X Series Service Manaul Rev A April 2012.pdf
Page 1
X Series Service Manual
™
9650-001356-01 Rev. A
Page 2
The issue date for the X Series Service Manual (REF 9650-001356-01 Rev. A) is April, 2012.
If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to determine if
additional product information updates are available.
Copyright © 2012 ZOLL Medical Corporation. All rights reserved. Rectilinear Biphasic, SurePower, X Series,
and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or
other countries.
Masimo and Rainbow are trademarks or registered trademarks of Masimo Corporation in the United States
and/or other countries.
Oridion and Microstream are trademarks or registered trademarks of Oridion Systems, Ltd. in the United States
and/or other countries.
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA USA
01824-4105
ZOLL International Holding B.V.
Newtonweg 18
6662 PV ELST
The Netherlands
0197
Page 3
Table of Contents
Preface
Overview................................................................................................................................ 1
Safety Considerations............................................................................................................ 2
Additional Reference Material................................................................................................ 2
Conventions ........................................................................................................................... 3
Service Policy Warranty......................................................................................................... 3
Technical Service................................................................................................................... 3
Technical Service for International Customers ...................................................................... 4
Chapter 1
Maintenance Tests
X Series Overview ................................................................................................................. 5
Calibration and Calibration Checks ....................................................................................... 7
NIBP Valve, Pump, and Tubing Integrity Tests .....................................................................11
Operational Tests ................................................................................................................. 14
Functional Verification Tests ................................................................................................ 21
Chapter 2
Functional Description
Overview.............................................................................................................................. 41
Appendix A
Overview.............................................................................................................................. 47
X Series Electrical Hardware Interconnect Diagram............................................................ 48
X Series Signal and Power Interconnect Schematic ........................................................... 49
Appendix B
Overview.............................................................................................................................. 51
Maintenance Test Checklist
Overview.............................................................................................................................. 55
9650-001356-01 Rev. A
ZOLL X Series Service Manual
i
Page 4
TABLE OF CONTENTS
ii
9650-001356-01 Rev. A
Page 5
Preface
Overview
ZOLL® Medical Corporation’s X Series™ Service Manual is intended for the trained
biomedical technician whose responsibility is to identify malfunctions and/or make repairs at
the subassembly level. The X Series Service Manual has two main sections and two
appendixes.
Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to
review this section thoroughly before attempting to use or service the X Series unit.
Chapter 1—Maintenance Tests describes the Annual Calibration Tests for NIBP and EtCO2
modules, Functional Tests that need to be run when a module has been repaired, and
Operational Tests that need to be performed at regular intervals to augment the automatic
testing the X Series does to ensure readiness.
Chapter 2—Functional Description provides technical descriptions for the X Series major
subassembly modules.
Appendix A—X Series Electrical Hardware Interconnect Diagram and X Series Signal and
Power Interconnect Schematic.
Appendix B—Contains simulators and settings that may be used to assess the performance of
the NIBP module.
9650-001356-01 Rev. A
ZOLL X Series Service Manual
1
Page 6
PREFACE
Safety Considerations
The following section describes general warnings and safety considerations for operators and
patients. Service technicians should review the safety considerations prior to servicing any
equipment and read the manual carefully before attempting to disassemble the unit. Only
qualified personnel should service the X Series unit.
Federal (U.S.A.) law restricts this unit for use by or on the order of a physician.
Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug
Administration (FDA) under section 510(K) of the Medical Device Act to obtain approval to
market is based upon the use of ZOLL accessories such as disposable electrodes, patient cables
and batteries. The use of external pacing/defibrillation electrodes and adapter units from
sources other than ZOLL is not recommended. ZOLL makes no representations or warranties
regarding the performance or effectiveness of its products when used in conjunction with
pacing/defibrillation electrodes and adapter units from other sources. If unit failure is
attributable to pacing/defibrillation electrodes or adapter units not manufactured by ZOLL, this
may void ZOLL's warranty.
