Training Course Overview
32 Pages

Preview
Page 1
Force Triad Technical Overview
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
Integrity, Honesty, Safety and Quality
We will conduct business with integrity and honesty in compliance with all laws and Company policy. We are committed to produce products that are safe and of the highest quality for our customers.
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
Introduction
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
3
Product Overview •
Combined Electro-Surgery and Vessel Sealing • •
•
Multiple language selection Single model for all line voltages
Electrosurgery • • •
2 independent ES outputs 1 separate monopolar foot-switching port 1 bipolar port •
•
•
Auto-bipolar Current meter
Vessel Sealing • •
2 independent VS outputs Hand or foot switching option
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
4
Technology Innovations •
Third monopolar mode ( •
•
Constant power across a wide range of impedance
Improved LigaSure™ characteristics •
•
Provides hemostasis with dissection
Closed Loop Coag •
•
)
Faster seal cycles
Faster tissue response in all modes • •
Tissue sensing technology 3333 decisions per second vs. 200 decisions per second
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
5
Technology Innovations • • •
Automatic device recognition Automatic generator setup Field software upgrades •
Internet download via PC • Software version 3.6 • •
Touch-screen technology New foot-switch capability • •
Three-mode capable Footswitch Adapter p/n : 1017577
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
6
System Characteristics •
•
•
Dimensions and Weight • •
18 in (45.7 cm) x 20 in (50.8 cm) x 10 in (25.4 cm) 30 lbs (13.6 Kg)
•
5 to 135 Ω or up to 40% increase in the initial measured contact resistance (whichever is less)
REM™
Internal Memory • •
•
Lithium Battery 120mAH Storage capacity 256Kb
Equipment Class • •
Class 1 Equipment (IEC 60601-1) Type CF Equipment/Defibrillator Proof
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
7
•
The Force TriVerse™ pen provides •
New mode for enhanced hemostasis with dissection
•
Control of all three tissue effect modes from the sterile field.
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
8
Front and Rear Panel
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
9
Front Panel
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
10
Back Panel
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
11
System Setup Functions
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
12
System Setup Functions •
Service Screens • •
Password protected (423213) Multiple levels
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
13
Enter password 1
2
3
* * * *
Main Menu Activation Log Service (423213) Restore Setup Demo Mode
Diagnostics
Disable errors?
Perform selftest Disable errors Debug mode Show messages
Debug mode
4 Bksp
Service Diagnostics Maintenance Statistics Configuration
Maintenance Calibration Show cal data Clear history
Statistics Display error list Display error histogram Display error log Display activation histogram log
Yes
Cancel
Mode selection Bip Standard Bip Macro Bip Low Ligasure Mono Fulg Mono Spray Mono Blend Mono DWH Port Selection: Mono1 Port Mono2 Port Acc Port Biploar Port Lig1 Port Lig2 Port Open/Closed Loop Level
Calibration
V cal I cal REM Autobipolar Touch Screen Scanner
Clear History Configuration
DSP1 version DSP2 version Host version Comm. Protocol version FPGA version Board IDs
Clear Error List
Clear Error histogram
Clear Event Log Clear Activation Histogram Clear All
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
14
System Setup Functions •DEMO Mode •REM and Dual instrument error deactivated but RF power is still delivered •Exit by powering off or exiting DEMO mode
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
15
Internal Components
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
16
Block Diagram
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
17
HVDC POWER SUPPLY •Regulates Output DC based on 0 to 5V analogue control voltage(ECON) •AC input range is 85 VAC to 264 VAC with line frequencies from 47 to 63 Hz •AC Section •Rectified +BUS and –BUS Voltage •If Line voltage 150VAC or less Rectified AC is doubled. •Safety Feature (400VAC) •Soft start circuit reduce inrush AC •DC Section •Phase shifted full bridge UCC3895 •Fault detection; Over current, voltage, power protection, short circuit •Active discharge circuit
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
18
RF Board PCBA •Convert DC voltage from HVDC PCBA into 472 kHz RF signal
• REM circuit monitors the resistance between the two return areas on REM electrode using the 80kHz signal generated by the controller board •Auto Bipolar uses 80kHz signal. Used to monitor the bipolar output impedance •Leakage Current Monitor measures the active and return currents, producing a difference voltage – if leakage limit is exceeded the RF reduces to prevent excess leakage current •Dual Sensor Circuit provides redundant output voltage and current monitoring. Fail-safe mitigation in event of circuit failure. Error msg to user.
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
19
Steering Relay PCBA •Multiple
functions on Steering Relay
•Routing
the 472kHz from RF board to one of the six outputs: •Lig1, Lig2, Bipolar, Mono1, Mono2, UFP •Only 1 output can be active at a time •Routing
Mono return and REM signal
•Routing
Auto Bipolar sensing signal
•Hand-switching detection •Individually isolated
and
opto-coupled •Five isolated, transformer based, +8V or +5V power supplies •Instrument detection •Universal Footswitching •Bipolar Receptacle
Port
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
20
Display PCBA •Hotlink transceiver U1 between FPGA U28 and Controller board •LCD Display driver (within FPGA); receives video data and outputs to displays (acts as FIFO buffer) •Touchscreen driver (within FPGA) polls screens; user input read by ADC U4 •DAC U24; controls brightness of LCDs •Barcode driver (within FPGA) communicates with scanners through UART U14; data read and sent to controller board via FPGA •Power Supply U16; +5V input DC to DC converter 3.3, 1.8V output COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
21
Footswitch/Audio PCBA •Supplied 5V and 12V delivered by RF PCBA connector •Produces 2.5V, 3.3V and isolated ±12V and 5V •SMbus/I²C data link communication •Audio Controlled by controller board through FPGA, DAC and audio amplifier •Footswitch data; Footswitch detection and activation through FPGA •Expansion port •Analogue outputs through FPGA •RS232 directly from controller board •ECG/blanking port directly controlled by the controller board COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
22
Controller PCBA •Digital Signal Processors •DSP1 Controls HVPS control voltage (ECON) Controls RF keying signals (T_ON, T_ON_180) Monitors primary voltage and current sensors •DSP2 Monitors secondary voltage and current sensors Dosage Error processor • Interface Control Logic (ICL) - Footswitch/Audio PCBA - Display PCBA - LVPS power fail circuit - REM and HVPS - RF relays •Host Processor - SDRAM - SRAM / RTC (snap hat battery and crystal) - Flash - Ethernet, USB, RS-232 COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
23
Other Components - LVPS •+5.1V 8.0A •+12V 2.5A •-12V 2.5A
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
24
System Fans
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
25
Output Receptacles
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
26
Common Error Messages
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
27
•
Dual Insertion Error Popup
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
28
LigaSure Alert •Reactivate
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
29
Other Common Errors? Error 300 – ERR_NR_BAD_SCANNER Bad or malfunctioning scanner Recommended Action: Recalibrate scanner. Replace scanner and scanner ribbon cable.
Gray screens Recommended Action: Check the Ethernet cable, Display cable (Steering Relay and Display).
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
30
Other Common Errors? Error 288 – ERR_NR_SENSOR_CLIP System has developed a DSP sensor clipping error Recommended Action: Replace HVPS PCBA , Recalibrate unit.
Error 289 – ERR_NR_SEN DSP sensor comparison error Recommended Action: Replace Controller PCBA, RF Board and HVDC Board
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.
31
COVIDIEN, COVIDIEN with logo, Covidien logo, positive results for life and ™ marked brands are U.S. and/or internationally registered trademarks of Covidien AG.