Alaris
Alaris GP Technical Service Manual Issue 2 Aug 2019
Technical Service Manual
84 Pages
Preview
Page 1
Alaris GP Volumetric Pump TM
Technical Service Manual
Alaris™ GP Volumetric Pump
Contents General Information... 5 Introduction... 5 Product Familiarity... 5 Purpose of this Manual... 5 Conventions Used in this Manual... 6 General Precautions... 6 Features of the Pump... 7 Controls and indicators... 8
Configuration and Calibration... 9 Service Mode... 9 Entering Service Mode... 9 Factory Defaults... 10 Service Configuration... 12 Near End Of Infusion (NEOI) Warning Configuration... 16 Data Set Transfer... 17 Calibration... 19
Preventative Maintenance... 23 Preventative Maintenance... 23 Visual Inspection... 23 Recommended Cleaning and Storage... 24 Updates... 26 Upgrading firmware... 26 Software Upload... 27 Calibration, Performance Verification and Testing... 29 Lever Release V-Seal... 30 Battery Test and Replacement... 31 Replace the Main Battery... 31 Test Verification/PVP (Service Mode)... 32 PVP Work Flow... 32 Alarms Functionality... 33 User Interface... 33 Power Supplies... 35 Sensor Tests... 36 COMMS Test... 39 Occlusion Test... 41 Occlusion Test (Optional)... 41 Volumetric Accuracy... 42 Performance Verification Procedure... 43
Troubleshooting... 44 Log Downloads... 44 Alaris™ GP Volumetric Pumps (with Plus software)... 44 Alaris™ GP Volumetric Pumps (Classic)... 44 View Downloaded Event Log... 45 Introduction... 47 Software Fault Codes... 47 BDTM00001 Issue 2
2 / 84
Alaris™ GP Volumetric Pump
General Fault Diagnosis... 49 Exception Error Handling... 49
Circuit Descriptions... 50 Functional Module Block Diagram... 50 Module Overview Functional Description... 51 Air In Line(AIL)/Safety Clamp Housing... 51 Battery Pack... 51 Comms PCB and IrDA Flexible PCB... 51 Control PCB... 51 Display PCB... 52 Door Detect Flexible PCB... 52 GP Pressure PCB... 52 IEC AC Power Inlet... 52 Interface PCB... 52 Membrane Keypad... 53 Motor... 53 Motor Encoder PCB... 53 Power Supply... 53
Corrective Maintenance... 54 Torque Guide... 55 Access to Pump... 56 Rear Case and Subassemblies... 58 Power Supply Unit (PSU) and Speaker... 58 AC power inlet, IrDA PCB, PE stud and magnet... 59 Pole Clamp... 60 Rail Cam... 61 RS232 Connector and Comms PCB... 62 Handle... 63 Front Case and Subassemblies... 64 Door... 64 Chassis Assembly... 66 Chassis Assembly Breakdown... 67 Control PCB and Interface PCB... 68 Display PCB... 69 AIL/Safety Clamp Housing, Top Retainer, Door Sensor Flexible Circuit... 70 Keypad and Labels... 71
Appendix... 73 Electromagnetic Compatibility... 73 Guidance and Manufacturer’s Declaration... 73 Guidance and Manufacturer’s Declaration... 74 Guidance and Manufacturer’s Declaration... 75 Recommended Separation Distances... 76 Disposal... 77 Information on Disposal for Users of Waste Electrical and Electronic Equipment... 77 Information on Disposal in Countries outside the European Union... 77 Battery Removal... 77 Spare Parts Listing... 78 Door Labels... 78 BDTM00001 Issue 2
3 / 84
Alaris™ GP Volumetric Pump
Individual Components... 78 Keypads... 78 Label Kits... 78 Software... 79 Spare Parts Kits... 79 Test Equipment... 80 Contact Us... 81 Customer Service Information... 81 Document History... 82 Software Upgrade Record... 83
BDTM00001 Issue 2
4 / 84
Alaris™ GP Volumetric Pump General Information
1 General Information Introduction
The Alaris™ GP Volumetric Pump series (hereinafter referred to as Pump) are small lightweight volumetric infusion pumps that provide accurate and reliable infusions over a range of rates. The Pump is designed to meet the infusion requirements as specified in the Directions For Use (DFU) for all hospital departments including general wards, critical and intensive care, operating rooms and accident and emergency rooms. This Pump is suitable for use by appropriately trained clinicians or nurses. This Pump can be used for intravenous infusion modes, supporting fluid and drug therapy, blood transfusions and parenteral feeding.
