ZIMMER™ Tissue Bank Dermatome FOR CADAVER USE ONLY Instruction and Service Manual REF: 8831-001  STORAGE The Zimmer Tissue Bank Dermatome has an operating temperature range of 50° to 88°F (10° to 31°C)... he following are environmental conditions for transportation and storage: A... mbient temperature range... 1°F to 149°F (–17°C to 65°C) B... elative humidity range... 10% to 80% C... tmospheric pressure range... 00 hPa to 1060 hPa  GENERAL DESCRIPTION The Zimmer Tissue Bank Dermatome is an electrically powered instrument designed specifically for tissue retrieval procedures... t is intended for use on cadavers only, and not for living patients requiring skin grafts... RECEIVING INSPECTION  This instrument is designed for retrieving tissue in thicknesses ranging from 0 to ... 0 in... 0 to 1...  mm) in ... 4 in... 2 mm) increments... t also allows a choice of either 3 or 4 in... 7... and 10... cm) skin widths using the two width plates included... wo stainless steel machine screws secure the plates to the underside of the instrument... he plates are easily fastened and removed with the screwdriver provided... After receiving the Zimmer Tissue Bank Dermatome, examine the unit for external signs of damage... etain all packing material until the contents have been verified and an initial operational check has been performed... f the device has been damaged, forward an immediate request to the delivering carrier to perform an inspection and prepare a "Concealed Damage" report... he terms of all Zimmer Orthopaedic Surgical Products sales agreements concerning this unit are F... B... Dover, Ohio, U... A... and the responsibility of Zimmer Orthopaedic Surgical Products ends with delivery to the first carrier... hereafter, the shipment becomes the property of the customer... ny claims for loss, damage, or non-delivery must be made against the delivering carrier within 10 days of shipment... This dermatome is powered by an ironless rotor, low inertia motor, which provides nearly vibration-free power... SPECIFICATIONS 1... OWER SUPPLY I... hysical: Weight: Dimensions:  II... lectrical: Power Requirements: Power Output:  Your Zimmer distributor or local Zimmer representative will assist you in determining the cost of replacement or repair so that a proper claim may be filed with the delivering carrier... 2... lb... 1...  kg) 9...  in... 23... cm) Length 5...  in... 13... cm) Height 6...  in... 17... cm) Depth  Note: Do not attempt to return damaged goods without first securing proper authorization from the delivering carrier... 100-240 V~ (Auto Switching) 150 VA, 50/60 Hz Single Phase 14... V , 4... A Maximum  GENERAL PRECAUTIONS This equipment is for use only by qualified tissue bank personnel... o avoid serious injury to the tissue bank staff when using the Zimmer Tissue Bank Dermatome, the operator must be thoroughly familiar with its function, application, and instructions before use... he handpiece and blade must be inspected for proper operation before use... f you have any questions, please contact your Zimmer representative... 2... ANDPIECE I... hysical: Weight: Dimensions: Vibration and Shock: Cutting Speed: II... lectrical: Power Requirements:  2... lb... 1...  kg) 9...  in... 24... cm) Length 5... in... 13... cm) Width Standard Commercial Practice 5000–6000 Cycles/Minute  Ground reliability can only be achieved when the hospital grade power cord is connected to a proper receptacle... EVER connect the Zimmer Tissue Bank Dermatome Handpiece to any source other than the Zimmer Tissue Bank Dermatome Power Supply... t has been factory calibrated to provide optimum cutting performance and maximum safety... t is also designed to meet or exceed specific medical electrical safety standards... 14... V , Fully Regulated and Isolated, 4... A Maximum  UL 60601-1 Classification: Type of protection against electrical shock  For continued protection against fire hazard, replace only with the same type and rating of fuse... Refer to FUSES section...  Class I  Degree of protection against electrical shock: Type BF applied part Classification according to the degree of protection against ingress of water:  IPXO  Mode of operation:  50% Duty Cycle, 10 minute maximum ON 88°F (31°C)  Maximum operating ambient:  Possible explosion hazard exists if this instrument is used in the presence of flammable anesthetics or gases... EVER sterilize the power supply... isconnect the dermatome handpiece from the power supply before sterilization... 1-   WARNING AND CAUTION LABELS  OPERATING PRECAUTIONS  See Fig... The on/off lever must be in the SAFE position before changing blades when the instrument is not in use, or when connecting power to the instrument... ccidental activation of the instrument during these procedures may injure the tissue bank personnel... SYMBOL DEFINITIONS See Fig... Zimmer TISSUE BANK Dermatome Components (REF 8831)  To avoid any injury, use extreme caution when handling the blade... 1... 0... 1... Power Supply (REF 8831-06) Handpiece with cord (REF 8831-01) Screwdriver (REF 8803) Autoclave case (REF 8831-03) Width plate, 3 in... 7... cm) (REF 8831-04) Width plate, 4 in... 10... cm) (REF 8831-05) Fuse drawer, 5 mm x 20 mm Fuses, 1...  A (Qty 2) (Installed) Fuses, 1...  A (Qty 2) (5 mm x 20 mm included) Power Cord Instruction and Service Manual  Use extreme caution when handling the dermatome with the blade installed... efore using, always inspect the cord for signs of wear or damage...  worn or damaged cord must be returned for service immediately for continued safety... ZIMMER TISSUE BANK DERMATOME ILLUSTRATIONS  1... lace the power supply in the OFF (O) position... lace the dermatome in the SAFE position by sliding the safety lock on the on/off lever toward the blade end of the instrument to the SAFE position... nly the word SAFE should be visible... Use only Zimmer Tissue Bank Dermatome Blades (REF 8831-000-10)... immer Tissue Bank Dermatome Blades have been specifically designed for use with the Zimmer Tissue Bank Dermatome... ther blades will not fit properly in the dermatome and may cause deeper grafts than selected... OPERATING INSTRUCTIONS  See Fig... –7... BEFORE OPERATION  2... hoose the proper width plate to satisfy cutting requirements... wo width plates are supplied: 3 and 4 in... 7...  and 10... cm)... Connect the Zimmer Tissue Bank Dermatome to its power supply by inserting the connector plug into the connector receptacle... See Fig... Disconnect the power supply from the power source before inserting... emove any kinks or twists from the cord... lign the connector plug to the receptacle by rotating the plug while gently pushing inward... hen aligned, push in until the plug clicks into the receptacle... 3... nstall the blade... Refer to Blade Placement diagram, (Fig... Use caution when inserting the blade to avoid nicking it, which may result in an uneven cut... se a new sterile blade for each procedure... lace the on/off lever in the SAFE position... To remove the connector plug, pull outward on the outer collar portion of the plug to separate the plug from the receptacle... b... sing the Zimmer screwdriver (REF 8803), loosen the width plate screws approximately two turns... O NOT remove screws from handpiece... To replace input fuses, disconnect the power supply from the power source... emove the fuse drawer by lifting up on the tab using a tool and pulling the drawer outward... See Fig... Replace the fuse drawer with the correct fuses installed... he correct fuse type is listed in the SERVICE INFORMATION section... he unit is shipped with a 0...  in... 6...  mm) x 1...  in... 31...  mm) fuse drawer installed...  separate fuse drawer is supplied for 0... 7 in... 5 mm) x 0... 7 in... 20 mm) fuses... c... emove width plate and used blade... lace a new blade onto the carriage... lign the two holes in the blade with the two pins on the carriage... lace width plate over blade and tighten screws... Before connecting the power supply to a power source, be sure that the switch on the front of the power supply is in the OFF (O) position, and that the safety lock is in the SAFE position... onnect to a power source via the "IEC" connector in the rear of the power supply  4... ontrol lever adjustment determines graft thickness... See Fig... 0... Factory markings indicate 0... 4 in... 0... 2 mm) divisions... et control knob pointer to desired thickness...  semi-annual calibration check by an authorized Zimmer Orthopaedic Surgical Products service center is recommended to verify continued accuracy... Turn power on to the power supply by moving the rocker switch on the front of the power supply to the ON (I) position... he rocker switch should illuminate... ove the safety lock to the ON position and briefly depress the on/off lever to ensure that power is being applied to the handpiece, and that the carriage is oscillating... eturn the safety lock to the SAFE position... 5... o activate the dermatome, place the power supply in the ON (I) position... ift the on/off lever and slide the safety lock back to the ON position... epress the on/off lever to start the motor... O NOT run the dermatome without cutting for an extended time, as this needlessly wears the blade... 2-
File Type: PDF
File Size: 617KB
File Name: zimmer_tissue_bank_dermatone_instruction_and_servic.pdf
Available for download

