0047-TM3 Elite Technical Manual Ver 2008-1.1
Technical Manual
18 Pages
![ Arcoma-Elite Version 2008-1.1 0047-TM3 NC](https://public.bioclinicalservices.com.au/library/representations/proxy/eyJfcmFpbHMiOnsiZGF0YSI6MTAwMTksInB1ciI6ImJsb2JfaWQifX0=--805fca02a2f9722d8a57c2f722dc1127acd9b6c5/eyJfcmFpbHMiOnsiZGF0YSI6eyJyZXNpemUiOiIzMjB4In0sInB1ciI6InZhcmlhdGlvbiJ9fQ==--4214356078d1f110e449cda512a9432562da3e70/0047-tm3_elite_technical_manual_ver_2008-1-1.pdf.png)
Preview
Page 1
Arcoma-Elite Version 2008-1.1
0047-TM3 NC
Contents
Contents Contents………………………………...…A-1 Introduction………………………………..…1 Important note……………………………………….1
System modes of operation.……………….2 General……………………………………………….2 Range of application……...……………………...…2
Safety……………………..………………….3 General…..…………………………………………..3 Maintenance…………………………………………3 Checks, test………………………………………….3 Installation, repair……………………………………3
Cleaning………………………………………4 Function and safety checks…………...……5 Daily checks……………………….…………………5 Monthly checks………………………………………5 Annual maintenance……………..…………………5
Operating the table………………….………6 General……………………………………………….6 Controlling the table…………………………………6 Controlling the table using the remote……….……7 Controlling the table using the foot switch ……….8
Protective measures…………………...……9 Mechanical safety…………………………...………9 Safety-relevant parts subject to wear……………10 Waste disposal…………………………………….10
Location of Identification label.………..…11 0047 TABLES
Arcoma-Elite Version 2008-1.1
A-1
0047-TM3 NC
Contents Technical Specifications..…………………12 External Power Supply…….……………………...12 General…..…………………………………………12 Operating environment…….……………………...13 Classifications….………………………………..…13 Bill of materials……………………………………..14 How to use the wrench…….……………………..24 Spare parts………………….……………………...25 Electrical schematics.…..…………………………26
A-2
Arcoma-Elite Version 2008-1.1
0047 TABLES
0047-TM3 NC
Introduction
Introduction Important note Using the Arcoma “Medstone-Elite” Tables presupposes that the operating personnel is thoroughly familiar with the operating instructions, in particular: There are many different versions ranging from a single movement to several movements. This manual covers all types. •
Safety (page 3)
•
Function and safety checks (page 5)
•
Protective measures (page 9)
0047 TABLES
Arcoma-Elite Version 2008-1.1
1
0047-TM3 NC
System modes of operation
System modes of operation General 0047-Series mobile tables are used for imaging applications. The tables, depending on your option are provided with several actuators and motorized column giving the tables good movement range.
B
A
D C
Figure 1. 0047-TM4 (shown) A Table Top
C Actuator
B Table Pad
D Column
Range of application The intended use of the table is for examinations and imaging (‘C; Arm required)only. Medical staff must supervise the equipment when it is in use near the patient.
2
Arcoma-Elite Version 2008-1.1
0047 TABLES
0047-TM3 NC
Safety
Safety General CAUTION! This device is restricted to use by or on the order of a physician. The device is intended for use in radiographic examinations under the guidance of trained health care professionals. Operating personnel must be familiar with the equipment and the instructions given in this manual before using it.
Maintenance To ensure the safety of the patients, the operators, and third parties, the equipment must be checked every 12 months to maintain its reliability and serviceability.
Checks, test Checks and tests that shall be performed with specified intervals are described in Function and safety checks (page 5)
Installation, repair Modifications of, or additions to the product must be made in accordance with the legal regulations and generally accepted engineering standards. The supplier cannot assume responsibility for the safety features and for the reliability and performance of the equipment, if: •
installation, equipment expansions or modifications are not carried out by persons authorized by us.
•
components affecting safe operation of the unit are not replaced by original spare parts in case of a malfunction.
•
the product is not used in accordance with the operating instructions.
On request, we will make technical documents for the product available to you at a charge that will be invoiced. This does not imply an authorization to undertake repairs. We accept no responsibility for repairs performed without our expressed written approval. We suggest that you obtain a report indicating the nature and extent of the work performed from the persons carrying out such work, and specify that the report shows any changes in rated parameters or operating ranges, as well as date, the name of the firm and a signature.
0047 TABLES
Arcoma-Elite Version 2008-1.1
3
0047-TM3 NC
Cleaning
Cleaning CAUTION! Surfaces that are in contact with the patient shall be cleaned with a lint free cloth dampened with a small amount of soapy water or cleaning solution.
