstrumentations must not be exposed to direct sunlight, ionising radiation, extreme temperatures nor particular contamination... nstrumentations must be handled with care to preserve integrity of their  packaging... nstruments delivered sterile must be stored in their sealed packaging of origin...  PACKAGING AND STERILIZATION: Instruments delivered sterile are sterilized by gamma irradiation... he expiration date for sterilization and integrity of the packaging must be checked... n instrument whose packaging is open or damaged or whose expiration date has passed must not be used... very precaution must be taken to ensure sterility when opening the packaging of the instrument... OTES 1... rench rule DGS/RI3/2001/449 of December 1st, 2011 updating recommendations to reduce the risk of transmitting non-conventional transmissible agents during invasive procedures and French Circular DGS/SD5C/DHOS/2005/435 of 23 September 2005 regarding recommendations for treatment of medical devices used on patients having received labile blood products retrospectively from donors that suffered suffering from a variant of Creutzfeld-Jakob Disease (CJD)... ood Pharmaceutical Practice – 22 June 2001 French regulation... D S 98-135 : April 2005 – Guide for the sterilisation of medical devices – Treatments applied to reusable medical devices... F EN ISO 17664 – August 2004 – Sterilization of medical devices – Information to be provided by the manufacturer for the processing of resterilizable medical devices... 2/3
File Type: PDF
File Size: 311KB
File Name: TORNIER Instrumentation Instructions for Use Jan 2013.pdf
Download

If your hospital with access, please login via your intranet.

Contact Us to Register Your InterestAlready a member? Log in hereAbout Us

TORNIER Instrumentation Instructions for Use Jan 2013.pdf

Page 1

UTBI2013.1

TORNIER INSTRUMENTATION

INSTRUCTIONS FOR USE

The manufacturer recommends that all personnel responsible for

handling and implanting the devices read and understand this

information before use. The use of surgical instrumentation

requires knowledge of anatomy, biomechanics, and reconstructive

surgery of the musculo-skeletal system. Surgical instrumentation

must be used only by a qualified surgeon operating in accordance

with current information on the state of scientific progress and the

art of surgery.

The user must ensure the adequate condition and function of

surgical instrumentation before use.

IMPORTANT: When the hospital does not own the surgical

instrumentation, it accepts invoicing and delivers payment in the

following cases:

- when the instrumentation must be destroyed according to sanitary

rules

- when the instrumentation has been damaged due to non-respect of

Tornier instructions mentioned herein.

1. DESCRIPTION

The surgical instrumentation consists of ancillary instruments,

packaging trays as well as containers. The instrumentation type is

inscribed on the metal container or, if the instrument is delivered

individually, on the packaging. The exact designation of each

instrument is given on the instrumentation list supplied or, if the

instrument is delivered individually, on the package label.

Symbols are sometimes used to identify instruments (labeling or

marking) and they have the following meanings:

XS = Extra-small;

S or SM = Small;

S+ or SM+ = Small+;

M or ME = Medium;

M+ or ME+ = Medium+;

L or LA = Large;

L+ or LA+ = Large+;

XL = Extra-large;

2XL = Extra-extra-large;

3XL = Extra-extra-extra-large;

L = Left;

R = Right;

"-" = Short neck; "0" = Medium neck; "+" = Long neck.

Tornier surgical instrumentation has been specially designed to

facilitate the implantation of Tornier implants and must be used

solely for this purpose. It is important to refer to the technical

documentation prior to the operation, or contact your Tornier

representative for a more detailed description of how to use the

instrumentation. Under no circumstance should an instrument be

implanted.

2. INTENDED USE:

The surgical instrumentation is used by surgeons for the

implantation of orthopedic implants.

Do not modify the instruments.

3. INSTRUMENT DELIVERED NON-STERILE

When surgical instrumentation is delivered non-sterile, the hospital

is responsible for its cleaning and sterilization prior to use, in

accordance

with

validated

methods.

