Digital Blood Pressure Monitor Model UA-851 Instruction Manual Original
Manuel d’instructions Traduction
Manual de Instrucciones Traducción
Manuale di Istruzioni Traduzione
Bedienungsanleitung Übersetzung
使用手冊 翻譯
1WMPD4001437C
Contents Dear Customers... 1 Preliminary Remarks ... 2 Precautions... 2 Parts Identification... 3 Symbols... 4 Operation Mode... 5 Using The Monitor ... 6 Installing / Changing The Batteries ... 6 Connecting The Air Hose ... 6 Connecting The AC Adapter ... 6 Selecting the Correct Cuff Size ... 7 Applying The Arm Cuff ... 8 How To Take Proper Measurements ... 8 Measurement ... 8 After Measurement ... 8 Measurements ... 9 Normal Measurement... 9 Measurement With The Desired Systolic Pressure ... 10 Notes For Proper Measurement ... 10 Recalling Memory Data ... 11 Clearing the memory data... 11 What Is An Irregular Heartbeat ... 12 Pressure Bar Indicator ... 12 WHO Classification Indicator ... 12 About Blood Pressure... 13 What Is Blood Pressure? ... 13 What Is Hypertension And How Is It Controlled? ... 13 Why Measure Blood Pressure At Home? ... 13 WHO Blood Pressure Classification ... 13 Blood Pressure Variations... 13 Troubleshooting ... 14 Maintenance ... 15 Technical Data ... 15
Dear Customers Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this device will facilitate your daily blood pressure regimen. We recommend that you read through this manual carefully before using the device for the first time. English
1
Preliminary Remarks This device conforms to the European Directive 93/42 EEC for Medical Products. This is made evident by the mark of conformity. (0123: The reference number to the involved notified body) The device is designed for use on adults only, not newborns or infants. Environment for use: The device is for use in the home healthcare environment. This device is designed to measure blood pressure and pulse rate of people for diagnosis.
Precautions Precision components are used in the construction of this device. Extremes in temperature, humidity, direct sunlight, shock or dust should be avoided. Clean the device and cuff with a dry, soft cloth or a cloth dampened with water and a neutral detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean the device or cuff. Avoid tightly folding the cuff or storing the hose tightly twisted for long periods, as such treatment may shorten the life of the components. The device and cuff are not water resistant. Prevent rain, sweat and water from soiling the device and cuff. Measurements may be distorted if the device is used close to televisions, microwave ovens, cellular telephones, X-ray or other devices with strong electrical fields. Used equipment, parts and batteries are not treated as ordinary household waste, and must be disposed of according to the applicable local regulations. When the AC adapter is used, make sure that the AC adapter can be readily removed from the electrical outlet when necessary. When reusing the device, confirm that the device is clean. Do not modify the device. It may cause accidents or damage to the device. To measure blood pressure, the arm must be squeezed by the cuff hard enough to temporarily stop blood flow through the artery. This may cause pain, numbness or a temporary red mark to the arm. This condition will appear especially when measurement is repeated successively. Any pain, numbness, or red marks will disappear with time.