Product Familiarity Ensure that you are fully familiar with the Pump by carefully studying the Directions for Use prior to operation and prior to attempting any repairs or servicing. As part of continuous improvement, product enhancements and changes are introduced from time to time.
Purpose of this Manual This Technical Service Manual (TSM) shows how to set up, test and maintain the following Alaris™ GP Plus Volumetric Pump models: Pump
Model number
Alaris™ GP Volumetric Pump
80063UN01 80263UN01
Alaris™ GP Guardrails™ Volumetric Pump
80263UN01-G
Alaris™ GP Volumetric Pump (with Plus software)
9002TIG01 9002MED01 9002TIG03
Alaris™ GP Guardrails™ Volumetric Pump (with Plus software)
9002TIG01-G 9002MED01-G 9002TIG03-G
It is intended for use by personnel experienced in medical equipment testing and maintenance procedures. This manual has been prepared for use by qualified service personnel only. BD recommends that all servicing should only be carried out by BD certified service personnel.
w
It is important to ensure that you only refer to the most recent version of the Directions for Use and Technical Service Manual for your BD products. These documents are referenced on www.bd.com. Copies can be obtained by contacting your local BD representative.
BDTM00001 Issue 2
5 / 84
Alaris™ GP Volumetric Pump General Information
Conventions Used in this Manual Bold
Used for Display names, self-test codes, controls and indicators referenced in this manual, for example, Battery Indicator, access code 212, ON/OFF button.
‘Single quotes’
Used to indicate cross-references made to another section of this manual. For example, see Chapter 2, ‘Configuration and Calibration’.
Italics
Used to refer to other documents or manuals. For example, refer to the relevant Directions For Use (DFU) for further information. Also used to define custom terminology specific to a manual e.g. The Alaris™ CC Plus Guardrails™ Syringe Pump (hereinafter referred to as Pump) … Wherever this symbol is shown a Hints and Tips note is found. These notes provide useful advice or information that may help to perform the task more effectively. Wherever this symbol is shown an Update note is found. A typical example is drawing attention to a software upgrade that should be confirmed has been installed. Warning symbol. A warning is a statement that alerts the user to the possibility of injury, death, or other serious adverse reactions associated with the use or misuse of the Pump.
w
Caution symbol. A caution is a statement that alerts the user to the possibility of a problem with the Pump associated with its use or misuse. Such problems may include Pump malfunction, Pump failure, damage to the Pump or damage to other property. The caution statement includes the precaution(s) that should be taken to avoid the hazard.
General Precautions Attention consult accompanying documents: Prior to using this Pump, carefully read the Operating Precautions described in the Directions for Use (DFU).
V
This Pump contains static-sensitive components. Observe strict precautions for the protection of static sensitive components when attempting to repair and service the Pump. An explosion hazard exists if the Pump is used in the presence of flammable anaesthetics. Exercise care to locate the Pump away from any such hazardous sources.
Ala
GP ris®
A M d
Dangerous Voltage. An electrical shock hazard exists if the casing of the Pump is opened or removed. Refer all servicing to qualified service personnel. This Pump is protected against the effects of high energy radio frequency emissions and is designed to be fail safe if extremely high levels of interference are encountered. Should false alarm conditions be encountered, either remove the source of the interference or regulate the infusion by another appropriate means. If the Pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise suspected to have been damaged, remove it from service for inspection by qualified service personnel. When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If the integrity of the external protective conductor in the installation or its arrangement is in doubt, the Pump should be operated from the battery.
