28th June 2017  To whom it may concern;  Sterilisation of Olympus Bougie Tube Set (WA33027A) with an autoclave Please follow the instructions in the IFU for WA33027A. This product is compatible with the reprocessing methods listed in the table. For details of the procedures, refer to the Olympus Endoscopy System Guide. + compatible − incompatible  When using an Autoclave to steam sterilise the equipment following cleaning please sterilise according to the Australian standards and your hospital protocol. As a guide the following settings should be utilised: – Temperature: 134°C – Steriliser pressure: 203kPA – Sterilisation time = 3.5 - 4 minutes – Dry time = as per hospital/machine protocol  Regards  Melanie O’Farrell Product Manager – Urology & Gynaecology  3 Acacia Place, Notting Hill. VIC 3168 PO Box 985 Mt Waverley. VIC 3149 Phone: 1300 132 992 Fax: 03 9543 1350 Web: www.olympusaustralia.com.au ABN 90 078 493 295