C3 WAVE
CONNECT / CAPTURE / CONFIRM
QUICK REFERENCE GUIDE
P-Wave and Catheter Placement “Position the tip of a CVAD in the lower segment of the superior vena cava at or near the CAJ for adults and children” (INS 2016)
FPO - need hi res Pulled from Word doc. Need to replace here and in #6309
No evident P wave changes indicates catheter is not in acceptable position
1
A P wave at its max height will indicate the catheter is in the lower 1/3 of the SVC
A downward deflection edge of P wave indicated the catheter is entering the RA
A biphasic P wave indicated the catheter is in the RA
An inverted P wave indicated catheter is approaching the right ventricle
Alexander M. Infusion Standards. Journal of Infusion Nursing. 2016;39(4):181-182.
QUICK REFERENCE GUIDE
C3 WAVE
Bluetooth® Pairing
HUB BLUETOOTH® PAIRING
REMOTE BLUETOOTH® PAIRING
Press and hold the power button for about 8 seconds. The blue power LED will flash rapidly to indicate that the hub has erased pairing information.
Press and hold all four arrow buttons for about 5 seconds. This will remove all pairing information from the remote.
On the new iPad go to Settings & Support and press the “Pair” icon under Hub Serial # ®
On the hub press and hold the power button for 2 seconds until the blue LED begins to flash slowly. On the iPad press the area marked “Chest Hub”. This will pair the hub with the iPad®. ®
You can verify that the hub has paired correctly by checking that the serial numbers match.
QUICK REFERENCE GUIDE
On the new iPad® go to Settings & Support and press the “Pair” icon under Remote Serial # On the remote press any button and the blue remote LEDs will begin to flash slowly. On the iPad® press the area marked “Remote”. This will pair the remote with the iPad®. You can verify that the remote has paired correctly by checking that the serial numbers match.
C3 WAVE
Warning Screen Pop-Ups WARNING
CONDITION
PERSISTENCE
PICC Lead Off
PICC lead disconnected from the patient. App in "internal mode".
Skippable cleared by pressing "Dismiss"
ECG Lead Off
ECG lead disconnected from the patient.
Skippable cleared by pressing "Dismiss"
Internal System Error
I2C bus failure
Unskippable cleared by power cycling the device
Charger Error
Battery charger failure
Unskippable cleared by removing and re-inserting the DC power connector
Internal Power Error
3.3V power rail goes below 3.1V
Unskippable cleared by power cycling the device
Hub Disconnected
Wireless connection with the hub is lost
Unskippable cleared by reconnecting the hub
QUICK REFERENCE GUIDE
SCREENSHOT
C3 WAVE
Warning Screen Pop-Ups WARNING
CONDITION
PERSISTENCE
Hub Low Battery
Hub battery is below 50%
Skippable cleared by pressing “Dismiss”
Hub Critical Battery
Hub battery is below 20%
Skippable cleared by pressing “Dismiss”
Remote Low Battery
Remote battery level is below 50%
Skippable cleared by pressing “Dismiss”
Remote Critical Battery Remote battery level is below 20%
Skippable cleared by pressing “Dismiss”
Hub Charging
Hub is charging via charging base or power supply
Unskippable cleared by unplugging or removing hub from charging base
Informational Message: “…confirm entry…”
Low [Grey]
Confirm selected action, such as “Finish Case”, or “Delete File”
QUICK REFERENCE GUIDE
SCREENSHOT
N/A
C3 WAVE
Troubleshooting PROBLEM
CAUSE
SOLUTION
Hub will not power on
Battery fully discharged
Charge the hub
Power button not held long enough
Press and hold the power button for 2 - 5 seconds until the button begins to flash
Hub is charging and connected to the app for a procedure
Charge the hub and remove it from the charger before going into the app to start a procedure
Hub will not turn off
Power button not held long enough
Press and hold the power button for 2 - 5 seconds until the button begins to flash
Hub will not connect
Hub is off
Press and hold the power button for 2 - 5 seconds
Not in a procedure
The hub and remote only connect on a new procedure
Hub not paired
See “Pairing Hub with App” for instructions
Bluetooth® turned off
In the iPad settings screen make sure that Bluetooth® is on
Hub power connector not inserted
Make sure the hub power connector is securely inserted
Power adapter not plugged into the wall
Make sure the power adapter is plugged into the wall
Outlet does not have power
Make sure AC power is present at the outlet
Hub not seated on the charging cradle
Make sure the hub is fully seated in the charging cradle
Batteries over-discharged
If the LED shows solid red when charging replace the batteries
No batteries present
Verify that approved medical grade batteries have been installed
Hub will not charge – no charge LED
Hub will not charge – solid red charge LED
QUICK REFERENCE GUIDE
C3 WAVE
Troubleshooting PROBLEM
CAUSE
SOLUTION
Remote will not power on
Batteries drained
Replace batteries
Remote powers on but will not connect (continuous flashing)
Remote not paired
See “Pairing Remote with App” for instructions
Not in a procedure
The hub and remote only connect on a new procedure
Bluetooth® turned off
In the iPad settings screen make sure that Bluetooth® is on
ECG lead disconnected
Check that each ECG lead is connected to the hub and to an ECG pad
ECG pads loose
Check that the ECG pads are firmly attached to the patient
Old ECG pads
Replace with new ECG pads
Mismatched ECG pads
Make sure the ECG pads are from the same package and are placed at the same time
Misplacement of pads
Verify that the pads have been placed in the correct location
Improper skin preparation
Clean the ECG placement location with an alcohol pad before placing the electrodes
Bad skin to electrode contact
When applying ECG pads press firmly against the patient to ensure good contact between the electrode and the skin
No ECG signal (with a warning message E205)
Poor quality ECG or PICC signal
QUICK REFERENCE GUIDE
C3 WAVE
Troubleshooting PROBLEM
CAUSE
SOLUTION
No PICC signal (with a warning message E202)
ECG lead disconnected
Check that each ECG lead is connected to the hub and to an ECG pad
ECG pads loose
Check that the ECG pads are firmly attached to the patient
PICC lead disconnected
Check that the ECG pads are firmly attached to the patient
Poor blood contact
Flush the PICC and wait about 30 seconds for signal capture
No PICC or ECG Signal (no warning message)
Hub incorrectly paired
Exit the app and open the iPad settings. Select Bluetooth® and under “My Devices” press the next to each item marked “Chest hub”. Select “Forget Device” for each item. On the hub press and hold the power button until it flashes rapidly (about 8 seconds). Re-pair the hub (see “Pairing the Hub” for instructions).