BDTM00001 Issue 2
6 / 84
Alaris™ GP Volumetric Pump General Information
Features of the Pump Alarm indicator Alaris®
GP
Door Display
Softkeys Chevrons Controls and Indicators (see overleaf)
Door Lever
Handle Release lever for rotating cam
Flow sensor connector (cover removed for clarity) RS232/Nursecall connector (cover removed for clarity)
Rotating cam to lock onto horizontal rectangular bars.
Folded pole clamp
Medical device interface (MDI)
AC power fuses AC power inlet IR communications port Potential Equalisation (PE) Connector
BDTM00001 Issue 2
7 / 84
Alaris™ GP Volumetric Pump General Information
Controls and indicators Controls Symbol
Description
a b h
ON/OFF button – Press once to switch the Pump on. Press and hold down for approximately three seconds to switch the Pump off. Note: Logs are maintained for power down events including when the Pump is powered down or unexpected power loss. RUN button – Press to start the infusion. The green LED will flash during infusion.
HOLD button – Press to put the infusion on hold. The amber LED will be lit while on hold.
R
MUTE button – Press to silence active alarm for two minutes. The alarm will re-sound after this time. To re-enable the alarm audio press the MUTE button a second time. Note: Attention alarm only: when not in alarm press and hold until four audible beeps are heard to extend the silence period to 15 minutes.
i
PRIME/BOLUS button – Press to access PRIME or BOLUS softkey. Press and hold down softkey to operate. PRIME – primes the infusion set with fluid when setting up an infusion for the first time. • Pump is on hold. • Infusion set is not connected to a patient. • Volume infused (VI) is not added to the total volume infused displayed. BOLUS – delivers fluid or drug at an accelerated rate. • Pump is infusing. • Infusion set is connected to patient. • Volume infused (VI) is added to the total volume infused displayed.
d e f g
OPTION button – Press to access optional features. PRESSURE button – Use this button to display the pumping pressure, and adjust the downstream occlusion alarm limit. CHEVRON keys – Double or single for faster/slower increase or decrease of values shown on display. BLANK SOFTKEYS – Use in conjunction with the prompts shown on the display.
Indicators Symbol
S j
Description AC POWER indicator – When illuminated the Pump is connected to an AC power supply and the battery is being charged. BATTERY indicator – When illuminated the Pump is running on the internal battery. When flashing, the battery power is low with less than 30 minutes of use remaining.
BDTM00001 Issue 2
8 / 84
Alaris™ GP Volumetric Pump Configuration and Calibration
2 Configuration and Calibration Service Mode Entering Service Mode
At no time should Service Mode be entered while the Pump is connected to a patient. Service Mode should only be accessed by qualified and trained personnel. Service Mode can be accessed via a three-digit access code that is entered using the following procedure: 1. Hold down b and turn the Pump on a. 2. Enter the access code 212 using the f keys and the NEXT softkey. 3. When the code shows on screen, press OK to confirm. Select the required option using the f keys and the OK softkey. For the Alaris™ GP Volumetric Pump the options will be as follows: Factory Defaults
Load a default data set. Confirm to perform a cold start.
Configuration
This menu comprises a list of options which are configurable by the user.
Data Set Transfer
Upload a data set to Pump.
Calibration
This menu comprises a list of calibrations which can be performed by the user.
Test Verification/PVP
Performance Verification Procedure Tests.
SERVICE
Factory Defaults Configuration Data Set Transfer Calibration Test Verification/PVP SELECT WITH
QUIT
OK
For the Guardrails™ and Plus software Pumps the options will be as follows: CQI Events Download
For future implementation
Data Set Transfer
Upload a data set to Pump.
Configuration
This menu comprises a list of options which are configurable by the user.
Calibration
This menu comprises a list of calibrations which can be performed by the user.
