SurePowerTM II Battery Pack Guide
?
0+
1:0
0+
2:0
0+
3:0
M
T
M
T
®
®
9650-000840-01 Rev. B
Page 2
The issue date for the SurePower™ II Battery Pack Guide (REF 9650-000840-01 Rev. B)
is December, 2012.
If more than three years have elapsed since the issue date, contact ZOLL Medical Corporation to
determine if additional product information updates are available.
SurePower, X Series, and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in
the United States and/or other countries.
All other trademarks are the property of their respective owners.
Copyright © 2012 ZOLL Medical Corporation. All rights reserved.
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA, 01824
USA
Tel.: +1 978-421-9655
Fax: +1 978-421-0001
ZOLL International Holding B.V.
Newtonweg 18
6662 PV ELST
Nederland
Tel.: +31 (0) 481 366410
Fax: +31 (0) 481 366411
Page 3
Introduction
The SurePowerTM II Battery Pack is a rechargable, sealed lithium ion battery for use in
compatible ZOLL ® defibrillator/monitors and chargers.
It is important that you read and understand the information in this manual. The proper use of
the SurePower II Battery Pack is critical to the reliable operation of your ZOLL defibrillator/
monitor.
Conventions
This manual uses the following conventions:
Warning statements describe conditions or actions that can result in personal injury or
death.
Caution
Caution statements describe conditions or actions that can result in damage to the SurePower II
Battery Pack.
Note:
Notes contain additional information on using and maintaining the SurePower II
Battery Pack.
The manual displays button and LED labels in bold face type.
Safety
The following sections provide important warnings, cautions, and notes regarding the ZOLL
SurePower II Battery Pack.
Warnings
• Locate ZOLL defibrillator/monitor units where they cannot harm the patient or medical
•
•
•
•
•
•
personnel if they fall -- accidental sliding or drop of the SurePower II Battery Pack may
cause physical injury. Damaged batteries may expose users to a safety hazard. Should the
SurePower II Battery Pack be dropped, inspect and test the unit prior to further use
following the procedures described in the Maintenance section of this Guide.
Avoid prolonged exposure of the SurePower II Battery Pack to direct sunlight to minimize
heating and damage that may occur from exposure to ultraviolet light.
SurePower II Battery Packs have been tested against interference from radio frequency
emissions typical of two-way radios used in emergency service/public safety activities.
Refer to the appropriate ZOLL defibrillator/monitor operator’s guide to determine
recommended operating distances from RF transmission equipment.
Do not use SurePower II Battery Packs unless they have been visually inspected, tested, and
recalibrated by appropriate personnel and you have been adequately trained in the proper
use of these batteries. See the Maintenance section of this guide for more information.
Do not disassemble the SurePower II Battery Pack.
Charge the SurePower II Battery Pack in a well-ventilated area.
You must perform periodic inspection, capacity testing, and functional testing of
rechargable defibrillator batteries to ensure the availability of safe and reliable batteries. See
the Maintenance section of this guide for more information.
9650-000840-01 Rev. B
SurePowerTM ll Battery Pack Guide
1
Page 4
INTRODUCTION
• When in use on a defibrillator/monitor, if the battery pack indicates a fault condition or any
•
•
•
•
•
operating state that can prevent the effective use of the defibrillator/monitor, replace that
battery pack as soon as possible.
Do not place foreign objects into battery sockets.
Do not use the SurePower II Battery Pack in the presence of flammable agents (such as
gasoline), oxygen-rich atmospheres, or flammable anesthetics. Using the device in the
presence of flammable agents can cause an explosion.
Do not immerse or set the SurePower II Battery Pack in liquid. Using the battery near or
within puddles of water may present a shock hazard to the operator, patient, and nearby
personnel.
Use only ZOLL chargers to charge ZOLL batteries.
If the display indicators on the SurePower II Battery Pack are not consistent with the
indicators on the SurePowerTM Charger Station, a fault condition may be present. Remove
the battery from service as soon as practical, and verify the correct operation of the charger,
defibrillator, and battery.
Cautions
• Follow all recommended maintenance instructions. If a problem occurs, obtain service
immediately.
Do not use alcohol or ketones (MEK, acetone, etc.) to clean SurePower II Battery Packs.
Do not sterilize or autoclave SurePower II Battery Packs.
Avoid storage of fully depleted battery packs.
Do not install batteries into a defibrillator/monitor unit that is not plugged into live AC
mains when storage may exceed 90 days. Battery damage may occur.
• If you are disposing of a SurePower II Battery Pack, contact your local authorities to
determine the requirements for the recycling and disposal of electrical equipment and follow
those requirements.
•
•
•
•
Notes
• SurePower II Battery Packs may feel warm to the touch following a large number of shocks
delivered at maximum energy or after recharging.
• You can get additional information about the SurePower II Battery Pack and its use from the
appropriate ZOLL defibrillator/monitor operator’s guide (for example, labeling information
-- symbols used, Serial Number format, Date of Manufacture, and so on).
2
www.zoll.com
9650-000840-01 Rev. B
Page 5
Using the SurePower II Battery Pack
The SurePower II Battery Pack provides enhanced on-board processing that allows the battery
to communicate its specific charging requirements to the SurePower™ Charger Station or
ZOLL defibrillator/monitor unit. The battery’s on-board processing also allows it to provide the
following information:
2:00+
3:00+
?
1:00+
• The battery’s run time or capacity.
• The battery’s calibration status.
• Battery Fault Indications.
3URE0OWER4- ))
8 3ERIES40ROPAQ --$
®
Figure 1-1 ZOLL SurePower II Battery Pack
Note:
ZOLL ships the SurePower II Battery Pack in a partially charged state and you must
fully charge the battery pack before first use.
It is very important that you recharge a depleted SurePower II Battery Pack as soon as possible.
Caution
Do not store fully depleted battery packs or damage can result.
Caution
Do not store fully charged batteries for more than 90 days. Should you need to store a battery for
90 days or longer, see the Extended Storage section of this guide for information on how to do
so.
9650-000840-01 Rev. B
SurePowerTM ll Battery Pack Guide
3
Page 6
USING THE SurePower II Battery Pack
Displaying Battery Runtime
The SurePower II Battery Packcan display its present status.
To display the status of the SurePower II Battery Pack, press and
release the Display button (momentary press) on the top of the
battery.
When you press and release the Display button, all LEDs blink to
indicate that they work. The battery then lights specific LEDs for
approximately 10 seconds to indicate the battery’s runtime, calibration
status, and fault state.
We describe what these LEDs indicate in the following sections.
Note:
The SurePower II Battery Pack’s LEDs light only when the Display button is pressed.
SurePower II Battery Pack
Icon
LED Indications
This is the runtime indicator that displays the estimated
runtime of the SurePower II Battery Pack in your defibrillator.
When you momentarily press and release the Display
button on the SurePower II Battery Pack, the LEDs above
these icons light for approximately 10 seconds to indicate
the battery’s estimated run time.
If the battery is fully charged, the LED over each of the five
runtime indicator icons illuminates. As the battery
discharges, the number of illuminated LEDs decreases.
As the SurePower II Battery Pack discharges, pressing the
Display button will light, from left to right, the LED’s that
indicate the battery’s run time. For example, if four LED’s
light, the battery can power a defibrillator/monitor for at least
3 hours (displayed as 3:00+); if three LED’s light, the battery
can power the defibrillator/monitor for at least 2 hours
(displayed as 2:00+), and so on.
When you press the Display button on a battery that is
installed and powering a defibrillator/monitor, the runtime
indicator displays the run time for the battery in that specific
unit.
When a SurePower II Battery Pack is not installed in a
ZOLL defibrillator/monitor unit, the runtime indicator displays
the runtime for the battery at a nominal drain rate for that
unit.
Note: If the left-most LED of the runtime indicator blinks,
the battery has less than fifteen minutes of use left,
and defibrillator shutdown will soon occur. Recharge
the battery as soon as possible.
4
www.zoll.com
9650-000840-01 Rev. B
Page 7
SurePower II Battery Pack
Icon
LED Indications
This is the Recalibration icon.
This LED lights for approximately 10 seconds (after you
press and release the Display button) if the battery needs to
be calibrated. If the Recalibration LED lights, the runtime
indicator will not display run time for that battery. For best
performance of the battery, you should recalibrate the
battery as soon as possible.
To manually recalibrate the SurePower II Battery Pack, you
can insert the battery into the SurePower Charger Station
(using the SurePowerTM II Battery Adaptor) and perform a
ManualTest (for more information, see the ZOLL
SurePower Charger Station Operator’s Guide). After you
recalibrate the battery, the Recalibration LED will only flash
momentarily when you press the Display button.
This is the Fault Icon.
If the SurePower II Battery Pack detects a battery fault, this
LED lights for approximately 10 seconds after you press the
Display button.
If the Fault LED lights, the runtime indicator will not display
run time for that battery.
See the following section for more information on battery
faults.
Cleaning
Clean the SurePower II Battery Pack with a soft damp cloth and mild soap.
Do not immerse or set the SurePower II Battery Pack in fluids of any kind. Do not use alcohol
or ketones to clean the SurePower II Battery Pack.
