Alere

PT-INR Professional Monitoring Systems

INRatio PTINR Monitoring System Error Message Troubleshooting Guide ver 02 March 2014

Troubleshooting Guide

4 Pages

®  Alere INRatio PT/INR Monitoring System Error Message Troubleshooting Guide 	
   ®  The Alere INRatio PT/INR Monitor can report abnormal operating conditions by displaying error or warning messages appropriate to the condition on the meter display. The following provides a list of the ® error messages generated by the Alere INRatio PT/INR Monitor with a description of the conditions which cause the message to be displayed. Some errors allow the user to recover from the error condition whereas others are non-recoverable. In most cases the user can recover from the error condition by repeating the test.	
   The following table describes both recoverable and non-recoverable error message. Recovery procedures are described for each recoverable error.  Display Screen 	
    What’s the problem?	
   The sample was applied during the warm up period before the “Apply Sample” screen appeared.  What to do?	
   Begin the process again with a new test strip. Be sure to wait until the “Apply Sample” screen appears before applying the sample.  A used strip was inserted.  Be sure to use a new strip for every test.  	
   Not enough blood reached the test area.  Begin the process again with a new test strip and perform another fingerstick.  The possible causes for this are: o Not enough blood applied to the sample well. o The strip channel is blocked.	
    Make sure to apply a large hanging drop of blood. Consult the Fingerstick Tip Sheet if you are having problems obtaining a good sample. Do not try to add more blood from the first fingerstick.  Only certain combinations of numbers and letters are recognized by the monitor. The code entered was not recognized by the monitor.	
    or	
  combination	
  of	
   three	
  digits.	
  Examples:	
   ERROR	
  411	
   ERROR	
  114	
   ERROR	
  123	
   Etc.	
    	
   Re-enter the correct code at the “Strip Code” prompt. If the message repeats, call Alere Technical Support.  The monitor had a data processing problem while calculating the results from some or all of the three channels. This may be caused by the following: o A problem with the individual test. o Not enough blood was applied to the sample well.  Begin the process again with a new test strip and perform another fingerstick.  The error appears as a three digit number. Each digit represents one of the channels on the strip and may appear as 1, 2, 3, or 4. o Example: ERROR 411  Do not try to add more blood from the first fingerstick.  Make sure to apply a large hanging drop of blood. Consult the Fingerstick Tip Sheet if you are having problems obtaining a good sample.  If the error message repeats, call Alere Technical Support
File Type: PDF
File Size: 218 KB
File Name: inratio_ptinr_monitoring_system_error_message_troub.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions