Neonatal Resuscitator 082
Operating manual
Sa - Wi - 82
ver : 01
Page 3
Contents
Introduction
•Equipment introduction .....................................................
•Different parts of equipment .............................................
•Keyboard of equipment .....................................................
1
2
3
Installation and starting .............................................................. 5
•The method of equipment installation and starting ........... 7
•Installation place conditions .............................................. 9
•Power supply conditions ................................................... 10
•The method of using .......................................................... 11
Maintenance
•Cleaning and disinfection of equipment ............................ 15
Trouble – shooting ..................................................................... 16
Technical specifications ............................................................. 17
Signs and labels .......................................................................... 18
Address and specifications of European representative ............. 19
Page 5
Preface
The company of Manufacturing Neonatal Resuscitation Systems (Tosan) is a research,
Production, industrial unit in neonatal medical equipment field, works about designing,
producing, and installation and after sales service of this equipment.
The company has worked in this field since 1360 until now and products of this company
are as follow:
Model 110a
Infant incubator
Model 120
Infant portable incubator
Model 082
Neonatal Resuscitator
Model 022
8 lamps cylindrical phototherapy
Model 024
12 lamps cylindrical phototherapy
Model 025
16 lamps cylindrical phototherapy
Model 023
Neonatal table phototherapy
Model 051
Neonatal Radiant simple warmer
Model 052
Neonatal Radiant servo warmer
Model 100
Oxygen tent
Model 031
Neonatal simple cot
Model 032
Neonatal cot with variable degree
Model 048
Head holder
Model 044
Average Oxygen Hood
Model 047
Adult Oxygen Hood
Model 041
Neonatal Oxygen Hood
Model 042
Neonatal Oxygen Hood
Model 070
Neonatal height measurement
Page 6
Preface
Tosan company, first designer and producer of infant medical equipment in IRAN, besides
of making use of systems as circuital customer and producing on the basis of demands of
customer, with standardizing and immunizing of equipment according to the updated standards
of world, provides safety conditions of users and patients .
Summaries of these accomplishments are as follow:
•
Having certification ISO 9001 : 2000 from Associated Partners Germany
•
Having certification ISO 13486 : 2003 from Associated Partners Germany
•
Having certification ISO 13485 : 2003 from QS Swiss
•
Having CE from QS Swiss
•
Having operating license standard for Incubator from institute of standards and
industrial research of IRAN (ISIRI)
•
Having wide sales agency and after sales service in the country and also abroad.
Page 7
Introduction
Equipment introduction
Neonatal Resuscitator
Is an instrument to infants operating, blood replacement, taking care of infant initially birth
near the delivery table.
This equipment includes manometer and flow meter oxygen capsule to give infant oxygen,
suction to operation or care at the time of birth, top thermal system to increase infant
temperature, lighting system and different alarms to notify user of disorder in equipment
controller or in infant temperature.
1
Page 8
Introduction
Different parts of equipment
1
2
3
8
9
4
5
10
11
6
12
7
13
14
15
1. Fan
2. Equipment head
3. Halogen lamp
4. Keyboard of equipment
5. Monitor stand (made – to – order)
6. Talc divider (movable)
7. Suction glass
2
8. IV pole
9. IV pole upright
10. Sensor connection to equipment
11. Pressure gauge of suction
12. Table
13. Oxygen capsule
14. Suction
15. Wheel
Page 9
Introduction
Keyboard of equipment
1
11
12
2
13
3
4
14
5
15
6
16
7
8
9
17
18
10
3
Page 10
Introduction
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
4
Keyboard of equipment
Sensor temperature display
Adjusted temperature display
Temperature adjustment key (up – down)
Heater activation display
Alarm for temperature more than allowed value
Sensor fault alarm
Temperature deviation fault alarm
Low battery alarm
Temperature deviation key
Key and display of selection kind of equipment Operation (Auto – manual)
Timer display
Timer activate key
Beginning and ending key of timer
Halogen lamps controller keys
Suction activate key and intended lamp
Table tilt adjustment keys
Table height adjustment keys
On/off key
Page 11
Installation and starting
Tosan’s neonatal resuscitator, packaging to three parts, to facilitate transport: (equipment head,
column, down part of equipment)
Also starting of equipment must be done by after sales service representatives of Tosan.
If equipment is installed by nonresponsible persons it won’t have 1 year warranty.
Before installation, make sure of existence of all instruments. These instruments have been
illustrated at equipment accompanied paper
(list of accompanied instrument).