Maintenance -- Testing and Inspection
Perform a battery check, which verifies that the battery is fully charged and fault free, as part of
any regularly scheduled unit test. Press the battery’s Display button to perform the battery
check.
As part of your defibrillator/monitor maintenance program, verify that the battery’s LEDs work
(press the Display button to test LEDs) -- if the LEDs do not work, remove the battery from
service. Also check the SurePower II Battery Pack casing for cracks, damaged contacts and
sockets, or other damage -- if the battery pack is cracked or damaged, remove it from service.
9650-000840-01 Rev. B
SurePowerTM ll Battery Pack Guide
5
Page 8
USING THE SurePower II Battery Pack
Extended Storage
Do not store the SurePower II Battery Pack for extended periods in a fully charged state. If you
must store the battery pack for an extended period
1. Charge the SurePower II Battery Pack until three runtime LEDs light (this takes
approximately 1.5 hours when charging a fully depleted battery).
2. Recharge the SurePower II Battery Pack until three runtime LEDs light every 4 months.
Checking Capacity Loss
To check battery capacity (the amount of charge the battery can hold), press and hold down the
Display button for at least 5 seconds.
Press and hold down the Display button for at least
5 seconds.
While you hold down the Display button, the
runtime indicator displays the battery runtime.
When you release the Display button, the
runtime indicator displays the battery capacity
for 2 seconds, then redisplays the battery
runtime.
During the 2 seconds after you release the
Display button, the runtime indicator displays
battery capacity and will light five LEDs to
indicate that the capacity is equal or close to
that of a new battery.
After displaying battery capacity for 2 seconds,
all LEDs blink and then display the battery’s
runtime.
Over time, as you use a SurePower II Battery
Pack, the battery capacity decreases and this
capacity check will light fewer LEDs. When the
capacity check lights only a single LED, the
battery is at or near 60% of its original capacity,
and you should remove the battery from service.
If the Recalibration or Fault LED lights after
you release the Display button, the runtime
indicator will not display the capacity or run time
for that battery.
6
www.zoll.com
9650-000840-01 Rev. B
Page 9
Understanding Battery Faults
The SurePower II Battery Pack can detect one or more simultaneously occurring faults that
render the battery unfit for use.
To determine the faults that affect a battery, follow this procedure when the fault light is on:
Press and hold down the Display button for at least
15 seconds.
If the battery does not detect any battery faults, all of the
LEDs will light for 10 seconds.
The Fault LED will blink to indicate the code that identifies
the battery fault (there will be a brief delay before the LED
blinks).
The fault codes number from 1 to 29. To indicate fault code
1, the Fault LED blinks once. To indicate fault code 2, the
LED blinks twice, and so on.
If the battery detects multiple battery faults, it displays the
first fault code, pauses briefly, and then displays the next
fault code, and so on.
The following table lists battery temperature faults and actions you can take to clear these fault
conditions. All other fault codes require you to remove the battery from service. For more
information on these fault codes, see the appropriate ZOLL defibrillator/monitor service
manual.
Fault ID
Fault Type
Recommended Action
13
Battery Over Temperature
During Charge.
Move the battery to a cooler room or out of
direct sunlight, if possible, and allow the
battery to cool before continuing to charge the
battery.
The battery is at or above its
maximum temperature for
charging (140°F/60°C).
14
Battery Under Temperature
during Charge.
The battery is at or below the
minimum temperature for
charging (32°F/0°C).
9650-000840-01 Rev. B
If the ambient temperature was low when the
battery reached or exceeded its maximum
temperature, do not use this battery.
Move the battery to a warmer room (above
32°F / 0°C) and allow the battery to warm up
before charging again.
SurePowerTM ll Battery Pack Guide
7
Page 10
USING THE SurePower II Battery Pack
SurePower II Battery Pack Specifications
Type
Lithium Ion
Size
7 (Length) x 3.7 (Width) x 1.2 (Height) in
17.8 (Length) x 9.3 (Width) x 3 (Height) cm
Weight
1.4 lb (.64 kg)
Voltage
11.1 V
Capacity
6.6 Ah when charged to 12.6V
Operating Temperature
32°F to 122°F
0°C to 50°C
Optimal Charging
Temperature
59°F to 95°F
Short Term Storage/
Transportation
Temperature *
-22°F to 158°F
Particle and Water Ingress
Protection
IP 55
Altitude of Operation
-557.7 ft through 15,000 ft (below and above sea level)
15°C to 35°C
-30°C to 70°C
-170 m through 4572 m (below and above sea level)
103 kPa through 57 kPa
*For extended storage periods, store the SurePower II Battery Pack between
60°F to 77°F (15°C to 25°C) to minimize capacity loss.
8
www.zoll.com
9650-000840-01 Rev. B