Test Verification/PVP
Performance Verification Procedure tests.
Factory Defaults
Load a default data set. Confirm to perform a cold start.
SERVICE
CQI Events Download Data Set Transfer Configuration Calibration Test Verification/PVP Factory Defaults SELECT WITH
QUIT
BDTM00001 Issue 2
9 / 84
OK
Alaris™ GP Volumetric Pump Configuration and Calibration
Factory Defaults Select the required option using the f keys and the OK softkey. Default Data Set
Replace the current data set with a default data set.
Cold Start Confirm
Confirm clearing and resetting the data set and calibration data to the factory defaults.
Clear CQI Log File*
Confirm clearing all Pump history and resetting the data set to the factory default.
* Alaris™ GP Guardrails™ Volumetric Pump and Alaris™ GP Guardrails™ Volumetric Pump (with Plus software) only. Note: For Pumps with Plus software the option order will be: Default Data Set Clear CQI Log File Cold Start Confirm
FACTORY DEFAULTS Default Data Set Cold Start Confirm Clear CQI Log File
SELECT WITH QUIT
OK
Default Data Set • Press OK to confirm loading the default data set.
DEFAULT DATA SET WARNING! ****** ****** Restoring factory
default data set will delete the clinically approved data set installed. Delete approved and restore factory default? QUIT OK
Cold Start Confirm • Press OK to confirm performing a cold start.
COLD START CONFIRM WARNING! ****** All data set and ****** cal
data will be reset to default! The pump will require a full calibration before returning to clinical use. QUIT
BDTM00001 Issue 2
10 / 84
OK
Alaris™ GP Volumetric Pump Configuration and Calibration
Clear CQI Log File • Press OK to confirm clearing the CQI Log File.
CLEAR CQI LOG FILE WARNING! ****** ****** This will remove all
instrument history and will restore the factory default data set by deleting the clinically approved installed data set. Clear current log?
QUIT
BDTM00001 Issue 2
11 / 84
OK
Alaris™ GP Volumetric Pump Configuration and Calibration
Service Configuration Select the required option using the f keys and the OK softkey.
SERVICE CONFIGURATION
Alarm Presets
Allows the preset alarm tones to be altered. (Applies only to models 9002TIG03 and 9002TIG03-G.)
Date/Time
Sets the current date and time used for event logging.
Software Versions
Displays the Pump software versions.
Serial Number
Configure the displayed serial number.
Pump Reference
Pump specific text to be displayed in user mode at start up. (20 characters max.)
Language
Configure the language used for display messages.
Backlight & Contrast
Adjust the backlight and contrast values.
Current Data Set File
Displays the current data set file details.
Alarm Presets Date/Time Software Versions Serial Number Pump Reference Language Backlight & Contrast Current Data Set File
SELECT WITH QUIT
OK
Alarm Presets 1. Select the required Alarm Preset using the f keys. The setting toggles between 3RD EDITION ALARMS and ORIGINAL ALARMS, and the active option is marked with a . 2. Once the correct option is selected, press OK to confirm. Note: Configurable alarm presets are only available on pump models 9002TIG03 and 9002TIG03-G. The alarm tone options available are: • ORIGINAL ALARMS: Low, medium and high priority alarm tones that sound identical to the auditory alarms and warnings from software versions prior to 2.3.4. • 3RD EDITION ALARMS: Low, medium and high priority alarm tones in accordance with IEC 60601-1 Edition 3.1. The default setting is ORIGINAL ALARMS. Changing the alarm tone settings should only be performed by Qualified Service Personnel. Note: All pumps in a single care area should be configured with the same alarm tones to avoid user confusion. The Hospital/ Facility is responsible for selecting and configuring the desired alarm scheme.
BDTM00001 Issue 2
12 / 84
ALARM PRESETS 3RD EDITION ALARMS ORIGINAL ALARMS
OK
Alaris™ GP Volumetric Pump Configuration and Calibration
Date/Time 1. Set the correct date and time using the f keys. 2. Press NEXT to continue to next item to change. 3. Press OK to confirm.