ECG Wave Inverted
Leads connected to pads in correctly
Verify that each lead is connected to the correct pad
Leads connected to hub incorrectly
Verify that each lead is connected to the correct port on the hub
Cannot find the printer wireless access point in the iPad settings
See “Configuring the Printer” for instructions
Cannot find the printer in the app print menu
Make sure that the iPad is connected to the printer. Wait 60 seconds and attempt to print again.
Unable to print
QUICK REFERENCE GUIDE
C3 WAVE
Troubleshooting PROBLEM
CAUSE
SOLUTION
iPad/App slow or crashing
Insufficient battery
Plug the printer into a wall outlet or charge the battery
No paper
Insert a label roll into the printer
Failed to retrieve printer status
Verify that a compatible printer is connected. Power cycle the printer
Overheating error
Allow the printer to cool down for 15 minutes
Paper jam
Open the lid and clear the paper jam
Replace paper cassette while printing.
Check that the label has been loaded correctly and that the printer has paper
Internal system error (E201) displayed
Feeding error or end of paper
Check that the label has been loaded correctly and that the printer has paper
Internal power error (E208) displayed
Please load the label roll (DK-2205)
Insert a label roll into the printer
US AND INTERNATIONAL DISTRIBUTORSHIPS
INTERNATIONAL DIRECT MARKETS
800-772-6446
+31 20 753 2949
Customer Service: Option 2 Clinical Specialists: Option 5 Hardware Services: Option 6
Customer Service: Option 1
Charging circuit error (E206) displayed
8AM-7PM EST, MON-FRI
QUICK REFERENCE GUIDE
8AM-5PM EST, MON-FRI
C3 WAVE
Consult your AngioDynamics representative for country specific product availability.
IMPORTANT RISK INFORMATION INTENDED USE/ INDICATIONS FOR USE: The C3 Wave System is indicated for use in the positioning of Peripherally Inserted Central Catheters (PICC). The C3 Wave provides real-time catheter tip location information by displaying changes in the patient’s cardiac electrical activity. The C3 Wave is indicated for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tip placement in adult patients. NOTE: Limiting, but not contraindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P-Wave: • Atrial fibrillation • Atrial flutter
• Severe tachycardia • Pacemaker-driven rhythm
• Chronic obstructive pulmonary disorder (COPD)
Such patients are easily identified prior to PICC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location. WARNINGS: The C3 Wave™ System works with the normal sinus rhythm of the heart. Do not rely on ECG signal detection for catheter tip positioning when interpretation of the external or intravascular ECG P-wave is difficult. For example, when: • P-wave is not present • P-wave is not identifiable • P-wave is intermittent •Place ECG adhesive electrodes carefully at locations indicated in these Instructions for Use and ensure good skin-electrode contact. Failure to do so may cause unstable ECG waveforms and/or ECG waveforms that are not described in these Instructions for Use • All components in the accessory pack are single use items. Do not reuse or reprocess • Monitor catheter tip placement during insertion procedure and verify catheter tip location placement using your institutions’ guidelines. • Failure to verify catheter placement may result in serious trauma or fatal complications. • Inspect package and product prior to use to verify that no damage has occurred during shipping. • Reuse or reprocessing may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness or death. • Reuse or reprocessing may also create a risk of contamination of the device and/ or cause patient infection or cross- infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another. • Contamination of the device may lead to injury, illness or death of the patient. • After use, dispose of product and packaging in accordance with hospital, administrative and/or local government policy. Refer to Directions for Use and/or User Manual provided with the product for complete Instructions, Warnings, Precautions, Possible Adverse Effects and Contraindications. Observe all instructions for use prior to use. Failure to do so may result in patient complications. CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician.
USA > 14 Plaza Drive, Latham, NY 12110 > tel: 800-772-6446 or 518-798-1215 > fax: 518-798-1360 International > Haaksbergweg 75 (Margriettoren), 1101 BR, Amsterdam Z-O > The Netherlands tel: +31 (0)20 753 2949 > fax: +31 (0)20 753 2939 www.angiodynamics.com
800-772-6446
8AM-7PM EST, MON-FRI Customer Service: Option 2 Clinical Specialists: Option 5 Hardware Services: Option 6
+31 20 753 2949 AngioDynamics, Inc. and the AngioDynamics logo are trademarks and/or registered trademarks of AngioDynamics, Inc., an affiliate or subsidiary. C3 Wave and the C3 Wave logo are trademarks and/or registered trademarks of AngioDynamics VA LLC. All other trademarks are property of their respective owners. © 2020 AngioDynamics VA LLC. GL/VA/MS/30 (v1.0) 01/2020
8AM-5PM EST, MON-FRI Customer Service: Option 1