Only qualified personnel should disassemble the X Series unit.
WARNING!
This unit can generate up to 2775 volts with sufficient current to cause lethal shocks.
All persons near the equipment must be warned to “STAND CLEAR” prior to discharging the
defibrillator.
Do not discharge the unit’s internal defibrillator energy more than three times in one minute or
damage to the unit may result.
Do not discharge a battery pack except in a ZOLL SurePowerTM Charging Station.
Do not use the X Series in the presence of flammable agents (such as gasoline), oxygen-rich
atmospheres, or flammable anesthetics. Using the unit near the site of a gasoline spill may
cause an explosion.
Do not use the unit near or within puddles of water.
Additional Reference Material
In addition to this guide, there is a X Series Operator’s Guide which is a comprehensive
reference work that describes all the user tasks needed to operate the X Series.
2
9650-001356-01 Rev. A
Page 7
Conventions
Conventions
WARNING! Warning statements describe conditions or actions that can result in personal injury or death.
Caution
Caution statements describe conditions or actions that can result in damage to the unit.
Note:
Notes contain additional information on using the defibrillator.
Service Policy Warranty
In North America: Consult your purchasing agreement for terms and conditions associated with
your warranty. Outside of North America, consult ZOLL authorized representative.
In order to maintain this warranty, the instructions and procedures contained in this manual
must be strictly followed. For additional information, please call the ZOLL Technical Service
Department 1-800-348-9011 in North America.
Technical Service
If the ZOLL X Series unit requires service, contact the ZOLL Technical Service Department:
Telephone: 1-978-421-9655; 1-800-348-9011
Fax: 1-978-421-0010
Email: techsupport@zoll.com
Have the following information available for the Technical Service representative:
• Unit serial number.
• Description of the problem.
• Department where equipment is used.
• Sample chart recorder strips or electronic log files documenting the problem, if applicable.
• Purchase Order to allow tracking of loan equipment.
• Purchase Order for a unit with an expired warranty.
If the unit needs to be sent to ZOLL Medical Corporation, obtain a service request number
(SR#) from the Technical Service representative. Return the unit in its original container to:
ZOLL Medical Corporation
269 Mill Road
Chelmsford, Massachusetts 01824-4105
Attn: Technical Service Department (SR#)
Telephone: 1-800-348-9011; 1-978-421-9655 FAX: 978-421-0010
9650-001356-01 Rev. A
ZOLL X Series Service Manual
3
Page 8
PREFACE
Technical Service for International Customers
International customers should return the unit in its original container to the nearest authorized
ZOLL Medical Corporation Service Center. To locate an authorized service center, contact the
International Sales Department at ZOLL Medical at the above address.
4
9650-001356-01 Rev. A
Page 9
Chapter 1
Maintenance Tests
X Series Overview
This chapter includes step-by-step procedures for the following types of maintenance tests:
•
•
•
•
Calibration and Calibration Checks
NIBP Valve, Pump, and Tubing Integrity Tests
Operational Tests
Functional Verification Tests
At a minimum, a trained biomedical technician must perform a CO2 calibration and an NIBP
calibration check once per year. A CO2 calibration check may be performed at any time to
verify the CO2 calibration. It is necessary to perform Functional Verification Tests after a
module is repaired to ensure that the functions of the X Series unit work properly. Use the
checklist at the back of this document (ZOLL X Series Maintenance Tests Checklist) to record
your results of the maintenance tests.
Because the X Series units must be maintained ready for immediate use, it is important for
users to conduct the Operator’s Shift Checklist procedure at the beginning of every shift. This
procedure can be completed in a few minutes and requires no additional test equipment. (See
the ZOLL X Series Operator’s Guide for the Operator’s Shift Checklist.) It is highly
recommended that the Operational Tests be performed at regular intervals to augment the
automatic testing the X Series unit does to ensure readiness.
9650-001356-01 Rev. A
ZOLL X Series Service Manual
5
Page 10
CHAPTER 1 MAINTENANCE TESTS
This chapter describes the following maintenance tests:
Calibration and Calibration Checks
• CO2 Calibration Procedure
• CO2 Calibration Check
• NIBP Calibration Check
NIBP Valve, Pump, and Tubing Integrity Tests
• Leak Test
• Deflation Test
• Inflation Test
Operational Tests
• Physical Inspection of the Unit
• Keypad Test
• LED Test
• LCD Test
• RFU Indicator Test
• Audio Test
• Printer Test
• Battery Test
• Fan Test
• NIBP Test
• USB Test
Functional Verification Tests
• Leads Test
• Heart Rate Display Test
• Heart Rate Alarm Test
• Leakage Current Test
• Shock Test
• Advisory Test
• Synchronized Cardioversion Test
• Pacer Test
• Printer Test
• USB File Transfer Test
• Treatment Report Test
• SpO2 Monitor Test
• EtCO2 Monitor Test
• IBP Monitor Test
• NIBP Monitor Test
• Temperature Test
6
9650-001356-01 Rev. A
Page 11
Calibration and Calibration Checks
Calibration and Calibration Checks
CO2 Calibration Procedure (required for annual calibration)
Equipment
5% CO2 calibration gas cylinder with CO2 sample line (kit available from Scott Medical
Products, PN 4653ORF-BD)
Procedure
1. Turn on the X Series unit.
2. On the X Series unit, select Supervisor>Service>Diagnostic>CO2>Calibration to enter
the CO2 Calibration control panel.
3. Connect the X Series unit to the 5% CO2 calibration gas cylinder using the sample line
included with the calibration kit.
4. From the CO2 Calibration Menu, select Calibration.
5. Activate the calibration gas flow by pressing the top valve on the calibration gas canister.
6. Continue to apply the calibration gas until the status in the CO2 Calibration control panel
displays: CALIBRATION GAS CAN BE REMOVED.
• When the calibration is complete, the status in the CO2 Calibration control panel
displays: CALIBRATION DONE - OK OR CALIBRATION DONE - ERROR.
7. To verify calibration, perform “CO2 Calibration Check Procedure” on page 8.
The calibration date will be updated upon exiting the CO2 Test control panel.
Corrective Action
If an error is reported, refer to the table below for possible causes.
If readings fail to meet the required accuracy, the CO2 module requires replacement or factory
analysis and realignment.
The following table describes possible calibration errors and likely causes.
Table 1-1.
Calibration Errors
Error Message
Cause
No gas or wrong gas
concentration
The concentration of supplied gas is not close to the expected
value.
No stable gas flow and
Calibration failure
The gas sample is not stable, i.e. due to leak or noise.
Measurement error
Occlusion detection during known gas sampling.
Measurement error
Sample line disconnection during gas sampling.
AZ fail
The Auto Zero that is part of the calibration process failed.
Calibration aborted by
user
Calibration mode canceled during gas sampling.
9650-001356-01 Rev. A
ZOLL X Series Service Manual
7
Page 12
CHAPTER 1 MAINTENANCE TESTS
CO2 Calibration Check Procedure
Equipment
5% CO2 calibration gas cylinder with CO2 sample line (kit available from Scott Medical
Products, PN 4653ORF-BD)
If you have previously run the CO2 Calibration Procedure, start at step 3.
Procedure
1. On the X Series unit, select Supervisor>Service>Diagnostic>CO2>Calibration to enter
the CO2 Calibration control panel.
2. Connect the X Series unit to the 5% CO2 calibration gas cylinder using the sample line
included with the calibration kit.
3. From the CO2 Calibration control panel, select Verify Calibration.
4. Activate the calibration gas flow by pressing the top valve on the calibration gas canister.
5. Continue to apply the calibration gas until the status in the CO2 Calibration control panel
displays: CALIBRATION GAS CAN BE REMOVED.
