manual_orthopaedic_surgical_instruments_rev_5.pdf
Page 1
Manual
Orthopaedic
Surgical
Instruments
Instructions for Care, Cleaning, Maintenance and Sterilization
Page 3
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
Table of Contents
1. Purpose............................................................................................................. 2
2. Scope................................................................................................................ 2
3. Glossary............................................................................................................ 3
4. Acronyms.......................................................................................................... 3
5. Symbols............................................................................................................ 3
6. Considerations.................................................................................................. 4
7. Processing Category Codes................................................................................ 5
8. Instructions....................................................................................................... 6
A. Warnings and Precautions............................................................................................... 6
B. Receiving Inspection – Content and Functionality Verification........................................... 7
C. Limitations and Restrictions on Reprocessing.................................................................. 7
D. Point of Use Preparation for Reprocessing........................................................................ 8
E. Preparation Before Cleaning............................................................................................ 8
F. Preparation of Cleaning Agents....................................................................................... 9
G. Manual Cleaning and Disinfection Instructions................................................................. 9
H. Combination Manual/Automated Cleaning and Disinfection Instructions......................... 10
I. Automated Cleaning and Disinfection Instructions.......................................................... 10
J. Inspection, Maintenance, Testing and Lubrication.......................................................... 10
K. Sterile Packaging.......................................................................................................... 11
L. Sterilization................................................................................................................. 12
M. Storage Recommendations........................................................................................... 13
9. Hospital Responsibilities for Zimmer Loaned Instrument Sets..........................13
10. Customer Service Information..........................................................................14
11. References......................................................................................................15
Table 1. Cleaning/Disinfection Options................................................................................. 8
Table 2. Typical U.S. Automated Washer/Disinfector Cycle for Surgical Instruments.............. 10
Table 3. Typical European Automated Washer/Disinfector Cycle for Surgical Instruments ......10
Table 4. Recommended Steam Sterilization Parameters....................................................... 13
Appendix 1. Cleaning/Disinfection/Sterilization Process..................................................... 16
1
Page 4
2
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
1. PURPOSE
These instructions are recommended for the care,
cleaning, maintenance and sterilization of reusable
Zimmer orthopaedic manual surgical instruments. This
document is intended to assist health care personnel
in safe handling practices, effective reprocessing and
maintenance of Zimmer reusable devices. It provides
information complementary to the instructions for use
in fulfillment of ISO 17664, ANSI/AAMI ST81, and the
European Council Directive 93/42/EEC, Annex 1, section
13.6 (h).
The instructions are intended to assist the hospital and
central supply management in developing procedures
for safe and effective reprocessing of Zimmer instrument
sets.
Hospital personnel, including those in receiving and
central sterile supply departments (CSSD), as well as
in the operating room (OR), may be directly involved
in handling instruments purchased from Zimmer or
on a loan basis as consignment instruments. Hospital
directors and other management in each of these
departments should be informed of these instructions
and recommendations to ensure safe and effective
reprocessing and to prevent damage or misuse of
reusable devices.
2. SCOPE
This instruction manual provides information on
the care, cleaning, disinfection, maintenance
and sterilization of manual surgical instruments
and is applicable to all reusable medical devices
manufactured and/or distributed by Zimmer, Inc.
This information is also applicable to single-use
medical devices manufactured by Zimmer that are
supplied non-sterile but are intended to be used in
a sterile state. These devices are single-use but can
be resterilized if not used (e.g. screws, plates, etc.).
This also includes single-use devices packaged and
sold sterile but removed from packaging and placed
in kits.
Note: not used refers to those single-use components that have not been in contact with blood,
bone, tissue or other body fluids. Any unused,
single-use device that has been exposed to blood,
bone, tissue or body fluids must not be reprocessed
or resterilized and must be discarded.
Devices that cannot be reused may be labeled with
the following symbol:
Do not reuse
This information is not applicable to singleuse devices that are sold sterile and cannot be
resterilized (e.g. osteotome blades).
Devices that cannot be resterilized may be labeled
with the following symbol:
Do not resterilize
This instruction manual is not applicable to air
driven or electrically powered equipment. However,
it is applicable to functional attachments (e.g.
reamers and drill bits) that are connected to
powered equipment for use.
Powered devices included in a manual device
case must be cleaned per specific manufacturer’s
instructions (e.g. Brasseler powered hand pieces).
Page 5
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
3. GLOSSARY
Chemical: a formulation of compounds intended for use in
reprocessing.
Note: Chemicals include detergents, surfactants, rinse aids,
disinfectants, enzymatic cleaners and sterilants.
Cleaning: the removal of contamination from an item to the
extent necessary for further processing or for the intended use.
Tray: basket, with or without a lid, that has perforated sides
or bottom, that holds instruments, and that is either enclosed
in sterilization wrap or a pouch or placed inside a container for
sterilization.
Washer/disinfector: a machine intended to clean and disinfect
medical devices and other articles used in the context of
medical, dental, pharmaceutical, and veterinary practice.
Contaminated: State of having been actually or potentially in
contact with microorganisms or infectious particles.
4. ACRONYMS
Containment device (case): reusable rigid sterilization container,
instrument case/cassette, or organizing tray and any reusable
accessories intended for use in health care facilities for the
purpose of containing reusable medical devices for sterilization.
BI = biological indicator
Decontamination: the use of physical or chemical means to
remove, inactivate, or destroy blood-borne pathogens on a
surface or item to the point where they are no longer capable
of transmitting infectious particles and the surface or item is
rendered safe for handling, use, or disposal.
OR = operating room
Disinfection: process used to reduce the number of viable
microorganisms on a product to a level previously specified as
appropriate for its further handling or use.
Note: Cleaning and disinfection are often conducted in the
same step (e.g. washer/disinfector equipment).
