CentriMag Circulatory Support Systems Alarms and Alerts Quick Reference Guide Rev 00 Dec 2019

Quick Reference Guide

2 Pages

CentriMagTM Circulatory Support System Alarms and Alerts Quick Reference Guide ID  S1  S2  M1  M2  M3  M4  M5  Alarm/ Alert  Text Message  System Status & OperatorResponse  Alarm  POWER ON TEST FAIL  The Pump will not start. An audible alarm will sound, which cannot be muted. Switch the Console OFF and ON again. If the alarm reappears, switch to the backup Console, Motor and Flow Probe, record the alarm message and contact your local Abbott Medical representative.  Alarm  SYSTEM FAULT (Run-Time System Failure)  The Pump will stop. An audible alarm will sound, which cannot be muted. Clamp the return tubing and switch to the backup Console, Motor and Flow Probe according the procedure described in Section 2. Resume support. Record the alarm message and contact your local Abbott Medical representative.  MOTOR STOPPED  The Pump will stop. An audible alarm will sound, which can be muted for 60 seconds.Clamp the return tubing and switch to the backup Console, Motor and Flow Probe according the procedure described in Section 2. Resume support. Record the alarm message and contact your local Abbott Medical representative.  MOTOR DISCONNECTED  The Pump will stop. An audible alarm will sound, which can be muted for 60 seconds. During setup of the System: Press the alarm acknowledge button and check that the Motor connector is fully inserted into the back of the Console. During support: Press the alarm acknowledge button and check that the connector of the Motor is fully inserted into the back of the Console. Resume support. If the visual alarm message does not disappear, clamp the return tubing and switch to the backup Console, Motor and Flow Probe according the procedure described in Section 2. Resume support. Record the alarm message and contact your local Abbott Medical representative.  Alarm  Alarm  Alarm  Alarm  Alert  PUMP NOT INSERTED  MOTOR ALARM  SET PUMP SPEED NOT REACHED  1  ID  B1  B2  F1  S3  F2  Alarm/ Alert  Text Message  System Status & OperatorResponse  Alert  BATTERY MODULE FAIL  The Console battery will not function. An audible alarm will sound. Switch to the backup Console, Motor and Flow Probe according to the procedure described in Section 2. Resume support. Record the alert message and contact your Abbott Medical representative.  Alert  BATTERY BELOW MINIMUM  The Pump will stop after a very short time. Plug the Console into a AC power outlet to charge the battery. If no AC outlet is available, switch to the backup Console, Motor and Flow Probe according to the procedure described in Section 2. Resume support.  Alert  FLOW PROBE DISCONNECTED  Check the Flow Probe connection on back of the Console. If necessary, reconnect the Flow Probe connector to the back of the Console. Press the alarm acknowledge button. Switch to the backup Flow Probe, if the alert message repeats.  SYSTEM ALERT  Press the alarm acknowledge button, if the message does not disappear, clamp the return tubing, stop the pump and switch to the backup Console, Motor and Flow Probe according the procedure described in Section 2. Resume support. Record the alert message and contact your local Abbott Medical representative.  Alert  Alert  System will not start. An audible alarm will sound, which can be muted for 60 seconds. Press the alarm acknowledge button. Insert or reinsert the pump and secure it with the locking screw. If the alarm repeats, switch to backup Console, Motor and Flow Probe according the procedure described in Section 2. An audible alarm will sound and the System will continue to operate. Press the alarm acknowledge button, if the visual alarm message does not disappear, clamp the return tubing, stop the pump and switch to the backup Console, Motor and Flow Probe according the procedure described in Section 2. Resume support. Record the alarm message and contact your local Abbott Medical representative. Check the Pump flow. If pump flow is satisfactory, reduce set speed while insuring that flow is maintained. Press the alarm acknowledge button. If the alert repeats, clamp the return tubing, stop the pump and switch to the backup Console, Motor and Flow Probe according the procedure described in Section 2. Resume support. Record the alert message and contact your Abbott Medical representative. If the pump flow is not satisfactory, clamp the return tubing and switch to the backup Console, Motor and Flow Probe according to the procedure described in Section 2. Resume support. Record the alert message and contact Abbott Medical representative.  