Brief Guide
2 Pages
Preview
Page 1
WILAmed AIRcon Gen2 – Brief guide Start therapy intermission or mute alarm.
Display of the actual temperatures of the patient outlet and the chamber. Pressing both buttons for approx. 3 secs resets to factory settings.
Setting the humidification levels in three stages.
39.0°C 38.8 37.0 °C °C
Expiratory heating power (only with heated double-hose system)
IV EXP.
WILAmed
Enter configuration mode or exit configuration and change mode.
Setpoint temperature of the patient outlet and the chamber.
Displayed if the warm-up phase is still active.
AIRcon
Enter change mode or confirm changes.
Display for the currently set mode (IV, NIV, FREE)
Configuration help Attention! This quick start guide does NOT replace the operating instructions. Specifications made by the doctor or hospital take priority over these recommendations. All warnings and guidelines in the operating instructions must be strictly observed. “IV” mode: Factory setting for invasively ventilated patients (setpoint temperature: near the patient = 39°C, chamber = 37°C max.) “NIV” mode: Factory setting for mask-ventilated patients (setpoint temperature: near the patient = 34°C, chamber = 31°C max.) “Free” mode: Enter individual operating parameters Condensate veil in the patient tube = sufficient humidification No condensate = increase humidification power Too much condensate = Reduce humidification Humidification power can be adjusted by water droplets from 1 drop (min.) to 3 drops (max.) or adjusted using the chamber temperature in the “FREE” mode.
To change the settings Enter the change mode by pressing the button for 3 secs. Select the element to be changed with the and buttons. By pressing the button, the selected element is highlighted in yellow and can be increased or decreased using the and buttons. Confirm the entered value with the button. Exit change mode by pressing the button for 3 s.
Display in the change mode
39.0°C
Setpoint temperature near the patient
36.0°C
Setpoint temperature in the chamber
If factory setting “IV” or “NIV” is changed, the system auto matically switches to “FREE” mode.
Expiratory heating power for double-hose systems
EXP.
Settings in configuration mode Enter configuration mode by pressing button
39.0°C
for 3 secs. Set setpoint temperatures for “FREE” mode.
36.0°C FREE IV Water Level Detection
NIV
ON
Set the operating mode
OFF Change the humidification power
Version 1.0
Stand: 01/05/2017
REF 8903103
Switch the water level detection on/off