FS-i Series Models 6Ki, 15Ki and 30Ki Check Weighing Scales Instruction Manual
Instruction Manual
34 Pages
Preview
Page 1
This Manual and Marks All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute: Product Safety Signs and Labels). The meanings are as follows: WARNING CAUTION
A potentially hazardous situation which, if not avoided, could result in death or serious injury. A potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
This is a hazard alert mark.
Note
This manual is subject to change without notice at any time to improve the product. No part of this manual may be photocopied, reproduced, or translated into another language without the prior written consent of the A&D Company. Product specifications are subject to change without any obligation on the part of the manufacture.
Copyright
2006
Contents 1. COMPLIANCE ... 2 2. INTRODUCTION ... 3 3. UNPACKING ... 3 4. CAUTION ... 4 4-1. Precautions for Installing the Scale ... 4 4-2. Precautions for Operating the Scale ... 4 4-3. Precautions for Storing the Scale ... 4 5. SETTING UP ... 5 5-1. Attaching a display pod to the display support column ... 5 5-2. Installing the scale... 5 6. NAMES AND FUNCTIONS ... 6 7. BASIC OPERATION ... 10 7-1. Turning the power ON and OFF ... 10 7-2. Selecting a weighing unit... 10 7-3. Basic operation ... 10 7-4. Weighing with Preset Tare...11 7-5. To Clear a Tare Weight ...11 7-6. Weight display resolution ... 12 7-7. Simplified operation mode... 12 7-8. LCD backlight... 12 8. CHECK WEIGHING ... 13 8-1. Target weight setting mode ... 13 8-2. Upper & Lower limits weight setting mode ... 16 8-3. The comparator memory ... 17 9. ANALOG SWEEP DISPLAY ... 19 9-1. Selecting a display mode ... 19 9-2. Example of the analog sweep display ... 20 10. CALIBRATION ... 22 10-1. Calibration using a weight ... 22 10-2. Gravity acceleration correction... 23 11. FUNCTIONS... 24 11-1. The procedure for setting parameters ... 24 11-2. Function list ... 25 12. SPECIFICATIONS... 27 12-1. Specifications ... 27 12-2. Dimensions ... 28 13. GRAVITY ACCELERATION MAP ... 29
1
1. COMPLIANCE Compliance with FCC rules Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when equipment is operated in a commercial environment. If this unit is operated in a residential area it might cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference. (FCC = Federal Communications Commission in the U.S.A.)
Classification of protection provided by enclosures The equipment is designed to comply with the IP Code of IEC 60529. The “IP65” is explained as follows: “IP”
International Protection.
“6”
Against ingress of solid foreign objects. Dust-tight. No ingress of dust.
“5”
Against ingress of water with harmful effects. Protected against water jets (no power full jets). Water projected in jets against the enclosure form any direction shall have no harmful effects.
NSF listed The equipment is certified and listed to NSF/ANSI Standard 169 by NSF International. NSF International evaluated the equipment and certified that it is compliant with food protection and sanitation requirements for the design, construction and materials.
2
2. INTRODUCTION This manual describes how this scale works and how to get the most out of it in terms of performance. FS-i series check weighing scales have the following features: The FS-i series has three kinds of weight display resolution, 1/3,000, 1/6,000 (~1/7,500) and 1/12,000 (~1/15,000) to cover various applications. There are 2 sizes of weighing pan. The FS-30Ki has a larger pan and the FS-6Ki / FS-15Ki has a smaller pan. Weighing units are kg kilogram, g gram, lb pound; oz ounce; and lb-oz pound and ounce. Waterproof to IP-65 specifications. Constructed from stainless steel for harsh working environment. Large liquid crystal display with back lighting and analog sweep display of 60 segments with back lighting. Scale may be operated by AC power source or an optional SLA (sealed lead acid) battery. Built in comparator with large and bright LED display of results. Three colors of comparator results for better visibility. Two modes of comparator operation, Target weight setting and Upper/Lower limits setting. Optional serial data and comparator relay interface.
3. UNPACKING Unpack the scale carefully and keep the packing material if you are likely to transport the scale again in the future. When unpacking, check whether all of the following items are included:
CAUTION Do not pull the load cell cable.
Instruction manual Display pod
FS-i series
Weighing pan
Main power cord
INSTRUCTION MANUAL
Please confirm that the main power type is correct for your local voltage and receptacle.