RESPONSE TO SYSTEM ALARMS OR ALERTS  When a System alarm or alert condition occurs, an alarm or alert tone sounds and text message appears on the display of the 2nd Generation CentriMag Primary Console. Most alarms and alerts require some action on the user’s part to correct the cause. The following table can be used to determine how to respond to an alarm or alert condition.  F3  F4  P1  P2  Alert  FLOW SIGNAL INTERRUPTED (Flow rate sensor error)  FLOW BELOW MINIMUM (Low Flow)  Manually disconnect, reposition and reconnect the Flow Probe transducer to the tubing. Press the alarm acknowledge button. Switch to the backup Flow Probe, if the alert message repeats. If problem still persists after switching to the backup Flow Probe, stop the pump and switch to a backup Console. Follow the instructions described in Section 2. Resume support. Record the alert message and contact your Abbott Medical representative.  Check for physiologic cause or circuit obstruction. Check minimum flow set point. Do not increase RPM without confirming that adequate blood volume is available. Common cause of this alert is inadequate blood volume at the drainage cannula site for the desired pump flow.  FLOW ABOVE MAXIMUM  Reduce Pump speed and check for cause.  Alert  PRESSURE 1 DISCONNECTED  Check the electrical connections on the pressure 1 transducer and recalibrate. If the problem persists disconnect, reconnect, and recalibrate the transducer. Consider changing the transducer and cable if the problem persists.  Alert  PRESSURE 2 DISCONNECTED  Check the electrical connections on the pressure 2 transducer and recalibrate. If the problem persists disconnect, reconnect, and recalibrate the transducer. Consider changing the transducer and cable if the problem persists.  Alert  Text Message  System Status & OperatorResponse  Alert  PRESSURE SYSTEM FAIL  The pressure monitoring system will not function. If pressure monitoring is needed then change to the backup Console, Motor and Flow Probe. Switch to the backup System according the procedure described in Section 2. Resume support. Record the alert message and contact your Abbott Medical representative.  Alert  PRESSURE 1 BELOW MINIMUM  Check for and resolve a physiological or mechanical cause. Ensure appropriate pressure alarms are set. Consider recalibrating the transducer, if alert cannot be explained by conventional troubleshooting.  Alert  PRESSURE 2 BELOW MINIMUM  Check for and resolve a physiological or mechanical cause. Ensure appropriate pressure alarms are set. Consider recalibrating the transducer, if alert cannot be explained by conventional troubleshooting.  Alert  PRESSURE 1 ABOVE MAXIMUM  Check for and resolve a physiological or mechanical cause. Ensure appropriate pressure alarms are set. Consider recalibrating the transducer, if alert cannot be explained by conventional troubleshooting. Consider reducing RPM to reduce the pressure if appropriate.  P7  Alert  PRESSURE 2 ABOVE MAXIMUM  Check for and resolve a physiological or mechanical cause. Ensure appropriate pressure alarms are set. Consider recalibrating the transducer, if alert cannot be explained by conventional troubleshooting. Consider reducing RPM to reduce the pressure if appropriate.  M6  Alert  MOTOR OVER TEMP  Switch to backup Console, Motor and Flow Probe according to the procedure described in Section 2. Verify that the backup Motor stands free and is not covered (e.g. blankets).  B3  Alert  BATTERY CHARGER FAIL  Press the alarm acknowledge button. If the alert message repeats, switch to backup System as described in Section 2. If this alert is associated with BATTERY MODULE FAIL then carry out the procedure associated with that alarm.  Alert  BATTERY MAINTENANCE REQUIRED  Do not use the Console. Perform battery maintenance according to the Battery Maintenance Procedure listed in the CentriMag Circulatory Support System Operation Manual.  Alert  LOW BATTERY  Plug the Console into a AC power outlet to charge the battery. If no AC outlet is available, switch to the backup Console, Motor and Flow Probe according to the procedure described in Section 2. Resume support.  Alert  ON BATTERY  Verify that the user wants the Console to be on battery. If so, carefully monitor the status of the battery charge indicator, while using the System on battery. Reconnect to AC-outlet, as soon as possible.  ID  P3  P4  P5  P6  B4  B5  B6  Alarm/ Alert  PL-0370 Rev 00 / December 2019 / Page 1
File Type: PDF
File Size: 400 KB
File Name: centrimag (1).pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions