Analytical Technology
PortaSens II Model C16 Operating and Maintenance Manual Rev I Sept 2014
Operating and Maintenance Manual
51 Pages
Preview
Page 1
PortaSens II Portable Gas Leak Detector Model C16
Home Office
European Office
Analytical Technology, Inc. 6 Iron Bridge Drive Collegeville, PA 19426 Ph: (800) 959-0299 Ph: (610) 917-0991 Fax: (610) 917-0992
ATI (UK) Limited Unit 1 & 2 Gatehead Business Park Delph New Road, Delph Saddleworth OL3 5DE Ph: 0800-018-4020 + 44 (0) 1457-873-318 Fax: + 44 (0) 1457-874-468
Email: [email protected]
Email: [email protected]
PortaSens II
Portable Gas Leak Detector, Model C16
PRODUCT WARRANTY Analytical Technology, Inc. (Manufacturer) warrants to the Customer that if any part(s) of the Manufacturer's products proves to be defective in materials or workmanship within the earlier of 18 months of the date of shipment or 12 months of the date of start-up, such defective parts will be repaired or replaced free of charge. Inspection and repairs to products thought to be defective within the warranty period will be completed at the Manufacturer's facilities in Collegeville, PA. Products on which warranty repairs are required shall be shipped freight prepaid to the Manufacturer. The product(s) will be returned freight prepaid and allowed if it is determined by the manufacturer that the part(s) failed due to defective materials or workmanship. This warranty does not cover consumable items, batteries, or wear items subject to periodic replacement including lamps and fuses. Gas sensors, except oxygen sensors, are covered by this warranty, but are subject to inspection for evidence of extended exposure to excessive gas concentrations. Should inspection indicate that sensors have been expended rather than failed prematurely, the warranty shall not apply. The Manufacturer assumes no liability for consequential damages of any kind, and the buyer by acceptance of this equipment will assume all liability for the consequences of its use or misuse by the Customer, his employees, or others. A defect within the meaning of this warranty is any part of any piece of a Manufacturer's product which shall, when such part is capable of being renewed, repaired, or replaced, operate to condemn such piece of equipment. This warranty is in lieu of all other warranties (including without limiting the generality of the foregoing warranties of merchantability and fitness for a particular purpose), guarantees, obligations or liabilities expressed or implied by the Manufacturer or its representatives and by statute or rule of law. This warranty is void if the Manufacturer's product(s) has been subject to misuse or abuse, or has not been operated or stored in accordance with instructions or if the serial number has been removed. Analytical Technology, Inc. makes no other warranty expressed or implied except as stated above.
O & M Manual Rev. I, September 2014
-2-
PortaSens II
Portable Gas Leak Detector, Model C16
Table of Contents PRODUCT WARRANTY ...2 SPECIFICATIONS ...4 INTRODUCTION ...5 PORTASENS II ...5 PACKAGE CONTENTS ...6 REFERENCE DRAWINGS ...6 SMART SENSOR ...11 CHANGING THE SENSOR ...11 SENSOR RESPONSE TIMES ...12 GAS INTERFERENCES...13 OPERATION ...15 QUICK START ...15 STARTUP SEQUENCE ...16 GAS CONCENTRATION READING ...17 SOFT KEYS ...17 SHUTDOWN...17 BATTERIES ...18 LED INDICATOR...19 RESPONSE TEST ...20 INLET AND OUTLET EXTENSIONS ...21 ANALOG OUTPUT ...21 OPERATING MODES ...22 Sample Mode ...22 ALARMS...24 LOW FLOW ALARM (“PUMP TROUBLE”) ...25 DEVICE CONFIGURATION ...27 VIEWING THE CONFIGURATION...27 CHANGING THE CONFIGURATION ...27 ALARM SETTINGS...28 Set Point (S.P.)...28 Function (FUNC) ...28 DISPLAY SETTINGS ...31 Range (RANGE) ...31 Blanking (BLANK) ...31 Averaging (AVG) ...31