Available for download in the BioClinical Services library

If you already belong to a hospital with access, please login via your intranet.

Contact Us to Register Your InterestAlready a member? Log in hereAbout Us

zimmer_tissue_bank_dermatone_instruction_and_servic.pdf

Page 1

ZIMMER™ Tissue Bank

Dermatome FOR CADAVER USE ONLY

Instruction and Service Manual

REF: 8831-001

STORAGE

The Zimmer Tissue Bank Dermatome has an operating

temperature range of 50° to 88°F (10° to 31°C).

The following are environmental conditions for

transportation and storage:

A. Ambient temperature

range . . . . . . . . . . . . . . .1°F to 149°F (–17°C to 65°C)

B. Relative humidity range . . . . . . . . . . . . . . .10% to 80%

C. Atmospheric pressure range . . . .500 hPa to 1060 hPa

GENERAL DESCRIPTION

The Zimmer Tissue Bank Dermatome is an electrically

powered instrument designed specifically for tissue

retrieval procedures. It is intended for use on cadavers

only, and not for living patients requiring skin grafts.

RECEIVING INSPECTION

This instrument is designed for retrieving tissue in

thicknesses ranging from 0 to .060 in. (0 to 1.52 mm) in

.004 in. (.102 mm) increments. It also allows a choice of

either 3 or 4 in. (7.6 and 10.2 cm) skin widths using the

two width plates included. Two stainless steel machine

screws secure the plates to the underside of the

instrument. The plates are easily fastened and removed

with the screwdriver provided.

After receiving the Zimmer Tissue Bank Dermatome,

examine the unit for external signs of damage. Retain all

packing material until the contents have been verified and

an initial operational check has been performed. If the

device has been damaged, forward an immediate request

to the delivering carrier to perform an inspection and

prepare a "Concealed Damage" report. The terms of all

Zimmer Orthopaedic Surgical Products sales agreements

concerning this unit are F.O.B., Dover, Ohio, U.S.A., and

the responsibility of Zimmer Orthopaedic Surgical Products

ends with delivery to the first carrier. Thereafter, the

shipment becomes the property of the customer. Any

claims for loss, damage, or non-delivery must be made

against the delivering carrier within 10 days of shipment.

This dermatome is powered by an ironless rotor, low

inertia motor, which provides nearly vibration-free power.

SPECIFICATIONS

1. POWER SUPPLY

I. Physical:

Weight:

Dimensions:

II. Electrical:

Power Requirements:

Power Output:

Your Zimmer distributor or local Zimmer representative will

assist you in determining the cost of replacement or repair

so that a proper claim may be filed with the delivering

carrier.

2.6 lb. (1.18 kg)

9.12 in. (23.2 cm) Length

5.44 in. (13.8 cm) Height

6.69 in. (17.0 cm) Depth

Note: Do not attempt to return damaged goods without first

securing proper authorization from the delivering carrier.

100-240 V~ (Auto Switching)

150 VA, 50/60 Hz

Single Phase

14.5 V

, 4.3 A Maximum

GENERAL PRECAUTIONS

This equipment is for use only by qualified tissue bank

personnel. To avoid serious injury to the tissue bank staff

when using the Zimmer Tissue Bank Dermatome, the

operator must be thoroughly familiar with its function,

application, and instructions before use. The handpiece

and blade must be inspected for proper operation before

use. If you have any questions, please contact your

Zimmer representative.

2. HANDPIECE

I. Physical:

Weight:

Dimensions:

Vibration and Shock:

Cutting Speed:

II. Electrical:

Power Requirements:

2.9 lb. (1.31 kg)

9.75 in. (24.8 cm) Length

5.2 in. (13.2 cm) Width

Standard Commercial

Practice

5000–6000 Cycles/Minute

Ground reliability can only be achieved when the hospital

grade power cord is connected to a proper receptacle.

NEVER connect the Zimmer Tissue Bank Dermatome

Handpiece to any source other than the Zimmer Tissue

Bank Dermatome Power Supply. It has been factory

calibrated to provide optimum cutting performance and

maximum safety. It is also designed to meet or exceed

specific medical electrical safety standards.

14.5 V

, Fully Regulated

and Isolated, 4.3 A Maximum

UL 60601-1 Classification:

Type of protection against electrical shock

For continued protection against fire hazard, replace only

with the same type and rating of fuse. (Refer to FUSES

section.)