CJD = Creutzfeldt-Jakob Disease
CSSD = Central Sterile Supply Department
PPE = personal protective equipment
SAL = sterility assurance level
TSE = Transmissible Spongiform Encephalopathy
5. SYMBOLS
Do not reuse
Manual cleaning: cleaning without the use of an automated
washer or washer/disinfector.
Processing/reprocessing: activity including cleaning,
disinfection and sterilization, necessary to prepare a new or
used medical device for its intended use.
Reusable rigid sterilization Container: sterilization containment
device designed to hold medical devices for sterilization,
storage, transportation, and aseptic presentation of contents.
Consult instructions
Do not resterilize
Sterile: free from all viable microorganisms.
Sterilization: a validated process used to render a device free
from all forms of viable microorganisms.
Note: In a sterilization process, the nature of microbiological
death is described by an exponential function. Therefore, the
presence of microorganisms on any individual item may be
expressed in terms of probability. While this probability may be
reduced to a very low number, it can never be reduced to zero.
This probability can only be assured for validated processes.
Caution or Instructions for Use
3
Page 6
4
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
6. CONSIDERATIONS
This instruction manual pertains to all hip, knee, trauma,
and extremity reusable medical devices manufactured
and/or distributed by Zimmer, Inc. This manual also
pertains to all hip, knee, trauma, and extremity single-use
medical devices manufactured by Zimmer that are supplied
nonsterile but are intended to be used in a sterile state.
This manual does not pertain to Zimmer spine or dental
devices. This information should be studied carefully.
This manual supercedes Zimmer, Centerpulse, and Implex
manual orthopaedic instrument reprocessing instructions
and instrument manuals published prior to Revision Date
2013.
The user/processor should comply with local laws and
ordinances in countries where reprocessing requirements
are more stringent than those detailed in this manual.
New and used instruments must be thoroughly processed
according to these instructions prior to use.
During musculoskeletal surgery, instruments become
contaminated from blood, tissue, bone fragments and
marrow. The instruments may also become contaminated
with body fluids containing hepatitis virus, HIV or other
etiological agents and pathogens. All health care workers
should become familiar with the necessary Universal
Precautions of preventing injuries caused by sharp
instruments when handling these devices during and after
surgical procedures and during reprocessing.
However, the next user must also inspect the set upon
receipt to verify that instruments have, in fact, been
adequately cleaned and decontaminated before repeating
reprocessing procedures to prepare the loaner set for
subsequent reuse. Zimmer cannot guarantee that sterility
was attained by the previous user and has been maintained
during transit. Zimmer representatives often open and
inspect instrument sets between users, which will, of
course, compromise cleanliness and sterility and require
complete reprocessing prior to subsequent use. Zimmer
requires certification of cleaning and disinfection prior to
return of loaner sets to Zimmer
This manual includes instructions for Zimmer reusable
devices marked with reprocessing category codes [a,
a+, b, b+, c]. See Section 7 of this manual for further
explanation of reprocessing codes. All Zimmer devices
may be safely and efficiently reprocessed using the manual
or combination manual/automated cleaning instructions
outlined in this manual. The combination with an automated
process is preferable.
Core orthopaedic instrument sets must be complete and in
good condition to be used correctly. Optional devices may
be available on request from your Zimmer representative.
To maintain instruments properly it is important to consider
the following information and processing instructions:
• Warnings and precautions
• Instrument set completeness and functionality
It should be noted that saline and other irrigation fluids
such as Ringers Solution are often used in copious amounts
during surgical procedures and may cause corrosion of the
instruments.
• Reprocessing limitations and or restrictions
Orthopaedic surgery requires instruments which are heavy
and have multiple components, articulating or rotating
parts, removable handles, plastic replacement parts, and
series of gauges or other measuring devices in graduated
sizes. Devices are usually supplied in sets and subdivided
into trays and cases in which the devices may be arranged
by size or in the order needed for a specific surgical
procedure.
• Cleaning, disinfection and drying
Hospitals must assume responsibility for cleaning,
disinfection, packaging and sterilization of all loaner
instrument sets before returning them to Zimmer.
• Preparation for reprocessing at the point of use
• Preparation for cleaning (including assembly/
disassembly as necessary)
• Maintenance, inspection, testing and lubrication
• Sterile packaging
• Sterilization
• Storage
Page 7
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
7. PROCESSING CATEGORY CODES
The following codes are etched on some devices and case components and provide information useful in the selection
of cleaning agents with appropriate pH as well as indications for disassembly as well as manual reprocessing.
Zimmer recommends that all reusable devices (regardless of etching) be processed in accordance with the manual or
combination manual/automated cleaning instructions contained in this reprocessing instruction.
Note: Codes on trays and cases apply only to those components noted and do not apply to the contents of the tray or
case.
a
Metal devices (excluding aluminum and titanium) and case components without features posing
a cleaning challenge or non-metal/polymer handles, or other components (e.g. retractors,
drills, testing trays, rasps, scissors, clamps, exploring hooks, compression forceps, skin bridge
elevators, guide wires, etc.). These devices are tolerant of alkaline cleaning agents when followed
by acidic neutralization and thorough rinsing. These devices can be cleaned with rust-removal
agents approved for surgical instruments if needed.
a+
Metal devices (excluding aluminum and titanium) and case components with features posing a
cleaning challenge but without non-metal/polymer handles or other components (e.g. drills with
elongated holes, belt tensioning pulleys, bone joint reamers, extractor cases). These devices are
tolerant of alkaline cleaning agents when followed by acidic neutralization and thorough rinsing.
These devices can be cleaned with rust-removal agents approved for surgical instruments if
needed. Features posing a cleaning challenge must be cleaned manually.
b
b+
c
Devices and case components without features posing a cleaning challenge made of polymers or
metal instruments paired with polymer components (e.g. testing trays for flat profiles, chisels with
non-metal handles, awls, dissectors, femur dilatators, pyramidal chisels/rasps). These devices
are tolerant of alkaline cleaning agents when followed by acidic neutralization and thorough
rinsing.