DATE / TIME AUG–04–2006 09:48 ADJUST WITH QUIT
NEXT
OK
Software Versions • Press OK to exit after verifying Software Version fitted. Display will vary depending on software version fitted.
SOFTWARE VERSIONS
SOFTWARE VERSIONS
SOFTWARE VERSIONS
SRP
001.005.004
SWV
001.009.002
SWV
002.001.012
MP FDP SP
001.005.004 001.005.004 001.005.004
PKG MPU MPT FDP SP LANGUAGE
001.009.000 001.009.000 001.009.000 001.007.016 001.005.000 001.004.005
PKG MPU MPT FDP SP LANGUAGE
009.004.002 002.001.012 002.001.012 001.007.016 001.005.000 001.005.001
QUIT
OK
QUIT
QUIT
OK
OK
Serial Number 1. Set the serial number required (maximum 9 characters) using the f keys. 2. Press NEXT to continue to next item to change. 3. Press OK to confirm.
SERIAL NUMBER 8026 - 00000 Write Success:
ADJUST WITH QUIT
BDTM00001 Issue 2
13 / 84
NEXT
. OK
Alaris™ GP Volumetric Pump Configuration and Calibration
Pump Reference 1. Set the Pump Reference required using the f keys. 2. Press NEXT to continue to next item to change. 3. Press OK to confirm.
PUMP REFERENCE ---------------------
ADJUST WITH QUIT
NEXT
OK
Language 1. Select the required Language using the f keys. 2. Press OK to confirm.
LANGUAGE
ENGLISH NORSK - NORWEGIAN DEUTSCH - GERMAN FRANCAIS - FRENCH NEDERLANDS - DUTCH SVENSKA - SWEDISH ESPAñOL - SPANISH ITALIANO - ITALIAN NONE - VANILLA
Note: Languages available will be dependent on the Pump software version.
SELECT WITH
QUIT
OK
Backlight & Contrast 1. Use the f keys to adjust Backlight and Contrast to required setting. 2. Press PARAM to change between Backlight and Contrast. 3. Press OK to confirm. Note: Select the Dimming parameter to see what the display would look like when dimmed.
BACKLIGHT & CONTRAST Backlight = 30% Contrast = 70% Dimming = 25%
ADJUST WITH QUIT
BDTM00001 Issue 2
14 / 84
PARAM
OK
Alaris™ GP Volumetric Pump Configuration and Calibration
Current Data Set File • Press OK to exit after verifying current data set information.
CURRENT DATA SET Data Set ID: ABCD Name: Data_set_EXAMPLE Version: 2 Last update: 01-Jun-2006 12:34 Hospital name: Basingstoke General
QUIT
BDTM00001 Issue 2
15 / 84
OK
Alaris™ GP Volumetric Pump Configuration and Calibration
Near End Of Infusion (NEOI) Warning Configuration NEOI Warning Configuration is separate from Service Mode and can be accessed via a three-digit access code that is entered using the following procedure: 1. Hold down b and turn the Pump on a. 2. Enter the access code 057 using the f keys and the NEXT softkey. 3. When the code shows on screen, press OK to confirm.
NEOI Warning 1. Select the required NEOI status can be enabled or disabled by pressing ON or OFF. Note: ON is shown when the NEOI is disabled and OFF is shown when the NEOI is enabled. 2. Press ACCEPT to confirm. Note: ACCEPT will not be shown until after the first toggle of the status.