Base unit
3
4. CAUTION 4-1. Precautions for Installing the Scale Ground the scale so that the user will not be subjected an electric shock. Do not handle the main power cord with wet hands. The AC plug is not water-resistant. Install it in an area where it does not get wet. Do not install the scale where there is flammable or corrosive gas present. Do not install the scale under water. Do not pull, fold or arrange cables forcibly.
Consider following conditions to get the most from your scale. Install the scale where the temperature and relative humidity is stable. There is no draft and a stable power source is available. Install the scale on a solid and level surface. Do not install the scale in direct sunlight. Do not install the scale near heaters or air conditioners. Do not install the scale where there is flammable or corrosive gas present. Do not install the scale near equipment which produces magnetic fields. Do not install the scale in a place where it is apt to be chraged with static electricity, or where the relative humidity is lower than 45% RH. Plastic and isolators are apt to be charged with static electricity. Do not use an unstable power source.
4-2. Precautions for Operating the Scale Periodically ensure that the weighing value is correct. Calibrate the scale periodically to keep weighing accuracy (See “10. CALIBRATION”). Calibrate the scale when you move it to another location. Do not place anything on the pan that is heavier than the weighing capacity. Do not apply a shock load to the scale. Do not use a sharp instrument such as a pencil or ball-point pen to press the keys. Make sure that the STABLE annunciator is ON whenever reading or storing a value. We recommend you to press the ZERO or TARE key before each weighing to prevent possible error.
4-3. Precautions for Storing the Scale Do not disassemble the scale. Do not use solvents to clean the scale. For best cleaning of the display pod, wipe with a lint free cloth that is dry or moistened with warm water and mild detergent. The base unit is can be cleaned with gentle water jet and brushing. Dry the unit well before use. Do not use a powerful water jet. 4
5. SETTING UP 5-1. Attaching a display pod to the display support column 1. Take out whole of the scale from the box taking care not to pull the load cell cable. 2. Remove 2 clamps and 2 M4x12 screws.
Display pod
3. Attach the display pod to the display support column and tighten the M4x12 screws removed at step 2. 4. Attach the clamps and tighten them after tilting the display pod where you like to use.
Clamp M4x12 screws
Put the excess part of the load cell cable into the display support column
Clamp
5-2. Installing the scale 1. Select the place for installing the scale. Refer to “Cautions for installing the scale” below. 2. Adjust the level of the base, using the spirit level and leveling feet. There is an extra foot under the display support column. Adjust this foot to reach floor after adjusting the level of the base. 3. Connect the main power cord to the outlet that has the earth wiring. You may use the earth terminal on the rear side of the display pod to ground the scale. 4. If necessary, adjust the viewing angle of the display by loosening the 2 clamps, changing the angle and re-tightening the clamps.
5
Earth terminal
Clamps
Spirit level
Leveling foot
Earth terminal
Column support foot
6. NAMES AND FUNCTIONS Display pod CAL switch cover Cable hook Display column
Earth terminal
Weighing pan
Column support foot Spirit level
Leveling foot
5
2
1 `
6
8 7 9 10
4
3
4
11
6
12
1
WEIGHT DATA DISPLAY. This display shows the weight on the weighing pan.
2
WEIGHING UNIT INDICATOR. This display area shows the weighing unit in use.
3
ANALOG WEIGHT DISPLAY. This scale has a 60 segments analog sweep display representing zero to full scale when the simple weighing mode display is selected. In the limit check weighing display mode, this display shows comparison limits and results.
4
OVER RANGE INDICATORS. This mark comes ON when the weight is outside the range of analog sweep display..
5
COMPARISON INDICATORS. The indicators LO (RED), OK (GREEN) and HI (YELLOW) show the results of the weight comparison.
6
STABLE ANNUNCIATOR. This annunciator will come ON when the weight reading is STABLE.
7
ZERO ANNUNCIATOR. This annunciator will come ON when the scale is showing the center of ZERO.
8
NET ANNUNCIATOR. This annunciator will come ON when the scale is displaying the NET weight on the weighing pan.
9
PT ANNUNCIATOR. This annunciator will come ON when the scale is displaying the preset tare weight.
10 PRINT ANNUNCIATOR. This annunciator will come ON for a moment when the scale sends out the weight data by pressing the PRINT key or auto-printing. 11 PRECAUTION AGAINST LOW BATTERY. This annunciator will come ON when the optional battery is getting close to low battery. 12 KEYPAD. A 13 key pad provides control and number keys.