Class I

Degree of protection against electrical shock: Type BF applied part

Classification according to the degree of

protection against ingress of water:

IPXO

Mode of operation:

50% Duty Cycle, 10

minute maximum ON

88°F (31°C)

Maximum operating ambient:

Possible explosion hazard exists if this instrument

is used in the presence of flammable anesthetics or

gases.

NEVER sterilize the power supply. Disconnect the

dermatome handpiece from the power supply before

sterilization.

-1-

Page 2

WARNING AND CAUTION LABELS

OPERATING PRECAUTIONS

See Fig. 3.

The on/off lever must be in the SAFE position before

changing blades when the instrument is not in use, or

when connecting power to the instrument. Accidental

activation of the instrument during these procedures may

injure the tissue bank personnel.

SYMBOL DEFINITIONS

See Fig. 4.

Zimmer TISSUE BANK Dermatome

Components (REF 8831)

To avoid any injury, use extreme caution when handling the

blade.

1.

2.

3.

4.

5.

6.

7.

8.

9.

10.

11.

Power Supply (REF 8831-06)

Handpiece with cord (REF 8831-01)

Screwdriver (REF 8803)

Autoclave case (REF 8831-03)

Width plate, 3 in. (7.6 cm) (REF 8831-04)

Width plate, 4 in. (10.2 cm) (REF 8831-05)

Fuse drawer, 5 mm x 20 mm

Fuses, 1.25 A (Qty 2) (Installed)

Fuses, 1.25 A (Qty 2) (5 mm x 20 mm included)

Power Cord

Instruction and Service Manual

Use extreme caution when handling the dermatome with

the blade installed. Before using, always inspect the cord

for signs of wear or damage. A worn or damaged cord

must be returned for service immediately for continued

safety.

ZIMMER TISSUE BANK DERMATOME

ILLUSTRATIONS

1. Place the power supply in the OFF (O) position. Place

the dermatome in the SAFE position by sliding the safety

lock on the on/off lever toward the blade end of the

instrument to the SAFE position. Only the word SAFE

should be visible.

Use only Zimmer Tissue Bank Dermatome Blades

(REF 8831-000-10). Zimmer Tissue Bank Dermatome

Blades have been specifically designed for use with the

Zimmer Tissue Bank Dermatome. Other blades will not

fit properly in the dermatome and may cause

deeper grafts than selected.

OPERATING INSTRUCTIONS

See Fig. 5–7.

BEFORE OPERATION

2. Choose the proper width plate to satisfy cutting

requirements. Two width plates are supplied: 3 and 4 in.

(7.6, and 10.2 cm).

Connect the Zimmer Tissue Bank Dermatome to its power

supply by inserting the connector plug into the connector

receptacle. (See Fig. 8.) Disconnect the power supply from

the power source before inserting. Remove any kinks or

twists from the cord. Align the connector plug to the

receptacle by rotating the plug while gently pushing inward.

When aligned, push in until the plug clicks into the

receptacle.

3. Install the blade. (Refer to Blade Placement diagram,

(Fig. 7.) Use caution when inserting the blade to avoid

nicking it, which may result in an uneven cut. Use a new

sterile blade for each procedure.

a. Place the on/off lever in the SAFE position.

To remove the connector plug, pull outward on the outer

collar portion of the plug to separate the plug from the

receptacle.

b. Using the Zimmer screwdriver (REF 8803), loosen the

width plate screws approximately two turns. DO NOT

remove screws from handpiece.

To replace input fuses, disconnect the power supply from

the power source. Remove the fuse drawer by lifting up on

the tab using a tool and pulling the drawer outward. (See

Fig. 9.) Replace the fuse drawer with the correct fuses

installed. The correct fuse type is listed in the SERVICE

INFORMATION section. The unit is shipped with a 0.25 in.

(6.35 mm) x 1.25 in. (31.75 mm) fuse drawer installed. A

separate fuse drawer is supplied for 0.197 in. (5 mm) x

0.787 in. (20 mm) fuses.

c. Remove width plate and used blade.

d. Place a new blade onto the carriage.

e. Align the two holes in the blade with the two pins on

the carriage.

f. Place width plate over blade and tighten screws.