1. Equipment upright
2. Equipment top column
3. Equipment heater enclosure
4. Capsule (made – to – order)
5. IV pole (made – to – order)
6. Flow meter (made – to – order)
7. Monometer (made – to – order)
8. Installation bolts
9. Pressure gauge installation upright (made – to – order)
10. Flow meter upright (made – to – order)
11. Pressure gauge (made – to – order)
12. Pressure gauge hose
13. Oxygen hose (made – to – order)
14. Temperature sensor
15. Suction hose (made – to – order)
16. Plastic joint
5
Page 12
Installation and starting
Attention
If there isn’t one of these items, please contact with sales unit or
representatives of Tosan.
Do installation and starting of equipment, after controlling
instrument inside the boxes.
6
Page 13
Installation and starting
The method of equipment installation and starting
Operator of equipment must be familiar with operation before equipment starting. Training
could be done by after sales representatives who have the charge of starting the equipment
and training the user, or by studying this operation instruction.
Do the following stages to install equipment.
1. Remove upper column from box, then place it on down column, so that installation strap
place inside upper column.
Attention
Connect electronic jacks, which are connections between upper and down column,
at their position.
2.Install pressure gauge upright to the right of column by two Allen bolt. (The side that IV
upright connected to column.)
3.Connect monometer upright to other side of column by other two Allen bolt 6.
4. Connect the other point of hose, connected to pressure gauge to down column hose and
fasten in with one of the fasteners.
5. Open the nut of flow meter and pass through flow meter upright hole, so that point of
hose would place at the back of equipment, then fasten its nut.
6. Place heater enclosure of warmer on upper column, so that four hole of enclosure upright
would settle at four hole of bolt side of upright. Nd fasten enclosure with four bolt,
accompanied with equipment.
7
Page 14
Installation and starting
The method of equipment installation and starting
Attention
Before fastening bolts, keep enclosure top to up.
• Connect jack which is intended to heater enclosure, to jack at upper column.
• Place plastic cap at top of the upright, on its position.
• Place IV pole at its position.
• After installation by representatives of company to make sure of proper installation,
control different parts of equipment to prevent of happening any problem at the time
of using equipment.
• Now equipment is ready to start, and do the following stages.
• Place temperature sensor jack at its position on upper upright and fasten its bolt.
• Open belt of capsule and place it on its position then fasten the belt.
• Connect monometer to capsule and fasten its bolt.
8
Page 15
Installation and starting
Installation place condition
Proper temperature of installation environment is between 25 to 30 centigrade degree. Increasing
temperature more than 30°c or exceeding decreasing will influence on thermal adjustment of
equipment.
Attention
• Never put the equipment under direct radiant of the sun.
• Don’t put equipment near cooler or heater instrument.
• Minimum proper distance between equipment and cooler or heater
instrument should be 60 centimeter.
• At the time of equipment operation, the head of equipment is too hot to
touch or place something on it.
• Maximum allowed weight on IV upright is 2 kilogram.
• Air movement could disorder infant thermal balance.
• To keeping fix equipment, wheels lock of equipment could be used.
9
Page 16
Installation and starting
Power supply conditions
To prevent of occurrence of any event, please observe the following items about electrical
cases:
1. Connect plug to a plug which is closed to equipment.
2. Use a separate plug for each of the equipment.
3. Plug must have ground wire.
10
Page 17
Installation and starting
The method of using
After installation, connect plug of equipment to plug, and then turn on on/off key of it.
Adjust the intended temperature with keys.
At this state, indicator will show the percent of turning on the sensor.
Green indicator indicates the sensed temperature by temperature sensor.
Equipment could be used at two moods.
1. Automatic mood
2. Manual mood
1. Automatic mood
If press automatic / manual key, so that automatic indicator will turn on. Equipment will
operate automatically and green indicator indicates sensed temperature by sensor.
Red indicator indicates adjusted temperature by user. At this state temperature control to
sensed temperature will be accomplished by sensor.
If temperature deviates by ± 2 C Air.S.D indicator will turn on with alarm. With pressing
A.S.D key audible alarm will stop for 20 minutes and if after passing this time equipment
couldn’t again reach infant temperature to adjusted temperature, audible alarm will start again
and all keys will be passive. Therefore turn off equipment and after trouble – shooting, turn on
equipment again.
