USA: CAUTION:  INSTRUCTIONS  Federal law restricts this device to sale by or on the order of a physician.  LEAKAGE TESTER  MB-155  1 Intended Use The Leakage Tester is designed to check watertightness of Olympus endoscopes in conjunction with Olympus Light Source or Maintenance Unit (MU-1).  2 Symbols The meaning(s) of the symbol(s) shown on the back cover of this instruction manual are as follows: Manufacturer  Authorized representative in the European Community  For US Customers only For a Symbols Glossary, visit us: http://www.olympus-global.com/en/common/pdf/symbolsglossary.pdf  3 Signal Words The following signal words are used throughout this manual: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.  4 Specifications Insufflation function  2.84 × 104 Pa or less  Performance  5 Inspection 1. Inspect that the Connector is not markedly deformed. 2. Inspect that the Coiled Tube is not cracked. 3. Inspect that the Connector Cap is not deformed, nor is it wet inside. Connector  Coiled Tube  Connector Cap  Water drops inside the Connector Cap may enter the endoscope through the Venting Connector of the endoscope’s Light Guide Connector Section or Water-resistant Cap and damage the internal construction. Remove them by shaking the Connector Cap several times.  –1–