User Manual
24 Pages
Preview
Page 1
NIOX VERO® Nasal Measurement Mode User Manual
NIOX VERO® Nasal User Manual April 2020 002346-04 English
Remember It is important to adhere to the following specified conditions: •
Ambient temperature: +10°C to +35°C
•
Humidity: 20% to 80% RH (non-condensing)
•
Mobile phones, cordless phones and gas emitting appliances might interfere with the instrument and should therefore be kept away from the instrument. Interference could make it impossible to perform a measurement.
•
Exhaled breath contains water vapour which can condense inside the instrument. When excessively used in a short period, there is a risk for condensation of water inside NIOX VERO®. Normally a maximum of 10 exhalations/hour can be performed with NIOX VERO® during continuous use. However, it is possible to perform 20 exhalations in one hour if the instrument is paused for a minimum of 30 minutes prior to the next session of exhalations. Exhalations include failed and successful measurements.
•
Avoid spilling water or other fluids on the instrument or sensor.
•
Always use a closed case or bag (NIOX VERO® bag recommended) for transportation and storage of NIOX VERO®.
•
It is recommended, after inserting a new sensor, to wait for three hours with the instrument switched on before performing a measurement.
•
Operational life-time for NIOX VERO® Instrument: Maximum 5 years in use or 15 000 measurements or the expiration date stated on the device, whichever comes first. 5 years in use shall be defined as starting from the first NO measurement taken using the device.
•
Operational life-time for NIOX VERO® Sensor: Maximum 12 months after opening package and installed in NIOX VERO® or expiration date as stated on the sensor, whichever comes first.
WARNING! Use of substances containing alcohol close to the NIOX VERO® instrument may cause erroneous measurement results. DO NOT clean the instrument or handle with alcohol or any spray or wipe containing alcohol! Do not use substances containing alcohol on or close to the NIOX VERO® instrument. This includes any cleaning agents used to clean the facility, or other equipment in the area, as well as alcohol wipes or sprays used on patients. Patient filters and nasal kits are labeled as single use only. Always use a new patient filter and nasal kit for each patient. Reuse between patients could increase the risk of cross-contamination or cross-infection. The same filter and nasal kit can be reused in one patient for multiple attempts in the same session. CAUTION!: Do not use NIOX VERO® in the proximity of areas where volatile substances such as organic fluids or disinfectants are being used. Special attention should be paid to aerosols and disinfection baths, either open vessels or ultrasonic baths. Refer to the NIOX VERO User Manual for complete review of all cautions, warnings, preventive care and troubleshooting.
Table of contents
1 NIOX VERO® Nasal measurement mode ...3 1.1 Before using NIOX VERO® ... 3 1.2 About this manual... 3 1.3 Compliance ... 3 1.4 Responsible manufacturer and contacts ... 3 1.5 Warnings ... 3 1.6 Intended Use ... 4
6.3 Precision ... 19 6.4 Accuracy ... 20 6.5 Essential performance ... 20
2 Product description ...5 2.1 Nasal accessories and parts ... 5 2.2 NIOX VERO® accessories and parts... 5 2.3 Main View... 6 3 Nasal NO measurements ...6 3.1 Preparation for Nasal (nNO) measurement... 6 3.2 Perform Nasal measurement ... 8 3.3 Expiration against Resistance (ER-nNO) ... 9 3.4 Tidal Breathing (TB-nNO)... 11 3.5 Alert codes and actions ... 11 4 Nasal measurements with NIOX® Panel ...12 4.1 Installation of NIOX® Panel ... 12 4.2 Setup ... 12 4.3 Perform Nasal measurements with NIOX® Panel ... 13 5 Cleaning procedure ...19 6 NIOX VERO® Nasal Technical Specification ...19 6.1 Performance data... 19 6.2 Linearity... 19
NIOX VERO® Nasal NO Measurements User Manual English 002346-04
1
Chapter 1 NIOX VERO® Nasal measurement mode
1
NIOX VERO® Nasal measurement mode
1.1
Before using NIOX VERO®
NIOX VERO® may only be operated as directed in this manual by trained healthcare professionals. Trained status is achieved only after careful reading of this manual. Read the entire instructions for use and make certain that you fully understand the safety information. This is an amendment to the NIOX VERO User Manual. Carefully read the NIOX VERO User Manual and the NIOX VERO® Nasal Measurement Mode User Manual before performing NIOX VERO Nasal measurements.
Symbol
Description
WARNING
Indicates a potentially hazardous situation that, if not avoided, can result in bodily harm or injury.
CAUTION
Indicates a potentially hazardous situation that, if not avoided, can damage a product or system, cause loss of data or harm to business.
Note
1.2
Alerts the reader to important information on the proper use of the product, user expectations, error situations, and actions related to these.
Information in this document is subject to change. Amendments will be made by Circassia as they occur. This User Manual provides instructions on how to operate NIOX VERO® in Nasal Measurement mode. It contains numbered step-by-step instructions with screens and illustrations. Choices within steps are displayed with bullet points.
1.3
Compliance
NIOX VERO® and NIOX® Panel are CE-marked according to In Vitro Diagnostics Device Directive 98/79/EC. NIOX VERO Patient filter and NIOX VERO Nasal kits are CE marked according to Medical Device Regulation NIOX VERO® is RoHS compliant.
1.4
Responsible manufacturer and contacts
Mailing address: Circassia AB, P.O. Box 3006 SE-750 03 Uppsala, Sweden Visiting address: Hansellisgatan13 SE-754 50 Uppsala www.circassia.com www.niox.com
About this manual
NIOX VERO® Nasal Measurement Mode User Manual - English 002346 version 04, April 2020. For instruction on how to view the software version number installed in the instrument, refer to the NIOX VERO User Manual.
002346-04 NIOX VERO® Nasal User Manual English
1.5 •
Warnings
Do not use substances containing alcohol on or close to the NIOX VERO® instrument. This includes any cleaning agents used to clean the facility, or other equipment in the area, as well as alcohol wipes or sprays used on patients.
3
Chapter 1 NIOX VERO® Nasal measurement mode • •
•
• • • • • • • • •
• • •
4
NIOX VERO® should only be operated by healthcare professionals. Operate NIOX VERO® as stated in this manual. Circassia accepts no responsibility for damaged equipment or faulty results, if the equipment is not handled according to this manual. When selecting an accessory for your NIOX VERO® product keep in mind that an accessory not recommended by Circassia may result in loss of performance, damage to your NIOX VERO® product, fire, electric shock, injury or damage to other property. The product warranty does not cover product failure or damage resulting from use with non approved accessories. Circassia takes no responsibility for health and safety problems or other problems caused by the use of accessories not approved by Circassia. NIOX VERO® should not be used adjacent to or stacked with other electronic equipment. Only use the power supply provided. Pull the plug when disconnecting NIOX VERO® from the power outlet. Use only the breathing handle supplied by Circassia. No modification of NIOX VERO® instrument, handle or sensor is allowed. Do not drop the instrument or subject it to strong impact. Do not use a damaged NIOX VERO® instrument or damaged components. Keep the instrument and sensor out of water. Ensure that no liquid is spilled or dropped on the instrument or sensor. Do not heat or dispose of the instrument or sensor in fire. NIOX VERO® and the NO scrubber in the breathing handle contains potassium permanganate. Used or expired instruments and breathing handles should be disposed of as hazardous waste in accordance with local waste disposal regulations. The breathing handle must not be used after expiration date. Patient filters should be used immediately after opening. NIOX VERO® Sensor contains chemicals that could be harmful if
• • • • •
•
• •
swallowed. Be careful when opening the sensor can. The inside of the opening may have sharp edges. Do not touch or clean the white sensor membrane. After inserting a new sensor it is recommended to wait for three hours with the instrument switched on before performing a measurement. Make sure to use the correct measurement mode, otherwise incorrect measurement results might be obtained. Patient filters are labeled for single use only. Always use a new patient filter for each patient. Reuse between patients could increase the risk of cross-contamination or cross-infection. The same patient filter can be reused in one patient for multiple attempts in the same session. Do not use NIOX VERO® in the proximity of areas where volatile substances such as organic fluids or disinfectants are being used. Special attention should be paid to aerosols and disinfection baths, either open vessels or ultrasonic baths. Do not use the instrument in the presence of flammable anesthetic, vapours or liquids. Nasal kit should be used immediately after opening. Nasal kits are labeled for single use only. Always use a new nasal kit for each patient. Reuse between patients could increase the risk of crosscontamination or cross-infection.The same nasal kit can be reused in one patient for multiple attempts in the same session.
1.6
Intended Use
NIOX VERO® measures Nitric Oxide in human breath (Fractional exhaled Nitric Oxide, FeNO) and Nasal Nitric Oxide (nNO) in the aspirated air from the nasal cavity.