Devices and case components with features posing a cleaning challenge, made of polymers or
metal instruments paired with polymer components (e.g. tibial mallets, flex screwdrivers, tibial
dilatators, etc.). These devices are tolerant of alkaline cleaning agents when followed by acidic
neutralization and thorough rinsing. Features posing a cleaning challenge must be cleaned
manually.
Devices and case components made of titanium or aluminum alloys and/or having assembly/
disassembly or other reprocessing aids (e.g. torque spanners, tibial aiming devices, pad cutters,
instrument cases, trays and sterilization containers). These devices should be cleaned using the
manual or combination manual/ automated cleaning procedures provided in this manual. These
devices should not be exposed to alkaline cleaning agents.
Note: Features posing a cleaning challenge include; lumens/cannulated bores, tightly mated surfaces, rough surfaces,
ball detents, springs, and multiple component designs.
5
Page 8
6
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
8. PROCESSING INSTRUCTIONS
These processing instructions are intended to assist the
hospital and central supply management in developing
procedures to attain safe and effective devices, both
for hospital-owned and for loaned instrument sets. This
information is based on Zimmer testing, experience
and material science, as well as widely accepted
recommendations of the following organizations:
• American National Standards Institute (ANSI)
• American Society for Testing and Materials (ASTM)
• Association for the Advancement of Medical
Instrumentation (AAMI)
• Association for Applied Hygiene (VAH)
• Association of Operating Room Nurses (AORN)
• Canadian Standards Association (CSA)
• Centers for Disease Control (CDC)
• Federal Institute for Drugs and Medical Devices
(BfArM, Bundesinstitut für Arzneimittel und
Medizinprodukte)
• German Instrument Working Group (AKI)
Arbeitskreis Instrumenten-Aufbereitung
• International Standards Organization (ISO)
• International Association of Healthcare Central
Service Material Management (IAHCSMM)
• National Health Service (NHS)
• Robert Koch Institute (RKI)
• Swissmedic
• World Health Organization (WHO)
Note: These instructions describe the necessary
processing steps that new and used instruments must
undergo to attain sterility.
A. Warnings and Precautions
• Universal Precautions should be observed by all
hospital personnel that work with contaminated or
potentially contaminated medical devices. Caution
should be exercised when handling devices with sharp
points or cutting edges.
• Personal Protective Equipment (PPE) should be
worn when handling or working with contaminated
or potentially contaminated materials, devices and
equipment. PPE includes gown, mask, goggles or face
shield, gloves and shoe covers.
• Metal brushes or scouring pads must not be used
during manual cleaning procedures. These materials
will damage the surface and finish of instruments. Softbristled, nylon brushes and pipe cleaners should be used.
• Cleaning agents with low foaming surfactants should be
used during manual cleaning procedures to ensure that
instruments are visible in the cleaning solution. Manual
scrubbing with brushes should always be performed
with the instrument below the surface of the cleaning
solution to prevent generation of aerosols and splashing
which may spread contaminants. Cleaning agents must
be completely rinsed from device surfaces to prevent
accumulation of detergent residue.
• Do not stack instruments or place heavy instruments on
top of delicate devices.
• Dry soiled surgical instruments are more difficult to
clean. Do not allow contaminated devices to dry prior to
reprocessing. All subsequent cleaning and sterilization
steps are facilitated by not allowing blood, body fluids,
bone and tissue debris, saline, or disinfectants to dry on
used instruments.
• Saline and cleaning/disinfection agents containing
aldehyde, mercury, active chlorine, chloride, bromine,
bromide, iodine or iodide are corrosive and should not
be used. Instruments must not be placed or soaked in
Ringers Solution.
• Mineral oil or silicone lubricants should not be used
because they: 1) coat microorganisms; 2) prevent direct
contact of the surface with steam; and 3) are difficult to
remove.
• Only devices manufactured and/or distributed by Zimmer
should be included in Zimmer instrument trays and
cases. These validated reprocessing instructions are not
applicable to Zimmer trays and cases that include devices
that are not manufactured and/or distributed by Zimmer.
• Descaling agents that include morpholine should not be
used in steam sterilizers. These agents leave residue
which can damage polymer instruments over time. Steam
Sterilizers should be descaled in accordance with the
manufacturer’s instructions.
Page 9
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
B. Receiving Inspection – Instrument set
content and functionality verification.
• Upon receipt in the hospital, instrument sets should
be inspected for completeness. Inspect for thumb,
wing, set, or other types of screws; screw-in or other
detachable handles; and auxiliary exchangeable
parts such as blades, right/left attachments or heads.
Many organizing cases have shadow graphs, outlines,
catalog numbers, and instrument names or sizes silkscreened or otherwise marked on the case or tray.
• Orthopaedic surgical procedures follow a precise
order in which the instruments are used. Also, many
instruments have dimensional features which govern
bone resections, determine implant sizes, and measure
intramedullary canal sizes, depth of drill holes,
angles of tube/plate, acetabular cup placements, etc.
Therefore, it is very important that all requested sizes
of a specific instrument series are available (specific
instruments are routinely omitted from instrument
sets due to infrequent use unless requested by the
user). Contact your Zimmer representative if requested
instruments have been omitted but are required for
surgery.
• Markings on instruments used for measuring
anatomical dimensions must be legible. These
may include gauge markings, angles, inner or outer
diameters, length or depth calibrations, and right/left
indications. Notify your Zimmer representative if scales
and other markings are not legible.
C. Limitations and Restrictions
• Neutral pH, enzymatic, and alkaline (pH ≤12) cleaning
agents are recommended and preferred for cleaning
Zimmer reusable devices. Alkaline agents with pH ≤
12 may be used to clean stainless steel and polymer
instruments in countries where required by law or
local ordinance; or where prion diseases such as
Transmissible Spongiform Encephalopathy (TSE) and
Creutzfeldt-Jakob Disease (CJD) are a concern. It is
critical that alkaline cleaning agents are thoroughly
neutralized and completely rinsed from devices.