O & M Manual Rev. I, September 2014
-3-
SAMPLE MODE SETTINGS ...33 Sampling Time (SAMP) ...33 Measuring Time (MEAS) ...33 Clearing Limit (CLEAR) ...33 SENSOR CALIBRATION ... 35 Zero Calibration (ZERO) ...35 Span Calibration (SPAN) ...35 PRESSURIZED GAS SOURCES...36 ZERO AND SPAN CALIBRATION STEPS ...36 DATA LOGGER ... 38 DATA LOGGER OPERATION AND SETTINGS ...38 Clear (CLEAR) ...38 Interval (INT) ...38 C16LOG ... 41 INSTALLATION ...41 Installing on Windows 7 & 8 ...41 Installing on Windows XP ...41 RS232-to-USB Adapter ...41 Connecting the Device ...41 STARTING C16LOG ...42 Version 2.0 and Higher ...42 Earlier Versions ...42 USING C16LOG...42 Configure Tab ...43 Important Note for Windows, Version 7 and Higher ...43 Download Tab ...45 OUTPUT FILES ...46 MAINTENANCE ... 48 Intake Pump Filter...48 TROUBLE ... 48 SPARE PARTS ... 49 SMART SENSOR MODULES (MODEL H10) ...50
PortaSens II
Portable Gas Leak Detector, Model C16
SPECIFICATIONS Gas Sensor
Accepts H10 “Smart Sensor” gas sensor modules (not included)
Range
Sensor dependent (see Smart Sensor Modules on page 50, or contact ATI)
Display
Backlighted, graphics LCD
Accuracy
Sensor dependent (typically ± 5% of value, limited by cal. gas)
Sensitivity
1% of sensor module range
Repeatability
± 1% of sensor module range
Input/Output
RS-232 for transfer of logged gas readings 0-1 VDC analog (requires optional cable)
Data Logger
1, 5, 10, and 15 minute intervals 12,000 samples (8 days@1 minute interval) C16Log application for transferring data (CD and Online)
Alarms
Three level alarms, configurable for high or low alarm (or disable) Low battery alarm Low flow rate alarm Alarm notifications provided by LCD and beeper
Power
Primary: D-cell battery, alkaline recommended, 75 hours operation Backup: Internal, rechargeable battery, 6 hours operation 115VAC (std.) battery charger (220VAC avail.)
Operating Temp.
-25º to +55º C
Humidity
0-95% Non-condensing
Construction
Glass filled nylon, PVC, and stainless steel
Accessories
10” Flexible Wand, Teflon lined Battery Charger (3) Intake Filters Outlet Barbed Fitting Flow Meter RS-232 Cable (DB9 Female) CD (C16Log downloader) Storage case Bypass-T Fitting (for calibration)
Shipping Weight
3 lbs. (1.4 Kg.)
O & M Manual Rev. I, September 2014
-4-
PortaSens II
Portable Gas Leak Detector, Model C16
INTRODUCTION PortaSens II The PortaSens II is a rugged, hand-held device used to detect leaks of toxic gas, and also to detect oxygen displacement or deprivation. It adapts quickly to measure different gasses by inserting the appropriate gas sensor module. The modules retain all of the calibration and configuration settings for the target gas, such as the full scale range and alarm settings. The device features a battery powered intake pump with a flexible wand for sampling. Readings and settings are presented on a backlighted, graphics LCD, with four “soft keys”, and a beeper to notify you about alarms and other events. Readings may be recorded and transferred to a PC using a supplied application, C16Log. The program outputs ASCII csv files for importing into applications, like Microsoft Excel®. The application operates on Microsoft Windows ® 98,XP, 2000, and Windows 7 & 8. For more information, see C16Log on page 41. Power is provided from one primary (non-rechargeable), D-cell battery, and switches automatically to a secondary (rechargeable) battery when the primary is removed, or its power drops below an acceptable level1.
1 The internal secondary battery should be charged prior to use if it has been more than three weeks after
leaving the factory. O & M Manual Rev. I, September 2014
-5-
PortaSens II
Portable Gas Leak Detector, Model C16
Package Contents The following items are included. o PortaSens II Gas Leak Detector (Figure 1) o Sampling Wand (Figure 2) o Outlet Tube Barbed Fitting (Figure 2) o Sensor Keeper2 (Figure 6) o Flow Meter (Figure 7) o Storage Case o Spare Alkaline D-Cell battery o Battery Charger o RS-232 Cable o C16Log CD ROM (PC version, only) o Calibration “T”
Reference Drawings
Figure 1. PortaSens II, Front View (ATI-0401)
2 H10 sensors are normally ordered separately.
O & M Manual Rev. I, September 2014
-6-
PortaSens II
Portable Gas Leak Detector, Model C16
Figure 2. PortaSens II,Rear View (ATI-0403)
O & M Manual Rev. I, September 2014
-7-
PortaSens II
Portable Gas Leak Detector, Model C16
Figure 3. PortaSens II, Side View (ATI-0402).
O & M Manual Rev. I, September 2014
-8-
PortaSens II
Portable Gas Leak Detector, Model C16
Figure 4. C16 Port (8-position MINI-DIN, female receptacle)
Table 1. C16 Port Pin Assignments Pin No.
Signal
I/O
Description
1
/CTS
Input
Clear to send - assert to suspend device output.
7
2
/RTS
Output
Request to send – asserted by device before transmit.
8
3
RX
Input
Receive data - device receives data on this pin.
3
4
Sig. Com.
Signal common (note: same as AOut Com.)
5
5
TX
Output
Transmit data – device transmits data on this pin.
2
6
AOut
Output
Analog output – 0-1 VDC output, proportional to gas concentration reading: AOut = gas_rdg/sensor_range * 1.0 VDC
N/C
7
N/C
Not connected.
N/C
8
AOut Com.
Analog output common (note: same as Sig. Com.)
N/C
O & M Manual Rev. I, September 2014
-9-
DB9 Pin on Cable 03-0201
PortaSens II
Portable Gas Leak Detector, Model C16
Figure 5. H10 Gas Sensor Module (not included)
00-0981 H10 Sensor Keeper Replace Battery every Six (6) Months
Figure 6. Sensor Keeper
Figure 7. Flow Meter
O & M Manual Rev. I, September 2014
- 10 -
PortaSens II
Portable Gas Leak Detector, Model C16
SMART SENSOR The PortaSens II measures gas concentration levels using “Smart Sensor” modules, which store their calibration and configuration settings, such as, zero, span, range, and alarm levels. This allows them to be quickly changed to detect leaks of different toxic gasses, or to detect oxygen displacement or deprivation. The modules are calibrated before leaving the factory and require only routine calibration, which may be performed by trained personnel on site using traceable gas standards and methods. A more economical solution may be to return them periodically to the factory for certified calibration. Contact ATI for details about the “Calibration Certification” program for Smart Sensor modules. Figure 8. H10 "Smart Sensor" module. Four modules may be kept in the battery-powered, “Sensor Keeper” (Figure 6 on page 10), which maintains the electro-chemical sensor cell, “on bias”, and ready for use (recommended). Gas sensor modules must be ordered separately. For a complete list of modules and their operating range, see Smart Sensor Modules (Model H10) on page 50. Note Sensors may take up to 12 hours to stabilize (bias) if not stored in the Sensor Keeper, or not installed in the device.
Changing the Sensor Referring to Figure 2 on page 7, loosen the two thumb screws and remove the flow manifold lid on the rear of the device. If a sensor module is already installed, grab the sides with your index finger and thumb, and pull straight up. Slide the new module into the recess, connector first, and rotate it to align the groove on the sensor with the guide in the recess. Press down on the sensor to seat the connector. The face of the sensor module should be flush with the outer surface of the recess. Replace the manifold lid and hand-tighten the thumb screws. Start the device and verify air flow through the sampling wand using the supplied flow meter. See Figure 7 on page 10. If power is on when the sensor is removed, the device displays “SENSOR MISSING”. At startup, and when the sensor is reinstalled with power applied, it displays the entire startup sequence, as detailed in the Startup Sequence on page 16.
O & M Manual Rev. I, September 2014
- 11 -
PortaSens II
Portable Gas Leak Detector, Model C16
Sensor Response Times The table below lists the time required to obtain readings of 66%, and 90%, of an instantaneous exposure to various gasses. The times listed are for the sensor only, and do not include travel time through the wand or tubing, or digital damping by the device. They are provided only for estimating the sampling period required for a representative reading. Table 2. Gas Response Times SENSOR TYPE
66% RESPONSE
90% RESPONSE
Oxidant Sensors (except H2O2)
20 seconds
60 seconds
Hydrogen Peroxide
40 seconds
120 seconds
Ammonia
30 seconds
120 seconds
Carbon Monoxide
10 seconds
30 seconds
Hydrogen
20 seconds
60 seconds
Oxygen
15 seconds
45 seconds
Nitric Oxide
10 seconds
20 seconds
Phosgene
70 seconds
300 seconds
Hydrogen Chloride
50 seconds
240 seconds
Hydrogen Fluoride
50 seconds
240 seconds
Hydrogen Cyanide
40 seconds
120 seconds
Hydrogen Sulfide
20 seconds
60 seconds
Nitrogen Dioxide
10 seconds
40 seconds
Sulfur Dioxide
10 seconds
40 seconds
Hydride Gases
30 seconds
70 seconds
Hydrocarbon Gases
40 seconds
90 seconds
O & M Manual Rev. I, September 2014
- 12 -
PortaSens II
Portable Gas Leak Detector, Model C16
Gas Interferences Some sensors exhibit a minor sensitivity to non-target gases and vapors. Table 3 lists the crosssensitivity factors of sensors to various, “interference” gases. To use the table, find the sensor type along the top, and then scan down the column for the factor corresponding to one of the interference gases listed on the left. Multiplying the factor times the concentration of the interference gas produces the reading that would be displayed by the device. For example, an SO 2 sensor exposed to 1 PPM of HCN would produce a reading of 0.15 PPM. See notes following the table for additional comments.
Table 3. Cross-sensitivity Factors
Sensor Type NH3
O3
HF
HCI
NH3
--
CO
0.01
0.005
H2
0.002
0.01
NO
0.08
1.5
O2
*
Cl2
-0.5
O3
Interference Gas
CI2
HCl
-0.5
HCN
0.1
HCN
-0.05
H2S
SO2
CO H2
H2O2 O2 NO
NO2
Hydride (4) SiH4
0.01
--
0.1
0.05
-0.05
(0.05)
0.1
--
0.03
-0.06
-0.005
*
0.01 0.01
0.4
0.08
0.1 *
0.09
2.0
*
*
*
*
-0.2
-0.4
1.0
-0.01 (-0.1)
-0.2
0.05
--
1
-0.05
-0.2
-0.4
1.0
(-0.1)
-0.2
-0.005
0.02
-0.06
(-0.01)
-0.2
0.05 (0.2)
HF
0.01
--
0.4
*
(0.04)
-0.05
-0.1
--
0.05
*
-0.5
*
0.014
2
--
*
--
0.3
1
0.5
*
0.0003
0.08
--
-0.08
COCl2 H-C
0.1
0.2
-0.1
0.8 *
0.1
0.05
0.2
0.5
0.1
2
2
--
H2S
1.0
-0.1
-0.1
-0.3
3
NO2
0.08
0.2
0.15
0.2
0.2
0.01
SO2
-0.1
-0.01
-0.01
1
0.5
0.4
0.08
Hydride
1.5
0.2
SiH4
1.5
0.2
--
6
-5.0
-0.3
-0.1
--
-0.001 (-0.1)
0.04
-1
0.1
--
0.42
-0.8
(0.5)
0.1
0.2
0.4
0.1
1
6
-3.5
--
5
2
0.1
0.3
6
2
0.005 (0.5)
--
2
2
-1.5
CO2 CH4 CH3SH
0.3
-0.04
-0.03
-0.1
1
C2H2
0.3 .0005
C2H4 C2H6O
0.015
0.05
0.02
0.002
1.7
0.1
0.1
0.002
0.3
0.1
0.1
0.015
0.015
COCl2
O & M Manual Rev. I, September 2014
0.02
0.1
0.005
1
-0.02
0.0001 (0.5)
0.05
1.2
-0.01
(0.1)
0.02
1
-0.07
(0.02)
0.1
0.7 --
- 13 -
PortaSens II
Portable Gas Leak Detector, Model C16
Notes on Interference Table: 1.
2. 3. 4. 5. 6. 7. 8.
Sensors marked with an asterisk (*) in the oxygen column are 3 electrode sensors that require a minimum of 5% oxygen to operate properly. Hydrogen sensors require oxygen levels at least two times the maximum percent hydrogen value to be measured. The data on the chlorine sensor also applies to bromine, chlorine dioxide, fluorine, and iodine sensors. Data on the hydride sensor refers to arsine, phosphine, diborane, hydrogen selenide, and germane sensors. Response is not exactly 1:1 for all hydrides. Contact ATI for details if exact response is needed. Data in ( ) refers to PPM versions of the hydride sensors relative to phosphine. Response is not exactly 1:1 for all hydrides. Contact ATI for details. The sensor column marked “H-C” stands for hydrocarbon sensors. These include ETO (ethylene oxide), formaldehyde, alcohol, acetylene, and vinyl chloride sensors. Data in this table represents exposures to low PPM levels of the interfering gas. Very high concentrations of any interfering gas may cause either short term or long term response from a sensor. Interference factors may vary over sensor lifetime. Calibration with interference gas is not recommended in most cases. Contact factory for more information. Empty cells in table indicate insignificant cross-sensitivity.
O & M Manual Rev. I, September 2014
- 14 -
PortaSens II
Portable Gas Leak Detector, Model C16
OPERATION Quick Start The PortaSens II is simple to operate. After removing the device from the storage case, install a gas sensor module (see Changing the Sensor on page 11) and plug the sampling wand into the inlet port of the manifold (Figure 2 on page 7). Note The “INLET” and “OUTLET” ports on the manifold must not be blocked.
To start the device, press and release button 1 (the green button below the display). You should hear a beep, and the internal pump should begin run. If a sensor is installed, the device reads the sensor memory, performs an internal test, and presents the sensor configuration on a number of review displays, before reaching the main display. Otherwise, the pump motor is turned off, and the display appears as that shown in Figure 10, below. During the startup sequence, you may press the Skip button to advance directly to the main display (the sequence is shown in Startup Sequence on page 16). The Main Display appears at the end of the startup sequence, and is depicted in Figure 9, below. The gas concentration reading appears in large, numeric digits, with the units of measurement on the right, and the gas symbol just below it. The example below shows an ammonia sensor is installed, but the units and symbol depend on the installed sensor. The display also indicates which battery is currently powering the device (see Batteries on page 18 for details about this indication). P
0 OFF
PPM NH3
MENU
SMPL
Figure 9. Main display with ammonia sensor installed. Figure 10 shows the display when a sensor is not installed. If a sensor is installed at this point, the device will present the startup sequence, as described above (and detailed in Startup Sequence on page 16).
P SENSOR MISSING OFF
Figure 10. Sensor missing display. Note A response test is recommended prior to use in the field (see Response Test on page 20).
