fabian Therapy evolution Test protocol for service & maintenance V08
Test protocol for service & maintenance
5 Pages
Preview
Page 1
Test protocol for service & maintenance Technik & Support
Process owner: G.Focic
Manufacturer: Acutronic Medical Systems AG Fabrik im Schiffli 8816 Hirzel / Switzerland Type: Modell:
SN prefix
fabian Therapy evolution
AT-
Customer:
Neonatal / Infant Respirator fabian Therapy evolution
fabian SN Inventory. No. Use in combination with “TS-AA-02e V08 TI fabian Therapy evolution”
1) Configuration
Working hours
Firmware Versions: FW / SW Version
Module
Device ID
Device
GUI Software Version
Battery
Hours
Monitor PIC Conductor PIC Blender PIC Accu PIC Kernel Version Language Version
Test equipment Tester used for
Type / Model
Serial no.
Version
Last cal.
Next cal.
Pressure Flow Oxygenation SpO2 Electrical Safety
TS-FB-12e V08 TPS fabian Therapy evolution
Page 1 von 5
Test protocol for service & maintenance Technik & Support
Process owner: G.Focic
fabian Therapy evolution
SN prefix AT-
2) Visual Inspection fabian is switched off
pass / fail
General device condition Housing without cracks or mechanical damage All cables without damages Check all connectors
3) Start-up / Self-test Connect fabian to mains and gas supply. Connect Test circuit and Test lung Switch on fabian
pass / fail
LCD screen backlight. LCD pixel. Up to 3 pixel failures are acceptable. Selftest ok Alarm sound Warning LED Red flashes two times at start up Disconnect mains cable Alarm sound on (Power Fail Alarm function) Orange LED on (Battery mode on) Connect mains cable Alarm sound off (Power Fail Alarm Off function) Orange LED off (Battery mode off)
4) Blender test Connect fabian to gas supply. For these test you have to work with the following gas standards:
Gas standardisation: Gas type:
20/1013 Air
Switch off fabian 4.1
Blender is not leaking. No flow coming out of the inspiration port
4.2
Tightness calibration valves.
TS-FB-12e V08 TPS fabian Therapy evolution
pass / fail
Page 1 von 5
Test protocol for service & maintenance Technik & Support
Process owner: G.Focic
fabian Therapy evolution
SN prefix AT-
5) Calibration Connect Air und O2, power on the device 5.1
Value
pass / fail
Set time and date
Set O2 to 60% and start fabian 5.2
Pressure Air
2400 ± 200
Pressure O2
2400 ± 200
Connect Air / O2 with AIR, restart fabian 5.3
21% calibrate to
9.5 l/min ± 0.1 l/min
100% calibrate to
9.5 l/min ± 0.1 l/min
Power off device and connect Air / O 2 restart device 5.4
P insp. ( p mix.) Sensor 0 mbar
570 ±10
P insp. ( p mix.) Sensor 60 mbar
2400 ±10
Pressure sensor ( prox. ) calibration 0 mbar 5.5
Pressure sensor ( prox. ) calibration 60 mbar
5.6
Touch screen calibration
5.7
Pressure relief valve
450 – 600 mbar
21% 5.8 5.9
100% Flow sensor calibration
optional
Close the device now! 6) O2 / Air blender All tests have to be done with the Gas Standard ATP Activate also “enable O2 comp. in “Service-Menu / Calibration” Connect fabian to gas supply and switch on the device Fastest way is to work with an open tube set. Setting O2 [%]
Reading fabian
Reading Test device
Tolerance
21
± 3 Vol%
40
± 3 Vol%
60
± 3 Vol%
100
± 3 Vol%
TS-FB-12e V08 TPS fabian Therapy evolution
pass / fail
Page 1 von 5
Test protocol for service & maintenance Technik & Support
Process owner: G.Focic
fabian Therapy evolution
SN prefix AT-
7) NCPAP Mode Be sure that the “leakage compensation” is activated This point is in the Ventilation - Menu Mode
CPAP
CPAP
5 mbar
Oxygen
21 Vol.%
Man. Breath
15 mbar
CPAP Settings
Reading fabian Pmean ± 1 mbar
Reading Testdevice Pmean ± 1 mbar
pass / fail
5 10 Manual breath 15 mbar ± 1 mbar
8) DUOPAP Mode Mode
DUOPAP
CPAP
3 mbar
P DUO
8 mbar
Frequenz
10 bpm
I - Time
2.00 sec.
Oxygen
21 Vol.%
Settings
Reading fabian
CPAP [mbar]
P-DUO [mbar]
3
8
5
10
10
15
Ppeak ±1
Pmean
Comparison from fabian to Tester
PEEP ±1
Ppeak ±2
Pmean ±1
pass / fail
PEEP ±1
9) System Alarms What to do
Expected alarm message
pass / fail
Disconnect Inlet pressure O2
Oxygen supply pressure (red blinking Info-bar)
Disconnect Inlet pressure Air
Air supply pressure (red blinking Info-bar)
Disconnect Inspiration Tube
Patient disconnected (red blinking Info-bar)
Disconnect Proximal Tube
Patient disconnected (red blinking Info-bar)
(if module installed) ☐
10) SpO2 functional check
The values that have to be reached are printed on the backside of the MASIMO Tester. Parameter
Tester value
Reading fabian
Tolerance
SpO2
±3%
PR
± 1 bpm
TS-FB-12e V08 TPS fabian Therapy evolution
pass / fail
Page 1 von 5
Test protocol for service & maintenance Technik & Support
Process owner: G.Focic
fabian Therapy evolution
SN prefix AT-
11) Maintenance Set the next maintenance date We recommend to use this function
12) Setting check Checklist
pass / fail
Load hospital settings or factory defaults Check if the Service Password is activated (Service Mode, System) Check if the O2 comp. is enabled
Restart fabian. fabian is working properly
13) Electrical Safety Test Check for the electric safety according to IEC 62353, or your local law. The printed results of the measurements should be attached to the inspection results.
14) Battery calibration Checklist
pass / fail
Battery calibration cycle performed
15) Completion Test results: Checked / Tested by: Date: Next Test: Remarks:
TS-FB-12e V08 TPS fabian Therapy evolution
Page 1 von 5