Note: Drill bits, reamers, rasps and other cutting devices
should be carefully inspected after processing with alkaline
detergents to ensure that cutting edges are fit for use.
Note: It is important to select enzymatic solutions intended
for breakdown of blood, body fluids and tissues. Some
enzymatic solutions are specifically for breakdown of fecal
matter or other organic contaminants and may not be
suitable for use with orthopaedic instruments.
• Repeated processing, according to the instructions in
this manual has minimal effect on Zimmer reusable
manual instruments unless otherwise noted. End of life
for stainless steel or other metal surgical instruments
is normally determined by wear and damage due to the
intended surgical use and not to reprocessing.
• Automated cleaning using a washer/disinfector alone
may not be effective for complex orthopaedic instruments
with lumens, cannulations, blind holes, mated surfaces
and other features. A thorough, manual or combination
manual/automated cleaning process is recommended.
The combination with an automated process is preferable.
• Where applicable, multi-component instruments should
be disassembled for cleaning. Disassembly, where
necessary, is generally self evident. More specific
instructions can be found in the instructions for use
and at www.zimmer.com. Care must be taken to avoid
losing small parts. If a part is lost, notify your Zimmer
representative when the instrument set is returned.
• At point of use, soiled instruments must be removed from
metal or polymer trays and moistened to prevent debris
from drying before transportation to the reprocessing
area for manual and/or automated cleaning procedures.
Do not clean soiled instruments while in polymer or metal
trays. Instrument trays, cases and lids must be cleaned
separately from soiled instruments.
• Non-sterile, single-use plate and screw implants should
not be cleaned but must be returned to the tray or caddy
for sterilization.
Note: Any unused, single-use device that has been
exposed to blood, bone, tissue or body fluids must not be
reprocessed or resterilized and must be discarded.
• Polymers used in Zimmer instrument sets can be
sterilized using steam/moist heat. Polymer materials
have a limited useful life. If polymer surfaces turn
“chalky,” show excessive surface damage (e.g. crazing
or delamination), or if polymer devices show excessive
distortion or are visibly warped, they should be replaced.
Notify your Zimmer representative if polymer devices
need to be replaced.
7
Page 10
8
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
• Most currently available polymers will not withstand
conditions in washers/sterilizers that operate at
temperatures equal to or greater than 141°C / 285°F,
and use live-steam jets as cleaning features. Severe
surface damage to polymer devices may occur under
these conditions.
• Soaking in disinfectants may be a necessary step to
control certain viruses. However, these agents may
discolor or corrode instruments (household bleach
contains or forms chlorine and chloride in solution and
has a corrosive effect similar to saline). Disinfectants
containing glutaraldehyde, or other aldehydes may
denature protein based contaminants, causing them
to harden and making them difficult to remove. Where
possible, soaking in disinfectants should be avoided.
• Steam/moist heat is the recommended sterilization
method for Zimmer instruments.
• Ethylene Oxide (EO), Gas Plasma Sterilization and dry
heat sterilization methods are not recommended for
sterilization of Zimmer reusable instruments.
• Instruments with removable polymer sleeves must be
disassembled for sterilization (e.g. acetabular reamer
shaft with sleeve, side cutters, etc.)
• During initial steam sterilization runs some
formaldehyde from polyformaldehyde surfaces may
vaporize and become noticeable. This should not
cause concern. After a few sterilization cycles, the odor
should be no longer evident.
• While ethylene oxide sterilization may prolong the
service life of certain polymers (e.g. polysulfone), this
method of sterilization is not recommended for Zimmer
surgical instruments. Large polyformaldehyde items
(Delrin®, Celcon®) have been found to require excessive
outgassing times (a minimum of five days at elevated
temperatures in a mechanical aerator); therefore,
gas sterilization for polyformaldehyde products is
contraindicated.
• Titanium and titanium alloy devices are especially
susceptible to discoloration from steam impurities and
detergent residues which form multi-colored surface
layers of oxide deposits. Upon repeated sterilization
these oxide layers, while not harmful to the patient,
may become dark and obscure graduation marks,
item and lot numbers, and other stamped or etched
information. Acidic, anti-corrosion agents may be used
to remove this discoloration as needed.
• Stainless steel instruments may be cleaned with rustremoval agents approved for surgical instruments if
needed.
• Use of hard water should be avoided. Softened tap water
may be used for initial rinsing. Purified water should be
used for final rinsing to eliminate mineral deposits on
instruments (e.g. ultra-filter (UF), reverse-osmosis (RO),
deionized (DI), or equivalent).
D. Point of Use Preparation for Reprocessing
• Remove excess body fluids and tissue from instruments
with a disposable, non-shedding wipe. Place instruments
in a basin of distilled water or in a tray covered with
damp towels. Do not allow saline, blood, body fluids,
tissue, bone fragments or other organic debris to dry on
instruments prior to cleaning.
Note: Soaking in proteolytic enzyme solutions or other
precleaning solutions facilitates cleaning, especially in
instruments with complex features and hard-to-reach
areas (e.g. cannulated and tubular designs, etc.). These
enzymatic solutions as well as enzymatic foam sprays
break down protein matter and prevent blood and
protein based materials from drying on instruments.
Manufacturer’s instructions for preparation and use of
these solutions should be explicitly followed.
• For optimal results, instruments should be cleaned
within 30 minutes of use or after removal from solution to
minimize the potential for drying prior to cleaning.
• Used instruments must be transported to the central
supply in closed or covered containers to prevent
unnecessary contamination risk.
E. Preparation Before Cleaning
• Symbols or specific instructions etched on instruments or
instrument trays and cases should be strictly followed.
• Where applicable multi-component instruments should be
disassembled for appropriate cleaning. Care should be
exercised to avoid losing small screws and components.
If a part is lost, notify your Zimmer representative when
the instrument set is returned.