Before connecting the power supply to a power source, be

sure that the switch on the front of the power supply is in

the OFF (O) position, and that the safety lock is in the

SAFE position. Connect to a power source via the "IEC"

connector in the rear of the power supply

4. Control lever adjustment determines graft thickness. (See

Fig. 10.) Factory markings indicate 0.004 in. (0.102 mm)

divisions. Set control knob pointer to desired thickness.

A semi-annual calibration check by an authorized

Zimmer Orthopaedic Surgical Products service center

is recommended to verify continued accuracy.

Turn power on to the power supply by moving the rocker

switch on the front of the power supply to the ON (I)

position. The rocker switch should illuminate. Move the

safety lock to the ON position and briefly depress the on/off

lever to ensure that power is being applied to the

handpiece, and that the carriage is oscillating. Return the

safety lock to the SAFE position.

5. To activate the dermatome, place the power supply in the

ON (I) position. Lift the on/off lever and slide the safety lock

back to the ON position. Depress the on/off lever to start

the motor. DO NOT run the dermatome without cutting

for an extended time, as this needlessly wears the blade.

-2-

Page 3

CUTTING THE GRAFT

Safely dispose of all used blades. Thoroughly scrub the

dermatome and width plate(s) with a soft brush and mild

detergent. All traces of blood, coagulated material, stains,

etc. should be removed.

1. Lubricating the donor site with sterile mineral oil or other

lubricant may ease the travel of the Zimmer Tissue Bank

Dermatome.

Rinse all detergent from the dermatome and width plate(s)

under a faucet. DO NOT IMMERSE IN LIQUID.

2. Hold the handpiece on the donor site at a 30–45 degree

angle.

Shake the dermatome and width plate free of water and

wipe with a clean, lint-free cloth. An alcohol wipe may be

used to disinfect the surface of the dermatome before

steam sterilization.

3. Depress the on/off lever to start the cut. Guide the unit

forward using a slight downward pressure to ensure that

the cutting edge remains in contact with the donor site.

4. Two methods of graft removal may be used:

Steam sterilize the Zimmer Tissue Bank Dermatome

handpiece, width plates, and screwdriver. Follow the

instructions in the STERILIZATION RECOMMENDATIONS

section.

a. Method I (see Fig. 11): Allow the cut graft to

accumulate in the pocket of the handpiece. Lift the

handpiece away from the donor site to end the graft

Return the on/off lever to the SAFE position and

carefully remove the graft.

Always inspect the handpiece carefully for possible

scratches, nicks, or burrs. Inspect the dermatome’s cord for

cuts or missing insulation caused by extended use or

mishandling. If necessary, return for service to an

authorized Zimmer Orthopaedic Surgical Products service

center.

b. Method II (see Fig. 12): Use tissue forceps to gently

lift the graft as it emerges from the pocket area. DO

NOT stretch or pull the graft as this causes irregular

edges and nonuniform cuts. Lift the handpiece away

from the donor site to end the graft. Return the on/off

lever to the SAFE position.

STERILIZATION RECOMMENDATIONS

5. DO NOT run the Zimmer Tissue Bank Dermatome

without cutting for an extended time. Release the on/off

lever and return the safety lock to the SAFE position

between cuts to remove the graft.

Steam sterilization is safe and effective, and is the only

recommended method for the dermatome. There are no

contraindications for steam sterilizing the Zimmer Tissue

Bank Dermatome handpiece.

CLEANING PRECAUTIONS

NEVER sterilize the power supply. Disconnect the

dermatome from the power supply before sterilization.

See Fig. 13–15.

Place cleaned instruments in an instrument tray or fully

perforated autoclave case. The Zimmer Tissue Bank

Dermatome Autoclave Case (REF 8831-03) is recommended.

DO NOT LUBRICATE THE ZIMMER TISSUE BANK

DERMATOME. The dermatome contains sealed bearings.

No lubrication is necessary.

DO NOT kink or crimp the dermatome’s cord when closing

the case lid.

NEVER IMMERSE THE ZIMMER TISSUE BANK

DERMATOME in liquid chemical disinfectants, detergent

solutions, or rinse water. Some solutions will corrode the

metal, delicate moving parts, and seals.

If the instruments are to be wrapped, two double thicknesses

of #140 thread count wrappers, or equivalent, should be

utilized. Exposure times are the same for wrapped or

unwrapped instruments.

DO NOT process the dermatome handpiece or cord in a

washer/sterilizer.

Follow the instructions for prevacuum steam sterilization or

gravity air displacement steam sterilization listed below.

