High Flow Oxygen Therapy – FD140 Guidance for Use* 1. Select correct size of AquaNASE® nasal cannulae. 2. Apply AquaNASE® nasal cannulae and secure neck strap as shown overleaf. 3. Insert humidification chamber onto heater plate and spike water bag. 4. Attach circuit as shown a-d.
5. Connect heater cable and temperature probes. 6. Turn on heater and select invasive mode. 7. Set flow at 20L/min and select prescribed FiO2 for 2 minutes. 8. Increase flow to max 70L/min as required monitoring SpO2 and respiratory rate.
FD140 Flow Driver (Supplied Separately)
AquaVENT® FD140 0
5 10 15 20 25
PAP
02
8 7 4a
1
2
4d
4c
5
Heater Humidifier (Supplied Separately)
4b 3
6
*This guidance should be read in conjunction with the IFU supplied with the specific Breathing Circuit, Heater Humidifier, AquaNASE® nasal cannulae and the Flow Driver.
PMA_AMFD140_HFOT_DSIFU_v1 UK 01/12/17_15639
Fitting guide Step 1 Fit the looped neck support strap by placing over patients head and sliding down around the neck as shown.
Step 2 Adjust the looped neck support strap by pulling one strap through the clamp as shown.
Step 4
Adjust the ends of the elasticated strap to secure the device.