• When the calibration check is complete, the status in the CO2 Calibration control panel
displays: VERIFICATION DONE - OK OR VERIFICATION DONE - ERROR.
• When the calibration check completes successfully, the measured gas value displays in
the CO2 Calibration control panel.
Acceptable Results
The message displays: VERIFICATION DONE - OK.
The measured gas reading passes if it is between 4.8 and 5.2 (Vol%).
6. Exit out of calibration by pressing the Back arrow key (
Display/Home button (
) three times or pressing the
).
Corrective Action
If an error is reported, refer to Table 1-1 on page 7 for possible causes.
If the measured gas reading is outside the acceptable range, or if calibration is due, complete
the following CO2 calibration procedure.
8
9650-001356-01 Rev. A
Page 13
Calibration and Calibration Checks
NIBP Calibration Check
Equipment
Double lumen BP hose, a 500 cc volume, and a calibrated manometer or pressure gauge
(< = 1mmHg accuracy). If a 500 cc volume is not available, an Adult BP cuff wrapped around a
rigid cylinder (such as a glass jar) will suffice.
Note:
This test is to be performed for the following target pressures: 50 +/- 5 mmHg,
150 +/- 5 mmHg, and 250 +/- 5 mmHg.
Procedure
1. Power on the X Series unit.
Note:
If the unit is in AED mode, press the Manual quick access key to go into manual
mode. Use the navigation keys to select the four digits in the Manual Mode pass code.
Press SAVE when you are finished. If the unit has not been configured to enter a pass
code, the message Exit to Manual Mode is displayed. Use the navigation keys to select
Yes.
2. Connect the X Series unit to the double lumen hose and the 500 cc volume. Alternatively,
replace the 500 cc volume with a standard Adult cuff wrapped around a rigid cylinder, such
as a glass jar. Connect a calibrated manometer to the air line using a T-fitting.
3. On the X Series unit, do the following to enter the NIBP Test control panel:
• Press the More (
) then the Setup (
) quick access keys.
• Use the navigation keys to select Supervisor>Service>Device Tests>NIBP Test.
Note:
You need a supervisor passcode to enter the Supervisor menu.
4. From the X Series NIBP Test control panel, use the navigation keys to select:
• Close Valve
• Turn Pump On
5. Monitor the pressure reported by the external pressure standard.
6. Once the pressure reaches the target pressure of 50 +/- 5 mmHg, use the navigation keys to
select Turn Pump Off.
7. Allow the system to stabilize for at least 10 seconds.
8. Record the pressure reported by the external pressure standard as Pa.
9. Record the pressures reported by the primary and safety transducers (displayed in the NIBP
Test control panel) as P1 and P2 respectively.
10. Calculate the absolute pressure differences between the applied pressure (Pa) and each of the
transducer pressures.
• Pd-Primary = abs(Pa-P1)
• Pd-safety = abs(Pa-P2)
11. From the NIBP Test control panel, select Open Valve to release the pressure.
9650-001356-01 Rev. A
ZOLL X Series Service Manual
9
Page 14
CHAPTER 1 MAINTENANCE TESTS
12. Repeat steps 3 to 11 to test for the remaining target pressures of 150 +/- 5 mmHg and
250 +/- 5 mmHg.
13. Once complete, use the navigation keys to select Update NIBP Cal Check Date to save the
date of the most recent calibration check.
14. Press
twice and select Diagnostic>NIBP>Calibration Info. Verify that the last NIBP
Cal date is today’s date.
Acceptable Results
The unit passes the test if Pd-Primary is less than or equal to 3.0 mmHg and Pd-Safety is less than
or equal to 4.0 mmHg for all three target pressures.
Corrective Action
If readings fail to meet the required accuracy, the NIBP module requires replacement or factory
analysis and realignment.
