zimmer_biomet_dental_instrument_kit_system_referenc.pdf
Page 1
Zimmer Biomet
Instrument Kit System
Reference Guide
Tapered Screw-Vent® (TSV™) Implant System
Trabecular Metal™ Dental Implants
3.1 mmD Eztetic™ Dental Implants
Staging Block
NP Surgical Module for
3.1 mmD Eztetic Implants
Page 2
Instrument Kit System For TSV,
Trabecular Metal And 3.1 mmD Eztetic
3
2
1
4
5
6
7
8
9
10
11
14
15
16
17
18
19
20
21
3
12
3
13
3
22
24
23
27
25
29
28
34
26
30
31
32
33
35
36
37
38
39
40
1
2
4
5
6
7
8
9
10
11
3
3.0 mmD
Round Bur
1203
2.3 mmD
Drill,
22 mmL
SV2.3DN
2.8 mmD
Drill,
22 mmL
SV2.8DN
3.4
2.8 mmD
Step Drill,
22 mmL
TSV3DN
3.4 mmD
Drill,
22 mmL
SV3.4DN
3.8/
3.4 mmD
Step Drill,
22 mmL
TSV3.8DN
3.8 mmD
Drill,
22 mmL
SV3.8DN
4.4/
3.8 mmD
Step Drill,
22 mmL
TSV4DN
5.1 mmD
Drill,
22 mmL
SV5.1DN
5.7/5.1
mmD
Step Drill,
22 mmL
TSV6DN
12
13
14
15
16
17
18
19
20
21
2.8 mmD
Drill,
22 mmL
SV2.8DN
3.4/
2.8 mmD
Step Drill,
16 mmL
TSV3DSN
3.4 mmD
Drill,
16 mmL
SV3.4DSN
4.4/
3.8 mmD
Step Drill,
16 mmL
TSV4DSN
5.1 mmD
Drill,
16 mmL
SV5.1DSN
5.7/
5.1 mmD
Step Drill,
16 mmL
TSV6DSN
Tapered
Pilot Drill,
2.1/
1.6 mmD,
8 mmL 11.5 mmL
0201DSN
(or 0201)
22
Drill
Extender
DE
2.3 mmD
Drill,
16 mmL
SV2.3DSN
Paralleling
Tool (Qty: 4)