The

following

recommendations do not substitute for the sanitary rules in force:

standards, guides, government notices,  ministerial  texts,  etc…  

Before any operation, it is necessary to remove wedging foam in

the metal containers as well as plastic bags if the instrument is

delivered individually.

Instruments made up of removable components must be dismantled

before pre-disinfection and cleaning, in accordance with

instructions provided in the inventory list. Articulated instruments

must be opened in order to allow the cleaning of all interstices.

The instructions hereafter must be followed in order to maintain

optimal efficiency and safety of instruments:

- The use of metallic brushes, scrub pads and other articles likely to

damage the instruments must be avoided.

- Chemicals such as chlorine or soda as well as organic or

ammoniated acids or solvents (e.g. acetone) which are likely to

damage the instruments must not be used.

- Chemicals including soda must not be used for metal containers.

- Phosphoric acid must not be used for the neutralization of alkaline

residues after the cycle of automated machine cleaning on

instrumentation packaging trays and on instruments made up of polymere

pieces (example : polymer handle).

Note: Orthopedic procedures are not considered at risk in relation to

NCTA (Non-Conventional Transmissible Agents). A complete cleaning

using molar sodium (1N) or sodium hypochlorite with a concentration of

2% active chlorine should only be reserved for instruments that have been

used on a patient with suspected or confirmed TSE (Transmissible

Spongiform Encephalopaties) before the invasive procedure.

3.1. STORAGE AND HANDLING: Surgical instrumentation must be

handled with care and stored in an appropriate, clean and dry location.

It is recommended to remove instruments from plastic bags before

storing them to avoid condensation. Instruments must not be stored in

contact with or near products that may have a corrosive effect.

3.2. PRE-DISINFECTION : Pre-disinfection aims to reduce the population of

micro-organisms and to make subsequent cleaning easier. It is also

intended to protect staff while handling instruments and avoid

contamination of the environment. All reusable devices must undergo

immediate pre-disinfection or be immediately treated in a washerdisinfector.

Pre-disinfection is achieved by dipping instruments, for a minimum of

15 minutes, in a neutral or alkaline decontaminant/disinfectant

bactericidal, fungicidal and possibly virucidal solution that does not

contain aldehyde nor ethanol. The use of brushes is authorized to

clean the parts from all soils that can potentially alter the action of

detergents and decontaminants. The instruments should then be

carefully rinsed in a controlled water to avoid interference between the

decontaminant/disinfectant and cleaning solutions. It is important to

refer to the instructions supplied by the manufacturer of these

products.

CAUTION : Packaging trays and baskets must not be in contact with

decontaminating solutions for a long time. Clean dirty areas and rinse

immediately.

3.3. CLEANING: The equipment is then thoroughly cleaned out of the

container (the efficiency of parts cleaned inside their loading container

is not ensured), after disassembly if necessary. A cleaning process

done out of qualification ranges can lead to sterility or toxicity issue.

Cleaning eliminates contamination of the material. It must be

performed in a washer-disinfector with a neutral or slightly alkaline

detergent used at a maximum of 60°C. The detergent shall be selected

for medical applications and present no known residual toxicity for the

patient. The use of a mechanical action through manual of ultra-sonic

means is recommended. In case the process cannot be done

automatically, a manual process shall be used by reproducing the

conditions described in the cleaning recommendations. The cleaning

cycle must include a final rinse with a controlled water. Time, water

flow and rinsing volumes shall be sufficient to produce a low level of

cleaning agent residues left on the product surface. Instruments should

be carefully dried to avoid recontamination.

A thermal decontamination at 93°C is recommended. The correct

operation of each instrument must be inspected thereafter per the

specific product documentation supplied.

The cleaning conditions validated by Tornier can be found in this

document.

In the case of patients with suspected or confirmed TSE, the cleaning

procedure for the washer-disinfector shall be done after a

decontamination process conform to the instruction DGS/RI3/2011/449

and 29CFR1910.1030.

The use of lubricants to optimize the use of the instruments is authorized.

The products used shall be selected specifically for a medical use.