O & M Manual Rev. I, September 2014
- 15 -
PortaSens II
Portable Gas Leak Detector, Model C16
Startup Sequence The startup sequence for a typical device with an ammonia sensor is shown in the table below. The gas symbol, range, and alarm settings will vary according to the installed sensor. No. D1
Display
Description Uploading sensor configuration and calibration settings.
LOADING SENSOR
Testing hardware and software.
D2 SELF CHECK
Tests completed successfully.
D3 SELF CHECK PASSED
D4
100 HOLD
D5
C
W
HOLD A
HOLD
D8
PPM NH3
NEXT
SKIP PPM NH3
NEXT
SKIP
D9
NEXT
P
Warning set point. The set point is 25 PPM, and the symbol, , indicates the alarm activates at, or above, the set point. Alarm operation is detailed on page 22. Alarm set point. The set point is 50 PPM, and the symbol,, indicates the alarm activates at, or above, the set point. Alarm operation is detailed on page 22.
Device and sensor revision levels.
x.xx y.yy
NEXT
SKIP
0 OFF
Caution set point. The line under C indicates the caution level is set at -20 PPM; the symbol, , indicates the alarm activates at, or below, the set point. Alarm operation is detailed on page 22.
SKIP
Firmware Rev Sensor Rev HOLD
Range setting (full scale). The “Soft Keys” appear along the bottom line (see Soft Keys on page 17).
Minimum and maximum range of the sensor.
Range 50 PPM to 500 PPM HOLD
D10
SKIP
50
SKIP
PPM NH3
NEXT
25
D7
MENU
20 HOLD
D6
NEXT
PPM NH3
PPM NH3
MENU
O & M Manual Rev. I, September 2014
SMPL
Main Display shows current gas concentration, units of measure, and gas symbol. The symbol in the upper left indicates which battery is providing power.
- 16 -
PortaSens II
Portable Gas Leak Detector, Model C16
Gas Concentration Reading The gas concentration reading appears on the device display Range Resolution with the units of measure, and the gas symbol. The range setting of 0.00 to 4.99 0.01 the device determines the resolution of the reading, as shown in the 5.0 to 49.9 0.1 table on the right, and is independent of the units of measure. It is not 50 and above 1 used to limit, or “clip”, readings, but does limit other settings. Averaging is used to stabilize the reading, and Blanking is used to filter small reading fluctuations near zero, and suppress negative readings. See Display Settings on page 31 for details about the operation and configuration of these settings.
Soft Keys The four push buttons on the front panel are context sensitive, which means their function changes according to the requirements of the user interface. The device normally displays a label just above the button to indicate the function that will be performed when the button is pushed3. The button, with its displayed label, is referred to as a “soft key”. Throughout this manual, the term “key” is used most often to refer to the labeled button, and “button” is used when no label appears above it. For example, during the startup sequence above, button 1 (the left most button) is labeled “HOLD”. It is referred to as the “HOLD key” and is used to halt the display sequence. Button 2 is labeled “NEXT”. It is referred to as the “NEXT key” and is used to advance to the next display. Finally, button 4 (the right-most button) is labeled “SKIP”. It is referred to as the “SKIP key” and is used to advance directly to the Main Display.
Shutdown To turn off the device, press and hold the OFF key on the Main Display, D10, until the device displays, “POWER DOWN”, then release the key. No. D10
Display P
0 OFF
Instructions Press and hold the OFF key until the display changes to D11 (approximately 3 seconds).
PPM NH3
MENU
SMPL
Release the key when this display appears.
D11 POWER DOWN
3 Exceptions to this are starting the device (the display is not yet active) and entering Configuration mode (no label
appears above the button).