• Instructions for instrument assembly/disassembly
and device specific cleaning aids can be found at www.
zimmer.com.
Page 11
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
F. Preparation of Cleaning Agents
• Neutral pH enzymatic and cleaning agents with low
foaming surfactants are recommended by Zimmer.
• Alkaline agents with pH ≤ 12 may be used in countries
where required by law or local ordinance. Alkaline
agents should be followed with a neutralizer and/or
thorough rinsing.
• Only agents with proven efficacy (FDA approved, VAH
listed, or CE mark) should be used. As a large variety
of cleaning agents and disinfectants exists around the
globe, Zimmer does not recommend any specific brand.
• Agents used during the validation of these processing
instructions are:
Steris®, Prolystica™ 2X Enzymatic Pre Soak and Cleaner,
Prolystica™ Ultra Concentrate Neutral Detergent
• All cleaning agents should be prepared at the usedilution and temperature recommended by the
manufacturer. Softened tap water may be used
to prepare cleaning agents. Use of recommended
temperatures is important for optimal performance of
cleaning agents.
• Dry powdered cleaning agents should be completely
dissolved prior to use to avoid staining or corrosion of
instruments and to ensure correct concentration.
• Fresh cleaning solutions should be prepared when
existing solutions become grossly contaminated
(bloody and/or turbid).
Table 1. Cleaning/Disinfection Options
Method
Description
Section
Manual
Enzymatic soak and scrub
followed by sonication.
G
Combination
Manual/
Automated
Enzymatic soak and scrub
followed by automated washer/
disinfector cycle
H
Automated
(Washer/
Disinfector)
Washer/disinfector cycle - not
recommended for use without
manual precleaning.
I
• The manual method is effective for all devices and may
be used when an automated option is not available.
Note: Manual cleaning may require onsite validation
by the healthcare facility and appropriate procedures/
documentation should be in place to avoid human factor
variability.
• The combination manual/automated method is
preferred and can be used for all devices.
• The automated method should only be used on simple
devices without multiple components, lumens/
cannulations, blind holes, mated surfaces, connectors
and internal mechanisms or other complex features
G. Manual Cleaning/Disinfection Instructions
1. Completely submerge instruments in an enzyme
or alkaline (pH ≤12) solution and allow to soak for
20 minutes. Use a soft-bristled, nylon brush to
gently scrub the device until all visible soil has been
removed. Particular attention must be given to
crevices, lumens, mated surfaces, connectors and
other hard-to-clean areas. Lumens should be cleaned
with a long, narrow, soft-bristled brush (i.e. pipe
cleaner brush).
2. Remove the instruments from the cleaning solution
and rinse in tap water for a minimum of 3 minutes.
Thoroughly and aggressively flush lumens, holes
and other difficult-to-reach areas.
3. Place prepared cleaning agents in a sonication unit.
Completely submerge device in cleaning solution and
sonicate for 10 minutes at 45-50 kHz.
4. Rinse instrument in purified water for at least 3
minutes or until there is no sign of blood or soil on
the device or in the rinse stream. Thoroughly and
aggressively flush lumens, holes and other difficultto-reach areas.
5. Repeat the sonication and rinse steps above.
6. Remove excess moisture from the instrument with a
clean, absorbent and non-shedding wipe.
Note: If stainless steel instruments are stained or
corroded, an acidic, anti-corrosion agent in an ultrasonic
cleaner may be sufficient to remove surface deposits.
Care must be taken to thoroughly rinse acid from devices.
Acidic, anti-corrosion agents should only be used on an
as needed basis.
9
Page 12
10
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
H. Combination Manual/Automated Cleaning
and Disinfection Instructions
Table 3. Typical European Automated Washer/
Disinfector Cycle for Surgical Instruments
1. Completely submerge the instruments in an enzyme
or alkaline (pH ≤12) solution and allow to soak
for 10 minutes. Use a soft nylon-bristled brush
to gently scrub the device until all visible soil has
been removed. Particular attention must be given
to crevices, lumens, mated surfaces, connectors
and other hard-to-clean areas. Lumens should be
cleaned with a long, narrow, soft nylon-bristled brush
(i.e. pipe cleaner).
Step
Description
1
5 min pre-rinse with cold tap water
2
10 min wash with 0,7% cleaning agent at
55°C
3
1 min rinse with cold tap water
4
Disinfection at 93°C with hot purified water
until A0 3000 is reached (approx. 10 min)
5
40 min hot air drying at 110°C
Note: Use of a sonicator at 45-50 kHz will aid in
thorough cleaning of devices.
Note: Use of a syringe or water jet will improve flushing
of difficult to reach areas and closely mated surfaces.
2. Remove instruments from the cleaning solution and
rinse in purified water for a minimum of 1 minute.
Thoroughly and aggressively flush lumens, blind
holes and other difficult-to-reach areas.
3. Place instruments in a suitable washer/disinfector
basket and process through a standard instrument
washer/disinfector cleaning cycle. The following
minimum parameters are essential for thorough
cleaning and disinfection.
Table 2. Typical U.S. Automated Washer/
Disinfector Cycle for Surgical Instruments
Step
Description
1
2 minute prewash with cold tap water
2
20 second enzyme spray with hot tap water
3
1 minute enzyme soak
4
15 second cold tap water rinse (X2)
5
2 minutes detergent wash with hot tap water
(64-66C/146-150F)
6
15 second hot tap water rinse
7
2 minute thermal rinse (80-93C/176-200F)
8
10 second purified water rinse with optional
lubricant (64-66C/146-150F)
9
7 to 30 minute hot air dry (116C/240F)
Note: The washer/disinfector manufacturer’s instructions
should be strictly adhered to. Use only cleaning agents
recommended for the specific type of automated washer/
disinfector. A washer/disinfector with approved efficacy
(e.g. CE mark, FDA approval, and validation according to
ISO 15883) should be used.
I. Automated Cleaning/Disinfection
Instructions
1. Automated washer/disinfector systems are not
recommended as the sole cleaning method for
surgical instruments. Orthopaedic instruments
should be cleaned following the manual or
combination manual/automated cleaning procedure
outlined in this manual except where specifically
indicated.
2. Simple instruments without multiple components,
lumens/cannulations, blind holes, mated surfaces,
connectors and internal mechanisms or other
complex features may be successfully cleaned and
disinfected using a typical washer/disinfector cycle
for surgical instruments as outlined in Table 2 of this
manual. Devices should be thoroughly inspected
prior to sterilization to ensure effective cleaning.
J. Inspection, Maintenance, Testing and
Lubrication
1. Carefully inspect each device to ensure that all visible
contamination has been removed. If contamination
is noted repeat the cleaning/disinfection process.
2. Visually inspect for completeness, damage and/or
excessive wear.
Page 13
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
Note: If damage or wear is noted that may compromise
the function of the instrument, contact your Zimmer
Representative for a replacement.
Note: If sterilization wraps are used they must be
free of detergent residues. Reusable wraps are not
recommended.
3. Check the action of moving parts (e.g. hinges, boxlocks, connectors, sliding parts, etc.) to ensure
smooth operation throughout the intended range of
motion.
Packaging instrument sets in rigid trays and cases with
lids
4. Hinged, rotating, or articulating instruments should
be lubricated with a water soluble product (e.g.
Instrument Milk or equivalent lubricant) intended
for surgical instruments that must be sterilized.
Some water-based instrument lubricants contain
bacteriostatic agents which are beneficial. To
remain effective, the expiration date specified by the
manufacturer should be adhered to for both stock
and use-dilution concentrations.
Note: Mineral oil or silicone lubricants should not be
used because they 1) coat microorganisms; 2) prevent
direct contact of the surface with steam; and 3) are
difficult to remove.
Note: These lubrication instructions are not applicable
to air-powered or electrical instruments. These devices
have different requirements and should be lubricated
according to the manufacturer’s instructions.
Safety Precaution: The total weight of a wrapped
instrument tray or case should not exceed 11.4kg/25lbs.
Instrument cases may be placed in an approved
sterilization container with gasketed lids at the user’s
discretion. The total weight of the instrument set,
case, and sterilization container, must not exceed
11.4kg/25lbs.
• Trays and cases with lids may be wrapped in standard
medical grade, steam sterilization wrap using the AAMI
double wrap method or equivalent.
• Trays and cases with lids may also be placed in an
approved sterilization container with gasketed lid for
sterilization.
Note: Follow the sterilization container manufacturer’s
instructions for inserting and replacing sterilization
filters in sterilization containers.
Instrument trays and cases with defined, preconfigured
layouts
5. Check instruments with long slender features
(particularly rotating instruments) for distortion.
• Areas designated for specific devices shall contain only
devices specifically intended for these areas.
6. Where instruments form part of a larger assembly,
check that the devices assemble readily with mating
components.
• Optional Zimmer instruments should not be added
to a preconfigured instrument tray or case unless a
dedicated universal space or compartment has been
included in the design and the guidelines described
below for trays and cases without defined layouts or
universal spaces can be applied.
K. Sterile Packaging
Packaging individual instruments
• Single devices should be packaged in a medical grade
sterilization pouch or wrap which conforms to the
recommended specifications for steam sterilization
provided in the table below. Ensure that the pouch
or wrap is large enough to contain the device without
stressing the seals or tearing the pouch or wrap.
• Standard medical grade, steam sterilization wrap
may be used to package individual instruments. The
package should be prepared using the AAMI double
wrap or equivalent method.
• Only devices manufactured and/or distributed by
Zimmer should be included in Zimmer instrument
trays. These validated reprocessing instructions are not
applicable to Zimmer trays that include devices that are
not manufactured and/or distributed by Zimmer.
Instrument trays with reconfigurable layouts
• Brackets designated for specific devices shall contain
only devices specifically intended for them.
11
Page 14
12
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
• Optional Zimmer instruments should not be added to a
reconfigurable tray unless a dedicated universal space
or compartment has been included in the layout and the
guidelines described below for universal trays without
defined layouts or universal spaces can be applied.
• Only devices manufactured and/or distributed by
Zimmer should be included in Zimmer instrument
trays. These validated reprocessing instructions are not
applicable to Zimmer trays that include devices that are
not manufactured and/or distributed by Zimmer.
• Brackets designed to force disassembly of a complex
device must not be altered to allow the assembled
device to be inserted into the tray or case.
• To ensure devices are fully seated in their
corresponding brackets and to prevent damage to tray
contents, individual brackets should not overlap one
another when inserted into the tray floor.
Note: Some individual brackets may be designated for
assembly onto other “host” brackets. In these instances,
the mating relationship between the brackets will be
graphically depicted on the face of the “host” bracket.
• Bracket fasteners should be fully engaged with the tray
floor to prevent unintended migration, damage and/or
loss of tray contents.
• Wave springs positioned over the shaft of the
bracket fasteners are intended to stabilize brackets
by minimizing free-play between them and the tray
floor. To ensure intended function, periodically
inspect brackets for damaged and/or missing springs
which can be replaced by contacting your Zimmer
representative.
• Identification tags and associated labels on trays
should correspond to tray contents to ensure correct
trays are available for use in surgery.
• Any manual tools provided by Zimmer to aid in the
removal of individual brackets must not remain in
the instrument trays during reprocessing and are not
intended for use in surgery.
Universal instrument trays and cases without defined,
preconfigured layouts or containing undefined universal
spaces or compartments should only be used under the
following conditions:
Safety Precaution: The total weight of a wrapped
instrument tray or case should not exceed 11.4kg/25lbs.
Instrument cases may be placed in an approved
sterilization container with gasketed lids at the user’s
discretion. The total weight of the instrument set, case
and container must not exceed 11.4kg/25lbs.
• Any device capable of disassembly must be
disassembled prior to placement in the case.