002346-04 NIOX VERO® Nasal User Manual English
Chapter 2 Product description FeNO FeNO is increased in some airway inflammatory processes such as asthma and decreases in response to anti-inflammatory treatment. FeNO measurements with NIOX VERO are quantitative, non-invasive, simple and safe and should be used as part of regular assessment and monitoring of patients with these conditions. NIOX VERO is suitable for patients age 4 and above for FeNO measurements. As measurement requires patient cooperation, some children below the age of 7 may require additional coaching and encouragement. NIOX VERO should be used as directed in the NIOX VERO User Manual. CAUTION! NIOX VERO for FeNO measurement can be operated with two different exhalation times, 10 seconds and 6 seconds. The 10 second test is the preferred mode. For children who are not able to perform the 10 second test, the 6 second test is an alternative. The 6 second test should be used with caution in patients over the age of 10 years. It should not be used in adult patients. Incorrect use of the 6 second exhalation mode may result in falsely low FeNO values, which can lead to incorrect clinical decisions. nNO
2 2.1
Product description Nasal accessories and parts
•
(A) NIOX VERO® Nasal Kit, Article No 12-1065 Pediatric, 12-1045 Adult, (B) NIOX VERO® Nasal Restrictor, Article No 12-1033 (for use together with NIOX VERO® Patient filter)
2.2
NIOX VERO® accessories and parts
Nasal Nitric Oxide has been shown to be decreased in patients with Primary Ciliary Dyskinesia (PCD), and measurement of nNO can assist in the identification of cases of PCD according to ERS guidelines1. Measurement of nNO with the NIOX VERO Nasal Measurement Mode is noninvasive, simple, safe and repeatable in patients age 5 and above when measured according to the NIOX VERO Nasal Measurement Mode User Manual. Suspected cases of PCD following screening with nNO should be confirmed according to published recommendations for PCD diagnosis and management. 1. Lucas JS, Barbato A, Collins SA, et al. European Respiratory Society guidelines for
the diagnosis of primary ciliary dyskinesia. Eur Respir J 2017; 49: 1601090
002346-04 NIOX VERO® Nasal User Manual English
(C) NIOX VERO® Breathing Handle, Article No 12-1010 (supplied with NIOX VERO®) (D) NIOX VERO® Patient Filter, Article No 12-1018 (supplied separately)
5
Chapter 3 Nasal NO measurements
2.3
Main View
3
Nasal NO measurements
Nasal measurements can be performed either using the Tidal Breathing (TB-nNO) method or by the velum closed with Expiration against Resistance (ER-nNO) method. Nasal measurements can be performed either on the left or the right nostril. WARNING! Always use a new nasal kit and patient filter (when using ERnNO method) for each patient. Reuse between patients could increase the risk of cross-contamination or cross-infection. The same filter and nasal kit can be reused in one patient for multiple attempts in the same session.
3.1
Preparation for Nasal (nNO) measurement
Verify proper preparations before performing a nasal measurement with NIOX VERO®. A basic preventive inspection is recommended before each use, refer to the NIOX VERO User Manual. (A) Status bar, (B) Instructive demonstration, (C) Patient ID, (D) Main menu, (E) Start measurement button
2.3.1
Main menu
Note:
Measurements can be performed without NIOX Panel software, however, if done, there is no graph available. Only the value will show on the display screen.
3.1.1
Attach the Nasal Kit to NIOX VERO instrument
1.
Place the device on its side on a level secure surface.
(a) Measurement mode 10s/6s, (b) Nasal measurement mode, (c) Demo, (d) Patient ID entry, (e) Settings
6
002346-04 NIOX VERO® Nasal User Manual English
Chapter 3 Nasal NO measurements 2.
Disconnect the breathing handle from the instrument by pushing the socket into the device and gently pull out the tube.
002346-04 NIOX VERO® Nasal User Manual English
3.
Take the nasal kit, remove the cap from the end of the tube and push it into the breathing handle port slowly until the triangle is no longer visible.
Note:
There is lubricant on the end of the tube for easier attachment to the NIOX VERO.
Note:
The triangle should not be visible when assembled correctly.
7
Chapter 3 Nasal NO measurements Note:
Performing a nasal measurement without using the NIOX Panel (see section 4.3, Perform Nasal measurements with NIOX® Panel) will not produce a graphical result, it will only give the numerical result of nasal nitric oxide measurement on the NIOX VERO screen.