Contraindications The following are precautions for proper use of the device. Do not apply the cuff to an arm with another medical electrical equipment attached. The equipment may not function properly. People who have a severe circulatory deficit in the arm must consult a doctor before using the device, to avoid medical problems. Do not self-diagnose the measurement results and start treatment by yourself. Always consult your doctor for evaluation of the results and treatment. Do not apply the cuff on an arm with an unhealed wound. Do not apply the cuff on an arm receiving an intravenous drip or blood transfusion. It may cause injury or accidents. Do not use the device where flammable gases such as anesthetic gases are present. It may cause an explosion. Do not use the device in highly concentrated oxygen environments, such as a high-pressure oxygen chamber or an oxygen tent. It may cause a fire or explosion. English 2
Parts Identification Artery Position Mark Proper Fit Range
Air Connector Plug
Display
Index Mark Arm Cuff Air Hose
Air Socket
START Button
(Blue illumination)
Battery Compartment
Battery Cover
1.5V Batteries (R6P, LR6 or AA)
DC Jack
Part of Display
MEMORY
Pressure Bar Indicator and WHO Classification Indicator Heart Mark Battery Indicator I.H.B. Indicator (Irregular heartbeat indicator) English 3
Systolic Pressure Diastolic Pressure Pulse Rate
Symbols Symbols
Function / Meaning Recommended Action Standby and turn the device on. Battery installation guide Direct current SN Serial number Date of manufacture Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks. The indicator that appears while Measurement is in measurement is in progress. It blinks progress. Remain as still when the pulse is detected. as possible. Irregular Heartbeat indicator. (I.H.B.) The indicator that appears when an irregular heartbeat or any excessive body movement is detected during the measurement. Previous measurements stored in MEMORY. The battery power indicator during Full Battery measurement. Replace all batteries with Low Battery The battery is low when it blinks. new ones, when the indicator blinks. Try the measurement Unstable blood pressure due to again. Remain very still movement during the measurement. during the measurement. The systolic and diastolic values are within 10 mmHg of each other. The pressure value did not increase Fasten the cuff correctly, during inflation. and try the measurement again. The cuff is not fastened correctly.
PUL. DISPLAY ERROR SYS. DIA. PUL./min.
The pulse is not detected correctly. Systolic blood pressure in mmHg Diastolic blood pressure in mmHg Pulse per minute EC directive medical device label WEEE label EU-representative English 4
Symbols
Function / Meaning Manufacturer Class II device
Recommended Action
Refer to instruction manual/booklet Polarity of DC jack
Operation Mode 1. Normal Measurement With Storing Data (Refer to page 9) Press the START button. Blood pressure is measured and data are stored within the device. The device can store the last ninety sets of data in memory automatically.
2. Recalling The Data (Refer to page 11) When nothing is displayed, press and hold the START button. After all of the display segments, the average of the stored data is displayed. Release the START button Then, the data number of the most recent measurement and the data are displayed. Automatically the older data are displayed in turn.
Press and hold START
Average Release the START button Data number
Stored Data
3. Clearing The Data (Refer to page 11) The stored data are cleared when the batteries are removed or drained.
4. Measurement With The Desired Systolic Pressure Refer to page 10 for measurement with a desired systolic pressure. English 5
Using The Monitor Installing / Changing The Batteries 1. Slide the battery cover to open it. 2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+) and (-) are correct. 3. Slide the battery cover to close it. Use only R6P, LR6 or AA batteries.
Step 2
Step 1 Step 3
CAUTION Insert the batteries as shown in the battery compartment. If not, the device will not work. When (LOW BATTERY mark) blinks on the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction. (LOW BATTERY mark) does not appear when the batteries are drained. Battery life varies with the ambient temperature and may be shorter at low temperatures. Use the specified batteries only. The batteries provided with the device are for testing monitor performance and may have a limited life. Remove the batteries if the device is not to be used for a long time. The batteries may leak and cause a malfunction. The stored data are cleared when the batteries are removed or drained.
Connecting The Air Hose Insert the air connector plug into the air socket firmly.
Air socket
Air connector plug
Connecting The AC Adapter Insert the AC adapter plug into the DC jack. Then, insert the AC adapter into an electrical outlet. The AC adapter, the model TB-233, is sold separately. English 6
DC jack AC adapter plug
Using The Monitor Selecting the Correct Cuff Size
Using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value. The arm size is printed on each cuff. The index and proper fit range, on the cuff, tell you if you are applying the correct cuff. (Refer to "Applying The Arm Cuff" on the next page. ) If the index points outside of the range, contact your local dealer to purchase a replacement cuff. The arm cuff is a consumable. If it becomes worn, purchase a new one. Arm Size Recommended Cuff Size Catalog Number 32 cm to 45 cm Large adult cuff CUF-D-LA 22 cm to 32 cm Adult Cuff CUF-D-A Arm size: The circumference of the biceps. Note: Model UA-851 is not designed for using a small adult cuff.
Symbols that are printed on the cuff Symbols
Function/Meaning
Recommended Action Set the mark on the artery of the upper arm or in line with the ring finger on the inside of the arm.
Artery position mark ▲ REF
Index Catalog number Proper fit range for the adult cuff. It's printed on the adult cuff. Range to use the large adult cuff. Over range printed on the adult cuff. Proper fit range for the large adult cuff. It's printed on the large adult cuff. Under range printed on the adult cuff. Range to use the adult cuff. It's printed on the large adult cuff. Lot number
Use the large adult cuff instead of the adult cuff.
Use the adult cuff instead of the large adult cuff.
Proper fit range
Large adult cuff Proper fit range
Adult cuff
English 7
Using The Monitor Applying The Arm Cuff
1. Wrap the cuff around the upper arm, about 1-2 cm above the inside of the elbow, as shown. Place the cuff directly against the skin, as clothing may cause a faint pulse, and result in a measurement error.
Do not roll up shirtsleeve tightly Cuff
Artery position mark
2. Constriction of the upper arm, caused by tightly rolling up a shirtsleeve, may prevent accurate readings. 3. Confirm that the index within the proper fit range.
1 to 2 cm
Secure the fabric fastener firmly Air hose
points Index
and Proper fit range
How To Take Proper Measurements For the most accurate blood pressure measurement: Sit comfortably on a chair. Rest your arm on the table. Do not cross your legs. Keep your feet on the floor and straighten your back. Relax before measurement. Place the center of the cuff at the same height as your heart. Remain still and keep quiet during measurement. Do not measure immediately after physical exercise or a bath. Rest before taking the measurement. Try to measure your blood pressure at the same time every day.
Measurement During measurement, it is normal for the cuff to feel very tight. (Do not be alarmed).
After Measurement After measurement, press the START button to turn off the power. Remove the cuff. Note: The device has an automatic power shut-off function, which turns the power off automatically one minute after measurement. Allow at least three minutes between measurements on the same person.
English 8
Measurements Model UA-851 is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. If your systolic pressure is expected to exceed 230 mmHg, read "Measurement With The Desired Systolic Pressure" on the next page.
Normal Measurement 1.
Place the cuff on the arm (preferably the left arm). Sit quietly during measurement.
2.
Press the START button. All of the display segments are displayed. Then, 0 (zero) is displayed blinking briefly. Then the display changes, as indicated in the figure at the right, as the measurement begins. The cuff starts to inflate. It is normal for the cuff to feel very tight. A pressure bar indicator is displayed, as in the figure at the right, during inflation. Note: If you wish to stop inflation at any time, press the START button again.
3.
4.
5.
When inflation is complete, deflation starts automatically and (heart mark) blinks, indicating that the measurement is in progress. Once the pulse is detected, the mark blinks with each pulse beat. Note: If an appropriate pressure is not obtained, the device starts to inflate again automatically. When the measurement is complete, the systolic and diastolic pressure readings and pulse rate are displayed. The cuff exhausts the remaining air and deflates completely. Press the START button again to turn off the power.
At heart level Press
START
All of the display segments displayed Zero display Starts inflation Pressurizing
Measurement in progress Systolic pressure Diastolic pressure WHO classification Pulse rate Exhausts remaining air automatically
Note: The device has an automatic power shut-off function, which turns the power off automatically one minute after measurement. Allow at least three minutes between measurements on the same person. English 9
Measurements Measurement With The Desired Systolic Pressure If your systolic pressure is expected to exceed 230 mmHg, use this procedure. 1.
Place the cuff on the arm (preferably the left arm).
2.
Press and hold the START button. Release the button when the average of all the data in memory is displayed. Press and hold the START button again until a number about 30 to 40 mmHg higher than your expected systolic pressure appears.
3.