7
Description of Key Operations ON/OFF Key The ON/OFF key turn the power on or off. When turned on, the scale will be automatically set to zero (power-on zero).
ZERO / ± Key The ZERO key zeros the scale when the weight is stable (STABLE annunciator ON). In the data entry sequence this key switches the sign “+” and “-”.
TARE Key The TARE key switches the scale to net weight mode and zeros the weight display when the weight is plus and stable. The ZERO and NET annunciators will come ON.
PT Key The PT pad.
key is used to enter a tare weight via the 10-key
SAMPLE / 9 Key The SAMPLE key will register the sample weight as the target. In the data entry sequence this key displays the number 9.
KEY / 8 Key The KEY key allows you to enter a target weight using the keypad. In the data entry sequence this key displays the number 8.
RECALL / 7 Key The RECALL key is used to recall target weight and/or HI/LO limits. In the data entry sequence this key displays the number 7.
HI / 6 Key The HI key allows you to enter the comparator HI limit using the numeric keypad. In the data entry sequence this key displays the number 6.
LO / 5 Key The LO key allows you to enter the comparator LO limit using the numeric keypad. In the data entry sequence this key displays the number 5.
8
STORE / 4 Key The STORE key is used to store target weight and/or HI/LO limits. In the data entry sequence this key displays the number 4.
DISP. / 3 Key When the DISP. key is pressed the analog weight display steps through the 4 available modes of simple weighing, target weighing, limit check weighing and display off. In the data entry sequence this key displays the number 3.
UNITS / 1 Key The UNITS key is used to select the desired weighing unit . In the data entry sequence this key displays the number 1.
PRINT / ENT Key The PRINT key is used when the serial data option OP-03 or OP-04 is installed and sends one data string. In the data entry sequence this key is used to ENTER the numeric data into the scale memory.
C Key The C key is used to clear an incorrect data entry from the numeric keys.
0 and 2 Keys These keys are used to display the number 0 or 2 during a data entry sequence.
9
7. BASIC OPERATION 7-1. Turning the power ON and OFF 1. Press the ON/OFF key to turn the power ON. All the display symbols are displayed and the scale waits for the weighing data to become stable. After the weighing value internally becomes stable, the display turns off for a moment and zero is shown with the ZERO annunciator (power-on zero). If the weighing value is unstable, the display will continue to “88888888”. Check if anything touches the weighing pan, or check if there is strong wind or vibration. The range for power-on zero is within ±50% of the weighing capacity (kg) around the calibrated zero point. If the power is switched ON while there is a load beyond this range, the scale shows “-------”. Remove the load on the weighing pan. 2. Pressing the
ON/OFF
key again, and the power will be switched OFF.
Auto power-off function It is possible to have the power automatically switched OFF, if zero is displayed for approximately 5 minutes. See “11-2. Function list” and set the function “F1-1”.
7-2. Selecting a weighing unit Press the
UNITS
key to select the weighing unit if necessary.
lb
oz
lb-oz
kg
g
It is possible to specify the display unit that will be shown first when the power is switched on. See the Function “F3”. In some countries or areas “kg” or “g” only is available and the not work. The “kg” or “g” is specified by the Function “F3”.
UNITS
key does
7-3. Basic operation 1. Press the
ON/OFF
key to switch on the scale.
2. Select a weighing unit using the
UNITS
key if necessary.
3. When the display doesn’t show zero, press the zero.
ZERO
key to set the display to
4. When using a tare (container), place the container on the weighing pan, and press the TARE key to set the display to zero (net weight display). 5. Place the item to be weighed on the pan or in the container, and wait for the STABLE annunciator to come on and read the value. 6. Remove the item from the weighing pan. 10
Note for the Legal for Trade Models The range for power-on zero is within ±10% of the weighing capacity around the calibrated zero point. The ZERO key will zero the scale if the weight is within ±2% of the weighing capacity around the power-on zero point. The ZERO annunciator turns on. When the weight exceeds ±2% of the weighing capacity, the key does not work.
7-4. Weighing with Preset Tare Using
PT
key, known tare weight can be entered via the 10-key pad.