CAUTION! Make sure to use the correct size olives, the seal between the olive and nostril must be airtight to ensure correct measurement result.
(A), Nasal Olive, (B) Nasal Filter, (C) Nasal Tube, (D) Cap
3.2
Perform Nasal measurement
Nasal measurements can be performed with either Expiration against Resistance (ER-nNO) method (see section 3.3) or Tidal Breathing (TB-nNO) method (see section 3.4). Note:
Make sure patient does not inhale by nose during the nasal measurement.
Note:
The patient should not perform nasal NO measurements if there is evidence of blood in the nostrils.
Note:
Prior to performing nasal NO measurements, the patient should be comfortably seated and have thoroughly cleared the nasal passage.
Note:
There are two phases of the nasal measurements. A measurement period where the instrument aspirates for 30 seconds, and an analysis period where the instrument analyses the sample for 30 seconds.
8
(A) Nasal Olive Adult, (B) Nasal Olive Pediatric
002346-04 NIOX VERO® Nasal User Manual English
Chapter 3 Nasal NO measurements CAUTION! Do not insert the nasal olive inside the nostril. The larger part of the nasal olive should remain outside the nostril.
3.3
Expiration against Resistance (ER-nNO)
ER-nNO Nasal measurements shall be performed while exhaling through the mouth towards a restrictor (breathing handle), this generates a pressure that makes sure the velum is closed. The NIOX VERO Breathing Handle with a patient filter and the NIOX VERO Restrictor is used as the restrictor in this breathing method. The breathing handle is NOT connected to the NIOX VERO when used in this manner. 1.
Open a new patient filter and insert a new nasal restrictor into the Patient filter, so that it rests on top of the white filter pad.
CAUTION! Ensure that the sampling hole in the nasal olive is aligned with the nasal passage and not blocked. A blocked sampling hole will influence the measurement result. Note:
If an error occurs see “Alert codes and actions” on page 11.
002346-04 NIOX VERO® Nasal User Manual English
Note:
Do not use the patient filter without the NIOX VERO Nasal Restrictor.
Note:
Use the patient filter and the nasal kit directly after it has been opened.
9
Chapter 3 Nasal NO measurements 2.
Attach the patient filter to the breathing handle until it clicks into place.
4.
Start nasal measurement by selecting the nasal icon (A)
5.
Enter patient ID (optional) and then start the nasal measurement by pressing the start measurement button (B)
Note:
3.
Give the breathing handle to the patient, put the olive to the selected nostril and instruct the patient to inhale to total lung capacity and to begin exhaling into the patient filter once their lips are tightly sealed around it. The patient will have to exhale steadily and continuously for 30 seconds.
6.
10
If measurement is started before patient is ready, there is an option to abort measurement, it is not recorded as a measurement on sensor or instrument.
The instrument will aspirate for 30 seconds, and then will start to analyse the sample. After the aspiration is complete, remove nasal olive from the nostril and wait for the result.
002346-04 NIOX VERO® Nasal User Manual English
Chapter 3 Nasal NO measurements
3.4
Tidal Breathing (TB-nNO)
3.
TB-nNO measurements shall be performed while breathing normally through an open mouth. 1.
Enter patient ID (optional) and then start the nasal measurement by pressing the start measurement button (B)
Put the olive to the selected nostril and instruct the patient to slowly inhale and exhale through the mouth. Note:
4.
If the measurement is started before patient is ready, there is an option to abort measurement, it is not recorded as a measurement on sensor or instrument.
The instrument will aspirate for 30 seconds and then will start to analyse the sample. After the aspiration is complete, remove nasal olive from the nostril and wait for the result.
3.5
Alert codes and actions
User alerts 2.
Start Nasal measurement by selecting the Nasal icon (A)
Screen
Action
A28
Internal hardware error recoverable Check that the sensor, battery and lid is in its correct position, also make sure that the tube is not folded. When finished restart the instrument.
A29
Analysis failure Ambient measurement failure. Click the OK button and obtain a new measurement.
For additional alert codes, refer to the NIOX VERO User Manual.
002346-04 NIOX VERO® Nasal User Manual English
11
Chapter 4 Nasal measurements with NIOX® Panel
4
Nasal measurements with NIOX® Panel
7. Follow the instructions and install the program. Note: Last step of installation “Removing backup files” takes a few minutes.
