Quick Reference Guide Urilyzer® 500 Pro
Getting Started Step 1:
Step 2:
Check for completeness
Connect the power supply
Step 3:
Insert the drop tray, the strip timer rake, and the test strip tray Step 4:
Step 5:
Load the printer paper
Power up the analyzer
GB
Quick Reference Guide Urilyzer® 500 Pro
Measurement Step 6:
Dip a CombiScreen® 11SYS Plus or CombiScreen® 11SYS test strip into the urine sample for about two seconds (). Make sure that you wet every reagent pad. Drag the edge of the strip against the tube when you remove it (). Tap the edge of the dipped strip lightly on a tissue to remove excess urine (). Do not dry the strip completely. Step 7:
When a continuous green light comes on at the strip delivery zone, place the dipped strip on the test strip tray. The analyzer detects strips automatically, and the measurement cycle begins.
Dip and place test strips on the ramp until all of your samples are processed. You can monitor the progress of the test strips in the left sidebar of the Measure menu.
GB
Quick Reference Guide Urilyzer® 500 Pro
After Measurement By default, the analyzer automatically prints the results of each finished measurement. Disable or enable autoprint: On the Measurement screen tap Main Menu, then User Options. Uncheck or check the autoprint option.To save changes tap , then tap to leave the menu.
Transfer results using several formats and output methods:
On the Measurement screen, tap Database. Select one or more results in the database (), go to further actions (), and tap Output () to transfer the selected result(s). Step 8:
When you have measured all your urine samples, remove the test strip ramp, dispose of the used test strips, and disinfect the test strip ramp/waste drawer (see the Care of the Instrument section of this guide).
GB
Quick Reference Guide Urilyzer® 500 Pro
Care of the instrument Step 9:
Always keep the exterior parts and surfaces of the instrument clean. With the analyzer turned off, wipe the outside (including the display) with a damp (not wet) cloth and a mild detergent. Make sure that no liquid enters the analyzer. At the end of each day, clean the internal elements. Wear protective gloves when handling the test strip tray, the strip timer rake, and the drop tray.
Rinse the test strip tray (), the strip timer rake (), and the drop tray () under running water, and then clean them with 70% isopropyl alcohol. Dry the internal elements with a piece of lint-free cloth before you re-insert them into the analyzer. Power supply: 100–240 V AC, 50/60 Hz with external mains adapter Operating conditions: 15°C to 32°C, 20% to 85% relative humidity (non-condensing), Indoor use only
For detailed information please read the User Manual! Download via QR Code or link below. PDF reader software is required. For a printed version, please contact: [email protected] Analyticon Biotechnologies GmbH Am Muehlenberg 10, 35104 Lichtenfels, Germany [email protected] www.analyticon-diagnostics.com Phone: +49 6454 7991 0
https://www.analyticon-diagnostics.com/en/service_und_ support/downloads.html#ca
IU500_en_27_001_04.01_20220729
All components of the urine test strip analyzer may come into contact with human urine and are therefore possible sources of infection. Use universal precautions, and consult your facility’s infection control policy.