10
9650-001356-01 Rev. A
Page 15
NIBP Valve, Pump, and Tubing Integrity Tests
NIBP Valve, Pump, and Tubing Integrity Tests
NIBP Leak Test
Equipment
Double lumen BP hose, a 500 cc volume, and a calibrated manometer.
Note:
If a 500 cc volume is not available, an Adult BP cuff wrapped around a rigid cylinder
(such as a glass jar) will suffice.
Procedure
1. Turn on the X Series unit.
2. Connect the X Series unit to the single lumen hose and the 500 cc volume. Alternatively,
replace the 500 cc volume with a standard Adult cuff wrapped around a rigid cylinder, such
as a glass jar. If available, connect a calibrated manometer to the air line using a T-fitting.
3. Using the X Series unit, do the following to enter the NIBP Test control panel:
• Press the More (
) then the Setup (
) quick access keys.
• Use the navigation keys to select Supervisor>Service>Device Tests>NIBP Test.
Note:
You need a supervisor passcode to enter the Supervisor menu.
4. From the X Series NIBP Test control panel, use the navigation keys to select:
• Close Valve
• Turn Pump On
5. Monitor the pressure reported by NIBP Test control panel or the external pressure standard.
6. Select Turn Pump Off as soon as the pressure reaches the target pressure of 250 +/- 5 mmHg.
7. Allow the system to stabilize for 10 seconds.
8. Select Mark P1 to mark the starting pressure. If available, measure the starting pressure
reported by the external pressure standard. Record the starting pressure as PS.
9. Wait 15 seconds.
10. Select Mark P2 to mark the ending pressure. If available, measure the ending pressure
reported by the external pressure standard. Record the ending pressure as PE.
11. From the NIBP Test control panel, select Open Valve to release the pressure.
12. Calculate the leak rate: L = (P1 - Pa) / (15 seconds).
Note:
If using the pressure reported in the NIBP Test control panel, calculate the Leak Rate
using the Primary transducer pressures.
13. Select Clear to clear Mark P1 and Mark P2.
Acceptable Results
The unit passes the test if the system leaks less than 5 mmHg over the 15-second time interval
(Leak Rate is less than 1/3).
9650-001356-01 Rev. A
ZOLL X Series Service Manual
11
Page 16
CHAPTER 1 MAINTENANCE TESTS
Corrective Action
If the unit fails the test, the most likely cause is damaged or disconnected internal tubing. The
device needs to be disassembled to replace or reattach the appropriate segment of NIBP tubing.
The NIBP valves should also be inspected.
NIBP Deflation Test
Equipment
Double lumen BP hose, a 500 cc volume, a timer, and a calibrated manometer (optional).
Note:
If a 500 cc volume is not available, an Adult BP cuff wrapped around a rigid cylinder
(such as a glass jar) will suffice.
If you have previously run the NIBP Leak Test, you can go to step 4.
Procedure
1. Turn on the X Series unit.
2. Connect the X Series unit to the double lumen hose and the 500 cc volume. Alternatively,
replace the 500 cc volume with a standard Adult cuff wrapped around a rigid cylinder, such
as a glass jar. If available, connect a calibrated manometer to the air line using a T-fitting.
3. Using the X Series unit, do the following to enter the NIBP Test control panel:
• Press the More (
) then the Setup (
) quick access keys.
• Use the navigation keys to select Supervisor>Service>Device Tests>NIBP Test.
Note:
You need a supervisor passcode to enter the Supervisor menu.
4. From the X Series NIBP Test control panel, use the navigation keys to select:
• Close Valve
• Turn Pump On
5. Monitor the pressure reported by the NIBP Test control panel or the external pressure
standard.
6. Select Turn Pump Off as soon as the pressure reaches the target pressure of 265 +/- 5 mmHg.
7. Allow the system to stabilize for 5 seconds.
8. Use the navigation keys to select Open Valve and immediately start the timer.
9. Stop the timer as soon as pressure drops below 15 mmHg.
10. Record the Deflation Time as TD.
Acceptable Results
The unit passes the test if the time to release pressure from 265 +/- 5 mmHg to 15 mmHg is less
than 10 seconds.
