AHA Diagnostic ECG Electrode Placement Guide Rev A

Guide

2 Pages

AHA Diagnostic ECG Electrode Placement The quality of an ECG is dependent on the preparation and the resistance between the skin and the electrode. To ensure a good quality ECG and minimize the skin/electrode resistance, remember the following points: 1.) Ensure that the patient is warm and relaxed 2.) Shave electrode area before cleaning 3.) Thoroughly clean the area with alcohol 4.) Let dry prior to applying electrode  Resting ECG RA and LA electrodes should be placed on the wrists. RL and LL electrodes should be placed a few inches above the ankle.  Lead  Electrode Placement  V1  Fourth intercostal space at the right border of the sternum  V2  Fourth intercostal space at the left border of the sternum  V3  Midway between locations V2 and V4  V4  At the mid-clavicular line in the fifth intercostal space  V5  At the anterior axillary line on the same horizontal level as V4  V6  At the mid-axillary line on the same horizontal level as V4 and V5  For service or technical support, please call (800) 535-6663. For supplies, please contact your authorized Welch Allyn distributor.  Advancing Frontline Care™ 71300-0000 Rev. A