PPAR
23
3.7 mmD
Bone Tap
TT3.7
24
4.1 mmD
Bone Tap
TT4.1
25
4.7 mmD
Bone Tap
TT4.7
3.8/
3.4 mmD
Step Drill,
16 mmL
TSV3.8DSN
26
6.0 mmD
Bone Tap
TT6.0
3.8 mmD
Drill,
16 mmL
SV3.8DSN
27
2.5 mm GemLock
Hex Drill
RHD2.5
28
2.5 mm GemLock
Hex Tool, Short
RH2.5
29
2.5 mm GemLock
Hex Tool, Long
RHL2.5
Page 3
The NP Surgical Module snaps into the Tapered Screw-Vent Surgical Kit
41
41
42
2.4 mmD Dríva™
Drill, 22 mmL
SV2.4DN
30
31
3.0 mm
Hex Insertion
Drill
HX3.0D
3.0 mm
Hex Insertion
Tool, 17 mmL
HX3.0-S
39
3.0 mm
Hex Insertion
Tool, 25 mmL
HXL3.0-S
Removal
Tool
TLRT2
48
NP Surgical Module
46
2.1 mm
GemLock Hex Drill,
Latch Lock, Long
CHDL2.1
47
2.1 mm
GemLock Hex Tool,
Ratchet, Short
CHR2.1
26
4.7 mmD
Bone Tap
TT4.7
33
47
46
45
25
4.1 mmD
Bone Tap
TT4.1
32
45
2.1 mm
GemLock® Hex
Drill, Latch Lock,
Short
CHD2.1
24
3.7 mmD
Bone Tap
TT3.7
44
44
3.0 mmD
Cortical Bone
Tap Tool
ZOPTT30
23
Drill
Extender
DE
43
43
2.8/2.4 mmD
Dríva Surgical
Step Drill, 22 mmL
ZOP28DN
22
42
27
6.0 mmD
Bone Tap
TT6.0
35
1.25 mm Hex
Driver with
GemLock
Retention,
22 mmL
HXGR1.25
1.25 mm Hex
Driver with
GemLock
Retention,
30 mmL
HXLGR1.25
2.1 mm
GemLock Hex Tool,
Ratchet, Long
CHRL2.1
28
2.5 mm GemLock
Hex Drill
RHD2.5
34
48
36
1.25 mm
Hex Tool,
17 mmL
HX1.25
Removal Tool for
Eztetic Abutments
CLRT2
29
2.5 mm GemLock
Hex Tool, Short
RH2.5
37
1.25 mm
Long
Hex Tool,
22 mmL
HXL1.25
2.5 mm GemLock
Hex Tool, Long
RHL2.5
38
1.25 mm
Hex Drill
23 mmL
HX1.25D
40
GemLock Retaining
Square Ratchet
RSR
Screwdriver Handle with Square
Connection
SSHS
Cleaning of Instruments
1. Disassemble two-piece components.
2. Rinse instruments in cool to lukewarm water for 2 and a 1/2 minutes.
3. For drills, use the cleaning wire to remove any debris from the irrigation channel. Using
a 25 gauge needle, flush the drill lumen with water to remove any remaining debris.
4. Sonicate the instruments for 10 minutes in an ultrasonic cleaner with a pH-neutral
enzymatic detergent diluted with tap water per the manufacturer’s instructions.
5. Rinse the instruments with tap water for 3 minutes.
6. Inspect the instruments for signs of wear, damage, or unrecognizable color
identification and replace the instruments accordingly.
Cleaning of Trays and Staging Block
1. Remove all parts and insert from the surgical tray.
2. Rinse the tray and tray insert with cool to lukewarm tap water.
3. Use a damp cloth to wipe and remove any excess soil from each part.
4. Wipe each part with a cloth dipped in an enzymatic detergent diluted with tap water
per the manufacturer’s instructions.
5. Rinse thoroughly with tap water for 3 minutes.
Sterilization
1. Dry parts thoroughly before sterilization.
2. Double-wrap kit with autoclave wrap and secure wrap with autoclave tape.
3. Recommended Sterilization Parameters:
Cycle Type
Temperature
Exposure Time
Dry Time
(only for kits)
Gravity
(steam)
132°C
270°F
15 mins
20 mins
Pre-vacuum
(steam)
132°C
270°F
4 mins
20 mins
Pre-vacuum
(steam)
134°C
273°F
3 mins
20 mins
Pre-vacuum
(steam)
134°C
273°F
18 mins
20 mins
3 For maximum cutting efficiency, replace drills frequently.
Page 4
Tapered Screw-Vent and Trabecular Metal
Implant Surgical Protocol
Intuitive Flow And Color-Coding – Its Beauty Is In Its Simplicity
A simple color-coding system identifies drills for each implant diameter, allowing you to easily follow any surgical sequence step-by-step. As an
example, surgical drills required for placement of the 3.7 mmD Tapered Screw-Vent Implant are represented by horizontal green bars on the
kit surface and are logically organized in the order you would use them from left to right. The color-coding also allows you to easily identify your
drill options for soft or dense bone protocols – a dotted color bar denotes a final soft bone drill, while the following solid color bar denotes a final
dense bone drill.
Tapered Screw-Vent And Trabecular Metal Implants Color-Coding
3.7 mm Implant Diameter
4.7 mm Implant Diameter
4.1 mm Implant Diameter
6.0 mm Implant Diameter
Step 1
Step 2
The 3.7 mmD Tapered ScrewVent and Trabecular Metal
Implant are color-coded in green.
Start with the first green bar on the
kit, which indicates the first drill to be
used in the drilling sequence for this
implant size.
Follow the green color bars
from left to right. In a soft bone
protocol, the dotted green bar
represents the final drill. For dense
bone, skip the dotted green bar and
move on directly to the next solid green
bar. The last solid bar in the sequence
represents the final drill for dense bone.
Step 3
When drilling in dense bone, you
can optionally use the 3.7 mmD
cortical bone tap located in a
green grommet directly below the
last solid green bar in the sequence.