Corrective Action
If the unit fails the test, the most likely cause is kinked or obstructed internal tubing, or
obstructed filter. The device needs to be disassembled to replace or reposition the appropriate
segment of NIBP tubing.
12
9650-001356-01 Rev. A
Page 17
NIBP Valve, Pump, and Tubing Integrity Tests
NIBP Inflation Test
Equipment
Double lumen BP hose, a 250 cc volume, a timer, and a calibrated manometer (optional).
Note:
If you have previously run the NIBP Deflation Test, you can go to step 6.
Procedure
1. Turn on the X Series unit.
2. Connect the X Series unit to the double lumen hose and the 250 cc volume. If available,
connect a calibrated manometer to the air line using a T-fitting.
3. Using the X Series unit, do the following to enter the NIBP Test control panel:
• Press the More (
) then the Setup (
) quick access keys.
• Use the navigation keys to select Supervisor>Service>Device Tests>NIBP Test.
Note:
You need a supervisor passcode to enter the Supervisor menu.
4. Wait 10 seconds to allow the system to vent any residual pressure.
5. Verify that the pressure reported by the NIBP Test control panel or the external pressure
standard system is less than 5 mmHg.
6. From the NIBP Test control panel, use the navigation keys to select Close Valve.
7. Select Turn Pump On and immediately start the timer.
8. Stop the timer as soon as pressure exceeds 210 mmHg.
9. Select Turn Pump Off, then Open Valve.
10. Record the Inflation Time as TI.
Note:
The pump will automatically turn off if pressure exceeds 310 mmHg during this test.
Acceptable Results
The unit passes the test if the time to increase pressure to 210 mmHg is less than 7 seconds.
Corrective Action
If the unit fails the test, the most likely cause is a faulty pump or a kinked internal hose. The
device needs to be disassembled to replace the pump or reposition the appropriate segment of
NIBP tubing.
11. Exit the Service menu and turn off the unit.
9650-001356-01 Rev. A
ZOLL X Series Service Manual
13
Page 18
CHAPTER 1 MAINTENANCE TESTS
Operational Tests
The X Series unit comes equipped with routine operational tests that can be performed
periodically. With the exception of the Physical Inspection of the Unit test, these tests contain
instructions within them that provide guidance during the tests.
Physical Inspection of the Unit
Observe this...
Pass/Fail
Housing
1.
Is the unit clean and undamaged?
o
o
2.
Does the unit show signs of excessive wear?
o
o
3.
Does the handle work properly?
o
o
4.
Does the recorder drawer open and close properly?
o
o
5.
Are input connectors clean and undamaged?
o
o
6.
Are there any cracks in the housing?
o
o
7.
Does the front panel have any damage or cracks?
o
o
8.
Are there any loose housing parts?
o
o
Do the adult and pedi plates have major scratches or show signs of damage?
o
o
10. Do the adult shoes slide on and off easily to expose the covered pedi plates?
o
o
11. Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
o
o
12. Are all cables free of cracks, cuts, exposed or broken wires?
o
o
13. Are all bend/strain reliefs undamaged and free of excessive cable wear?
o
o
14. Is the X Series battery fully charged?
o
o
15. Is the battery seated in the battery well correctly?
o
o
16. Is battery housing free of cracks and damage?
o
o
17. Are connectors free of damage and excessive wear?
o
o
Paddles (if applicable)
9.
Cables
Battery
14
9650-001356-01 Rev. A
Page 19
Operational Tests
Accessing the Operational Tests
The operational tests on the unit can be accessed by doing the following:
1. Turn on the X Series unit.
Note:
If the unit is in AED mode, press the Manual quick access key to go into manual
mode. Use the navigation keys to select the four digits in the Manual Mode pass code.