Using NIOX Panel provides the ability to see the nNO measurement graphically, produce the file and print the results.
8. When the installation is complete, click Close. The application is now available on the start menu.
NIOX® Panel System requirements • Windows® 7, Windows® 8 (RT versions excluded) or Windows® 10 • .NET Framework 4.5 • 1 GHz or faster processor • 256 MB RAM (512 MB RAM recommended) • 250 MB of video graphics RAM • 250 MB of available hard-disc space • 1024x768 screen resolution • Generic Microsoft Bluetooth driver*
4.2 1.
Setup Turn on the PC and monitor.
2.
Turn on the instrument.
3.
Select the Start or the Windows button normally found in the left lower hand corner of the monitor. Select NIOX® Panel from the program list. Plug the USB cable into the USB port on the NIOX VERO® and connect it to the USB port on the PC or connect by Bluetooth. This icon is shown on the display to indicate that a connection is established and NIOX VERO® is running remote controlled.
4. 5.
* Needed for Bluetooth communication
4.1
Installation of NIOX® Panel
The NIOX® Panel software is supplied on a USB storage device. 1.
Insert the USB storage device in the computer’s USB port.
2. 3.
Select the file named setup.exe. If .NET Framework 4.5, VC++ 2013 or SQL Server Compact is not installed, an installation wizard for each of the programs opens, one at the time. Select to accept license agreement for the programs. Follow the instructions and wait for the programs to install. The Installation wizard for NIOX® Panel Nasal opens.
4. 5. 6.
12
002346-04 NIOX VERO® Nasal User Manual English
Chapter 4 Nasal measurements with NIOX® Panel 6.
The NIOX® Panel application opens and it is possible to perform nasal NO measurements.
4.3
Perform Nasal measurements with NIOX® Panel
WARNING! Always use a new nasal kit and patient filter (when using ER-nNO method) for each patient. Reuse between patients could increase the risk of cross contamination or cross-infection. The same filter and nasal kit can be reused in one patient for multiple attempts in the same session. Note: 1.
2.
Select nasal button. (B)
3.
Select which expiration method to use, ER-nNO (see section 4.3.1) or TBnNO (see section 4.3.2) and which nostril, left or right, patient age is optional.
4.
Select which file format to save measurement curve, .xlsx, pdf or both. If the pdf file is chosen, an excel file is saved in the NIOX Nasal Measurements folder on the computers Desktop.
The buttons, symbols and views are similar on NIOX® Panel and on NIOX VERO®.
To enter patient ID, select patient id button. (A)
002346-04 NIOX VERO® Nasal User Manual English
13
Chapter 4 Nasal measurements with NIOX® Panel
4.3.1
Expiration against Resistance (ER-nNO)
ER-nNO Nasal measurements shall be performed while exhaling through the mouth towards a restrictor (breathing handle), this generates a pressure that makes sure the velum is closed. The NIOX VERO Breathing Handle with a patient filter and the NIOX VERO Restrictor is used as the restrictor in this breathing method. The breathing handle is NOT connected to the NIOX VERO when used in this manner. 1.
Open a new patient filter and insert a new nasal restrictor into the Patient Filter, so that it rests on top of the white filter pad.
2.
Attach the patient filter to the breathing handle until it clicks into place
3.
Give the breathing handle to the patient, put the olive to the selected nostril and instruct the patient to inhale to total lung capacity and to begin exhaling into the patient filter once their lips are tightly sealed around it. The patient will have to exhale steadily and continuously for 30 seconds.
4.
Select start measurement button (A) to start a ER-nNO Nasal measurement with NIOX Panel.
Note:
If measurement is started before patient is ready, there is an option to abort measurement, it is not recorded as a measurement on sensor or instrument.
Note:
The y-axis does not show a nitric oxide value during the measurement and analysis periods. This is due to the entire sample having to cycle through the sensor (delayed analysis).
14
5.
The instrument will aspirate for 30 seconds and then will start to analyse the sample. After the aspiration is complete, remove nasal olive from the nostril and wait for the result.
Note:
By selecting the abort button (A) the measurement is cancelled. This can be pressed in the measurement or analysis period to abort the measurement.
002346-04 NIOX VERO® Nasal User Manual English
Chapter 4 Nasal measurements with NIOX® Panel Note:
6.
When selecting the OK button (B) an option to save the file will be shown.
Note:
The measurement result will always be saved in the instrument memory.
002346-04 NIOX VERO® Nasal User Manual English
7.