NEAR END OF INFUSION Enabled
Press ON/OFF to toggle
QUIT
BDTM00001 Issue 2
16 / 84
ACCEPT
OFF
Alaris™ GP Volumetric Pump Configuration and Calibration
Data Set Transfer Upload data set to an Alaris™ GP Volumetric Pump Equipment required: • Alaris™ GP Editor Software Kit (1000SP01310) – includes the Alaris™ GP Transfer Tool • RS232 cable (1000SP01183) • USB to Serial Adapter Kit (12273817) – optional • USB to RS232 converter 4 way hub (1000EL00980) – optional • PC – for requirements see Upgrading Firmware PC requirements Using the Alaris™ GP Transfer Tool allows a released data set to be uploaded to an Alaris™ GP Volumetric Pump. At no time should the Alaris™ GP Transfer Tool be used to upload to an Alaris™ GP Volumetric Pump while the Pump is connected to a patient. In Service Mode select Data Set Transfer using the f keys and the OK softkey. 1. Using the Alaris™ GP Transfer Tool select data set to be uploaded. 2. Press the RS232 or IrDA softkey to select the Comms mode being used. 3. Connect the Pump to PC. 4. Press the START softkey to begin transfer. 5. Please ensure the data set ID shown on the Pump is identical to the one transferred. 6. Press PASS softkey to confirm correct transfer and exit. 7. To transfer the data set to another Pump repeat steps 2 to 6.
Upload data set to an Alaris™ GP Guardrails™ Volumetric Pump Equipment required: • Guardrails™ Editor v3.1 Software Kit (1000SP01389) or Guardrails™ Editor v3.1 Transfer Tool Software Kit (1000SP01390) • RS232 cable (1000SP01183) • USB to Serial Adapter Kit (12273817) • USB to RS232 converter 4 way hub (1000EL00980) – optional • PC – for requirements see Upgrading Firmware PC requirements Using the Guardrails™ Editor v3.1 Transfer Tool allows an approved data set to be uploaded to an Alaris™ GP Guardrails™ Volumetric Pump. At no time should the Guardrails™ Editor v3.1 Transfer Tool be used to upload to an Alaris™ GP Guardrails™ Volumetric Pump while the Pump is connected to a patient. In Service Mode select Data Set Transfer using the f keys and the OK softkey. 1. Using the Guardrails™ Editor v3.1 Transfer Tool Transfer Tool select data set to be uploaded. 2. Connect the Pump to PC. 3. Press the START softkey to begin transfer. 4. Please ensure the data set ID shown on the Pump is identical to the one transferred. 5. Press ACCEPT softkey to confirm correct transfer and exit. 6. To transfer the data set to another Pump repeat steps 2 to 5.
w
Note: For more information relating to the Alaris™ GP Editor Software or the Guardrails™ Editor v3.1 Software refer to the relevant Directions For Use supplied with the software.
Loading the Data Set Transfer Tool software is considered a non-clinical service activity. Interconnecting the Pump with a PC may cause the safety or electromagnetic environment to change while the connection exists. The threat of higher leakage currents or EMI disturbances may be present. Disconnect the IrDA or RS232 cable connection at both ends following software upload activities.
BDTM00001 Issue 2
17 / 84
Alaris™ GP Volumetric Pump Configuration and Calibration
Upload data set to an Alaris™ GP (Guardrails™) Volumetric Pump (with Plus software) Equipment required: • Alaris™ Editor v4.4.1 Software Kit (1000SP01534) or Alaris™ Transfer Tool Software Kit (1000SP01535) • RS232 cable (1000SP01183) • USB to Serial Adapter Kit (12273817) • USB to RS232 converter 4 way hub (1000EL00980) – optional • PC – for requirements see Upgrading Firmware PC requirements Using the Alaris™ Transfer Tool allows an approved data set to be uploaded to a Pump (with Plus software). At no time should the Alaris™ Transfer Tool be used to upload to a Pump while the Pump is connected to a patient. In Service Mode select Data Set Transfer using the f keys and the OK softkey. 1. Select the Data Set Upload button (Ctrl+U). The Data Set Upload Wizard is displayed. Click Next to continue. 2. Select the serial port(s) required to transfer the data set to the Pump(s). Click Next to continue. 3. Click the Select Dataset button and Open the required data set. The selected data set and path are displayed. Click Next to continue. 4. Enter the data set ID and click Validate. If the data set ID entered matches the ID of the selected data set the Next button will become available. Click Next to continue. 5. The data set selection is now complete. Click Finish. The file information is updated to show the File used, Data Set and Supported Pump Types. 6. Connect the Pump to PC. 7. Press the START softkey to begin transfer. 8. When the transfer is complete, the Pump will validate the data then display Transfer Complete. The transfer status will be shown as Data Set Deployed to the Pump. Note: Please ensure the data set ID shown on the Pump is identical to the one transferred. 9. Press ACCEPT softkey to confirm correct transfer and exit. The transfer status will be shown as Deployed. 10. Switch off the Pump by holding the power button for 3 seconds and disconnect from the PC. 11. If multiple uploads are required, connect the next Pump and repeat from step 5. 12. When all uploads are complete, click Menu > Exit to close the Transfer Tool.