Press SAVE when you are finished. If the unit has not been configured to enter a pass
code, the message Exit to Manual Mode is displayed. Use the navigation keys to select
Yes.
2. Press the More (
) then the Setup (
) quick access keys.
3. Use the navigation keys to select Supervisor>Service>Device Tests. The operational tests
are displayed. Use the navigation keys to select one of the tests.
Note:
You need a supervisor passcode to enter the Supervisor menu.
9650-001356-01 Rev. A
ZOLL X Series Service Manual
15
Page 20
CHAPTER 1 MAINTENANCE TESTS
Keypad Test
This test checks all the front panel buttons on the unit to make sure they are working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
LED Test
This test checks all the Light Emitting Diodes in the unit to make sure they are working
correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
Note:
16
The speed of the flash varies by color.
9650-001356-01 Rev. A
Page 21
Operational Tests
LCD Test
This test checks the colors of the Liquid Crystal Display on the unit.
Follow the instructions and use the navigation keys to move around the display and make
selections.
RFU Indicator Test
This test checks the Ready For Use Indicator on the unit to make sure it is working correctly.
Follow the instructions shown on the display.
9650-001356-01 Rev. A
ZOLL X Series Service Manual
17
Page 22
CHAPTER 1 MAINTENANCE TESTS
Audio Test
This test checks the audio tones on the unit to make sure they are working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
Printer Test
This test checks the printer to make sure it is working correctly.
Follow the instructions shown on the display.
18
9650-001356-01 Rev. A
Page 23
Operational Tests
Battery Test
This test checks the battery to make sure it is working correctly.
Follow the instructions on the display.
Fan Test
This test checks the fan in the unit to make sure it is working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
9650-001356-01 Rev. A
ZOLL X Series Service Manual
19
Page 24
CHAPTER 1 MAINTENANCE TESTS
NIBP Test
This test allows the operator to perform various checks of the NIBP module to make sure it is
working correctly.
Follow the instructions in the NIBP Calibration Check procedure and other procedures outlined
in this manual. Use the navigation keys to move around the display and make selections as
required to execute the procedures.
USB Test (Optional)
Note:
Before starting this test, contact the ZOLL Technical Service Department to obtain a
test file.
This test allows the operator to perform the USB test to make sure it is working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
When these tests are completed, use the navigation keys to exit out of the Service menu.
20
9650-001356-01 Rev. A
Page 25
Functional Verification Tests
Functional Verification Tests
Before You Begin the Functional Verification Tests
• Assemble the tools listed in following section, “Equipment You Need to Perform the
•
•
•
•
•
Functional Verification Tests.”
Keep an extra fully charged X Series battery available.
Schedule an hour to conduct the entire maintenance test.
Perform the tests in the order presented.
Perform all the steps of each test procedure.
Complete all the steps of the procedure before evaluating the test results.
Equipment You Need to Perform the Functional Verification Tests
For testing purposes, you can substitute an equivalent device.
• Fluke Impulse 7000DP Analyzer, or equivalent, with appropriate interface adapter
(Fluke Biomedical BP Pump/Dynatech Nevada Inc. part number 3010-0378)
IBP/Temp Simulator, or equivalent
NIBP Simulator
Single or dual lumen NIBP hose
Adult NIBP cuff and a rigid cylinder to wrap it around
5% CO2 calibration gas cylinder with CO2 sample line (kit available from Scott Medical
Products, PN 4653ORF-BD)
• ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201)
•
•
•
•
•
Accessories You Need for the X Series Options Functional Verification Tests
• 3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is
•
•
•
•
•
•
•
•
installed.)
NIBP cuff and hose
IBP cable
SpO2 cable and sensor
EtCO2 cable and sensor
Paddles
Printer paper
Battery
Auxiliary power source
9650-001356-01 Rev. A
ZOLL X Series Service Manual
21