If the option “No” is selected the file won’t be saved.
A save dialogue will be shown and an option to view the file.
Note:
The Excel file has options to adjust width of the calculation area, or change the range of seconds used to calculate the measurement result.
Note:
The pdf-file shows the graph.
15
Chapter 4 Nasal measurements with NIOX® Panel
(A) Complete nNO curve data • • •
Blue line of graph 30 second aspiration measurement 30 second baseline measurement
(B) nNO calculation Range • •
Orange section of graph nNO as calculated by NIOX VERO using clinically validated method
(C) Adjustable Calculation Area •
Grey section of graph
(D) nNO Value and Measurement information • • •
16
nNO in ppb Measurement Method (TB or ER) Nostril (L or R)
(E) User defined nNO Value • •
Can change the adjustable calculation area Output is displayed in ‘red’ text
(F) Patient Information • • • •
Patient ID Age Aspiration time (always 30sec) Date of Measurement
002346-04 NIOX VERO® Nasal User Manual English
Chapter 4 Nasal measurements with NIOX® Panel
4.3.2
Tidal Breathing (TB-nNO)
TB-nNO measurements shall be performed while breathing normally through an open mouth. 1.
Before selecting the start measurement button make sure to instruct the patient to slowly inhale and exhale though the mouth.
2.
Select start measurement button measurement with NIOX Panel.
to start a TB-nNO Nasal
Note:
If measurement is started before patient is ready, there is an option to abort measurement, it is not recorded as a measurement on sensor or instrument.
Note:
The y-axis does not show a nitric oxide value during the measurement and analysis periods. This is due to the entire sample having to cycle through the sensor (delayed analysis).
3.
4.
When selecting the OK button (B) an option to save the file will be shown.
5.
A save dialogue will be shown and an option to view the file.
The instrument will aspirate for 30 seconds and then will start to analyse the sample. After the aspiration is complete, remove nasal olive from the nostril and wait for the result.
Note:
By selecting the abort button (A) the measurement is cancelled. This can be pressed in the measurement or analysis period to abort the measurement.
002346-04 NIOX VERO® Nasal User Manual English
17
Chapter 4 Nasal measurements with NIOX® Panel Note: Note:
The measurement result will always be saved in the instrument memory. The pdf-file shows the graph.
(D) nNO Value and Measurement information • • •
nNO in ppb Measurement Method (TB or ER) Nostril (L or R)
(A) Complete nNO curve data • • •
Blue line of graph 30 second aspiration measurement 30 second baseline measurement
(B) nNO calculation Range • •
Orange section of graph nNO as calculated by VERO using clinically validated method
(C) Adjustable Calculation Area •
18
Grey section of graph
(E) User defined nNO Value • •
Can change the adjustable calculation area Output is displayed in ‘red’ text
(F) Patient Information • • • •
Patient ID Age Aspiration time (always 30sec) Date of Measurement
002346-04 NIOX VERO® Nasal User Manual English
Chapter 5 Cleaning procedure
5
Cleaning procedure
WARNING! DO NOT clean the instrument or handle with products containing alcohol. This includes sprays or wipes containing alcohol! WARNING! DO NOT clean area immediately surrounding the NIOX VERO® with products containing alcohol. This includes sprays or wipes containing alcohol. For cleaning of the instrument see NIOX VERO User Manual. Note: • •
6
NIOX VERO® Nasal Technical Specification
6.0.1
Electromagnetic emissions
CAUTION! This equipment has been designed and tested to CISPR 11 Class A. In a domestic environment it may cause radio interference, in which case, you may need to take measures to mitigate the interference. For guidance and manufacturer’s declaration - electromagnetic emissions see www.niox.com
6.1
Performance data
Measurement range: 5 to 2000 ppb Lowest detection limit: 5 ppb Sampling flow rates: 5 ml/s
NIOX Nasal Kit are disposable items for single patient use only.
If the any part of the nasal kit appears clogged, replace the nasal kit The NIOX Nasal Kit is for single patient use and do not need any cleaning.
6.2
Linearity
Squared correlation coefficient r2 ≥ 0.998
6.3
Precision
< 3ppb of measured value for values < 30 ppb, < 10% of measured value for values ≥ 30 ppb. Expressed as one standard deviation for replicate measurements with the same instrument, using a certified gas concentration of nitric oxide reference standard.
002346-04 NIOX VERO® Nasal User Manual English
19