If temperature deviates by ± 2 C Air.S.D indicator will turn on with alarm.With pressing A.S.D
key audible alarm will stop for 20 minutes and if after passing this time equipment couldn’t
again reach infant temperature to adjusted temperature, audible alarm will start again and all
keys will be passive.
Therefore turn off equipment and after trouble – shooting, turn on
equipment again.
11
Page 18
Installation and starting
The method of using
Recommended that temperature sensor places on infant’s belly at liver region.
When equipment operates automatically, if for any reason temperature sensor stopped, equipment
would operate manually so stopped sensor and manual display would continue flashing to
notify the nurse. And green display because of lacking or damaging sensor would indicate 45
numbers.
2. Manual mood
If press automatic/ manual key, so that manual indicator turned on. Equipment will be
controlled manually and green display just would indicate infant temperature and
temperature will be controlled according to adjusted degree by user.
So in order to attend the user, manual indicator will flash.
12
Page 19
Installation and starting
The method of using
Attention
To use timer of equipment, you should press set key and timer will be zero. With
pressing start/stop, timer will begin to work. With pressing start/stop key again, time
will stop.
• Press light key, to use lamp of equipment.
• Press suction key, to activate suction of equipment, and indicator of suction activeness
will turn on.
• To change the angle of table, press keys.
• To adjust height of table, press keys.
• Height of desk is adjustable up to 18 centimeters and the angle of desk surface to horizontal
is adjustable up to 12 degrees. Lighting system is used to recognize blood vessel of infant or
surery operation. Beside suction and Oxygen are used to resuscitate the infant.
• Nurse can adjust infant temperature of 28 to 32 degrees by adjusting temperature and
installing sensor to infant skin.
• Nurse can adjust the flow of oxygen by flow meter. In order to comfort nurse and doctor,
the walls of desk are folding. Nurse can also use of central oxygen.
• If sensed temperature by sensor is more than 39/3°c, equipment will give audible and visible
alarm and over heat indicator will turn on and heater of equipment will stop.
• If thermal sensor stops or its circuit has problem:
While equipment indicates alarm (sensor interruption) on display, will give audible alarm
for 5 times and will continue to operate manually.
• If back up battery is low, intended display will flash that battery should be changed.
• If main electricity interrupted for any reason while equipment was operating, electricity
interruption alarm would activate.
13
Page 20
Installation and starting
The method of using
Attention
1. Never place temperature sensor outside of table region.
2. Recommended that temperature sensor places on infants belly (liver region). If sensor
doesn’t place on infant body, sensor will sense ambient temperature and won’t have exact
control.
14
Page 21
Maintenance
Cleaning and disinfecting of equipment
Attention
• Don’t put something on heater enclosure.
• Prevent of liquid splash on equipment heater.
• Prevent of exceeding pressure on in front of desk.
• For cleaning use of a cloth saturated with disinfectant.
Effective life
Effective life of equipment is 10 years that instruments and accessories device will be
calibrated by lawful representatives of Tosan.
15
Page 22
Trouble – shooting
Attention
Attention if there is any disorder at equipment operation, please inspect
following items, before contact with Tosan company repair service group.
Symptom of disorder
Equipment doesn’t turn on
Equipment gives A.S.D alarm
16
Possible cause(s) and recommended
action(s)
On/off key doesn’t turn on
Plug doesn’t connect to plug
The electricity of equipment
Settlement is interrupted
Equipment couldn’t adjust the
temperature for any reason
Page 23
Technical specifications
Power supply
220 to 240 volts – AC – 50 hertz, 450 watt
3A
Classification
Class I- BF- Normal equipment
Alternative operation
Temperature range
28 to 37 centigrade degrees
Height of table
98 Maximum
80 Minimum
Tray tilting
12 degrees
Power of suction
0.6 Bar
Lamps
4, 20 watt halogen lamps
Weight of equipment
120 kg
Life time of equipment
10 years
• Life time of equipment is determined by Tosan Company according to power
of accessories instruments.
17
Page 24
Technical specifications
Label sign
Label description
Indicates maximum weight tolerance of
monitor stand
MAX 12Kg
!
A warning of hot surface
:¾]¼U
½I«Twj ·¼ÃwHoLòI¨ » ÁnHk¿«º ,K~º S¿] DŸ‰²
.j¼{ ̼]n Iµ¹ÀHn ¾aoTÎj ¾M
!
A warning that before installation starting
and using, study instruction manual
Technical specifications label, serial
number, equipment power producing
date and …
MAX 3 kg
18
A warning of maximum weight tolerance
of I.V.Stand