NEVER CLEAN THE DERMATOME IN AN ULTRASONIC

CLEANER. Ultrasonic cleaning will dislodge oil from the

internal bearings and render the instrument inoperative.

Prevacuum Steam Sterilization

HANDLE THE ZIMMER TISSUE BANK DERMATOME

CAREFULLY. Should it be dropped or damaged, it should

be returned to an authorized Zimmer Orthopaedic Surgical

Products service center to ensure the integrity of

calibration.

If your prevacuum steam sterilizer has prefixed cycles, use

the "hard goods" cycle.

If it does not have a prefixed cycle:

1. Set temperature at 270–272°F (132–133°C).

2. Set exposure time for 4 minutes.

3. Set drying time for 8 minutes minimum.

NEVER STERILIZE OR IMMERSE THE POWER SUPPLY.

Gravity Air Displacement Steam Sterilization

1. Set the temperature and corresponding exposure time as

follows:

270–272°F (132–133°C) 35 minutes

250–254°F (121–123°C) 60 minutes

2. Set the drying time for 8 minutes minimum.

DO NOT IMMERSE in liquid to cool. Cool by exposure to

room temperature or cover with a cold, sterile towel.

CLEANING INSTRUCTIONS

Clean the power supply with a damp, lint-free cloth using a

mild detergent. An alcohol wipe may be used to disinfect

the surface of the power supply. NEVER IMMERSE the

power supply.

-3-

Page 4

SERVICE INFORMATION

WARRANTY (U.S.A.)

Caution: Risk of electric shock. Do not remove cover.

Return to an authorized Zimmer Orthopaedic Surgical

Products service center for servicing. Do not attempt to

disassemble the handpiece. It is a factory sealed unit with

no user serviceable parts inside. Refer servicing to

qualified service personnel.

Zimmer Orthopaedic Surgical Products warrants that the

Zimmer Tissue Bank Dermatome, along with all parts and

accessories, have been tested and inspected, and have

left the factory in proper working condition, free from

visible defects.

Zimmer Orthopaedic Surgical Products warrants to the first

purchaser of the Zimmer Tissue Bank Dermatome, Width

Plates, Screwdriver, and Power Supply that these products

and accessories will, under normal and reasonable use,

be free from defects in material and workmanship for one

(1) year after the date of shipment from the factory. The

Zimmer Tissue Bank Dermatome Blades are warranted to

be free from defects in material and workmanship upon

delivery. During the warranty period, Zimmer Orthopaedic

Surgical Products shall repair (or at its sole option replace)

the defective product or part without cost to the purchaser.

Defective parts replaced under this warranty shall become

the property of Zimmer Orthopaedic Surgical Products.

This warranty does not cover damage caused by misuse,

abuse, accident, neglect, or any uses not prescribed in

this manual. If the unit becomes defective because of misuse or abnormal conditions of operation, repairs will be

billed at our current rate.

DO NOT attempt to adjust the power supply. It has been

factory calibrated to provide optimum cutting performance

and maximum safety. It has been designed to meet or

exceed specific medical electrical safety standards.

The Zimmer Tissue Bank Dermatome should be returned

to an authorized Zimmer Orthopaedic Surgical Products

service center every 6 months for inspection, preventive

maintenance, and calibration.

FUSES

Input fuses are located in the fuse drawer in the power

entry module at the rear of the power supply.

120 V~1.25 A time lag/delay

TROUBLESHOOTING GUIDE (TABLE 1)

PROBLEM

ALL OTHER WARRANTIES, EXPRESS, IMPLIED, OR

STATUTORY, INCLUDING, BUT NOT LIMITED TO,

IMPLIED WARRANTIES OF MERCHANTABILITY OR

FITNESS FOR A PARTICULAR PURPOSE, ARE

LIMITED IN TIME TO THE PERIOD OF THE WARRANTY

GRANTED HEREBY. SOME STATES DO NOT ALLOW

LIMITATIONS ON THE DURATION OF AN IMPLIED

WARRANTY, SO THE ABOVE LIMITATIONS MAY NOT

APPLY TO YOU.

ACTION

DERMATOME NOT

OPERATING.

POWER LIGHT OFF.

CHECK INPUT FUSES.

CHECK POWER CORD

CONNECTIONS.

RETURN FOR SERVICE.

DERMATOME NOT

OPERATING.