1/3

Page 2

Wrapped

Immersion

Rinsing 1

Rinsing 2

Water

Drying

Solution

Time

(min)

T°C

Water

1. Pre-disinfection

/ Decontamination

(FR / UK

conditions)

MEDICLEAN

0.5%

or

ALKAZYME 0.5%

pH10-11

15

Ambient

Tap water 1’ in

ambient

temperature –

12.5l minimum

2. Manual

precleaning on

ultrasonic bath

(FR / UK

conditions)

MEDICLEAN

FORTE 2%

pH10.4-10.8

5

20-30°C

Tap water 1’ in

ambient

temperature –

12.5l minimum

3. Automatic

cleaning process

(FR / UK

conditions)

MEDICLEAN

0.6%

pH 8.5-9.5

10

93°C

Softened water in

ambient

temperature –

12.5l minimum

1. Pre-disinfection

/ Decontamination

(US conditions)

ALKAZYME 0.5%

pH10-11

15

Ambient

Tap water 1’ in

ambient

temperature –

12.5l minimum

2. Manual

precleaning on

ultrasonic bath

(US conditions)

CIDEX 28

7.5-8.1

5

20-30°C

Tap water 1’ in

ambient

temperature –

12.5l minimum

3. Automatic

cleaning process

(US condition)

Prolystica 0.7%

pH1.2

10

93°C

Softened water in

ambient

temperature –

12.5l minimum

STERILIZATION PARAMETERS RECOMMENDED IN THE UK6,7:

The heaviest containers must be placed at the bottom of the autoclave.

In order to avoid residual water in containers after sterilization we

advise that a folded paper or non woven sheet is placed on the back of

the container before sterilization, to improve vaporization during final

drying.

Tornier recommends to sterilize through this method:

Purified water 1’ in

ambient

temperature –

12.5l minimum

Y

T°C

Incertainty

Time

Drying

Wrapped

Purified water 1’ in

ambient

temperature –

12.5l minimum

After sterilization, the wrapped trays shall be handled with care

to the point of use to prevent any disrupture of the sterile

barrier.

The sterilization conditions validated by Tornier can be found in

this document.

STERILIZATION PARAMETERS RECOMMENDED IN THE USA5,7

The sterilization may be performed using the following method.

Tornier recommends to sterilize through this method :

Pre-vacuum

method

132°C (269.6°F)

+3°, -0°

4 min

20 min

yes

STERILIZATION PARAMETERS RECOMMENDED IN FRANCE1,2,3,4,7

The heaviest containers must be placed at the bottom of the

autoclave.

In order to avoid residual water in containers after sterilization

we advise that a folded paper or non woven sheet is placed on

the back of the container before sterilization, to improve

vaporization during final drying.

Tornier recommends to sterilize through this method :

T°C

Incertainty

Time

Drying

Pre-vacuum method

134°C (273.2°F)

+3°, -0°

18 min

20 min

Pre-vacuum method

134°C (273.2°F)

+3°, -0°

3 min

20 min

yes

Y

3.4. STERILIZATION: Instruments, packaging trays and baskets are

adapted for steam sterilization at a temperature not exceeding

140°C. Sterilization is mandatory for Tornier instruments,

disinfection is not sufficient. It shall be done using the containers

wrapped per the l’ANSI AAMI ST 79, Fig.6.

The instruments’ function shall be controlled before the

sterilization per the specific product check-lists supplied.

Generally speaking, the instruments must be sterilized when

assembled, but certain instruments consisting of components

that can be disassembled must be disassembled before predisinfection and cleaning if assembly/disassembly instructions

are provided with the product check-list.

T°C

Incertainty

Time

Drying

Wrapped

yes

3.5 AFTER USE:

After each use and before returning to Tornier, the instrumentation (entire

box or isolated instrument) must be pre-disinfected, cleaned and sterilized

according to the aforementioned recommendations. Instruments that

appear to be non-functional must immediately be sent to Tornier for

maintenance or exchange. The nature of dysfunction must be clearly

indicated. The instrumentation must be correctly packaged before being

returned, and the original positioning of the components in corresponding

containers should be respected.