1. Remove everything on weighing pan. 2. Press the PT key. The display will blink with the preset tare weight previously entered, or will show blank if there is no preset tare entered. 3. Use the 10-key pad to display the known tare weight. If you hit the wrong key, press the again.
C
key to return to the display at step 2 and try
4. Press the ENT key. The weight display changes to net weight mode and the NET annunciator comes on. 5. Place the item together with the tare (container) to display the net weight.
7-5. To Clear a Tare Weight Either: 1. Remove all items from the weighing pan. 2. Press the
ZERO
key. Then the display will go to zero and tare will be cleared.
The NET annunciator goes off. In some countries or areas, the ZERO key does not clear the tare weight. Press the TARE key after zeroing at step 2. Then the tare weight will be cleared.
Or: 1. Press the PT key. The display will blink with the preset tare weight previously entered, or will show blank if there is no preset tare entered. 2. Press the
0
key and .press the
ENT
key.
3. The tare weight is cleared and the NET annunciator goes off.
11
7-6. Weight display resolution The FS-i series has three kinds of weight display resolution, NORMAL, HIGH and HIGHER. The following is about “kg” and “lb” display for reference. See the “12. SPECIFICATIONS” in detail. NORMAL: HIGH: HIGHER:
1/3,000 1/6,000 ~ 1/7,500 (depending on capacity) 1/12,000 ~ 1/15,000 (depending on capacity)
The factory setting is the NORMAL resolution, but it is possible to change by Function “F2”. Set this function according to the application. The resolution of the Legal for Trade Models is fixed to NORMAL, and the setting F2 cannot be changed.
7-7. Simplified operation mode If desired, the FS-i scale can be set in a Simplified Operation Mode. There are two kinds of mode according to the Function F14-1 and F14-2. In this mode, the following keys are active.
f14-1:
Set the upper/lower limits in the setting F14-0 fist and change it to F14-1. Then, the target weight only can be set using the SAMPLE key. This setting will be used with Function F7-1 or F7-2.
f14-2:
Set the target weight and upper/lower limits in the setting F14-0, and change it to F14-2. Then, The settings can not be changed accidentally. Be sure to set the Function F3 (weighing unit to be used) and F16 (analog sweep display mode) together with the setting above. Because the weighing unit and analog display cannot be changed in the simplified operation mode.
7-8. LCD backlight The Function F17 controls how the LCD backlight turns off. If F17-2 or F17-3 is selected, the backlight will automatically turn off after the weight display continues to be stable for 30 or 60 seconds. It will turn on when the weight changes more than 4d (d= minimum display division) or any key operation is pressed.
12
8. CHECK WEIGHING The FS-i series provides for easy check weighing of products. There are two comparison modes according to number of parameters to set, “Target weight setting mode” and ”Upper & lower limits weight setting mode”. The comparator results are indicated by HI (yellow), OK (green) or LO (red) lights on the display. The results are also audible by the buzzer inside the scale. The analog sweep display will help to see where the weight is in the OK range or how far from the OK range. See “19. ANALOG SWEEP DISPLAY” in detail. If the optional OP-03 or OP-04 is installed, comparator relay output is also available. It is necessary to set the "comparator function (F8-0 ~ F8-6)" parameters, upper limit (HI), lower limit (LO) and target weight (Target weight setting mode only) in advance to use comparator function. Select the operating conditions to compare (see the Function F8). F8-0: Comparator OFF. F8-1: Compare when weight is stable or moving. F8-2: Compare when weight is stable. (STABLE annunciator lit.) F8-3: Compare when weight is stable or moving and more than +4d or less than -4d. F8-4: Compare when weight is stable and more than +4d or less than -4d. F8-5: Compare when weight is stable or moving and more than +4d above zero. F8-6: Compare when weight is stable and more than +4d above zero. d = minimum weight display in “kg”. (see “12-1 Specifications”) If necessary, use the buzzer sound for comparator results. The Function set the buzzer to sound with which result (see the Function F9). The brightness of comparator lights is adjustable. See the Function F15 to adjust.
