GETINGE FLOW-i Anaesthesia Systems

FLOW-i Servo Duo Guard Cleaning Procedure July 2019

Cleaning Procedure

5 Pages

Wipe off/ discard  Flow-i/c Cleaning procedure when Servo Duo Guard filter is used Important If using a Servo Duo Guard (or filter with equivalent performance) on the expiratory connection, the system and removable parts can normally be wiped with a soft lint free cloth. This cloth should be moistened with soap and water or a detergent-based disinfectant. Note In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.  Hazardous Waste (Infectious) Whenever possible, clean the equipment immediately after use. Do not leave foreign material, such as blood or saliva, to dry on the equipment  www.getinge.com/anz ANZ_CC_V1_050_0719  Caution All personnel should be aware of the risk of parts being infected when disassembling and cleaning the system.