3.7 mmD Tapered Screw-Vent and Trabecular Metal Implant (3.5 mmD Platform)
1
3.7 mmD
SV2.3D
2.3 mmD
Drill
2
2
FOR SOFT BONE
SV2.8D
2.8 mmD
Drill
FOR DENSE BONE
TSV3D
3.4/2.8 mmD
Drill
3
OPTIONAL FOR
DENSE BONE
TT3.7
3.7 mmD
Cortical Bone Tap
Page 5
4.1 mmD Tapered Screw-Vent and Trabecular Metal Implant (3.5 mmD Platform)
1
2
SV2.3D
2.3 mmD
Drill
4.1 mmD
3
SV2.8D
2.8 mmD
Drill
3
FOR SOFT BONE
SV3.4D
3.4 mmD
Drill
4
FOR DENSE BONE
TSV3.8D
3.8/3.4 mmD
Drill
OPTIONAL FOR
DENSE BONE
TT4.1
4.1 mmD
Cortical Bone Tap
4.7 mmD Tapered Screw-Vent and Trabecular Metal Implant (4.5 mmD Platform)
1
2
SV2.3D
2.3 mmD
Drill
4.7 mmD
3
TSV3D
3.4/2.8 mmD
Drill
3
FOR SOFT BONE
SV3.8D
3.8 mmD
Drill
4
FOR DENSE BONE
TSV4D
4.4/3.8 mmD
Drill
OPTIONAL FOR
DENSE BONE
TT4.7
4.7 mmD
Cortical Bone Tap
6.0 mmD Tapered Screw-Vent and Trabecular Metal Implant (5.7 mmD Platform)
1
6.0 mmD
2
SV2.3D
2.3 mmD
Drill
3
TSV3D
3.4/2.8 mmD
Drill
4
TSV4D
4.4/3.8 mmD
Drill
FOR SOFT BONE
SV5.1D
5.1 mmD
Drill
For detailed cleaning and sterilization instructions, refer to the Instructions for Use provided with the Zimmer Instrument Kit System.
4
FOR DENSE BONE
TSV6D
5.7/5.1 mmD
Drill
5
OPTIONAL FOR
DENSE BONE
TT6.0
6.0 mmD
Cortical Bone Tap
Page 6
Eztetic Implant Surgical Protocol
Intuitive Flow And Color-Coding – Its Beauty Is In Its Simplicity
A simple color-coding system identifies drills for each implant diameter, allowing you to easily follow any surgical sequence step-by-step. As an
example, surgical drills required for placement of the 3.1 mmD Eztetic Implant are represented by horizontal blue bars on the kit and NP Module
and are logically organized in the order you would use them from left to right and up to the NP Module. The color-coding also allows you to easily
identify your drill options for soft or dense bone protocols – a dotted color bar denotes a final soft-bone drill, while the following solid color bar
denotes a final dense-bone drill.
3.1 mmD Eztetic Implants Color-Coding
With the instrument kit system comes a simple way of working. Its unique, color-coded surgical protocol labeling system helps to guide you
effortlessly through each drilling sequence.
3.1 mm Implant Diameter
Step 2
Step 1
The 3.1 mmD Eztetic
Implant is color-coded
in blue. Start with the first
blue bar in the main kit.
Follow the blue color
bars from left to right
for the 2.3 mmD Drill,
and up to the NP Surgical
Module. Utilize the 2.4 mmD
Drill as the final drill in soft bone. The
last blue solid bar in the sequence is
located in the NP Surgical Module and
represents the final drill in dense bone
(2.8/2.4 mmD).
Step 3
When drilling in dense
bone, you can optionally
use the 3.0 mmD cortical
bone tap located in the
blue grommet that follows the
2.8/2.4 mm drill.
Page 7
3.1 mmD Eztetic Implants
1
0201DSN
2.1/1.6 mmD,
8 mmL - 11.5 mmL
Drill
2
3
SV2.3DN
2.3 mmD
Drill
3
SV2.4DN
2.4 mmD
Drill
ZOP28DN
2.8/2.4 mmD
Drill
4
OPTIONAL FOR
DENSE BONE
ZOPTT30
3.0 mmD
Cortical Bone Tap
Page 8
Contact us at 1-800-342-5454 or visit
zimmerbiometdental.com
Zimmer Biomet Dental
Global Headquarters
4555 Riverside Drive
Palm Beach Gardens, FL 33410
Phone: +1-561-776-6700
Fax: +1-561-776-1272
Unless otherwise indicated, as referenced herein, all trademarks are the property of Zimmer Biomet; and all products are manufactured by one or more of the dental
subsidiaries of Zimmer Biomet Holdings, Inc., and distributed and marketed by Zimmer Biomet Dental (and, in the case of distribution and marketing, its authorized
marketing partners). For additional product information, please refer to the individual product labeling or instructions for use. Product clearance and availability may
be limited to certain countries/regions. This material is intended for clinicians only and does not comprise medical advice or recommendations. This material may
not be copied or reprinted without the express written consent of Zimmer Biomet Dental. ZB0105 REV A 08/17 ©2017 Zimmer Biomet. All rights reserved.