Release the START button to start measurement, when the desired number is reached. Then continue to measure your blood pressure as described on the previous page.
Notes For Proper Measurement
At heart level
Press and hold the button
START
Release the button when the average is displayed. Press and hold the button
START
Starts inflation
Release the button at the desired systolic pressure
Sit down in a comfortable position. Place Refer to the previous page the arm to be used for the measurement on a table or other support so that the for measurement center of the cuff will be at the same height as your heart. Relax for about five or ten minutes before taking a measurement. If you are excited or depressed by emotional stress, the measurement will reflect this stress as a higher (or lower) than normal blood pressure reading and the pulse reading will usually be faster than normal. An individual's blood pressure varies constantly, depending on what you are doing and what you have eaten. What you drink can have a very strong and rapid effect on your blood pressure. This device bases its measurements on the heartbeat. If you have a very weak or irregular heartbeat, the device may have difficulty determining your blood pressure. Should the device detect a condition that is abnormal, it will stop the measurement and display an error symbol. Refer to page 4 for the description of the symbols. This blood pressure monitor is intended for use by adults only. Consult with your physician before using this device on a child. A child should not use this device unattended. English
10
Recalling Memory Data Note: This device stores the last ninety measurements in memory. The data are retained as long as the batteries are in the device. When you remove the batteries or the batteries are drained, the data will be erased.
START
Press and hold
1. Press and hold the START button until the average of all the data stored in memory is displayed. Release the START button.
2. Then, the data number of the most recent measurement and the data are displayed. Automatically the older data are displayed in turn.
Averaged systolic Averaged diastolic Number of data Release the button Data number of the most recent data Systolic pressure Diastolic pressure Pulse rate of the most recent data
3. After the last data are displayed, the power is turned off automatically. Note: Pressing the START button during the recalling operation will turn the power off. To cancel the recalling operation and enter the measurement mode, press and hold the START button.
Clearing the memory data The stored data are cleared when the batteries are removed or drained.
English
11
What Is An Irregular Heartbeat The device provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you be relaxed, remain still and do not talk during measurements. Note: We recommend contacting your physician if you see this indicator frequently.
Pressure Bar Indicator The indicator monitors the progress of pressure during measurement. Pressurizing
Inflation in progress
Releasing air
Measurement in progress Inflation complete
WHO Classification Indicator Each segment of the bar indicator corresponds to the WHO blood pressure classification described on the next page. WHO Classification Indicator Severe hypertension Moderate hypertension Mild hypertension High normal Normal Optimal
Moderate hypertension
Mild hypertension
: The indicator displays a segment, based on the current data, corresponding to the WHO classification.
English
Example:
12
High normal
About Blood Pressure What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.
What Is Hypertension And How Is It Controlled? Hypertension, an abnormally high arterial blood pressure, if left unattended, can cause many health problems including stroke and heart attack. Hypertension can be controlled by altering lifestyle, avoiding stress, and with medication under a doctor’s supervision. To prevent hypertension or keep it under control: Do not smoke Exercise regularly Reduce salt and fat intake Have regular physical checkups Maintain proper weight
Why Measure Blood Pressure At Home? Blood pressure measured at a clinic or doctor's office may cause apprehension and can produce an elevated reading, 25 to 30 mmHg higher than that measured at home. Home measurement reduces the effects of outside influences on blood pressure readings, supplements the doctor's readings and provides a more accurate, complete blood pressure history.
WHO Blood Pressure Classification Diastolic blood pressure
Standards to assess high Reference Material: Journal of Hypertension 1999, Vol 17 No.2 blood pressure, without Grade 3 hypertension (severe) regard to age, have been Grade 2 hypertension (moderate) established by the World Health Organization (WHO), Grade 1 hypertension (mid) as shown in the chart. High-normal Normal Blood Pressure Variations Optimal An individual’s blood pressure varies greatly on a daily and Systolic blood pressure seasonal basis. It may vary by 30 to 50 mmHg due to various conditions during the day. In hypertensive individuals, variations are even more pronounced. Normally, the blood pressure rises while at work or play and falls to its lowest levels during sleep. So, do not be overly concerned by the results of one measurement. Take measurements at the same time every day using the procedure English
13
Blood Pressure
described in this manual to get to know your normal blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note date and time when recording your blood pressure. Consult your doctor to interpret your blood pressure data.