Instrumentation must be returned with the Count Sheet filled in and duly

signed by authorized hospital personnel (with respect to position,

qualification or authority).

4. SINGLE USE INSTRUMENT DELIVERED STERILE

4.1. PRECAUTIONS OF USE:

- Never re-use an instrument designated for single-use, even if it

appears to be in perfect condition, to prevent any risks of crosscontamination or a risk of reduced performances.

- Never re-sterilize an instrument designated for single-use.

4.2. STORAGE AND HANDLING: The instrumentation must be stored in an

appropriate, dry and clean location to prevent any loss of sterility.

Instrumentations must not be exposed to direct sunlight, ionising

radiation, extreme temperatures nor particular contamination.

Instrumentations must be handled with care to preserve integrity of their

packaging.

Instruments delivered sterile must be stored in their sealed packaging of

origin.

4.3. PACKAGING AND STERILIZATION: Instruments delivered sterile are

sterilized by gamma irradiation. The expiration date for sterilization and

integrity of the packaging must be checked.

An instrument whose packaging is open or damaged or whose expiration

date has passed must not be used.

Every precaution must be taken to ensure sterility when opening the

packaging of the instrument.

NOTES

1. French rule DGS/RI3/2001/449 of December 1st, 2011 updating

recommendations to reduce the risk of transmitting non-conventional

transmissible agents during invasive procedures and French Circular

DGS/SD5C/DHOS/2005/435 of 23 September 2005 regarding

recommendations for treatment of medical devices used on patients

having received labile blood products retrospectively from donors that

suffered suffering from a variant of Creutzfeld-Jakob Disease (CJD).

2. Good Pharmaceutical Practice – 22 June 2001 French regulation.

3. FD S 98-135 : April 2005 – Guide for the sterilisation of medical

devices – Treatments applied to reusable medical devices.

4. NF EN ISO 17664 – August 2004 – Sterilization of medical devices –

Information to be provided by the manufacturer for the processing of

resterilizable medical devices.

2/3

Page 3

5. “Recommended   practices   for   sterilization   in   perioperative  

practice   settings,”   in   Standards, Recommended Practices, and

Guidelines (Denver: AORN, Inc, 2007) 673 – 677.

Comprehensive guide to steam sterilization and sterility assurance

in health care facilities - ANSI/AAMI ST79-2010.

AAMI TIR12 : Designing, testing, and labelling reusable medical

devices for reprocessing in health care facilities : a guide for

medical device manufacturers.

AAMI TIR30 : A compendium of processes, materials, test

methods, and acceptance criterias for cleaning reusable devices.

6.   “Health   Technical   Memorandum   2010”   – Part 2: design

considerations – Sterilization – London: HMSO – NHS Estates

7. NF EN ISO 17665-1 - Sterilization of health care products.

Moist heat. Requirements for the development, validation and

routine control of a sterilization process for medical devices

Interpretation of terms and symbols

Do not reuse

Use by

Sterilized using irradiation

Batch code (sterilization)

Serial number (manufacturing)

Catalogue number

Caution, consult accompanying documents

Consult instructions for use

Non-sterile

Do not use if package is damaged

Do not re-sterilize

Not made with natural rubber latex

Rev. 02/2013

The French text is the reference text.

Manufacturer :

TORNIER SAS

161, rue Lavoisier

38330 MONTBONNOT SAINT MARTIN - FRANCE

Tél: +33 (0)4.76.61.35.00 – Fax: +33 (0)4.76.61.35.33

USA Business address:

TORNIER, Inc.

10801 Nesbitt Avenue South

Bloomington, MN 55437 USA

Tel +1 952 921 7100

Fax +1 952 236 4007

= Reusable instrument

0459

= Instrument depending on a

source of electrical energy /

single-use instrument

3/3

Previous Page Next Page
© 2021 BioClinical Services Ptd Ltd
Terms of Use
Privacy Statement