• All devices must be arranged to ensure steam
penetration to all instrument surfaces. Instruments
should not be stacked or placed in close contact.
• The user must ensure that the instrument case is not
tipped or the contents shifted once the devices are
arranged in the case. Silicone mats may be used to
keep devices in place.
• Only devices manufactured and/or distributed by
Zimmer should be included in Zimmer instrument
trays. Zimmer validated reprocessing instructions are
not applicable to Zimmer trays that include devices
that are not manufactured and/or distributed by
Zimmer.
L. Sterilization Instructions
• See Table for recommended minimum sterilization
parameters that have been validated by Zimmer to
provide a 10-6 sterility assurance level (SAL).
• The hospital is responsible for in-house procedures
for the reassembly, inspection, and packaging of the
instruments after they are thoroughly cleaned in a
manner that will ensure steam sterilant penetration
and adequate drying. Provisions for protection of
any sharp or potentially dangerous areas of the
instruments should also be recommended by the
hospital.
• Moist heat/steam sterilization is the preferred and
recommended method for Zimmer orthopaedic
instrument sets.
• Sterilizer manufacturer recommendations should
always be followed. When sterilizing multiple
instrument sets in one sterilization cycle, ensure that
the manufacturer’s maximum load is not exceeded.
Page 15
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
• Instrument sets should be properly prepared and
packaged in trays and/or cases that will allow steam to
penetrate and make direct contract with all surfaces.
• Ethylene oxide or gas plasma sterilization methods
should not be used unless package inserts for the
applicable product specifically provide instructions for
sterilization using these methods.
• Gravity displacement sterilization cycles are not
recommended because cycle times are too long to be
practical.
Table 4. Recommended Steam Sterilization
Parameters
Cycle Type
Temperature 2
Exposure Time 1
Wrapped7,8 and
Unwrapped9
UK Prevacuum/
Pulsating Vacuum3
134°C / 273°F
3 minutes
Prevacuum/
Pulsating Vacuum2,3
132°C / 270°F
4 minutes
Prevacuum/Pulsating
134°C / 273°F
Vacuum4,12,13
18 minutes
Minimum Minimum
Dry Time10 Cool Time11
132°C / 270°F
8 minutes
Gravity/Gravity
Displacement
121°C / 250°F
90 minutes
Drying times vary according to load size and should be increased for
larger loads.
11
Cooling times vary according to the type of sterilizer used, device
design, temperature and humidity of ambient environment, and type
of packaging used. Cooling process should comply with ANSI/AAMI
ST79.
12
This cycle is not for use in the United States.
13
This cycle is not to be used for the inactivation of prions.
Note: The Sterilizer Manufacturer’s instructions for
operation and load configuration should be followed
explicitly.
M. Storage Instructions
• Sterile, packaged instruments should be stored in a
designated, limited access area that is well ventilated
and provides protection from dust, moisture, insects,
vermin, and temperature/humidity extremes.
30 minutes
Prevacuum/
Pulsating Vacuum5
10
30 minutes
40 minutes
1
Validated exposure time required to achieve a 10-6 sterility assurance
level (SAL).
2
Validated exposure temperature required to achieve a 10-6 sterility
assurance level (SAL).
3
Local or national specifications should be followed where steam
sterilization requirements are stricter or more conservative than those
listed in this table.
4
Disinfection/steam sterilization parameters recommended by the
World Health Organization (WHO) for reprocessing instruments where
there is concern regarding TSE/CJD contamination.
5
For Universal Instrument Cases without defined load configurations.
6
AAMI/AORN steam sterilization cycles with longer times than those
listed are also acceptable.
7
Medical grade steam sterilization compatible wrap equivalent to four
thicknesses of 140-thread-count muslin.
8
Rigid sterilization container that complies with ANSI/AAMI ST77.
9
Flash (immediate-use) steam sterilization by exposure at 132134C / 270-273F should only be used as an emergency procedure.
Instruments must be cleaned and disassembled.
• Sterile instrument packages should be carefully
examined prior to opening to ensure that package
integrity has not been compromised.
Note: Maintenance of sterile package integrity is
generally event related. If a sterile wrap is torn,
perforated, shows any evidence of tampering or has
been exposed to moisture, the instrument set must be
cleaned, repackaged and sterilized.
Note: If there is any evidence that the lid seal or filters
on a sterilization container have been opened or
compromised, the sterile filters must be replaced and
the instrument set resterilized.
9. HOSPITAL RESPONSIBILITIES FOR ZIMMER
LOANER SETS
• Orthopaedic surgical instruments generally have a
long service life; however, mishandling or inadequate
protection can quickly diminish their life expectancy.
Instruments which no longer perform properly because
of long use, mishandling, or improper care should
be returned to Zimmer to be discarded. Notify your
Zimmer representative of any instrument problems.
13
Page 16
14
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
• Loaner sets should undergo all steps of
decontamination, cleaning, disinfection, inspection,
and terminal sterilization before being returned to
Zimmer. Documentation of decontamination should
be provided with instruments being returned to
Zimmer.
• Missing or damaged instruments from loaner sets
should be brought to the attention of the operating
room supervisor, to the director of the central supply
department, and to your Zimmer representative to
ensure that the next hospital will receive a complete
set of instruments in good working condition.
• The instructions provided in this manual have been
validated by Zimmer in the laboratory and are
capable of preparing orthopaedic devices for use. It
is the responsibility of the Hospital to ensure that
reprocessing is performed using the appropriate
equipment and materials, and that personnel in the
reprocessing facility have been adequately trained
in order to achieve the desired result. Equipment
and processes should be validated and routinely
monitored. Any deviation by the processor from
these instructions should be properly evaluated
for effectiveness to avoid potential adverse
consequences.
10. CUSTOMER SERVICE INFORMATION
Mailing Address
Telephone
Zimmer, Inc.