8-1. Target weight setting mode The TARGET WEIGHT SETTING MODE uses a target weight and +/- deviation from the target. The target weight is entered via the 10-key pad or a sample product to be weighed. The upper (HI) and lower (LO) limits are entered via the 10-key pad. Those will be deviation weight or % of the target weight. Set the Function setting “F7-1 Target weight setting mode with HI/LO limit weight” or “F7-2 Target weight setting mode with HI/LO limit % of the target weight”. The formula to compare is as follows: LO < Lower limit value ≤ OK ≤ Upper limit value < HI Lower limit value = Target weight – LO limit Upper limit value = Target weight + HI limit
13
Setting the target weight by sample 1. Press the ZERO necessary.
key to zero the weight display if
2. Place the sample on the weighing pan. The display shows the sample item’s weight. 3. Press the SAMPLE key. The sample weight flashes and the central part of analog display turns on. Press the ON/OFF key or without anything changed.
SAMPLE
-000. 000 ◄ -003.000 -003.000
key to exit
4. Wait for the STABLE annunciator to come on and press the ENT key to store the sample weight in memory. for a few seconds and
-5tored
1. Press the KEY key and the last target weight entered will flash and the central part of analog display turns on.
-003.500
2. Enter the target weight via the 10-key pad.
-003.000
5. The display shows 5tored returns to weighing mode.
Setting the target weight by key
If you hit the wrong key, press the C key and the old target weight is re-displayed. Continue data entry again. You can place minus value using the key toggles the sign. 3. Press the
ENT
±
key. This
key to store the target weight in memory.
4. The display shows 5tored returns to weighing mode.
for a few seconds and
-5tored
Setting the HI limit weight and LO limit weight Set the Function setting “F7-1 Target weight setting mode with HI/LO limit weight”. 1. Press the HI key. Then, the last HI limit weight entered will flash and the right part of analog display turns on.
2. Enter the HI limit weight via the 10-key pad. If you hit the wrong key, press the C key and the old HI limit weight is re-displayed. Continue data entry again. You cannot enter the “+” or “-” sign for the LO limit. 14
-000.060 -000.050
3. Press the
ENT
key to store the HI limit weight in memory. for a few seconds and
-5tored
5. Press the LO key. Then, the last LO limit weight entered will flash and the left part of analog display turns on.
-000.040
4. The display shows 5tored returns to weighing mode.
6. Enter the LO limit weight via the 10-key pad. If you hit the wrong key, press the C key and the LO limit weight is re-displayed. Continue data entry again.
-000.030
You cannot enter the “+” or “-” sign for the LO limit. 7. Press the
ENT
key to store the LO limit weight in memory.
8. The display shows 5tored returns to weighing mode.
for a few seconds and
-5tored
Example: Target weight: 3.000 kg, HI limit weight: 0.050 kg, LO limit weight: 0.030 kg Then, Upper limit = 3.000 kg + 0.050 kg = 3.050 kg Lower limit = 3.000 kg - 0.030 kg = 2.970 kg LO: weight < 2.970 kg OK: 2.970 kg ≤ weight ≤ 3.050 kg HI: 3.050 kg < weight
Setting the HI limit % and LO limit % Set the Function setting “F7-2 Target weight setting mode with HI/LO limit %”. 1. Press the HI key. Then, the last HI limit % entered will flash and the right part of analog display turns on.
2. Enter the HI limit % via the 10-key pad. If you hit the wrong key, press the C key and the old HI limit % is re-displayed. Continue data entry again.
-0001.50 -0001.00
You cannot enter the “+” or “-” sign for the HI limit. 3. Press the
ENT
key to store the HI limit % in memory.
4. The display shows 5tored returns to weighing mode.
for a few seconds and
15
-5tored
5. Press the LO key. Then, the last LO % weight entered will flash and the left part of analog display turns on.
6. Enter the LO limit weight via the 10-key pad. If you hit the wrong key, press the C key and the LO limit % is re-displayed. Continue data entry again.
-0000.60 -0000.50
You can not enter the “+” or “-” sign for the LO limit. 7. Press the
ENT
key to store the LO limit % in memory.
8. The display shows 5tored returns to weighing mode.
for a few seconds and
-5tored
The HI/LO limit % has 2 decimal places fixed in the setting mode above. The HI/LO limit weight entered in the previous section will not be converted to %, and vice versa. Example: Target weight: 3.000 kg, HI limit weight: 1.00 %, LO limit weight: 0.50 % Then, Upper limit = 3.000 kg + 3.000kg x 1.00 % = 3.030 kg Lower limit = 3.000 kg - 3.000kg x 0.50 % = 2.985 kg
8-2. Upper & Lower limits weight setting mode The UPPER & LOWER LIMITS WEIGHT SETTING MODE uses an upper limit weight and a lower limit weight. The target weight is not used. These upper (HI) and lower (LO) limits are entered via the 10-key pad directly. Set the Function “F7-0 Upper & Lower limit weight setting”. The formula to compare is as follows: LO < Lower limit value ≤ OK ≤ Upper limit value < HI
Setting the upper (HI) limit weight and lower (LO) limit weight Set the Function “F7-0 Upper & Lower limit weight setting”. 1. Press the HI key and the last HI limit weight entered will flash and the right part of analog display turns on.