w
Loading the Transfer Tool software is considered a non-clinical service activity. Interconnecting the Pump with a PC may cause the safety or electromagnetic environment to change while the connection exists. The threat of higher leakage currents or EMI disturbances may be present. Disconnect the IrDA or RS232 cable connection at both ends following software upload activities.
Note: For more information relating to the Alaris™ Editor Software refer to the relevant Directions For Use supplied with the software.
BDTM00001 Issue 2
18 / 84
Alaris™ GP Volumetric Pump Configuration and Calibration
Calibration Zero Point Calibration Select the required option using the f keys and the OK softkey. 1. Wait for the pressure sensors to park. 2. Ensure that an infusion set is not installed. 3. Press the START softkey. 4. Pump will countdown for 15 seconds. 5. Press the ACCEPT softkey. 6. Press the PASS softkey.
CALIBRATION Zero Point Cal Pressure Cal Volumetric Cal Battery Cal
SELECT WITH
QUIT
BDTM00001 Issue 2
19 / 84
OK
Alaris™ GP Volumetric Pump Configuration and Calibration
Pressure Calibration Equipment required: • Calibrated Pressure Gauge, minimum specification of: -- Accuracy = 0.10% of full scale -- Full scale = 1500mmHg • Pressure Calibration Set (1000SP01422) Use to calibrate 10 pumps and then change Set up equipment as per Figure 2–1 and allow 30 seconds before proceeding. Select the required option using the f keys and the OK softkey. 1. Turn 3-way tap to close from atmosphere. 2. Press the START softkey. 3. Apply a pressure of 200mmHg and Pump will countdown for 15 seconds. 4. Press the ACCEPT softkey. 5. Apply a pressure of 800mmHg and Pump will countdown for 15 seconds. 6. Press the ACCEPT softkey. 7. Turn 3-way tap to vent to atmosphere. 8. Turn 3-way tap to close from atmosphere. 9. Press the NEXT softkey to proceed to Verification Procedure.
CALIBRATION Zero Point Cal Pressure Cal Volumetric Cal Battery Cal
SELECT WITH
QUIT
OK
Verification Procedure: Software version v1.7.x and below: 10. Apply a pressure of 200mmHg and wait for 5 seconds. 11. Press the NEXT softkey. 12. Apply a pressure of 400mmHg and wait for 5 seconds. 13. Press the NEXT softkey. 14. Apply a pressure of 600mmHg and wait for 5 seconds. 15. Press the NEXT softkey. 16. Apply a pressure of 800mmHg and wait for 5 seconds. 17. Press the NEXT softkey. 18. Turn 3-way tap to vent to atmosphere. 19. Press the PASS softkey.
Software version v1.9.x and above: 10. Apply a pressure of 500mmHg and wait for 5 seconds. 11. Press the NEXT softkey. 12. Turn 3-way tap to vent to atmosphere. 13. Press the PASS softkey.
Pressure source (50ml/100ml syringe or similar device)
Pressure Gauge
Vent to atmosphere
Alaris®
GP
Figure 2–1 Pressure Calibration Equipment Set Up
BDTM00001 Issue 2
20 / 84