O & M Manual Rev. I, September 2014
- 17 -
PortaSens II
Portable Gas Leak Detector, Model C16
Batteries The primary battery is a standard, alkaline D-cell. It is housed in the handle of the device and will operate the device continuously for approximately 75 hours. If the primary battery becomes discharged, or disconnected, a secondary (rechargeable) battery will supply power for up to 6 hours. The device indicates which battery is currently supplying power by the symbol that appears in the upper left corner of the display. P
0 OFF
PPM NH3
MENU
SMPL
…primary battery …secondary battery …battery charger is connected Power is drawn from the primary battery first. When the primary battery is low, the ‘P’ symbol will flash for about one hour. The device will eventually begin to beep once per minute, before switching over to the secondary battery. When the secondary batter is low, the ‘S’ symbol will flash for about an hour, and again the device will beep once per minute, before the secondary battery is discharged. When both batteries are discharged, the instrument will shut off and not restart until one or both of the batteries have been restored. It is not unusual for power to switch between batteries for a period of time. The primary battery recovers after being switched off, and eventually, recovers enough to be switched back on. This process repeats for some time, until the primary battery is fully discharged. An A/C charger is supplied for the secondary battery. The jack is located under the rubber seal on the side of the device (see Figure 3 on page 8). When charging, the LED will light steady. The device may be operated continuously while charging.
Warning Use only the battery charger supplied by ATI for this device. Connecting a different charger may result in permanent damage to the battery or other electronic components.
O & M Manual Rev. I, September 2014
- 18 -
PortaSens II
Portable Gas Leak Detector, Model C16
LED Indicator The LED on the panel indicates battery status. It lights steady when the charge cable is connected. When the cable is disconnected, the LED flashes once every six seconds to indicate the device is capable of starting (one, or both, batteries are ready). If the LED is off, both batteries are discharged, and the device will not start. In this condition, any sensor installed in the device may lose bias, and will require several hours to stabilize after restoring battery power. Table 4 summarizes the battery status indications. Table 4. LED Battery Status Indicator
On (steady)
Device is off Charge cable connected Secondary battery charging
Flashing
Device is off Charge cable disconnected Battery power is available for start up
Off (steady)
Device is on, or, Device is off, charge cable is disconnected, and no battery power is available.
Reminder Keep the sensor modules in the Sensor Keeper when not in use.
O & M Manual Rev. I, September 2014
- 19 -
PortaSens II
Portable Gas Leak Detector, Model C16
Response Test A response test is recommended before each use. Table 5 lists methods for generating gas that are appropriate for a target gas. Contact ATI for details on other methods. Table 5. Gas Generation Methods. Target Gas
Quick Test Material and Method
Oxidant Sensors (except H2O2)
Dry calcium hypochlorite or liquid bleach
Hydrogen Peroxide
Sodium bisulfite in plastic bottle
Ammonia Sensor
Household ammonia solution
Carbon Monoxide
Cigarette smoke
Hydrogen
No simple test. Must test with cylinder hydrogen
Oxygen
Unit should read air levels
Nitric Oxide
No simple test. Must test with nitric oxide
Phosgene
No simple test. Must test with phosgene
Hydrogen Chloride
(2) drops of conc. HCl in plastic bottle
Hydrogen Fluoride
Dry calcium hypochlorite or liquid bleach
Hydrogen Cyanide
Sodium bisulfite in plastic bottle
Hydrogen Sulfide
Sodium sulfide in plastic bottle
Nitrogen Dioxide
Dry calcium hypochlorite or liquid bleach
Sulfur Dioxide
Sodium bisulfite in plastic bottle
Hydride Gases
No simple test. Must test with hydride gas.
Hydrocarbon Gases
Rubbing alcohol
Many of these methods produce high gas levels in closed containers. It is best to approach the opening of the container, slowly, with only the tip of the wand, to reduce the level of exposure at the sensor. While a high exposure will not harm the sensor, it may take several minutes for it to recover. Warning Do not draw liquid into the wand - only sample the air space just above the liquid.
O & M Manual Rev. I, September 2014
- 20 -