2. Enter the HI limit weight via the 10-key pad. If you hit the wrong key, press the C key and the old HI limit weight is re-displayed. Continue data entry again. You can enter minus value using the key toggles the sign. 16
±
key. This
-002.340 -003.050
3. Press the
ENT
key to store the HI limit weight in memory. for a few seconds and
-5tored
5. Press the LO key and the last LO limit weight entered will flash.
-003.560
4. The display shows 5tored returns to weighing mode.
6. Enter the LO limit weight via the 10-key pad. If you hit the wrong key, press the C key and the LO limit weight is re-displayed. Continue data entry again. You can enter minus value using the key toggles the sign. 7. Press the
ENT
±
-002.950
key. This
key to set the LO limit weight in memory.
8. The display shows 5tored returns to weighing mode.
for a few seconds and
-5tored
Example: Upper (HI) limit weight: 3.050 kg, Lower (LO) limit weight: 2.950 kg Then, LO: weight < 2.950 kg OK: 2.950 kg ≤ weight ≤ 3.050 kg HI: 3.050 kg < weight
8-3. The comparator memory The FS-i scale can store up to 100 comparator limits by 2 digit numbers, form 00 to 99. To use this function, the scale must not be set to the simplified operation mode.
Storing a set of limits in memory 1. Set the comparator data according to the previous sections “8-1. Target weight setting mode” (target weight, HI limit and LO limit) or “8-2. Upper & Lower limits weight setting mode” (HI limit and LO limit). 2. Press the STORE show 5d 00 .
key. Then the display will
3. Press the memory number, for example 6.
-f5d-00
key, the scale
-f5d-06
4. Press the ENT key to store the data to into the memory. The display shows 5tored for a few seconds and returns to weighing mode.
-5tored
If you stop to store, press the returns to the weighing mode.
C
17
Recalling a set of limits from memory 1. Press the RECALL show rd 00 .
key. Then the display will
2. Press the memory number, for example 3. If you stop to recall, press the returns to the weighing mode. 3. Press the
ENT
C
key, the scale
-frd-00 -frd-03
key to recall the data from the memory.
4. The display shows a set of data for a few seconds one by one and returns to weighing mode.
TARGET WEIGHT (Target weight setting mode only)
-003.000
HI LIMIT
-000.050
LO LIMIT
-000.030 WEIGHING MODE
18
9. ANALOG SWEEP DISPLAY The FS-i series has a 60 segments analog sweep display. The its display mode.
DISP.
key switches
Simple weighing display Target weighing display
Except the UPPER & LOWER LIMIT WEIGHT SETTING MODE (F07-0).
Limit check weighing display Analog display OFF It is possible to specify the display mode that will be shown first when the power is switched on. See the Function F16. Use the analog sweep display with positive target and limit weights.
9-1. Selecting a display mode Simple weighing display mode The 60 segments represent zero to full scale. When the scale is in the net mode, the analog sweep display shows net weight. In the TARGET WEIGHT SETTING MODE (F07-1 or F07-2), the target weight will be shown on the display as a flashing segment. The UPPER & LOWER LIMIT WEIGHT SETTING MODE (F07-0) does not show the target weight segment.
Target weighing display mode The 60 segments represent zero to full scale. When the scale is in the net mode, the analog sweep display shows net weight. In the TARGET WEIGHT SETTING MODE, the 40th segment will flash as the target weight. Thus the 60 segments represent the relative weight that the target weight is scaled to the 40 segments. As the weight increases, the analog segments approach the flashing segment. This will provide a simple way of filling the container to the target weight. The UPPER & LOWER LIMIT WEIGHT SETTING MODE (F07-0) does not have this display.
Limit check weighing display mode The display shows comparison limits and results. The lower limit is shown at the 21st segment flashing and the upper limit is shown at the 40th segment flashing. The results will be known by the segments turned on.
19