4
Arcoma-Elite Version 2008-1.1
0047 TABLES
0047-TM3 NC
Function and safety checks
Function and safety checks If any malfunction is detected, the entire equipment must be taken out of use until a service engineer from the supplier eliminates the malfunction or by local technical staff trained by the supplier.
Daily checks When necessary, clean according to Cleaning (page 4).
Monthly checks •
Check the legibility of the label according to section Location of Identification label (page 11). If any label must be replaced, please notify the supplier customer service.
•
Power up the stand and check all functions according to Technical Specifications (page 12).
Annual maintenance The following maintenance is necessary to perform every 12 months to keep the equipment safe and operational. •
Power up the stand and check all functions according to Technical Specifications (page 12).
•
Check parts according to Safety-relevant parts subject to wear (page 10).
0047 TABLES
Arcoma-Elite Version 2008-1.1
5
0047-TM3 NC
Operating the table
Operating the table General CAUTION ! Make sure that all wheels are locked before the patient is placed on the table. USE OF PATIENT RESTRAINT IS REQUIRED DURING USE.
Controlling the table The table movement is controlled with the hand switch in accordance with the following figures. Figure 2
TRENDELENBURG
LIFT
LATERAL TILT
6
Arcoma-Elite Version 2008-1.1
0047 TABLES
0047-TM3 NC
Protective Measures
Protective measures Mechanical safety It is the operator’s duty to ensure that any danger to the patient or third parties is precluded before movements are released. Possible hazard areas are indicated in the figure below.
Figure 3. Possible Hazard Areas 0047 TABLES
Arcoma-Elite Version 2008-1.1
7
0047-TM3 NC
Protective measures Safety-relevant parts subject to wear This unit contains the following safety-relevant parts, which should be checked by service personnel according to the following: •
Tightening of bolts: Bolt Size
Torque Tightness
M6 M8 M12
6 Ft Lbs 16 Ft Lbs 60 Ft Lbs - Normal 30 Ft Lbs - Table Top 21 In Lbs
M4
8
Arcoma-Elite Version 2008-1.1
0047 TABLES
0047-TM3 NC
Location of Identification label
Location of Identification label
LOCATION OF IDENTIFICATION LABEL
Figure 4. Location of Identification label
0047 TABLES
Arcoma-Elite Version 2008-1.1
9
0047-TM3 NC
Technical specifications
Technical specifications General Table top size
22” (559 mm) x82” (2083 mm)
Model number Table top lateral rotation Table top movement, vertical Table top movement, Trendelenburg Travel Longitudinal Travel Lateral Weight
0047-TE N/A 300mm N/A N/A N/A 500 lbs
0047-TT N/A 300mm +15°/-15º N/A N/A 550 lbs
0047-TL ± 15°, 300mm N/A N/A N/A 550 lbs
0047TM3 ± 15°, 300mm +15°/-15º N/A N/A 600 lbs
0047-TM4 ± 15°, 300mm +15°/-15º 400mm N/A 650 lbs
0047-TM5 ± 15°, 300mm +15°/-15º 400mm 200mm 650 lbs
400mm Travel
200mm Travel
± 15o o
± 15
300mm Travel
Figure 5. Table Movement Range
10
Arcoma-Elite Version 2008-1.1
0047 TABLES
0047-TM3 NC
Technical specifications
Technical specifications External power supply
Power supply Power consumption
120 VAC, 50 – 60 Hz or 230VAC, 50 Hz 9 Amps, 4.5 Amps
Operating environment
Operating temperature Storing temperature Operating and storing humidity Patient weight limit
+10° - +30° C -20° - +40° C 30 – 80 % relative humidity, non condensing 500 lbs, evenly distributed
Classifications
Protection against electrical shock
Duty factor Mode of operation Degree of safety of apllication in the presence of flammable anaesthetic mixture with air or nitrous oxygen
0047 TABLES
Arcoma-Elite Version 2008-1.1
Type B according to degree of protection (IEC 601-1) Class 1 according to type of protection (IEC 601-1) EMC-disturbances can appear even if the product complies with the EMC demands 15% Continuous operation with intermittent loading Not suitable for use in the presence of flammable anaesthetic mixtures.
11
0047-TM3 NC
12
Arcoma-Elite Version 2008-1.1
0047 TABLES
0047-TM3 NC
Recommended Spare Parts
Page 13
Arcoma-Elite Version 2008-1.1
0047-TM3 NC
14
Arcoma-Elite Version 2008-1.1
0047 TABLES
0047-TM3 NC
0047 TABLES
Arcoma-Elite Version 2008-1.1
15
0047-TM3 NC