1800 West Center Street
Warsaw, Indiana 46580
USA
Inside USA: 1-800-348-2759
Outside USA: local international
access code +1-574-267-6131
Zimmer GmbH
Sulzer-Allee 8
CH-8404 Winterthur,
Switzerland
+41 (0) 52 262 60 70
This Zimmer reprocessing manual and device specific cleaning and
assembly/disassembly instructions can be found at www.zimmer.
com under the “Medical Professional” heading.
Page 17
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
11. REFERENCES
1. AAMI TIR12, Designing, testing, and labeling reusable
medical devices for reprocessing in health care facilities: A
guide for medical device manufacturers
2. AAMI TIR13, Principles of industrial moist heat sterilization
3. AAMI TIR30, A compendium of processes, materials, test
methods, and acceptance criteria for cleaning reusable
medical devices
4. AAMI TIR34, Water for the reprocessing of medical devices
5. ANSI/AAMI ST67, Sterilization of health care products –
Requirements for products labeled “Sterile”
6. ANSI/AAMI ST77, Containment devices for reusable medical
device sterilization
7. ANSI/AAMI ST79, Comprehensive guide to steam
sterilization and sterility assurance in health care facilities
8. ANSI/AAMI ST81, Sterilization of medical devices –
Information to be provided by the manufacturer for the
processing of resterilizable medical devices
9. ANSI/AAMI/ISO 15223-1, Medical device – Symbols to be
used with medical device labels, labeling, and information
to be supplied - Part 1: General Requirements
10. AORN, Standards, Recommended Practices and Guidelines
11. Association for Applied Hygiene (VAH) Verbund für
Angewandte Hygiene, List of Disinfectants
12. ASTM F 565, Standard Practice for Care and Handling of
Orthopedic Implants and Instruments
13. European Commission, Council Directive 93/42/EEC of 14
June 1993 concerning medical devices
14. German Instrument Working Group (AKI) Arbeitskreis
Instrumenten-Aufbereitung, Proper Maintenance of
Instruments, 8th Ed, 2004.
15. IAHCSMM, Central Service Technical Manual
16. ISO 15883, Washer/Disinfectors: General Requirements,
Terms and Definitions and Tests
17. ISO 17664, Sterilization of medical devices –
Information to be provided by the manufacturer for the
processing of resterilizable medical devices
18. ISO 17665-1, Sterilization of health care products –moist
heat, Part 1
19. ISO 17665-2, Sterilization of health care products –moist
heat, Part 2
Note: All trademarks are the property of their respective owners.
20. Robert Koch Institute (RKI), Hygiene Requirements for
Reprocessing Medical Devices Federal Health Gazette,
10/2012
21. Robert Koch Institute (RKI), Hospital Supplies and
Instrument Sterilization in Light of CJD Patients and
Suspected CJD Cases, Federal Health Gazette, 7/1998
22. UK Department of Health, Health Technical Memorandum
(HTM) 2010, Sterilization, Part 5 – Good Practice Guide
23. UK Department of Health, Health Technical Memorandum
(HTM) 2030, Washer-Disinfectors – Validation and
Verification
24. World Health Organization (WHO), WHO/CDS/CSR/APH
200.3, WHO Infection Control Guidelines for TSE
15
Page 18
16
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
Appendix 1- Cleaning/Disinfection/Sterilization Validation Process
Zimmer manual and combination manual/automated cleaning and disinfection methods are developed using spore
logarithmic reduction and visual inspection criteria to determine critical cleaning parameters. The reusable devices
or device features are exposed to blood soil containing approximately 106 spores per ml. The samples are allowed to
dry for 30 minutes prior to cleaning/disinfection. After cleaning the test samples are visually inspected for any signs of
remaining blood soil. Bioburden extractions are performed to determine the number of spores remaining on thhe test
samples. The data is compared to positive controls to determine if the acceptance criteria is met. The cleaning/
disinfection processes outlined in this manual have been validated per the following standards:
AAMI TIR 12, Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A
guide for device manufacturers
AAMI TIR 30, A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable
medical devices
All reusable devices are reviewed to ensure that the device features are not more challenging than those validated.
As new, more challenging features emerge, the features are validated or the device is modified to a less challenging
design.
Zimmer prevacuum (pressure pulse) and gravity moist heat sterilization cycles are developed using F0 data to establish
critical cycle parameters. Calibrated thermocouples are placed in the most challenging locations within the instrument
cases for heat penetration studies. Moist heat cycles are validated using microbial challenge half-cycle studies to prove
lethality via the “overkill” method to demonstrate a 10-6 or better sterility assurance level (SAL). This method utilizes
biological indicators (BI) inoculated with more than one million (106) resistant spores (Geobacillus stearothermophilus)
placed in the most challenging locations within the instrument set. Studies are conducted using disposable moist heat
sterilization wrap and/or rigid steam sterilization containers (Aesculap). Successful half-cycle microbial challenge
studies result in total kill of all BIs demonstrating a spore logarithm reduction of at least twelve (10-6 SAL or better)
for the most resistant microorganism at a population of greater than one million. Recommended dry time is validated
by demonstrating that pre-sterilization and post sterilization case weight is within 3 percent with no visible moisture
at the end of the full sterilization cycle. Durability studies are also performed using multiple cycles at 135°C for 18
minutes to ensure that the instrument case and contents can withstand repeated processing at maximum parameters.
The sterilization parameters outlined in this manual are validated per the following standards:
AAMI TIR 12, Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A
guide for device manufacturers
ANSI/AAMI/ISO 17665, Sterilization of health care products – Moist heat – Part 1: Requirements for the development,
validation, and routine control of a sterilization process for medical devices
Page 19
Zimmer Instrument Care, Cleaning, Maintenance and Sterilization Instructions
17
Page 20
Contact your Zimmer representative or visit us at www.zimmer.com
The CE mark is valid only if it is also printed on the product label.
97-5000-170-00 Rev. 5, ©1987, 1988, 2002, 2006, 2010, 2012 and 2013