Typical fluctuation within a day (Measured every five minutes) Systolic
Diastolic
Sleep
Troubleshooting Problem
Possible Reason
Recommended Action Replace all batteries with new Nothing Batteries are drained. ones. appears on the Reinstall the batteries with display, even Battery terminals are not negative and positive terminals when the power in the correct position. matching those indicated on the is turned on. battery compartment. Battery voltage is too low. (LOW BATTERY mark) The cuff does Replace all batteries with new blinks. [ If the batteries are not inflate. ones. drained completely, the mark does not appear.] The cuff is not fastened Fasten the cuff correctly. properly. You moved your arm or Make sure you remain very still body during the and quiet during the The unit does measurement. measurement. not measure. Sit comfortably and still. Readings are The cuff position is not Raise your hand so that the cuff too high or too correct. is at the same level as your heart. low. If you have a very weak or irregular heat beat, the device may have difficulty in determining your blood pressure. The value is different Refer to “Why measure blood from that measured at a pressure at home”. clinic or doctor’s office. Other Remove the batteries. Place them back properly and try the measurement again. Note: If the actions described above do not solve the problem, contact the dealer. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid. English
14
Maintenance Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, request service from your dealer or from the A&D service group. The A&D service group will provide technical information, spare parts and units to authorized dealers. The device was designed and manufactured for a long service life. However it is generally recommended to have the monitor inspected every 2 years, to ensure proper functioning and accuracy. Please contact either your authorized dealer or A&D for maintenance.
Technical Data Type Measurement method Measurement range
UA-851 Oscillometric measurement Pressure: 20 - 280 mmHg Pulse: 40 - 180 beats / minute Measurement accuracy Pressure: ±3 mmHg Pulse: ±5% Power supply 4 x 1.5V batteries (R6P, LR6 or AA) or AC adapter (TB-233) (Not included) Approx. 450 measurements, when AA Alkaline Number of measurements batteries are used, with pressure value of 180 mmHg at room temperature of 23°C Classification Internally powered ME equipment (Supplied by batteries) Class II (Supplied by adapter) Continuous operation mode Clinical test According to ANSI / AAMI SP-10 1992 EMC IEC 60601-1-2: 2007 Memory Last 90 measurements Operating conditions +10°C to +40°C / 15%RH to 85 %RH 800 hPa to 1060 hPa Transport / Storage conditions -10°C to +60°C / 15%RH to 85 %RH Dimensions Approx. 150 [W] x 156 [H] x 120 [D] mm Weight Approx. 480 g, excluding the batteries Applied part Cuff Type BF Useful life Device: 5 years (when used six times a day) Cuff: 2 years (when used six times a day) English
15
Accessory AC adapter TB-233
TB-233C
The adapter is to connect the blood pressure monitor to a power source at home. Please contact your local A&D dealer for purchasing. The AC adapter is required to be inspected or replaced periodically. Input: 100-240V Output: 6V 500mA 139°C
TB-233BF
Input: 240V Output: 6V
2A
500mA 139°C
2A
Accessories sold separately Cuff Catalog Number Cuff Size CUF-D-LA Large adult cuff CUF-D-A Adult cuff AC adapter Catalog Number Plug TB-233C Type C TB-233BF Type BF
Arm Size 32 cm to 45 cm 22 cm to 32 cm
Note: Specifications are subject to change without prior notice. EMC table information is listed on our website: http://www.aandd.jp/products/manual/medical/emc_acadapter_en.pdf
English
16
1-243 Asahi , Kitamoto-shi, Saitama 364-8585, JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119