POWER LIGHT ON.

CHECK BLADE FIT.

CHECK HANDPIECE

CONNECTION TO

POWER SUPPLY.

RETURN FOR SERVICE.

DERMATOME

OPERATING TOO

SLOWLY.

CHECK HANDPIECE

CONNECTION TO

POWER SUPPLY.

ERRATIC SPEED

CHANGES.

CHECK HANDPIECE

CONNECTION TO

POWER SUPPLY.

Neither Zimmer Orthopaedic Surgical Products nor the

Zimmer distributor who sells the Zimmer Tissue Bank

Dermatome is responsible for indirect, incidental, or

consequential damages. Some states do not allow the

exclusion of incidental or consequential damages, so the

above limitations or exclusions may not apply to you.

Zimmer Tissue Bank Dermatomes requiring service or

repair may be returned to Zimmer Orthopaedic Surgical

Products, 200 West Ohio Avenue, Dover, Ohio, U.S.A.

44622. A purchase order and return goods authorization

number must accompany all equipment for repair. The

customer will be responsible for all shipping charges. For

further information concerning warranties and repairs,

contact your Zimmer sales representative or the Zimmer

Customer Service Department at 800-348-2759 (inside the

U.S.A.) or 574-267-6131 (outside the U.S.A.).

WARRANTY (Outside U.S.A.)

ACCESSORIES

ZIMMER TISSUE BANK DERMATOME

Please contact your local Zimmer representative for

warranty information.

See Fig. 16–23.

For further information contact your Zimmer sales

representative or the Zimmer Customer Service

Department at 800-348-2759 (inside the U.S.A.) or

574-267-6131 (outside the U.S.A.).

-4-

Page 5

1

2

A

B

C

E

D

A.

B.

C.

D.

E.

Autoclave Case

Power Supply

Screwdriver

Width Plates

Handpiece

-5-

Page 6

3

WARNING AND

CAUTION LABELS

SYMBOL

DEFINITIONS

4

A. Type of Equipment

B. Direct Current

C. Year of Manufacture

D. OFF (Power: Disconnection from the Mains)

E. On (Power: Connection to the Mains)

F. Alternating Current

WARNING

G. Protective Earth (Ground)

THE TISSUE BANK DERMATOME IS INTENDED FOR

USE ON CADAVERS ONLY, AND NOT FOR LIVING

PATIENTS REQUIRING SKIN GRAFTS.

H. Potential Equalizing Conductor (Ground)

I.

Zimmer Dermatome Handpiece

J. Attention, Consult Accompanying

Documents

K. Conformity Marking for Underwriters

Laboratories, Inc. UL 60601-1 Standard

5

B

C

D

A

E

Power Supply

A.

B.

C.

D.

E.

Front

Mains Switch

Connector Receptacle

Power Entry Module

Back

-6-

Page 7

6

8

B

C

A

D

E

Handpiece

A.

B.

C.

D.

E.

On/Off Safety Lock

On/Off Lever

Handpiece

Connector Plug

Thickness Control Lever

9

7

A

B

C

D

10

Blade Replacement

A.

B.

C.

D.

Screwdriver

Width Plate

Blade

Carriage

-7-

Page 8

11

13

A

B

14

METHOD I

A. Handpiece

B. Pocket

12

B

15

A

METHOD II

A. Handpiece

B. Forceps

-8-

Page 9

16

17

REF 8831

DESCRIPTION:

Zimmer Tissue Bank Dermatome Complete

(includes 8831-01, 8821-06, 8831-03)

REF 8831-01

DESCRIPTION:

Handpiece

(includes 8831-04, 8831-05, 8803)

18

19

REF 8831-06

DESCRIPTION: Power Supply

REF 8831-03

DESCRIPTION: Autoclave Case

-9 -

Page 10

20

REF 8831-000-10

DESCRIPTION:

Zimmer Tissue Bank Dermatome

Blades, Sterile

(10 per box)

21

REF 8803

DESCRIPTION:

Screwdriver

22

REF 8831-04, 8831-05

DESCRIPTION:

Width Plates

23

REF 8803-001-10

DESCRIPTION:

Width Plate Screws

(10 per pack)

- 10 -

Page 11

- 11 -

Page 12

Revised: 05-05

© 2005 Zimmer, Inc.

Printed in U.S.A.

18-10620

Printed on Recycled Paper